RFP A22-1089 Addendum 3 with Appendix B dated 2022-03-17.pdf

PDF 281 KB Posted

Attached to
Request for Proposal for a Pulsed Magnetizer with Addendums 1, 2 & 3
Solicitation number
RFP-A22-1089
Issued by

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

For the best experience, open this PDF portfolio in

Acrobat X or Adobe Reader X, or later.

Get Adobe Reader Now!

http://www.adobe.com/go/reader v2022-03-17

RFP A22-1089 Specifications for Pulsed Magnetizer

Ames Laboratory is requesting quotes on a pulsed magnetizing system that can be operated in a glove box supplied by customer. The system should be compatible with commercially available with appropriate feedthrough. The capacitor band and other electronics, with the exception of magnetizing fixture, should be housed outside of glove box.

• Required to magnetize materials in inert chambers, e.g. glovebox.

i. Describe approach for integrating into an inert atmosphere.

ii. Describe feedthrough requirement for the inert chamber. Preferred that feedthroughs are compatible with standard Roxtec or similar transit.

• Minimum magnetizing fixture active volume of 4 inches diameter and 6 inches length.

• System with two readily interchangeable coils: (1) 4” ID, 10” length, >6 T field and (2) 6” ID, 12” length, >4 T field. Please specify if lengths are different than noted.

• Fixture 1: Low Field Coil for Aligning Powders

The rubber die dimension is 4x4x4 inch cube, within the die, we have 3x3x3 inch cube volume of powders to be aligned.

To accommodate this rubber die, we will need a coil with 6 inch ID.

Within this 6 inch ID coil, the required volume with uniform 4T field (cylindrical) is diameter 4.5 inch and length 4 inch.

Fixture 2: High Field Coil for Charging Sintered Magnet The sintered sample dimension is 2x2x2 inch cube.

To accommodate this sample, we will need a coil with >3 inch ID.

Within this >3 inch ID coil, to charge the sintered 2x2x2 inch cube magnet, the required volume with uniform 7T field (cylindrical) is diameter 3 inch and length 3 inch.

• With the magnetizing fixture in an inert chamber, describe if the unit can be operated outside of the inert chamber.

• Minimum magnetic field in the active volume of 6 T. Higher field strengths are preferred. Please specify the uniformity of the field over the active volume.

• Specify the following for the system

i. Input power requirement: voltage, # of phases, current, etc.

ii. Number of capacitors and total capacitance.

iii. Maximum energy

iv. Pulse rate with and without water cooling.

v. Cooling requirement in flow rate and necessary temperature of coolant.

vi. Magnetizing part holder dimensions

• Describe the magnetizing part holder.

• Data acquisition, storage and transfer.

• Describe power drain capability.

• Describe interlocks and other safety mechanisms.

v2022-03-17

• Explain temperature monitoring mechanism and shut-off controls.

• Warranties

• Quote the following as a separate line item

i. On-site (Ames Laboratory) installation and training as a separate.

Ames Laboratory RFP A22-1089 ADDENDUM NUMBER 3

RFP A22-1089 ADDENDUM NUMBER 3

ADDENDUM DATE: 3/17/2022

PROJECT: RFP A22-1089 Pulsed Magnetizer PROPOSAL DUE DATE: 3/25/2022 at 4:00 PM

ADDENDUM ISSUED DATE: 3/17/2022

The information contained in this Addendum modifies, supplements or replaces information contained in the Bidding Documents dated 3/3/2022 and is hereby made a part of the Contract Documents.

Acknowledge receipt of this Addendum in the proposal submitted.

Addendum consists of 3 pages.

ADDENDUM INDEX

GENERAL ITEMS APPLICABLE TO THE SPECIFICATIONS Item 1

GENERAL ITEMS

1. Appendix B – Specifications for Pulsed Magnetizer has been updated and the 3rd bullet point has been struck and replaced with the following:

“Fixture 1: Low Field Coil for Aligning Powders The rubber die dimension is 4x4x4 inch cube, within the die, we have 3x3x3 inch cube volume of powders to be aligned.

To accommodate this rubber die, we will need a coil with 6 inch ID.

Within this 6 inch ID coil, the required volume with uniform 4T field (cylindrical) is diameter 4.5 inch and length 4 inch.

Fixture 2: High Field Coil for Charging Sintered Magnet The sintered sample dimension is 2x2x2 inch cube.

To accommodate this sample, we will need a coil with >3 inch ID.

Within this >3 inch ID coil, to charge the sintered 2x2x2 inch cube magnet, the required volume with uniform 7T field (cylindrical) is diameter 3 inch and length 3 inch.”

File details come from the government source that posted it. Updated .