This is a Minimum Requirements Document for a federal procurement of radiation detection equipment by the USAF School of Aerospace Medicine, Occupational and Environmental Health (USAFSAM/OEC).
The acquisition requires one (1) RSX-1 Sodium Iodide Radiation Detector measuring 4" x 4" x 16" to enable directional radiation source identification and Health Risk Assessments for Air Force personnel protection. USAFSAM/OEC currently operates a single RSX-1 unit lacking directional capability; the additional unit will create a networked two-detector array allowing triangulation and pinpointing of radiation sources. The RSX-1 model and specified dimensions are mandatory for seamless integration with existing equipment and multi-detector array functionality. Required specifications include: Thallium-activated Sodium Iodide (NaI(Tl)) scintillation crystal with rectangular geometry of 4" x 4" x 16" and minimum active volume of 4.0 liters; energy resolution less than 8.5% with detection range from 20 keV to 3.0 MeV; 1024-channel fully linearized spectral output; automatic multi-peak gain stabilization using naturally occurring background isotopes; internally integrated Advanced Digital Spectrometer (ADS) or equivalent digital signal processing hardware capable of processing up to 250,000 counts per second with effectively zero dead-time under typical conditions and less than 20% dead-time at maximum saturation; ruggedized, weather-sealed enclosure (carbon fiber or heavy-duty aluminum) not exceeding approximately 750 mm (L) x 170 mm (W) x 180 mm (H) and 25 kg (55 lbs); continuous operation from -30°C to +45°C; and 9–40 VDC power compatibility drawing no more than 6 Watts under normal operation. Deliverables include the detector assembly, manufacturer calibration certificates, quality assurance/resolution test results, integration documentation, mechanical interface drawings, and standard digital interface cables (Ethernet/USB/Serial as applicable).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SSJ (Redacted).pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum Requirements Document
RSX-1 Radiation Detector
29 July 2026
Background & Scope: The USAF School of Aerospace Medicine, Occupational and
Environmental Health (USAFSAM/OEC), requires the procurement of one (1) RSX-1 Sodium
Iodide Radiation Detector, measuring 4" x 4" x 16". This acquisition is critical for enabling directional radiation source identification and for conducting comprehensive Health Risk
Assessments (HRAs) to protect Air Force personnel from high levels of ionizing radiation.
USAFSAM/OEC currently possesses a single RSX-1 unit. While this detector provides essential radiation identification and detection capabilities, it lacks the intrinsic ability to determine the directionality of a radiation source. This limitation necessitates the use of secondary equipment and time-consuming procedures to locate the origin of radioactive contamination, reducing operational efficiency and potentially increasing exposure risk.
The acquisition of an additional RSX-1 unit is required to create a networked sensor system. By operating two identical detectors in tandem, personnel can triangulate and pinpoint the exact location of radiation sources with speed and precision. This directional capability is mission-essential for effective contamination control and accurate risk assessment.
The specified 4" x 4" x 16" dimensions and the RSX-1 model are the only specifications that will ensure seamless integration with existing equipment. Compatibility is paramount for creating the multi-detector array necessary for achieving the required directional functionality. The use of any other model would fail to meet this critical interoperability requirement.
Requirements/Salient Characteristics: The proposed system must meet or exceed the following salient characteristics:
1. Detector Physics and Geometry
• Detector Material: Thallium-activated Sodium Iodide (NaI(Tl)) scintillation crystal.
• Crystal Geometry: Rectangular prism measuring 4 inches x 4 inches x 16 inches (101.6 mm x 101.6 mm x 406.4 mm).
• Active Volume: Minimum of 4.0 Liters (approx. 256 cubic inches) of detection volume per unit to ensure high geometric sensitivity.
2. Spectroscopic Performance
• Energy Resolution: Must be <8.5% for accurate nuclide identification.
• Energy Range: Must cover a minimum detection range from 20 keV to 3.0 MeV.
• Spectral Output: Must output a fully linearized, high-resolution 1024-channel energy spectrum.
• Gain Stabilization: Must feature automatic multi-peak gain stabilization utilizing naturally occurring background isotopes (no external radioactive check sources required for routine stabilization).
3. Signal Processing and Electronics
• Digital Integration: The detector pack must contain an internally integrated Advanced
Digital Spectrometer (ADS) or equivalent digital signal processing (DSP) hardware.
• Throughput & Dead Time: Must be capable of processing high count rates up to
250,000 counts per second (cps) per crystal. System must maintain effectively zero dead-time under typical environmental background conditions and <20% dead-time at maximum saturation.
• Linearization: Must support fully linearized spectral summation to allow multiple detector packs to be combined without degrading the spectrum resolution or distorting the
Poisson distribution.
4. Physical and Environmental Characteristics
• Enclosure: Must be housed in a ruggedized, weather-sealed enclosure (e.g., carbon fiber, heavy-duty aluminum) designed to withstand severe vibrations and shocks associated with off-road vehicle or aviation environments.
• Dimensions: Must not exceed approximately 750 mm (L) x 170 mm (W) x 180 mm (H).
• Weight: Total assembly weight must not exceed 25 kg (approx. 55 lbs).
• Operating Temperature: Must operate continuously in extreme field conditions from -
30°C to +45°C.
• Power Consumption: Must be compatible with standard vehicle/aircraft power, operating on 9–40 VDC and drawing no more than 6 Watts under normal operation.
Deliverables:
• One (1) 4L NaI(Tl) Gamma-Ray Scintillation Detector assembly meeting the specifications above.
• Manufacturer calibration certificates and initial quality assurance/resolution test results.
• Integration documentation, mechanical interface drawings, and standard digital interface cables (e.g., Ethernet/USB/Serial as applicable).