About this file

This document is Appendix 'C' from the City of Ocala's Laboratory Deionized Water System solicitation (RFQ# WRS/250636), specifically detailing standard methods and maximum acceptance limits for water quality testing. The appendix outlines precise contaminant thresholds using Standard Methods 23rd Edition Table 9020: II, which establishes rigorous quality control parameters for laboratory deionized water.

The table specifies maximum acceptable limits for various contaminants, including conductivity (< 2 µmhos/cm at 25°C), total heavy metals (< 0.10 mg/L), individual heavy metals like cadmium and lead (< 0.05 mg/L), total organic carbon (< 1.0 mg/L), total chlorine residual (< 0.10 mg/L), and heterotrophic plate count (< 500 CFU/mL). These stringent specifications ensure the high-purity water required for accurate laboratory testing and regulatory compliance at the City of Ocala's Water Quality Laboratory, with a statistical validation method (Student's t ≤ 2.78) to verify water quality standards.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Appendix ‘C’

Standard Methods 23rd Edition Table 9020: II

Contaminants Maximum Acceptance Limits Conductivity < 2 µmhos/cm at 25 ºC

Heavy metals, Total <0.10 mg/L Heavy metals, singles:

Cadmium Chromium Copper Lead Nickel Zinc

<0.05 mg/L

Organic Carbon, Total (TOC) <1.0 mg/L

Total chlorine residual <0.10 mg/L

T. Nitrogen Ammonia/organic nitrogen <0.10 mg/L

Heterotrophic Plate Count <500 CFU/mL

Use Test Student’s t ≤ 2.78