Coagulation SOW - Enclosure A Site Specific information Amendment 0008.xlsx

XLSX spreadsheet 20 KB Posted

Attached to
FY26 Lab Consolidation for Coagulation CPRR Effort
Solicitation number
HT941026Q2004
Issued by

About this file

This is a Site Specific Information document (Enclosure A) detailing requirements for coagulation analyzer deployment at Madigan Army Medical Center, Joint Base Lewis-McChord, Washington, under the FY26 Lab Consolidation for Coagulation CPRR Effort (Solicitation HT941026Q2004). The document specifies that the anticipated delivery start date is contingent on the Government's exercise of remaining Option Periods under the existing contract, with a transition period of 2-4 months anticipated for analyzer installation. Madigan Army Medical Center requires a single coagulation analyzer with maximum space dimensions of 4 feet by 3 feet by 6 feet and power requirements of 100-240V at 50-60Hz with maximum power consumption of 65W, or alternatively 95V, 115V, or 230V at 50/60Hz. The delivery start date is listed as October 1, 2027, October 1, 2028, or October 1, 2029, depending on whether the MTF exercises Option 2. The site can accommodate the new equipment while existing equipment remains in place.

The document includes detailed questions regarding laboratory operations, including DIC panel composition and test counts, batch testing frequencies and QC protocols, routine test QC frequency and levels, reagent-grade rinse water availability, requirement for onsite vendor support for lot-to-lot assistance and regression reports, and whether APTT testing is used for heparin level monitoring. These site-specific requirements are part of a cost-per-reportable-result (CPRR) contract solicitation issued by the Defense Health Agency for commercial coagulation analyzer products and services to support military treatment facilities. This document is Amendment 0008 to the original solicitation, with revised pricing requested by September 17, 2026.

View the file

Other files

File Type Posted
HT941026Q2004 Amendment 0008 Cover Letter.pdf PDF
Combined Synopsis Solicitation HT941026Q2004 Amendment 0008.pdf PDF
CSS Coagulation - Attachment 3 Quote Sheet Amendment 0008.xlsx XLSX spreadsheet
CSS Coagulation - Attachment 1 SOW Amendment 0008.pdf PDF
CSS Coagulation - Attachment 1 SOW Amendment 0007.docx DOCX document
Combined Synopsis Solicitation HT941026Q2004 Amendment 0007.docx DOCX document
HT941026Q2004 Amendment Cover Letter 0007 V3.pdf PDF
HT941026Q2004 Amendment 0006 Cover Letter.pdf PDF
CSS Coagulation - Attachment 2 Technical Submission Form Amend 0006.xlsx XLSX spreadsheet
HT941026Q2004 Amendment 0005 Cover Letter.pdf PDF
HT9410Q2004 0005 Questions and Answers.pdf PDF
Combined Synopsis Solicitation HT941026Q2004 Amendment 0005.pdf PDF
CSS Coagulation - Attachment 3 Quote Sheet Amendment 0003.xlsx XLSX spreadsheet
HT9410Q2004 0004 Questions and Answers.docx DOCX document
HT941026Q2004 Amendment Amendment 0004 Cover Letter.pdf PDF
CSS Coagulation - Attachment 3 Quote Sheet Amendment 0002.xlsx XLSX spreadsheet
HT941026Q2004 Amendment 0002 Cover Letter.pdf PDF
Combined Synopsis Solicitation HT941026Q2004 0001.pdf PDF
HT941026Q2004 Amendment Cover Letter.pdf PDF
Coagulation SOW - Enclosure A Site Specific information.xlsx XLSX spreadsheet
CSS Coagulation - Attachment 4 Pre-Solication Q and A.pdf PDF
CSS Coagulation - Attachment 1 SOW.pdf PDF
CSS Coagulation - Attachment 3 Quote Sheet.xlsx XLSX spreadsheet
Combined Synopsis Solicitation HT941026Q2004.pdf PDF
Coagulation SOW - Enclosure B CyberLOG-ICS.pdf PDF
CSS Coagulation - Attachment 2 Technical Submission Form.xlsx XLSX spreadsheet
Show all 26

