The SOW outlines the requirements for the subcontractor to provide dosimeters, dosimetry reports, and associated equipment and materials to support Fermilab's External Dosimetry Program. Key details include the subcontractor maintaining DOELAP accreditation for whole-body and extremity dosimeters, providing online systems for managing personnel dosimetry data, supplying control dosimeters, and processing and reporting dose data within specified timeframes. The subcontract is an indefinite-quantity type with an initial 1-year term and four 1-year option periods. The related federal contract opportunity is a Request for Proposal (RFP) for these Professional External Radiation Dosimetry Services, with a response due date of June 2024.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
for
External Radiation Dosimetry Services at
Fermi National Accelerator Laboratory
25 April 2024
I. Overview
This Statement of Work (SOW) defines the requirements for external radiation dosimetry services provided by the subcontractor to Fermi National Accelerator Laboratory (Fermilab).
Fermilab is managed by Fermi Research Alliance LLC (FRA) for the U.S. Department of Energy (DOE) Office of Science.
II. Contacts
Administration of these services will be handled by Fermilab's Dosimetry Program Office, primarily by the Dosimetry Program Manager (DPM), Michael Vincent. All technical correspondence and inquiries and shall be directed to him at vincent@fnal.gov or 630-840-3692. Billing and contractual matters shall be directed to Fermilab's Procurement Department, which can be reached at 630-840-3521. Electronic invoices shall be sent to apinvoices@fnal.gov.
Should any of the preceding contact information change, the subcontractor shall be informed of said changes in writing.
III. Schedule
This agreement shall commence on 01 July 2024 and continue through 30 June 2025, with additional four option years available, which extend through 30 June 2029.
IV. Applicable Documents
A. Standards
1) Department of Energy Laboratory Accreditation Program for Personnel Dosimetry, DOE-STD-1095-2018, U.S. Department of Energy, 30 October 2018
2) Department of Energy Laboratory Accreditation Program Administration, DOE-STD-1111-2018, U.S. Department of Energy, 07 September 2018
3) Personnel Dosimetry Performance--Criteria for Testing, ANSI/HPS N13.11-2022, American National Standards Institute, Inc., 01 December 2022
4) Performance Testing of Extremity Dosimeters, ANSI/HPS N13.32-2018, American National Standards Institute, Inc., 27 July 2018
5) National Voluntary Laboratory Accreditation Program Ionizing Radiation Dosimetry, NIST Handbook 150-4, D. Ho, National Institute of Standards and Technology, November 2019 mailto:vincent@fnal.gov mailto:apinvoices@fnal.gov
B. Checklists
1) DOE-STD-1095 DOELAP Checklist for External Dosimetry Assessments, Revision
2022-01
V. Requirements and Deliverables
The subcontractor shall provide the following products and services.
A. General
1) The subcontractor shall supply dosimeters, dosimetry reports, and any necessary equipment and materials required to support Fermilab's External Dosimetry Program. The subcontractor shall provide a price listing for such equipment and materials, e.g. report binders, spare dosimetry badge clips, etc.
2) Fermilab's dosimetry personnel shall have the ability to make changes (e.g. additions and deactivations) at any time to the list of individuals being supplied with dosimeters. These types of changes shall not require communication through the Fermilab Procurement Department.
3) An online system shall be provided by the subcontractor to allow Fermilab's DPM or designee to make personnel additions, deactivations, and transfers directly to the subcontractor database. This system shall have the ability to export data to confirm such changes. It shall also have the ability to export dose data in Excel, CSV, or similar format. It shall also have the ability to provide missing dosimeter lists. The subcontractor shall provide support for this online system as needed.
4) Upon Fermilab dosimetry program personnel request, the subcontractor shall provide relevant documents, including but not limited to the following:
a. Quality assurance (QA) and/or quality control (QC) documentation and data
b. Blind audit data
c. Dosimetry technical basis documents
d. Policies and procedures
e. External dosimetry data validation and verification documents
5) The subcontractor shall have a documented privacy policy in place forbidding the disclosure of any personal data, including dose data, of individuals monitored for radiation at Fermilab to anyone other than members of the Fermilab Dosimetry Program Office. The subcontractor's employees shall be specifically trained in this policy.
6) The subcontractor shall attend meetings with Fermilab dosimetry personnel on an as-needed basis to discuss issues related to services being provided by the subcontractor. These meetings may be virtual or in-person.
