This is a Performance Work Statement (PWS) for calibration services of Test, Measurement, and Diagnostic Equipment (TMDE) for 3D Medical Logistics Company (3D MEDLOG) Biomedical Equipment Department.
The contractor shall perform off-site calibration services on 189 serialized pieces of TMDE across two locations: 122 units at 3D MEDLOG in Camp Kinser, Japan, and 67 units at Combat Logistics Company 33 in Kaneohe Bay, Hawaii. Services must be executed in accordance with Original Equipment Manufacturer (OEM) specifications and National Institute of Standards and Technology (NIST) standards. The contract requires two calibration cycles: an initial service within 60 days of contract award, with equipment returned within 60 days of receipt, followed by a second calibration service six months after completion of the initial service. The Period of Performance spans one base year (30 September 2026 to 29 September 2027) plus four optional 12-month extension years. Calibration requirements are divided into two tiers: Tier 1 (NIST-traceable calibration at contractor or accredited subcontractor facilities) and Tier 2 (OEM-authorized calibration requiring direct manufacturer or certified service center submission). The contractor bears all costs for shipping, handling, subcontracting, and incidental expenses under a firm-fixed-price contract line item. Deliverables include receipt confirmation, detailed calibration status reports within seven business days of completion, Field Service Reports or Calibration Completion Certificates within three business days of service completion, and weekly status updates. The contractor must provide a 90-day warranty on all TMDE from return date and a 12-month warranty on calibration services. Equipment includes biomedical test simulators (Fluke IMPULSE, PROSIM, IDA-5 series), electrical safety analyzers (ESA 609/620), vital signs simulators (VT650), network analyzers (Fluke Networks X2 Kit), temperature sensors (Newport 1936-R), gas analyzers (TSI 4081 series), and various other diagnostic equipment.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Source Selection Information—See FAR 2.101 and 3.104
Performance Work Statement (PWS) for the
Calibration of Test, Measurement, and Diagnostic Equipment (TMDE)
1.0 General. The 3D Medical Logistics Company (3D MEDLOG) Biomedical Equipment Department (BIOMED) requires periodic calibration of TMDE to ensure accurate inspection, preventive maintenance, calibration, and repair of 3D MEDLOG Medical and Dental Equipment.
1.1. Description of Services/Introduction. Periodic TMDE calibration is crucial for the effective maintenance and upkeep of Medical and Dental Equipment (MDE) in compliance with Original Equipment Manufacturer (OEM) specifications and other relevant references defined in this PWS. TMDEs that are not consistently calibrated and used during inspection, preventive maintenance, repair, or calibration activities invalidate procedures and pose safety risks for patient use. The consequences of inaccurate metric data can be catastrophic and potentially fatal for both patients and users.
1.1.1. The contractor shall provide all personnel, equipment, supplies, facilities, transportation, tools, materials, management, labor, and other items and non-personal services necessary to perform scheduled periodic calibration of TMDE in accordance with manufacturer specification and other applicable references as defined in this PWS.
1.2. Scope of Work and Objective. The contractor shall execute calibration services as defined in this PWS. The contractor shall perform (2) off-site calibration services, the initial service will begin within 60 days of the contract award. The final calibration service will be (6) months after the completion of the initial calibration service, this is to ensure that all TMDE is not down at the same time.
1.3. Period of Performance. The period of performance shall be for one (1) Base Year of 12 months and four (4)
12-month option years. The Period of Performance reads as follows:
Base Year: 30 September 2026 to 29 September 2027.
Option Year I: 30 September 2027 to 29 September 2028.
Option Year II: 30 September 2028 to 29 September 2029.
Option Year III: 30 September 2029 to 29 September 2030.
1.3.1. Federal Holidays. The contractor shall not be required to provide services on federal holidays unless the holiday is particularly specified otherwise in this PWS.
Federal Holiday Occurrence
New Year’s Day 1 January*
Martin Luther King Jr.’s Birthday 3rd Monday of January
President’s Day 3rd Monday of February
Memorial Day last Monday of May
Juneteenth 19 June*
Independence Day 4 July*
Labor Day 1st Monday of September
Columbus Day 2nd Monday of October
Veterans Day 11 November*
Thanksgiving Day 4th Thursday of November
Christmas Day 25 December*
* When a federal holiday falls on Sunday, the following Monday will be observed as a holiday by the U.S.
government. When a federal holiday falls on a Saturday, the preceding Friday will be observed as a holiday.
