U.S. Department of Veterans Affairs VA Palo Alto Healthcare System 3801 Miranda Avenue Palo Alto, CA 94304
Li-Cor Odyssey® CLx Infrared Imaging System Data Integrity™ Bundle, Premium or Equal Salient Characteristics
| Project: | VAPAHCS Research VMU | |
| Transaction Number: | 640-19-3-522-0048 |
Features, Design and Construction Common to Infrared Imaging Systems:
· Laser lifetime shall be 40,000 hours, minimum
· 700-channel laser source: solid-state diode laser at 685 nm, maximum
· 800-channel laser source: solid-state diode laser at 784 nm, maximum.
· Dynamic range shall be at least 4 logs (Manual); 6 logs (Auto) minimum.
· Focusing range adjustable to be aligned to the top surface of the scan bed up to 4 mm minimum above the scan bed, to obtain the best signal-to-noise ratio of sample in membranes, gels, or microplates.
· Resolution shall be within range of 10-337 µm.
· Scanning speed shall be 5-50 cm/s.
· System shall include silicon avalanche photodiode detectors.
· Must be able to scan microscope slides (tissue imaging) , plates( in-cell and on-cell imaging) and Membrane
Warranty Common to Infrared Imaging Systems:
· The vendor shall provide a warranty of at least twelve months of parts and labor.
Certifications, Environmental Impact and Testing Common to Infrared Imaging Systems:
· Class 1 Laser Product Complies with IEC 60825-1:2014
Critical Dimensions Unique to Infrared Imaging Systems (max)
· Width: 50-80cm
· Depth: 60-70cm
· Height: 35-50 cm
· Height (with hood fully open): 70-75 cm
· Weight: no more than 65 kg