7) The subcontractor shall maintain Fermilab's historical dosimetry records (personnel exposure histories) indefinitely and will provide copies of individual records as requested. Fermilab dosimetry personnel shall have the right to request the transfer of these personnel exposure history records to Fermilab or to a third party; in the unlikely event that such a transfer is requested, the subcontractor shall work with the Dosimetry Program Office to securely and completely transfer these records.
8) Dosimeter unreturned fees will not be refunded if a dosimeter is returned 90 days after the end of the dosimeter wear date
9) This is an indefinite-quantity type of subcontract and is effective for the period stated in Section III above. The quantities of services specified are estimates only and are not necessarily purchased by the subcontract. Accordingly, this does not require that a certain number of dosimeters and/or associated items will be ordered.
B. Dosimeter Types
1) The subcontractor shall provide whole-body dosimeters that conform to
Department of Energy Laboratory Accreditation Program (DOELAP) standards (see Section IV) with the following capabilities contained within a single holder (see Item VI. B 1 for specific DOELAP category information):
a. X-ray and gamma radiation detection
b. Beta radiation detection
c. Neutron radiation detection
2) The subcontractor shall provide extremity dosimeters that conform to DOELAP standards (see Section IV) with the following capabilities contained within a single holder (see Item VI. B 1 for specific DOELAP category information):
3) The subcontractor shall additionally provide area monitoring dosimeters that adhere to the National Voluntary Laboratory Accreditation Program (NVLAP) standard (see Section IV) with the following capabilities contained within a single holder:
c. Neutron radiation detection
4) All dosimeters shall be able to be used for a wear period of at least three (3) months and be able to be accurately processed and reported up to at least one (1) year after the end of that wear period.
C. Accounts
1) The subcontractor shall provide the following:
a. A minimum of six (6) separate primary dosimeter accounts for use at
Fermilab's Batavia site, each with their own subaccounts, with the ability to add more accounts if requested. These accounts will use the DOE-assigned Computerized Accident/Incident Reporting System (CAIRS) facility code for Fermilab in Batavia, Illinois, which shall be provided to the subcontractor by Fermilab dosimetry personnel. Subaccounts are to be designated by codes of up to three (3) letters and/or numbers, to be assigned by Fermilab dosimetry personnel.
i. One (1) of these primary accounts shall be reserved for Fermilab employee dosimeters.
ii. One (1) of these primary accounts shall be reserved for Fermilab visitor dosimeters.
iii. One (1) of these primary accounts shall be reserved for Fermilab contractor dosimeters.
iv. One (1) of these primary accounts shall be reserved for temporary (spare) dosimeters to be used at Fermilab in Batavia. See Section V. H below for more details about these spare dosimeters.
v. Two (2) of these primary accounts shall be reserved for dosimeters not worn by personnel and used for activities such as area monitoring or quality control. One (1) of these accounts shall be used to provide the NVLAP dosimeters described in Item V. B 3 above.
b. A minimum of two (2) separate primary dosimeter accounts for use at Fermilab's Long-Baseline Neutrino Facility Far Site (LBNF-FS) in Lead, South Dakota, each with their own subaccounts, with the ability to add more accounts if requested. These accounts will use the DOE-assigned CAIRS facility code for LBNF-FS, which shall be provided to the subcontractor by Fermilab dosimetry personnel. Subaccounts are to be designated by codes of up to three (3) letters and/or numbers, to be assigned by Fermilab dosimetry personnel.
i. One (1) of these primary accounts shall be reserved for Fermilab
LBNF-FS employee dosimeters.
ii. One (1) of these primary accounts shall be reserved for temporary (spare) dosimeters to be used at LBNF-FS. See Section V. H below for more details about these spare dosimeters.
2) All accounts shall default to using account-level control dosimeters for control dose subtraction. However, the option shall be provided that upon written request from Fermilab dosimetry personnel, certain subaccounts will instead utilize their own subaccount-level controls. The number of subaccount-level control dosimeters shall be specified by Fermilab dosimetry personnel in writing. See Item V. E 12 for more details on control dosimeters.
3) Modifications to dose histories in any account due to previous exposures, exposure investigations, or other reasons shall be completed upon written request by Fermilab dosimetry personnel to the subcontractor.
D. Dosimeter Numbering and Labels
1) Participants in the accounts reserved for Fermilab employees, visitors, and contractors shall be permanently assigned numbers corresponding to their five (5) digit Fermilab ID number.
