S02 - ATTACHMENT 2 - LINE-ITEM DESCRIPTION - IRRADIATOR - BRAND NAME OR EQUAL TO RAD SOURCE IAW SOW - 08-19-2026.pdf

PDF 54 KB Posted

Attached to
6640--IRRADIATOR FOR COLONY OF HUMANIZED MICE "BRAND NAME OR EQUAL TO RAD SOURCE" OEM LETTER FROM THE MANUFACTURER IS REQUIRED. IF NO OEM LETTER VENDOR WILL BE CONSIDERED NON-RESPONSIVE.
Solicitation number
36C24526Q0797
Issued by

About this file

This is a Price/Cost Schedule for a federal contract solicitation requiring the acquisition of X-ray irradiation equipment and related services. The procurement consists of two line items on a firm-fixed-price basis.

Item 0001 specifies a RAD SOURCE RS 2000-Q2 1 kW X-ray irradiator or equivalent product meeting defined specifications. Key technical requirements include non-isotopic operation requiring no nuclear licensing, a dose uniformity ratio of 95% or greater, self-shielded design, compliance with 21 CFR 1020.40 radiation safety standards, 1,000-4,000 W generator capacity, minimum 225 kV energy delivery output, 208-240 V single-phase electrical requirements (30-50 A, 50-60 Hz), equipment weight of approximately 2,750-3,300 pounds, and specified dimensional parameters. The equipment must include a data recording module and rodent cage in pie configuration. Item 0002 covers installation, operational training, and a 12-month extended service agreement including assembly, calibration, dose mapping, radiation survey, and one year or 3,200 hours of warranty coverage on parts, labor, and travel. Both items are classified under NAICS code 334516 (Analytical Laboratory Instrument Manufacturing) and product/service code 6640 (Laboratory Equipment and Supplies). The schedule allows for brand-name or equal products in accordance with the Statement of Work, with unit prices and amounts to be filled in by the contractor.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

B.4 PRICE/COST SCHEDULE

ITEM INFORMATION

ITEM

NUMBER

DESCRIPTION OF

SUPPLIES/SERVICES QUANTITY UNIT UNIT PRICE AMOUNT

1.00 EA __________________ __________________

RAD SOURCE RS 2000-Q2 1KW X- RAY IRRADIATOR WITH:

•Self-contained X-ray emitter with non-isotopic operation/no radioactive source, requiring no nuclear licensing •Dose uniformity ratio of 95% or greater •Self-shielded requiring no specialized shielding in the room it operates in •Meets all radiation safety requirements specified in Federal Regulation 21 CFR 1020.40 •1,000-4,000 W generator •Energy delivery output of at least 225 kV •Electrical Requirements: 208-240 V, Single Phase, 30-50 A, 50-60 Hz •Equipment Weight: Approx. 2,750-3,300 lbs •Equipment Dimensions: 43-56” W x 75-76” H x 31-36” D •Internal Chamber Dimensions: 17.5-25” W x 15.5-42” H x 16.75-27” D •Data Recording Module (User ID, Date, Time, and Total Dose) •Rodent cage- pie configuration preferred

"BRAND NAME OR EQUAL TO RAD SOURCE IN ACCORDANCE WITH

(IAW) SOW"

CONTRACT TYPE: FIRM-FIXED PRICE

PRINCIPAL NAICS CODE: 334516 - Analytical Laboratory Instrument Manufacturing PRODUCT/SERVICE CODE: 6640 - Laboratory Equipment and Supplies

MANUFACTURER PART NUMBER (MPN): RS 2000-Q2

LOCAL STOCK NUMBER: RS 2000-Q2

1.00 EA __________________ __________________

INSTALLATION, TRAINING AND WARRANTY & RS 2000-Q2 1KW

12-MONTH SERVICE EXTENDED AGREEMENT.

•Installation of the irradiator into room BB-120 once components are delivered. Installation includes assembly and calibration, dose mapping and radiation survey.

•Operational training for device usage after the installation is complete •1 year or 3,200 hours Warranty on Parts, Labor, and Travel

"BRAND NAME OR EQUAL TO RAD SOURCE IAW SOW"

CONTRACT TYPE: FIRM-FIXED PRICE

PRINCIPAL NAICS CODE: 334516 - Analytical Laboratory Instrument Manufacturing PRODUCT/SERVICE CODE: 6640 - Laboratory Equipment and Supplies

MANUFACTURER PART NUMBER (MPN): N/A

LOCAL STOCK NUMBER: N/A

GRAND TOTAL __________________