The document is a Salient Characteristics specification for an Auditory Brainstem Response (ABR)/Auditory Evoked Potentials Machine for audiology testing. The machine must be capable of performing multiple specialized auditory tests including ECochG, ABR, MLR, LLR, MMN, P300, ASSR, cVEMP, oVEMP, and electrically evoked ABR for cochlear implant users. Required technical specifications include IP 30 insert phone and B82 bone conduction oscillator transducers, OtoAccess database integration, and advanced features across ABR/EP, ASSR, and VEMP testing domains.
Specific technical requirements include auto peak picking, Bayesian weighting, objective waveform analysis measures, support for traditional and tone burst stimuli, CE chirp stimuli (narrow and broadband), ECochG area calculator, multi-frequency binaural testing, and EMG-related VEMP features such as patient monitoring, scaling, and controlled recording. The document appears to be part of a pre-solicitation for the Department of the Army Medical Command's Madigan Audiology procurement, with the solicitation number PANCHA25P0000008600, indicating a potential medical equipment acquisition for military audiology services.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sources Sought_Audology.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics ABR/Auditory Brainstem Response (Auditory Evoked Potentials) Machine
• Must be able to complete the following tests: ECochG, ABR, MLR, LLR, MMN, P300, ASSR, cVEMP & oVEMP, and electrically evoked ABR for cochlear implant users
• Must have IP 30 insert phone and B82 bone conduction oscillator transducers
• Mush have OtoAccess database
• Have the following ABR/EP features:
o Auto peak picking o Baysian weighting o Objective waveform analysis measures including FMP (function of multiple points), residual noise calculator, waveform reproducibility, and signal-to-noise ratio o Make use of traditional clinic and tone burst (250-8000 Hz) stimuli o Have available narrow band (500-4000 Hz) and broadband CE chirp stimuli o ECochG Area Calculator
• Have the following ASSR features o Automatic waveform analysis o Multi-frequency binaural simultaneous testing o Narrow band CE chirps (500-4000 Hz) o Masking calculator
• Have the following VEMP features:
o EMG VEMP patient monitor o EMG scaling o EMG controlled recording