2) Dosimeters in the accounts reserved for temporary dosimeters shall be numbered consecutively with five (5) digit numbers, with no number being reused; the initial sequence of these numbers shall be determined by Fermilab's Dosimetry Program Manager and provided to the subcontractor. These numbers shall be unique across all temporary dosimeter accounts, i.e., no numbers from the Fermilab in Batavia temporary dosimeters shall ever appear on LBNF-FS temporary dosimeters at any point in time.
3) Dosimeters in the two (2) accounts reserved for area monitoring, QC, etc. shall be assigned five (5) digit numbers as determined by Fermilab's Dosimetry Program Manager.
4) The six (6) accounts containing dosimeters worn by personnel shall have the ability to assign an additional six (6) digit identification number to each participant, corresponding to that individual's Fermilab-assigned CNAS number; this number shall be provided by Fermilab dosimetry personnel where appropriate and shall serve as that individual's unique Dose Reporting ID.
5) The five (5) digit numbers referred to in Items 1, 2, and 3 above, along with the participant names, account numbers, subaccount codes, and wear period start dates, shall be clearly printed on the labels of all whole-body dosimetry badges and, where possible due to the limited space available, extremity dosimeters.
6) Every DOELAP-accredited whole-body dosimeter, personnel or otherwise, shall also have a unique, non-repeating eight (8) digit badge ID number assigned to it; this number shall be clearly printed on the labels of all whole-body dosimeters, and a two-dimensional (2D) barcode version of this number shall also be printed on the labels of all dosimeters to enable scanning devices to read this number. The initial sequences of these barcode numbers shall be determined by Fermilab's Dosimetry Program Manager and provided to the subcontractor.
7) Every DOELAP-accredited extremity dosimeter shall also have a unique, non-repeating badge ID number assigned to it; this number shall be clearly printed on the labels of all extremity dosimeters, and a two-dimensional (2D) barcode version of this number shall also be printed on the labels of these dosimeters to enable scanning devices to read this number. The initial sequences of these barcode numbers shall be determined by Fermilab's Dosimetry Program Manager and provided to the subcontractor.
8) Dosimeter labels or holders shall alternate colors to readily differentiate their quarterly wear periods.
E. Dosimeter Shipments
1) Prior to the start of each quarter, Fermilab dosimetry personnel shall supply the subcontractor with an updated list of participants containing their account numbers, ID numbers, subaccount codes, and badge rack locations or mail stop recipient names and numbers. Rack monitoring dosimetry badges shall be included in this list.
2) Fermilab dosimetry personnel shall also supply the subcontractor with physical badge racks labeled with each participant's name and ID number.
3) All personnel and rack monitor dosimeters that are assigned to badge racks by Fermilab dosimetry personnel shall be loaded onto those badge racks in the proper order by the subcontractor. Dosimeters shall be placed directly onto the labeled badge rack prior to shipment to Fermilab.
4) All personnel and rack monitor dosimeters not assigned to badge racks shall be appropriately packaged according to the mail stop recipient names and numbers in the list supplied by Fermilab dosimetry personnel. The names and mail stop numbers associated with those dosimeters shall be written on the outside of their packaging by the subcontractor.
5) Should any dosimeters have no position on a rack nor any entry in the participant list with mail stop name and number, the subcontractor shall ship those dosimeters to Fermilab in bulk.
6) Extremity dosimeters shall be shipped in bulk.
7) Dosimeters from the account reserved for temporary dosimetry badges shall be packaged by the subcontractor and shipped in bulk. These shall be numerically sorted and attached to a suitable carrier to prevent mixing.
8) Dosimeters from the accounts reserved for area monitoring, QC, etc. shall be packaged appropriately by the subcontractor and shipped in bulk.
9) Electronic copies of the packing lists for each account shall be provided for each shipment of dosimeters. Where applicable, the packing list entries shall include account numbers, subaccount codes, dosimeter wear dates, badge ID numbers, and participant names, numbers, and Dose Reporting ID numbers.
10) Additional dosimeters shall be supplied by the subcontractor upon request at any point during a quarter. These shall be shipped in bulk, and electronic copies of the packing lists shall be provided.
11) Dosimeters shall be shipped to ensure arrival at Fermilab at least ten (10) working days prior to the start of the quarter. At the beginning of each year, Fermilab dosimetry personnel will contact the subcontractor to specify anticipated dates for badge racks to be shipped to the subcontractor and returned to Fermilab. Specific shipping dates shall be coordinated between the subcontractor and Fermilab dosimetry personnel around the time the shipment is ready each quarter.
