This is a Brand Name or Equal Worksheet documenting the salient characteristics of an automated external defibrillator (AED) and defibrillator device for federal procurement purposes.

The document specifies detailed technical and operational requirements for a multi-function defibrillator system capable of functioning as both an AED and manual defibrillator for adult, pediatric, and neonatal patients. Key physical specifications include maximum dimensions of 9.4 x 10.3 x 8.6 inches, weight of 13.2 pounds with battery, and a color LCD 800 x 600 touchscreen with integrated printer. The device must feature a rechargeable lithium-ion battery with less than 7 seconds ready charge time, integrated Wi-Fi and Bluetooth connectivity, and a 5-year hospital warranty. Clinical capabilities required include high-current rectilinear biphasic waveform (RBW) defibrillation, configurable shock protocols, adult and pediatric CPR feedback with compression artifact filtering, non-invasive blood pressure monitoring, mainstream end-tidal CO2 monitoring via Masimo, and pulse oximetry compatibility with both Masimo and Nellcor systems. The device must support non-invasive pacing with 40 millisecond pacing pulse capability, rapid shock analysis in under 5 seconds, ventilation volume indicators, audiovisual CPR coaching with numeric displays for compression depth and rate, and a Shock Conversion Estimator Module utilizing AMSA measurement technology. Additional features include color-coded pediatric-specific pads and electrodes, easily removable parameter pods for reduced repair downtime, extensive IT configuration options with password protection, and a secondary tablet-based display for real-time case data review and playback.

View the file

Other files

File Type Posted
36C26126Q0651 0003.pdf PDF
36C26126Q0651 Questions and Responses Revised.pdf PDF
36C26126Q0651 0002.pdf PDF
36C26126Q0651 Questions and Responses.pdf PDF
36C26126Q0651 0001.pdf PDF
36C26126Q0651.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Brand Name or Equal Worksheet (“Salient Characteristics”) v 2.0 Brand Name Item:

· Maximum dimensions: 9.4 in x 10.3 in x 8.6 in

· Weight: 13.2 pounds with battery

· Includes alternating current (AC) power cord for use in the United States

· Rechargeable Lithium-ion Battery Pack

· Integrated Wi-Fi and Bluetooth (functionality can also be disabled)

· 5-year hospital warranty

· Ability to function as adult or pediatric automated external defibrillator (AED)

· Operational manual included with each defibrillator which includes manufacturer’s instructions – Language, English Physical/operational features

· Color LCD 800 x 600 touchscreen

· Integrated printer

· Ability to complete self-test directly with specific pre-connected electrodes

· Adult and pediatric dual sensor CPR feedback technology designed to negate “mattress effect”

· Masimo and Nellcor Pulse Oximetry Compatibility

· Non-Invasive Blood Pressure

· For use with adult, pediatric and neonatal patients

· Mainstream End tidal CO2 – Masimo

· Lock screen button

· Ability to operate with touchscreen or rotary dial

· Easily removeable parameter pods designed to decrease downtime for repairs Configurations

· Offers extensive configuration options for customization including the ability to operate and power on either in AED Mode or Manual mode with the ability override and change with just a few button pushes

· Critical care transport (CCT) device (or configuration) offered

· Dedicated, IT configuration menu that is password-protected

· Configurable for either cardiopulmonary resuscitation (CPR) or shock-first driven protocols

· Energy sequences can be configured for single or multiple shocks with fixed or escalating energy levels Batteries

· Rechargeable Lithium lon Battery

· Run-Time and power level indicator

· Designed to automatically re-calibrate, Stores history of use and maintenance

· Less than 7 seconds of ready charge time for a new fully charged battery

Clinical features Defibrillation

· High-current Rectilinear Biphasic Waveform (RBW)

· Configurable option to power on and function as an automated external defibrillator (AED)

· Manual defibrillator to support pediatrics to adults including automatic safety features eliminating the risk of delivering an adult energy dose to a pediatric patient with the use of appropriate pads

· Pediatric specific AED analysis algorithm Electrodes

· Specific adult and pediatric CPR electrodes

· Specific, color-coded packaging for both patient types

· Pediatric pads which clearly differentiate from adult pads Audiovisual CPR Coaching Module

· Numeric display of chest compression depth and rate for adult and pediatric patients

· Visual and audio prompts to coach cardiopulmonary resuscitation (CPR) depth for adult patients

· Visual Release bar prompt to indicate appropriate release off the chest

· Metronome to coach rate for adult and pediatric patients.

· Elapsed time counter (since last detected chest compression) Ventilation Volume Indicators

· Numeric display of tidal volume and ventilation rate delivered by manual oxygen bag

· Visual indicators on screen and cable indicating in or out of range measurements

· Can be used with intubated and non-intubated patients Compression Artifact Filtering

· Algorithms which filters out chest compression artifact during on-going CPR to assist in reducing pause times Non-invasive Pacing Capabilities

· 40 msec pacing pulse

· 4:1 button which allows clinicians to visualize the patient’s intrinsic rhythm without losing capture Rapid Shock Capabilities

· Analyzes ECG rhythm by filtering out chest compression artifact in AED Mode

· Determines shock/no shock decision and enables ability to shock in under [5] seconds Vitals Display for TBI Patient

· Technology indicated to provide graphical trend data for SpO2, Systolic BP (SBP), and EtCO2 as well as ventilation assistance relevant to the management of a TBI patient.

Shock Conversion Estimator (SCE) Module

· Algorithm which measures the probability that a shockable rhythm will be successfully converted by immediate defibrillation

· Utilizes AMSA measure developed by Weil Institute of Critical Care Medicine Secondary Tablet

· Tablet based device and software which provides a secondary display of activity on the defibrillator screen

· Provides customized, real-time views of case data

· Ability to rewind, review, and playback data

File details come from the government source that posted it. Updated .