In addition to these federal holidays, if the President of the United States issues an executive order closing the federal government, that date shall also be considered a recognized federal holiday (Reference
Executive Order 11582 of February 11, 1971, and 5 U.S.C. 5546 and 6103(b) and other similar statues insofar as they relate to the pay and leave of employees of the United States).
2.0 Place of Performance. Calibration services will be performed at the contractor facility; the contractor will be responsible for all associated shipping cost of equipment through the appropriate means and providing the government with shipping documents such as prepaid shipping labels, sufficient insurance, invoice, delivery list, and any other incidental expenses associated with off-site services.
2.1. Return Shipment. The contractor is responsible for return serialized equipment to the correct location outlined within attachment 1 and 2. Return shipping for sterilized equipment in attachment 1 will be shipped to 3D
MEDLOG BLDG 502 Camp Kinser MF M97115 MEDLOG CO 3D SUP BN, CAMP KINSER, Japan 901-2100 upon completion of calibration. Return shipping for sterilized equipment in attachment 2 will be shipped to CLC33, 271 5TH STREET, KANEOHE BAY, HAWAII 96863 upon completion of calibration.
3.0 Technical Requirements (Description of Work / Tasks) The contractor shall perform calibration services on all TMDE defined in this PWS.
3.1. Calibration of Equipment. The contractor shall perform all off-site calibration services in accordance with the standards set forth by the OEM, NIST, and any other applicable regulatory references. The contractor shall bear full responsibility for all costs associated with the off-site calibration services, including subcontract costs, outbound and inbound shipping, handling, and any incidental expenses covered under a firm-fixed price of the contract line item.
The contractor is required to ensure that all equipment is calibrated in a timely manner and returned to its appropriate location within (60) days of receipt at contractors’ location, excluding any delays due to uncontrollable factors. The contractor shall provide a notice to the Government Representative (GR) notifying their receipt of all
TMDE.
3.1.1 Standards of Calibration. The Government categorizes TMDE calibration requirements into two tiers, defined as follows:
• Tier 1: Standard NIST-Traceable Calibration. Calibration utilizing standards traceable to the National
Institute of Standards and Technology (NIST) or other recognized national/international standards. This may be performed at the Contractor's or subcontractor’s accredited laboratory.
• Tier 2: OEM-Authorized Calibration and Proprietary Software Integration. Calibration that requires direct submission to the Original Equipment Manufacturer (OEM) or an OEM-certified service center. This tier includes proprietary software updates, firmware integration, and specialized sensor alignments that can only be executed utilizing OEM-controlled software and diagnostic equipment.
3.1.2 Equipment List Designations. The Contractor shall structure their firm-fixed-price quote to account for these specific designations of all items located in Attachment 1 and Attachment 2. The Contractor shall bear all subcontracting and direct factory shipping fees associated with Tier 2 OEM-Authorized calibration requirements.
3.2. Serialized Equipment. The contractor shall provide confirmation of receipt of all TMDE at their facility, the contractor will provide the GR with a report consisting of all TMDE and submitting a detailed calibration status report for 100% of the equipment. The review must accurately identify any discrepancies or calibration needs and provide actionable recommendations for corrective actions in line with regulatory standards. This report must demonstrate thoroughness and precision to ensure accurate equipment records. The contractor must submit an end-of-service report to the GR within (7) business days of completion, providing clear calibration results, any discrepancies, and a summary of actions taken. The report should be comprehensive, accurate, and contain actionable insights for future calibration cycles. This report shall also contain the calibration results, and the condition of the equipment upon return, ensuring compliance with the service-level expectations.
4.0 Timeline. The contractor shall perform two calibration services as specified in this PWS, with the initial calibration completed within (60) days of contract award, followed by the last calibration service (6) months after the completion of the initial service. All TMDE sent for calibration services must be returned within (60) days from the date of receipt. If any deadline extension is needed, the contractor must submit a request no later than (7) seven days before the established deadline to the GR. Weekly status updates on the progress of all TMDE, including tracking information, calibration progress, and any delays or issues encountered, must be provided by the contractor.