12) Control dosimeters shall be supplied by the contractor with each shipment of dosimeters. The number of control dosimeters needed for each quarterly shipment shall be determined by Fermilab's Dosimetry Program Manager and may be modified at any time by a written request to the subcontractor. For subsequent smaller shipments during a quarter, the subcontractor shall determine the appropriate number of control dosimeters to include.
13) The quarterly shipments of the racks and bulk dosimeters shall be shipped by the subcontractor and delivered to a location at Fermilab in Batavia as specified by the Dosimetry Program Manager. The subcontractor shall ensure that all dosimetry trunks and boxes are to be unloaded from the delivery vehicle by the driver and not by Fermilab personnel.
F. Processing
1) A positive system for identifying and tracking all dosimeters through the processing cycle shall be established by the subcontractor.
2) All dosimeters that are shown to have an exposure of greater than 1 rem or any neutron exposure with no accompanying gamma exposure shall be re-read, and their results shall be examined by a health physicist to have the initial reading verified.
3) All whole-body dosimeters shall have their neutron components processed and their neutron doses reported. If any neutron components are unable to be reported, a list containing, at minimum, the participant names, participant numbers, and wear periods of those dosimeters shall be provided in writing to the Fermilab DPM by the subcontractor.
4) All participant dose results shall be calculated using an algorithm compliant with the DOELAP standards referenced in Section II and consistent with the algorithm used during DOELAP performance testing (see Section VI. C).
5) Any changes to the established algorithms used to calculate doses shall be communicated to Fermilab's DPM in writing.
6) Fermilab dosimetry personnel shall have the option of selecting customer-specific calibration sources and associated correction factors. These selections shall be possible to make at the account level or for any specific subaccount, participant, or individual dosimeter. Fermilab's current selections for calibration sources and correction factors are as follows:
Dosimeter Calibration Source Correction Factor
Whole-body photon standard default no customer-specific correction factors
Whole-body beta Sr-90 no customer-specific correction factors
Whole-body neutron AmBe ICRP 60 fast neutron CF = 1.20, thermal neutron CF = 1.06
Extremity Sr-90 field-specific correction factor
"K" = 1.00
Fermilab dosimetry personnel shall communicate any future changes to these calibration sources to the subcontractor in writing.
7) The subcontractor shall provide an emergency processing option, with reports to be issued no more than 24 hours after receipt of dosimeters at the subcontractor. The subcontractor shall provide written instructions explaining how to request emergency processing along with pricing for such services.
8) The subcontractor shall process blind audit dosimeters, which will have been irradiated in known radiation fields to known doses by Fermilab dosimetry personnel or a third party. Fermilab dosimetry personnel shall make the results of these blind audits available to the subcontractor. As necessary, performance issues related to the results of these blind audit tests will be discussed during meetings between Fermilab dosimetry personnel and the subcontractor.
G. Reporting
1) The subcontractor shall immediately notify Fermilab's Dosimetry Program Manager in writing when any dosimeter indicates that a dose equivalent equal to or greater than any of the following were received:
Period Dose (mrem) Dose Type
Quarterly 250 Whole-Body (DDE)
Quarterly 3,750 Lens of Eye (LDE)
Quarterly 12,500 Shallow (SDE)
Quarterly 12,500 Extremity
Quarterly 20 Fetal
Annual (Year to Date) 1,000 Whole-Body (DDE)
Annual (Year to Date) 7,500 Lens of Eye (LDE)
Annual (Year to Date) 25,000 Shallow (SDE)
Annual (Year to Date) 25,000 Extremity
Annual (Year to Date) 250 Fetal
2) Quarterly Reports
a. Reports shall be made available to Fermilab dosimetry personnel electronically through a secured website or equivalent.
b. Using an email message or equivalent, the subcontractor shall notify Fermilab dosimetry personnel whenever a new or updated report is made available.
c. Two (2) copies of each electronic quarterly dose report shall be provided for every account, one (1) that includes sensitive identifying data such as birthdates, gender, etc., and one (1) that omits such data.
d. Dose reports for all accounts shall be sorted first by wear period and then by subaccount code. Different wear periods shall not appear on the same page, nor shall different subaccounts. Within subaccounts, individuals shall be sorted alphabetically by last name.
e. Dose reports shall contain equivalent dose data for skin, lens of the eye, and whole-body (gamma, beta, neutron, and total) doses for whole-body dosimeters and skin doses for extremity dosimeters.
