U.S. Department of Veterans Affairs VA Palo Alto Healthcare System 3801 Miranda Avenue Palo Alto, CA 94304

Li-Cor Pearl® Trilogy Small Animal Imaging System, Premium, Salient Characteristics

Project:VAPAHCS Research VMU
Transaction Number:640-19-3-522-0048

Features, Design and Construction Common to Small Animal Imaging Systems:

· The small animal imaging system dynamic range shall be 22 bit (>6 logs)

· The small animal imaging system detector type shall be CCD. Thermoelectrically cooled.

· The small animal imaging system scan time shall be 500 ms to 60 seconds, depending upon camera settings.

· The small animal imaging system shall include a 700 nm channel: excitation 685 nm, emission 720 nm.

· The small animal imaging system shall include an 800 nm channel: excitation 785 nm, emission 820 nm.

· The small animal imaging system shall include a bioluminescence channel: emission 450 nm – 640 nm.

· The small animal imaging system shall include a white channel: excitation white.

· The small animal imaging system shall include the following accessories: docking station and clean box.

· The small animal imaging system shall include dyes.

· The small animal imaging system shall include the following display options: pseudo-color, grayscale, single color (red, green, or blue), or two colors with overlapping fluorescence displayed in a third color.

· The small animal imaging system shall include imaging software.

· The small animal imaging system shall output imaging at 85, 170 or 255 micron.

Warranty Common to Small Animal Imaging Systems:

· The vendor shall provide a warranty of at least twelve months of parts and labor.

Certifications, Environmental Impact and Testing Common to Small Animal Imaging Systems:

· Class 1 Laser Product Complies with IEC 60825-1:2014

Critical Dimensions Unique to Small Animal Imaging Systems:

· Capacity: one animal, with linked Lookup tables for image normalization

· Imaging Bed:

· Width: 16.8 cm

· Depth: 12 cm

· Vertical clearance to top of drawer is 3.8 cm

· Field of View:

· Width: 11.2 cm

· o Depth: 8.4 cm

· Imaging Bed Temperature Range:

o 32-42 ̊C