Additionally, the contractor is responsible for ensuring all TMDE is returned to the designated facility without delay or issues.
5.0 Calibration Reports. The contractor shall provide a Field Service Report (FSR) or Calibration Completion
Certificate for each TMDE to the GR within (3) three business days of service completion. The reports must comply with all applicable standards as outlined in this PWS. Each TMDE must be furnished with a calibration certificate and a calibration sticker or label, including traceable calibration numbers, dates, expiration dates, technician details, and company contact information. Physical copies of all reports and certificates must be delivered to the GR, and electronic copies (PDF format) must be provided upon request. The contractor must ensure that all calibration documentation is complete, accurate, and submitted promptly to meet the specified timelines
6.0 Repair. The contractor is prohibited from conducting repair services on any unserviceable TMDE unless explicit authorization is granted by the contracting officer through a contract modification. If repairs are necessary, the contractor shall notify the contracting officer and GR via email, including a detailed repair quotation with labor and parts costs, as well as the proposed repair timeline. This information must be submitted for review and authorization prior to commencing any repair work. After the GR reviews the repair quotation, the contracting officer will notify the contractor to either return the item(s) without repair or proceed with repairs and calibration.
Any repairs or services identified during contract performance that are outside the original scope and not anticipated in the firm-fixed price agreement shall be considered "over-and-above work." Over-and-above work must be documented and approved through a contract modification, in accordance with DFARS Subpart 217.77, prior to any additional work being performed. Such modifications will be issued by the contracting officer to account for any additional costs or tasks that arise beyond the initial scope defined in the contract. Repair services may extend beyond the timelines specified in this PWS, with the contractor required to provide regular updates on any such delays or additional work.
7.0 Warranty. The contractor shall provide a warranty service of at least (90) days for all TMDE from the return ship date or calibration date, covering any defects identified during the warranty period. Calibration services performed by the contractor will be warranted for a period of (12) months from the date of completion. The contractor is responsible for providing documentation of the warranty period and related items to the GR. If a repair action falls under the manufacturer or OEM warranty provisions, the contractor must ensure compliance with these provisions and replace warranted parts at no additional cost to the government.
8.0 Accountability. The contractor shall be responsible for providing a receipt of custody for all TMDE upon delivery to the contractor’s facility. Upon completion of service or repair, the contractor must notify the GR that the necessary actions have been completed and provide a detailed shipping document of all serialized TMDE upon return shipment to its appropriate location.
9.0 Unauthorized Work Performance. The contractor shall not perform any work that deviates from the contract requirements and specifications outlined in this PWS. Any deviations from the specified work must have prior approval from the GR. If the contractor performs unauthorized work without such approval, the contractor shall be fully responsible for any associated costs, which shall be borne at the contractor's own risk. No additional costs resulting from unauthorized work will be reimbursed, and the contractor is required to absorb these expenses.
10.0 Testing Equipment. The contractor shall use TMDE testing equipment that are traceable and certified as defined in this PWS.
10.1. A copy of the test equipment calibration or certification must be submitted to the GR upon service completion.
11.0 Government Points of Contact. The Government points of contact are not warranted Contracting Officers and do not have the authority to take any action, either directly or indirectly, that would change the pricing, quantity, place of performance, delivery schedule, or any other terms and conditions of the contract.
11.1. Government Representative: Available after award
12.0 Delivery or Deliverables. The contractor shall provide a calibration status report within (7) business days upon request from the GR for the duration that the equipment is in the possession of the contractor.
13.0 Meetings and Status Reports. The contractor shall provide status updates regarding any delays, repairs, or issues pertaining to the equipment and services involved as defined in this PWS.
14.0 Applicable Documents / Governing References. In addition, requirements of regulations and policies applicable to this contract/order by virtue of clauses contained herein, the contractor shall, in the performance of this contract/order, also comply with all references applicable to the specific tasks of this Performance Work Statement, as listed below, including any amendments or subsequent issuance.