f. Dose report formats shall conform to DOELAP standards. In particular, each report must include the following:
• Processor name and address
• Name of accredited program
• Pertinent dates for the wear period and the identification of dosimeters
• Personnel monitoring results
• Processor and accredited program identification codes, as appropriate
• Explanations of any deviations from routine processing procedures if the deviation could affect the reported dose
• Signature of or reference to the technical lead (however named)
• Software version(s) of the dose algorithm(s) used
g. Where initials or other abbreviations are used to identify the technical lead (however named) on reports, the subcontractor shall provide Fermilab dosimetry personnel with a means of identifying the full names of the individuals from those initials or abbreviations. This does not have to appear on the reports themselves but can take the form of a reference sheet or equivalent; updated versions shall be provided as necessary.
h. Reports shall be made available to Fermilab dosimetry personnel no more than twenty-five (25) working days after a shipment of used dosimeters is received by the subcontractor for processing.
i. Dose results shall by default be reported in 10 mrem increments unless specifically requested otherwise by Fermilab dosimetry personnel.
j. The symbol SL shall be used to designate a dose equivalent below Fermilab's selected reporting limit of 10 mrem.
k. Control dose values used for control subtractions shall be indicated on all reports. When control dosimeters are omitted from a return shipment of used dosimeters from Fermilab or are otherwise unavailable to the subcontractor, an alternative method of control subtraction shall be used;
this method shall be clearly indicated on the affected report(s), as shall the control dose used.
l. The option shall also be provided for Fermilab dosimetry personnel to request in writing that certain accounts and/or subaccounts not subtract control dose values automatically but instead always report control-unsubtracted values; in this case, the subcontractor shall always automatically provide Fermilab dosimetry personnel with any relevant control dosimeter readings even though they were not used for subtractions.
m. Whenever reports are modified or corrected by the subcontractor, updated versions shall be issued that clearly indicate what changes or modifications were made.
n. Dosimeters returned to the subcontractor late, i.e., beyond ninety (90) from the end of the wear period, shall have their reported doses credited to the proper wear period and participant upon processing.
o. Dosimeters sent back to the subcontractor labeled as Unused shall not have their dose results appear on reports. The data field on the report that normally contains the participant's dose shall indicate that the dosimeter was Unused for that wear period. Should Fermilab dosimetry personnel determine that a dosimeter was incorrectly designated as Unused, the subcontractor shall adjust the participant's dose records accordingly and issue a corrected report.
3) Annual Reports
a. Each year, the subcontractor shall provide electronic summaries of annual and lifetime dose totals for all individuals issued a dosimeter in the preceding year. These data are to be used by Fermilab dosimetry personnel to generate annual dose reports to individuals and to the Department of Energy. This data shall be made available securely through the use of password protection or equivalent means.
b. Annual reports shall be provided for each specific facility separately as designated by that facility's CAIRS code. Personnel monitored for radiation at both Fermilab in Batavia and LBNF-FS shall appear separately in each annual report with their dose received at each facility specified.
c. Lifetime reports shall be provided that combine dose for each individual from all facilities and accounts into a single record for that individual, e.g. if someone was monitored in a given year at both Fermilab in Batavia and LBNF-FS, the total dose from both facilities shall be combined into a single entry for that individual in the lifetime report.
H. Spare Dosimeters
1) As noted in Item V. C 1 above, two (2) accounts shall be dedicated to spare dosimeters. Each dosimeter in this account will either be issued to a participant by Fermilab personnel or returned to the subcontractor as Unused; see Item V. G 2(o) for information on Unused dosimeters.
2) For spare dosimeters that were issued, Fermilab dosimetry personnel shall enter the data for those participants in the subcontractor's online database where possible, e.g. for participants who have never been issued a dosimeter at Fermilab, or who have only ever received spare dosimeters in the past. When a participant is issued subsequent spare dosimetry badges, their doses shall be combined into that participant's dose history record in the spare dosimeter account. As with all accounts for dosimeters issued to personnel, the subcontractor shall keep quarterly dose records and lifetime exposure histories for the participants in this spare dosimeter account.
3) In situations where a participant who was issued a spare dosimeter already exists in one of the other "permanent" accounts, Fermilab dosimetry personnel must send the spare dosimeter and participant information to the subcontractor in writing. The subcontractor shall then merge the dose data from the spare dosimeter with the pre-existing dose data from that participant's entry in their home account. This data shall be reported on the dose reports for the participant's home account.