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Site Specific Information

NOTE: Contingency of Delivery Start: The initial delivery start date (or commencement of supply) under any resulting Purchase Order is contingent upon the Government’s determination to exercise or not exercise remaining Option Period(s) under existing Order/Contract. The Government anticipates transition period of 2-4 months for analyzers.
MTFNetworkAnticipated Start Date of DeliveriesAnalyzer QuantityMaximum Space Dimensions (Height x Width x Depth) for Analyzer EquipmentPower Requirement (what is the maximum power available)Any site-specific requirements or additional equipment required?Does MTF have room for new equipment while old equipment still there?AddressLaboratory Hours of OperationDIC Panel Question 1: What individual tests do you anticipate making up this panel?DIC Panel Question 2: Are these individual tests included in the other test count totals provided or to be billed a s a CPRR separately?Does your labs already have access to reagent grade rinse water? Y / NOnsite vendor supported “lot to lot” assistance and regression reports are required for Naval MCSD. Does your site also require onsite vendor support for Lot-to-Lot assistance and reports? Y / NWhich tests are performed in batches, if any?If tests are performed in batches, how frequently is the QC performed?What frequency and number of levels of QC are performed on Routine tests? (Example: 3 levels once per day, 2 levels every 8 hours, or Only when ordered)Please list what tests are included in a "DIC Panel".Which MTFs, if any, use the “Activated Partial Thromboplastin Time (APTT)” test to monitor heparin levels in patients? (Does you MTF use APTT to monitor heparin levels in patients?)
Madigan Army Medical Center, Joint Base Lewis-McChord, WAIndo PacificEither 10/1/2027, 10/1/2028, or 10/1/2029* TBD - Depends on MTF requesting to exercise Option2Site can accommodate a large space - maximum allowable dimensions: 4’ x 3' x 6'100-240V with 50-60Hz and max power consumption 65W and 95V, 115V, or 230V at 50/60Hz and max power <110W (1400VA)N/ANo for one analyzer; yes for one analyzerDepartment of Pathology / Core Laboratory

9040 Jackson Ave.

Tacoma, WA 9843124 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerDIC are billed as a panel; further testing will be billed seperatelyY (currently through NERL water through Omnicell)NoAll Factors assays (V, VII, VIII, IX, X and XI) and Lupus Anticoagulant (LAC)Weekly2 levels every 8 hoursPT, PTT, FIB and D-DimerWe still use PTT for Heparin Therapy but as of October 2024, DHA has added Anti-Xa testing to our Heparin Therapy Power Plans.
Naval Hospital Camp Pendleton, CAPacific RimEither 2/1/2027 or 2/1/2028* TBD - Depends on MTF requesting to exercise Option2H 80" x W 50" x L 125"120V, 60 Hz.N/AYES, depending on the size of the new equipment.200 Mercy Circle
Oceanside, CA 9205524 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Y (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerNo
Naval Hospital Bremerton, WAPacific RimEither 8/1/2027 or 8/1/2028* TBD - Depends on MTF requesting to exercise Option120.5" x 44.5" x 19.5"120V, 60 Hz.Must NOT exceed defined maxiumum space dimensions for site. Shall NOT require an external water source or drains for operation. Must be Fully automated benchtop. Shall be capable of clot detection, automatically dilution, processing plasma patient samples, and positively identifying patient samples using bar codes.Yes, depending on the size of the new equipment.Laboratory Department

1 Boone Road Code O5L, Building 1

Bremerton, WA 98312-18980700 to 1600, 7 days a week (holidays included)N/AYes, all the test are included in the count provided.Yes (currently through NERL water through Omnicell)NoN/AN/A2 levels every 8 hoursN/AYes
Robert E. Bush Naval Hospital Twentynine Palms, CAPacific RimEither 10/1/2027, 10/1/2028, 10/1/2029, or 4/1/2030* TBD - Depends on MTF requesting to exercise Option2H 40" x W 96" x L 30"120V, 60 Hz.Does not require an external water source. See max dimension column.Yes, depending on the size of the new equipment.Laboratory Department

1145 Sturgis Rd., Bldg. 1343

Twentynine Palms, CA 9227824 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerWe do not have a DIC panelYes (currently through DMLSS or GPC)NoN/AN/A2 levels every 8 hours for PT, PTT. As needed QC for D-Dimer and Fibrinogen due to low volume.PT, PTT, FIB and D-DimerNo
99th Medical Group, Mike O'Callaghan Military Medical Center, Nellis AFB, NVContinental11/17/2026 OR 11/17/2027 * TBD - Depends on MTF requesting to exercise Option2Site cannot accommodate a large space: There is not extra space available at this time - H 27" x W 31" x D 35"; (one analyzer is countertop)120 V/ 60 HzConnection compatibility with the Aptio LineNoCore Lab

0079 MMDCN Lab 4420 N. Grissom Ave.

Nellis AFB, NV 8919124 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Yes. They are bought via ECAT. Preferably, reagent grade water be included in this contract along with the required consumables.On site support is preferred at first. Can be as needed remote support after.N/AN/A2 levels every 8 hours for PT and PTT. As needed QC for D-Dimer, Fibrinogen, Heparin due to low volume.N/ANo
60th Medical Group, David Grant Medical Center, Travis AFB, CACentral9/1/272130" x70" x 70"The facility is able to supply the maximum power requirements.NoBased on the laboratory layout with all laboratory equipment and when the other contracts are awarded. If multiple contracts are awarded at once the laboratory lacks sufficient space to old and new equipment.Core Lab