15.0 Government Furnished Facilities, Equipment, Services and/or Information. Government property identified herein, or by attachment, is subject to the requirements of FAR and DFARS part 45, pursuant to the enumerated property clauses incorporated in this contract/order.
16.0 Special Instructions and Administrative Information
16.1. Non-Personal Services. This is a non-personal services, performance-based contract. The contractor bears responsibility to manage its employees and guard against any actions that could be characterized as personal services as defined in FAR 37.104, or which convey the perception that personal services are being delivered. No provisions are implied nor provided in this contract for any benefits such as medical care, retirement, or worker’s compensation, other than those incorporated through enumerated clauses.
16.2. Identification of Contractor Personnel and Equipment. In accordance with DFARS 211.106, contractor personnel performing non-personal services on DoD installations or other government facilities shall ensure that they are readily identifiable as contractor employees (including subcontractors and consultants). All contractor employees shall comply with the following at all times:
16.2.1. Telephone Identification. Identify themselves as contractor personnel at the onset of every telephone call made from a Government telephone or any other phone if the call is made in support of Government-related requirements and activities.
16.2.2. Recorded Messages. Identify themselves as contractor personnel in all recorded messages including those, which are heard by callers attempting to contact contractor employees via answering machines or voicemail.
16.2.3. Meetings, Conferences and Events. Identify themselves as contractor personnel at the onset of every meeting, conference or any other gathering attended in support of any Government related requirements or activities.
Reference Location
NAVMC 4000.2A, “Marine Corps VIIIA Handbook” www.marcorsyscom.marines.mil
NAVMC 4000.3, “Marine Corps Medical Logistics Information Systems
Handbook” www.marcorsyscom.marines.mil
JOINT PUBLICATION 4-02, “Joint Health Service Support” www.marcorsyscom.marines.mil
MCO 4400.201 VOL. 12, “Marine Corps Class VIII Management and
Sustainment” www.marcorsyscom.marines.mil
DODI 6430.02, “Defense Medical Logistics Program, Vol. 7, Equipment
Management and Equipment Maintenance” www.health.mil
NAVMED P-5132, “Equipment Management Manual” www.med.navy.mil
National Institute of Standards and Technology (NIST) www.nist.gov http://www.marcorsyscom.marines.mil/ http://www.marcorsyscom.marines.mil/ http://www.marcorsyscom.marines.mil/ http://www.marcorsyscom.marines.mil/ http://www.health.mil/ http://www.med.navy.mil/ http://www.nist.gov/
16.2.4. Correspondence. Identify themselves as contractor personnel on any correspondence, documents, or reports accomplished or sent in support of any service provision to the Government, including, but not limited to, correspondence sent via the U.S. Mail, facsimile, or electronic mail (email) inclusive of out-of-office replies.
16.2.5. Personal and Corporate Identification. Wear or display contractor provided nametags, badges or attire which display, at a minimum, the name of the contractor and their respective company.
16.3. Subcontract Management. The contractor shall be responsible for any subcontract management necessary to integrate work performed on this requirement and shall be responsible and accountable for subcontractor performance on this requirement. The prime contractor will manage work distribution to ensure there are no
Organizational Conflict of Interest (OCI) considerations. Contractors may add subcontractors to their team after notification to the Procuring Contracting Officer (PCO) or Contracting Officer Representative (COR). Cross teaming may or may not be permitted.
16.4. Mission Essential Personnel / Tasks. Unless specified otherwise in this Performance Work Statement, contractors supporting this contract/order are not considered “Mission Essential” personnel (as defined in DFARS
237.76) who would be required during crisis or emergency operations.
16.5. Hazardous Materials and Safety. The Contractor and its subcontractors shall comply with all applicable laws, including Public Law 91-596 (Occupational Safety and Health Administration (OSHA)) and DOD Directive
4715.1E, Environmental, Safety, and Occupational Health (ESOH)). Contractor personnel performing services on a
DOD installation shall participate in local Voluntary Protection Programs (VPP). Information on the VPP is available at http://www.osha.gov/dcsp/vpp/index.html
16.6. Privacy Act. Pursuant to FAR clause 52.224-3, as incorporated, DoD requires agency-specific Privacy Act training as prescribed in DODM 5400.11-R. DODM 5400.11-R also prescribes required actions regarding information and privacy, which are applicable to the contractor and its personnel performing under this contract/order. The regulation may subject the contractor and/or contractor's personnel to criminal liability under
Title 18, Section 793 and 7908 of the United States Code.