I. Backup Services and Equipment
1) The subcontractor shall have a documented plan in place to maintain adequate backup equipment and systems for use in the event their primary systems fail.
2) Backup equipment and systems must be routinely tested by the vendor to demonstrate its capabilities.
3) Should a separate processing facility be designated for use as a backup, that facility must maintain DOELAP accreditation.
VI. Quality Assurance
A. Quality Assurance Program
1) The subcontractor shall have a documented QA program describing its internal management structure and system of procedures and practices used to ensure dosimetry results are accurate, repeatable, verifiable, and properly recorded.
2) The subcontractor shall have a QA manual or supporting documentation that includes the following:
a. A statement of quality policy and quality objectives
b. Documented processes, procedures, and instructions
c. Documents needed to ensure effective planning, operation, and control of processes
d. Records required to demonstrate compliance with the quality assurance program
e. Dosimetry specifications and technical basis documentation
f. Acceptance criteria for dosimeter materials and holders
g. Training objectives and processes for maintaining proficiency
h. Practices for handling and resolving contested dosimetry data and test reports
3) Software verification and validation (V&V) shall be performed in accordance with an appropriate, documented software quality assurance plan. V&V shall include process control software, dose algorithms, data processing, and record keeping. In addition, software version control shall be included in the documented control procedures.
4) All dosimeters shall be processed in accordance with the subcontractor's QA procedures. Any deviations from these procedures that could potentially affect Fermilab's dose results shall be communicated in writing, along with the corrective actions that were taken.
B. Accreditation
1) The subcontractor shall be accredited by the Department of Energy Laboratory
Accreditation Program for their personnel dosimeters in the DOELAP categories specified below:
Dosimeter Type DOELAP Category
Whole-body IA Accident, photons: General
Whole-body IIA Photons/photon mixtures: General
Whole-body IIIA Betas: General
Whole-body IIID Betas: Uranium slab
Whole-body IVA-1 Photon/beta mixtures: IIA + IIIA
Whole-body IVD-1 Photon/beta mixtures: IIA + IIID
Whole-body VA-1 Neutron/photon mixtures: IIA + General Neutrons
Dosimeter Type DOELAP Category
Extremity IA High-Dose, photons: General
Extremity IIA Photons: General
Extremity IIIA Betas: General
Extremity IIID Betas: Uranium slab
Fermilab dosimetry personnel shall have the right to make changes to these DOELAP categories at any time; such changes shall be communicated to the subcontractor in writing.
2) The subcontractor shall maintain their DOELAP accreditation throughout the course of the contract and shall adhere to all pertinent aspects of the DOELAP standards cited above.
3) The subcontractor shall notify Fermilab's DPM in writing of any potential deviations from the DOELAP standards.
4) Upon request, the subcontractor shall provide access to Fermilab dosimetry personnel to any documents required to demonstrate conformance to DOELAP standards.
5) In addition to the above, the subcontractor shall provide Fermilab dosimetry personnel with the following documentation regarding the whole-body and extremity dosimeters. Should any of these documents be revised, the subcontractor shall communicate this information to Fermilab in writing.
a. Lower limit of detectability (LLD)
b. Useful dose range
c. Background contribution to dose equivalent
d. Processing system measurement uncertainty
e. Repeatability/precision
f. Residual signal/fade
g. Angular dependence
h. Batch homogeneity
i. Energy response
j. Dosimeter algorithms, along with data used to create and test them, uncertainty, process controls, and attributes and limitations
k. Design specifications, including dosimeter holders, filter materials used, areal density (mg/cm2) of the materials, and positions of dosimeter material
6) The subcontractor shall have a procedure describing how they will ensure that fulfillment of all DOELAP requirements is maintained.
7) Per DOELAP requirements, Fermilab dosimetry personnel shall be entitled to perform at least one (1) onsite audit of the subcontractor every three (3) years. All aspects of the DOELAP onsite assessment checklist may be covered during the course of this onsite assessment.
8) For area monitoring dosimeters, the subcontractor shall maintain accreditation by the National Bureau of Standards through participation in the National Voluntary
Laboratory Accreditation Program.
C. Performance Testing
1) The subcontractor shall participate in the DOELAP performance testing program.
This testing shall conform to the criteria established in ANSI/HPS N13.11 and ANSI/HPS N13.32 and the DOELAP standards.
2) Results of all performance tests shall be made available upon request by Fermilab dosimetry personnel to the subcontractor.