RFQ_WRS250636_Laboratory_Deionized_Water_System_-_Exhibit_C_Appendix.pdf
PDF 105 KB Posted
- Attached to
- Laboratory Deionized Water System
- Solicitation number
- RFQ# WRS/250636
- Issued by
- Marion County, Orlando City, Florida
About this file
This document is Appendix 'C' from the City of Ocala's Laboratory Deionized Water System solicitation (RFQ# WRS/250636), specifically detailing standard methods and maximum acceptance limits for water quality testing. The appendix outlines precise contaminant thresholds using Standard Methods 23rd Edition Table 9020: II, which establishes rigorous quality control parameters for laboratory deionized water.
The table specifies maximum acceptable limits for various contaminants, including conductivity (< 2 µmhos/cm at 25°C), total heavy metals (< 0.10 mg/L), individual heavy metals like cadmium and lead (< 0.05 mg/L), total organic carbon (< 1.0 mg/L), total chlorine residual (< 0.10 mg/L), and heterotrophic plate count (< 500 CFU/mL). These stringent specifications ensure the high-purity water required for accurate laboratory testing and regulatory compliance at the City of Ocala's Water Quality Laboratory, with a statistical validation method (Student's t ≤ 2.78) to verify water quality standards.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Appendix ‘C’
Standard Methods 23rd Edition Table 9020: II
Contaminants Maximum Acceptance Limits Conductivity < 2 µmhos/cm at 25 ºC
Heavy metals, Total <0.10 mg/L Heavy metals, singles:
Cadmium Chromium Copper Lead Nickel Zinc
<0.05 mg/L
Organic Carbon, Total (TOC) <1.0 mg/L
Total chlorine residual <0.10 mg/L
T. Nitrogen Ammonia/organic nitrogen <0.10 mg/L
Heterotrophic Plate Count <500 CFU/mL
Use Test Student’s t ≤ 2.78