16.7. Quality Control and Quality Assurance. In accordance with FAR part 46, and the respective enumerated clauses incorporated in this contract/order, the contactor is reminded that the government may conduct Quality
Assurance actions, inspect the Contractor’s Quality Control Plan and records, and take remedial actions as necessary, at any time through the duration of this contract/order.
16.8. PERFORMANCE REQUIREMENT SUMMARY (PRS).
Task Performance
Objective
PWS
Reference
Performance Standard Acceptable
Quality Level
Surveillance
Method
Incentive
Calibration of Equipment
Perform calibration and assessment of all TMDE to ensure operability within acceptable standards.
3.1 The contractor shall
perform off site calibration service for all
TMDE and have them returned to the
Government within (60) days from the date of departure from the on-site location. The timeline shall not include delays caused by uncontrollable factors, including but not limited to lead times for shipping equipment to the OEM.
AQL = 100%
Frequency =
Random
Sampling
100% Inspection
Payment
Serialized Equipment The contractor shall provide conformation of receipt of all
3.2 Technical & Timeline:
Requires the Contractor to provide all transportation vehicles, AQL = 100%
Frequency =
Random
Sampling
100% Inspection
Payment http://www.osha.gov/dcsp/vpp/index.html
Attachment 1. SERIALIZED TMDE LIST, 3D MEDLOG CO, OKINAWA JAPAN.
LIN NSN Serial Model Manufacturer Required
Tier
1 6515015867368 4180628 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
2 6515015867368 2365063 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
3 6515015867368 4166046 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
4 6515015867368 4169006 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
5 6515015867368 2654220 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
6 6515015867368 2384002 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
7 6515015867368 6935041 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
8 6515016239999 5347501 PROSIM 8 FLUKE BIOMEDICAL Tier 1
9 6515016239999 5359054 PROSIM 8 FLUKE BIOMEDICAL Tier 1
10 6515016239999 5348033 PROSIM 8 FLUKE BIOMEDICAL Tier 1
11 6515016239999 5339500 PROSIM 8 FLUKE BIOMEDICAL Tier 1
12 6515016239999 5359052 PROSIM 8 FLUKE BIOMEDICAL Tier 1
13 6515016239999 5353013 PROSIM 8 FLUKE BIOMEDICAL Tier 1
14 6515016239999 5355647 PROSIM 8 FLUKE BIOMEDICAL Tier 1
15 6515016239999 5359053 PROSIM 8 FLUKE BIOMEDICAL Tier 1
16 6515016454093 4961610 IDA-5 FLUKE BIOMEDICAL Tier 1
17 6515016454093 5933070 IDA-5 FLUKE BIOMEDICAL Tier 1
18 6515016569370 6181059 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
19 6515016569370 6292045 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
20 6515016569370 6291069 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
21 6515016569370 6291072 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
22 6515016569370 6291074 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
23 6515016569370 5347076 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
24 6515016569370 5347085 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
25 6515016569370 5347084 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
26 6515016620048 5898023 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
27 6515016620048 3358023 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
TMDE at their facility ensuring all equipment arrives without incident.
Material Handling Equipment (MHE) and
Personal Protective
Equipment (PPE).
Calibration Reports The contractor shall submit end of service report listing actions taken to ensure all
TMDE is serviced to the OEM standard.
5.0 The contractor must
provide an end-of-service report for each calibration service, including delivery confirmation, within (7) business days after the return of the equipment to 3D MEDLOG.
