This document is a Salient Characteristics Brand Name or Equal (BNOE) specification for a digital fundus camera system.
The procurement specifies a Nikon Optos Silverstone P200TxE camera (M5510) with a minimum quantity of one unit. The equipment must provide nine imaging modalities combined with ultra-wide field (UWF) imaging and integrated Swept Source OCT capability. Required image capture features include non-mydriatic operation with minimum 2mm pupil capability, confocal scanning laser ophthalmoscopy (cSLO) imaging through cataracts, and 4-in-1 Color Depth Imaging. The UWF imaging modalities must include Color RG, Color RGB, Sensory (red-free), Choroidal, Autofluorescence Green and Blue, Fluorescein (FA), and Indocyanine Green (ICG), with optomap resolution of 20 μm standard or 14 μm plus, and exposure time under 0.4 seconds. OCT specifications require a 1050 nm Swept Source wavelength with axial resolution less than 7 microns, transverse resolution less than 20 microns, scan depth up to 2.5 mm, and A-scan rate up to 100k cycles per second, with scanning capabilities for line scans (6-23 mm width) and volume scans (3.5-9 mm height, 6.0-14 mm width). The equipment must include an adjustable table with space requirements of 887 mm width, 600 mm depth, and 725-1205 mm height, with unit dimensions of 23"D x 22"W x 28"H.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24W26Q0147.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALIENT CHARACTERISTICS BRAND NAME OR
EQUAL (BNOE)
M5510 – Camera, Fundus, Digital – Nikon, Optos Silverstone P200TxE
a) Unit must include minimum Image Capture Features:
a. Nine imaging modalities and UWF with integrated Swept Source OCT, facilitate detailed examination of the retina-vitreous to sclera
b. UWF guided, Swept Source OCT, images pathology anywhere on the optomap
c. 1050 nm OCT light source, provides deeper tissue penetration for clear, detailed choroidal imaging
d. Non-mydriatic, with a minimum 2mm pupil.
e. cSLO imaging through most cataracts.
f. 4-in-1 Color Depth ImagingTM provides important clinical data from the retinal surface through the choroid
b) UWF Imaging:
a. Imaging Modalities: Color RG, Color RGB, Sensory (red-free), Choroidal, Autofluorescence Green (Green AF), Autofluorescence Blue (Blue AF), Fluorescein (FA), Indocyanine Green (ICG)
b. Resolution: optomap: 20 μm, optomap plus: 14 μm
c. Laser Wavelengths: Blue Laser: 488 nm (for FA), Red laser: 635 nm, Green laser: 532 nm (for AF), Infra-red: 802 nm (for ICG)
d. Exposure Time: Less than 0.4 seconds
c) OCT Imaging:
a. Signal Type: Optical scattering from tissue
b. Signal Source: Swept Source OCT, Wavelength 1050 nm
c. Axial Resolution: < 7 micron
d. Transverse Resolution: < 20 micron
e. Scanners: Galavanometric X, Y pair
f. Scan Depth: Up to 2.5 mm
g. A-Scan Rate: Up to 100k cycles/sec
h. Scan Types: Line Scans Width: 6, 14, 23 mm; Volume Scan Height: Min 3.5 mm; Max 9 mm; Volume Scan Width: Min 6.0 mm; Max 14 mm
d) Must include table (Space Requirements: Width: 887 mm / 35 in, Depth: 600 mm / 24 in, Height:
725 to 1205 mm / 29 - 48 in)
e) Dimension: 23”D x 22”W x 28”H
f) Quantity: 1