POL Flashpoint tester Pensky Martens ASTM D93.pdf

PDF 78 KB Posted

Attached to
Flashpoint Tester Pensky Martens ASTM D93
Solicitation number
W50S9926Q1003
Issued by

About this file

This document is a requirements specification for a Pensky Martens flashpoint tester needed by the Fuels flight organization. The tester must be capable of automated operation using the ASTM D93 method and accurately detecting flashpoint temperatures up to 400 degrees Celsius for petroleum products. The equipment must feature an electric igniter to eliminate the need for propane or gas, along with a heater band for uniform heating of the heater block. A color LCD screen is required for displaying test results and inputting sample information including sample number and PASS/FAIL status, with visual representation of flash temperature.

The specification requires automatic detection of method or safety violations, built-in calibration equipment for in-house calibration capability, and 120-volt operation compatible with standard wall outlets. Additional functional requirements include easy sample cup cleaning, overheating protection, automatic diagnostic capabilities for troubleshooting, and built-in fire containment. The document does not provide information regarding procurement timeline, budget, quantities needed, evaluation criteria, response deadlines, or contracting vehicle type.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

POL Flashpoint tester Pensky Martens ASTM D93

The Fuels flight needs a flashpoint tester that meets the following requirements.

-Automated Pensky Martens ASTM D93 method -Accurately detect flashpoint up to 400 degrees Celsius for petroleum products.

-Electric igniter no need for propane or gas -Heater band for uniform heating of the heater block -Color LCD screen for displaying results and inputting sample information (sample number, PASS/FAIL) Visual of flash temperature.

-Automatic detection of method or safety violations.

-Calibration equipment included to perform calibrations in house -120 voltage for standard wall outlet -Easy cleaning for the sample cup -Over heating protection -Automatic diagnostic for troubleshooting -Built in fire containment