AQL = 100%
Frequency = Random
Sampling
100%
Inspection
Payment
28 6515016620048 6297068 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
29 6515016620048 6297069 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
30 6515016620048 5898022 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
31 6515016620048 5898021 IMPULSE 7010 FLUKE BIOMEDICAL Tier 1
32 6515017089175 22K281A025
VILC2S
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
33 6515017089175 22K281A01V
LKTH3M
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
34 6515017089175 22K281A025
VIMM33
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
35 6515017089175 22K281A01V
LKUWK3
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
36 6515017089175 24G04-
1A0366F4Q10
PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
37 6515017089175 24G04-
1A0366F2D1
K
PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
38 6515016940335 5387003 QA-ES III FLUKE BIOMEDICAL Tier 1
39 6515016940335 5381062 QA-ES III FLUKE BIOMEDICAL Tier 1
40 6515016940335 5387002 QA-ES III FLUKE BIOMEDICAL Tier 1
41 6515016940335 6256117 QA-ES III FLUKE BIOMEDICAL Tier 1
42 6515016940335 6252037 QA-ES III FLUKE BIOMEDICAL Tier 1
43 6515016940335 6250002 QA-ES III FLUKE BIOMEDICAL Tier 1
44 6625015927436 4910022 ESA 620 FLUKE BIOMEDICAL Tier 1
45 6625015927436 642004 ESA 620 FLUKE BIOMEDICAL Tier 1
46 6625015927436 4899028 ESA 620 FLUKE BIOMEDICAL Tier 1
47 6625015927436 4884523 ESA 620 FLUKE BIOMEDICAL Tier 1
48 6625015927436 4900003 ESA 620 FLUKE BIOMEDICAL Tier 1
49 6625015927436 4884521 ESA 620 FLUKE BIOMEDICAL Tier 1
50 6625016568416 4518026 ESA 609 FLUKE BIOMEDICAL Tier 1
51 6625016568416 4920033 ESA 609 FLUKE BIOMEDICAL Tier 1
52 6625016568416 4920037 ESA 609 FLUKE BIOMEDICAL Tier 1
53 6625016568416 3520044 ESA 609 FLUKE BIOMEDICAL Tier 1
54 6625016568416 3519029 ESA 609 FLUKE BIOMEDICAL Tier 1
55 6625016568416 4518025 ESA 609 FLUKE BIOMEDICAL Tier 1
56 6630016791832 6455700 VT650 FLUKE BIOMEDICAL Tier 1
57 6630016791832 6433723 VT650 FLUKE BIOMEDICAL Tier 1
58 6630016791832 6458708 VT650 FLUKE BIOMEDICAL Tier 1
59 6630016791832 6343703 VT650 FLUKE BIOMEDICAL Tier 1
60 6630016791832 5866081 VT650 FLUKE BIOMEDICAL Tier 1
61 6630016791832 5838034 VT650 FLUKE BIOMEDICAL Tier 1
62 6630016791832 5828064 VT650 FLUKE BIOMEDICAL Tier 1
63 6685016719273 4967219 700G04 FLUKE BIOMEDICAL Tier 2
64 6685016719273 4967225 700G04 FLUKE BIOMEDICAL Tier 2
65 6685016719273 4568572 700G04 FLUKE BIOMEDICAL Tier 2
66 6685016719273 4575577 700G04 FLUKE BIOMEDICAL Tier 2
67 6685016719273 4575587 700G04 FLUKE BIOMEDICAL Tier 2
68 6685016719274 6822050 700G04 FLUKE BIOMEDICAL Tier 2
69 6525016930527 323618 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
70 6525016930527 320918 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
71 6525016930527 323190 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
72 6525016930527 283483 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
73 6525016930527 283502 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
74 6525016930527 283489 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
75 6515016689919 67526 1936-R NEWPORT
ELECTRONICS INC
Tier 2
76 6515016689919 67505 1936-R NEWPORT
ELECTRONICS INC
Tier 2
77 6515016689919 67498 1936-R NEWPORT
ELECTRONICS INC
Tier 2
78 6515016689919 17533 1936-R NEWPORT
ELECTRONICS INC
Tier 2
79 6515016689919 17531 1936-R NEWPORT
ELECTRONICS INC
Tier 2
80 6515016689919 64924 1936-R NEWPORT
ELECTRONICS INC
Tier 2
81 6515016689919 64927 1936-R NEWPORT
ELECTRONICS INC
Tier 2
82 6515016689919 63734 1936-R NEWPORT
ELECTRONICS INC
Tier 2
83 6515016689919 81918 1936-R NEWPORT
ELECTRONICS INC
Tier 2
