Sources Sought Q and A.docx

DOCX document 16 KB Posted

Attached to
DNA Sequencing Service
Solicitation number
PCA-NIEHS-00080
Issued by

About this file

This file is a Questions and Answers document clarifying requirements for a DNA Sequencing Service sources sought solicitation (PCA-NIEHS-00080) issued by the National Institutes of Health, Office of Logistics and Acquisition Operations under the Department of Health and Human Services.

The solicitation requires high-throughput Sanger sequencing services using automated instrumentation with appropriate sample tracking, quality control, and laboratory practices. Sequencing must achieve Phred-equivalent quality scores greater than 20 within the reportable region and cover user-specified regions of PCR products and plasmid DNA. Target region sizes average 700-1,000 base pairs with single-read coverage per sample, with users able to specify single-strand or bi-directional sequencing at submission. Deliverables must include .ab1, .seq, .scf, .phd.1, .txt, and .fasta/.fa file formats. The contractor must investigate and repeat failed testing within 2-5 days and retain all samples for a minimum of 180 days with documented disposal records. This document does not establish pricing terms, response dates, award dates, or contract vehicles, as it serves solely to clarify technical requirements identified during the sources sought phase.

View the file

Other files

File Type Posted
Sources Sought PCA-NIEHS-00080.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought: PCA-NIEHS-00080 DNA Sequencing Service Questions & Answers

1. Sequencing performance: Perform high-throughput Sanger sequencing with appropriate automated instrumentation, sample tracking, quality control, and laboratory practices. Successful sequencing must cover the user-specified region of PCR products and plasmid DNA, with a Phred-equivalent quality score greater than 20 within the reportable region. Question: what coverage level is being requested for coverage of the target region? 1x coverage Single strand, bi-directional, and how many reads per strand? We would like to be able to define what the user wants (single or bi-directional) at the time of submission. 1 read per sample is sufficient. What is the expected average target region size? 700-1000 base pairs

1. Deliverables: Sanger sequencing files: .ab1, .seq, and qscrl files. Question: qscrl does not appear to be a file format associated with Sanger data quality analysis. Perhaps it should be .phd.1 files? Please confirm We want.ab1, .seq, .scf, .phd.1, .txt, and .fasta/.fa

1. Question: what is the timeframe for failure investigations and repeat testing? 2-5 days

1. Sample retention: please confirm that all samples must be retained for at least 180 days, and that disposal must be recorded. That is confirmed