84 6515016689919 81901 1936-R NEWPORT
ELECTRONICS INC
Tier 2
85 6625015809615 11450 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
86 6625015809615 11449 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
87 6625015809615 11440 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
88 6625015809615 11443 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
89 6625015809615 11446 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
90 6625015809615 11448 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
91 6625016626924 7535 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
92 6625016626924 7538 SL-8 PRONK TECHNOLOGIES
Tier 1
93 6625016626924 7541 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
94 6625016626924 7542 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
95 6625016626924 7533 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
96 6625016626924 7540 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
97 6625016640608 OX10555 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
98 6625016640608 OX10559 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
99 6625016640608 OX10558 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
100 6625016640608 OX10560 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
101 6625016640608 OX10551 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
102 6625016640608 OX10550 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
103 6630016713921 51M0017601-
FI-8000 RIKEN KEIKI CO LTD Tier 1
104 6630016713921 51M0017601-
FI-8000 RIKEN KEIKI CO LTD Tier 1
105 6630016713921 51M0017601-
FI-8000 RIKEN KEIKI CO LTD Tier 1
106 6630016713921 51N0068901-7 F1-8000 RIKEN KEIKI CO LTD Tier 1
107 6630016713921 51N0068901-6 FI-8000 RIKEN KEIKI CO LTD Tier 1
108 6630016713921 52H0044201-
FI-8000 RIKEN KEIKI CO LTD Tier 1
109 6530016775459 40812140035 4081 K TSI INC Tier 2
110 6530016775459 40812141002 4081 K TSI INC Tier 2
111 6530016775459 40811605013 4081 H TSI INC Tier 2
112 6530016775459 40812021014 4081 K TSI INC Tier 2
113 6530016775459 40812021034 4081 K TSI INC Tier 2
114 6530016775459 40812021007 4081 K TSI INC Tier 2
115 6530016775459 40812021022 4081 K TSI INC Tier 2
116 6530016775459 40812121009 4081 K TSI INC Tier 2
117 6625017057614 63008706 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
118 6625017057614 58578706 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
119 6625017057614 63008705 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
120 6625017057614 63008703 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
121 6625017057614 61928702 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
122 6625017057614 61748705 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
Attachment 2. SERIALIZED TMDE LIST, COMBAT LOGISTICS COMPANY 33, KANEOHE BAY, HAWAII.
LIN NSN Serial Model Manufacturer Required
Tier
1 6515015867368 3990559 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
2 6515015867368 6370032 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
3 6515015867368 6361128 IMPULSE 7000DP FLUKE BIOMEDICAL Tier 1
4 6515016239999 2499012 PROSIM 8 FLUKE BIOMEDICAL Tier 1
5 6515016239999 6333100 PROSIM 8 FLUKE BIOMEDICAL Tier 1
6 6515016239999 6334087 PROSIM 8 FLUKE BIOMEDICAL Tier 1
7 6515016454093 6290068 IDA-5 FLUKE BIOMEDICAL Tier 1
8 6515016454093 6307075 IDA-5 FLUKE BIOMEDICAL Tier 1
9 6515016454093 6272035 IDA-5 FLUKE BIOMEDICAL Tier 1
10 6515016569370 5347077 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
11 6515016569370 6292048 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
12 6515016569370 6290035 PROSIM SPOT FLUKE BIOMEDICAL Tier 1
13 6515017089175 22K24-
1A025VIR134
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
14 6515017089175 22K28-
1A025VIKTJ9
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
15 6515017089175 23E24-
1A02KXEY83
E
PTS PORTABLE
CALIBRATION
SYSTEM
DELFI MEDICAL
INNOVATIONS INC
Tier 1
16 6515016940335 6166060 QA-ES III FLUKE BIOMEDICAL Tier 1
17 6515016940335 6250052 QA-ES III FLUKE BIOMEDICAL Tier 1
18 6515016940335 6255036 QA-ES III FLUKE BIOMEDICAL Tier 1
19 6625015927436 7014074 ESA 620 FLUKE BIOMEDICAL Tier 1
20 6625015927436 6445007 ESA 620 FLUKE BIOMEDICAL Tier 1
21 6625015927436 6393139 ESA 620 FLUKE BIOMEDICAL Tier 1
22 6630016791832 5837811 VT650 FLUKE BIOMEDICAL Tier 1
23 6630016791832 7000843 VT650 FLUKE BIOMEDICAL Tier 1
24 6630016791832 6466731 VT650 FLUKE BIOMEDICAL Tier 1
25 6685016719273 5177099 700G04 FLUKE BIOMEDICAL Tier 2
26 6685016719273 6857216 700G04 FLUKE BIOMEDICAL Tier 2
27 6685016719273 6507357 700G04 FLUKE BIOMEDICAL Tier 2
28 6525016930527 283399 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
29 6525016930527 323408 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
30 6525016930527 323422 X2 KIT W/ RF SENSOR FLUKE NETWORKS Tier 2
31 6515016689919 81145 1936-R NEWPORT ELECTRONICS
Tier 2
32 6515016689919 81288 1936-R NEWPORT ELECTRONICS
Tier 2
33 6515016689919 81492 1936-R NEWPORT ELECTRONICS
INC
Tier 2
34 6625015809615 9517 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
35 6625015809615 14513 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
36 6625015809615 14457 SC-5 PRONK TECHNOLOGIES
INC
Tier 1
37 6625016626924 5827 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
38 6625016626924 10550 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
39 6625016626924 10531 SL-8 PRONK TECHNOLOGIES
INC
Tier 1
40 6625016640608 11765 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
41 6625016640608 12490 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
42 6625016640608 12483 OX-1 PRONK TECHNOLOGIES
INC
Tier 1
43 6630016713921 51M0069901-
FI-8000 RIKEN KEIKI CO LTD Tier 1
44 6630016713921 51N0068801-
FI-8000 RIKEN KEIKI CO LTD Tier 1
45 6630016713921 51N0068801-
FI-8000 RIKEN KEIKI CO LTD Tier 1
46 6630016713921 51N0203601-
F1-8000 RIKEN KEIKI CO LTD Tier 1
47 6630016713921 51N0068801-
FI-8000 RIKEN KEIKI CO LTD Tier 1
48 6530016775459 40812121043 4081 K TSI INC Tier 2
49 6530016775459 40812145090 4081 K TSI INC Tier 2 Tier
50 6530016775459 40812204017 4081 H TSI INC Tier 1
51 6695016060366 A064545 LIGHT METER EXTECH INSTRUMENTS
CORP
Tier 1
52 6695016060366 A065860 LIGHT METER EXTECH INSTRUMENTS
CORP
Tier 1
53 6695016060366 A064556 LIGHT METER EXTECH INSTRUMENTS
CORP
Tier 1
54 6625017057614 58538713 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
55 6625017057614 63398704 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
56 6625017057614 58538707 190 M-2
OSCILLOSCOPE
FLUKE BIOMEDICAL Tier 1
57 6625015098205 62570259 287 MULTIMETER FLUKE BIOMEDICAL Tier 1
58 6625015098205 62320002 287 MULTIMETER FLUKE BIOMEDICAL Tier 1
59 6625015098205 62340256 287 MULTIMETER FLUKE BIOMEDICAL Tier 1
60 6685015399758 64452429MV 54 II B
THERMOMETER
FLUKE BIOMEDICAL Tier 1
61 6685015399758 65040119MV 54 II B
THERMOMETER
FLUKE BIOMEDICAL Tier 1
62 6685015399758 64600821MV 54 II B
THERMOMETER
FLUKE BIOMEDICAL Tier 1
63 6680014696040 H471951 TACHOMETER EXTECH INSTRUMENT
CORP
Tier 1
64 6680014696040 H471917 TACHOMETER EXTECH INSTRUMENT
CORP
Tier 1
65 6680014696040 H471951 TACHOMETER EXTECH INSTRUMENT
CORP
Tier 1
66 6680014696040 H445228 TACHOMETER EXTECH INSTRUMENT
CORP
Tier 1
67 6680014696040 H445233 TACHOMETER EXTECH INSTRUMENT
CORP
Tier 1
End Performance Work Statement
File details come from the government source that posted it. Updated .