101 Bodin Circle, Bldg. 777

Travis AFB, CA 9453524 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Y (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hours for PT, PTT. As needed QC for D-Dimer and Fibrinogen due to low volume.PT, PTT, FIB and D-DimerNo
Weed Army Community Hospital, Fort Irwin, CAWest12/30/261L 130" X W 31 ½"110VN/AYesAttention: LAB

390 N Loop Rd.

Fort Irwin, CA 923100730-1630 for phlebotomy and 24hrs for inpatient/ER careN/AN/AY (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerNo
Naval Medical Center San Diego, CA
- Core LabPacific RimEither 10/1/2027, 10/1/2028, or 10/1/2029* TBD - Depends on MTF requesting to exercise Option2L 72" x W 36" x H 48"The facility is able to supply the maximum power requirements.Shall NOT require an external water source or drains for operation. Must be Fully automated benchtop. Shall be capable of clot detection, automatic dilution, processing plasma patient samples, positively identifying patient samples using bar codes, providing onboard QC data base for current and historical lots, provide onboard historical patient database of results to include patient demographics, lot info, etc. (for duration of 5 yr contract), and provide lot to lot calibration support with inclusion of final report with easy-to-read data regression and acceptability interpretation.YesNMCSD Core Lab

34800 Bob Wilson Dr.

San Diego, CA 9213424 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.YesYesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerYes
Tripler Army Medical Center, Fort Shafter, HIIndo Pacific10/1/2026, 2/1/2027, or 8/1/2027* TBD - Depends on MTF requesting to exercise Option2Site can accommodate a large space - maximum allowable dimensions: H 3' x W 12' x D 2'8"95V, 115V, or 230V at 50/60Hz and max power <110W (1400VA)NoYesDepartment of Pathology

1 Jarrett White Road

Tripler AMC, HI 968590730-1630N/AN/AYesWe do not require on site assistance, but do require assistance with calculating geo mean and normal ranges.Prior to inital testing of patient samples and if exceeds 8 hours of the last acceptable QC results.Prior to inital testing of patient samples and if exceeds 8 hours of the last acceptable QC results.2 levels every 8 hoursN/ANo
673rd Medical Group, Joint Base Elmendorf-Richardson (JBER), AKCentralEither 8/1/2027, 8/1/2028, 8/1/2029, or 8/1/2030* TBD - Depends on MTF requesting to exercise Option140 x 100 x 50120V, 60 Hz.NoYES, depending on the size of the new equipment.Lab Department (Core)

5955 Zeamer Ave.

Elmendorf AFB, AK 995060730-1630 for phlebotomy and 24hrs for inpatient/ER careN/AN/AY (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hoursN/ANo
56th Medical Group, Luke AFB, AZCentralN/A1L 72 x W 36" x H48"
Luke does not have a space restriction.120V, 60 Hz.NoYes7219 N Litchfield Rd.
Luke AFB, AZ 8530924 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerDIC are billed as a panel; further testing will be billed seperatelyY (currently through NERL water through Omnicell)NoN/AN/A2 levels every 8 hoursN/ANo

Estimated Annual Workload

NOTE: Contingency of Delivery Start: The initial delivery start date (or commencement of supply) under any resulting Purchase Order is contingent upon the Government’s determination to exercise or not exercise remaining Option Period(s) under existing Order/Contract. The Government anticipates transition period of 2-4 months for analyzers.
MTFProthrombin Time (PT)Activated Partial Thromboplastin Time (APTT)FibrinogenD-DimerHeperinLupus AnticoagulantDIC PanelPatient Thrombin TimeMixing StudiesReptilase TimeFactor V AssayFactor VII AssayFactor VIII AssayFactor IX AssayFactor X AssayFactor XI AssayAntithrombin IIIVon Willebrand Factor AntigenFactor II Assay
Madigan Army Medical Center, Joint Base Lewis-McChord, WA125006625750225075188013131363191301313130025
Naval Hospital Camp Pendleton, CA2000800310600000000000000000
Naval Hospital Bremerton, WA3352111950000000000000000
Robert E. Bush Naval Hospital Twentynine Palms, CA1000800300600000000000000000
99th Medical Group, Mike O'Callaghan Military Medical Center, Nellis AFB, NV810235392701631215200000000000000
60th Medical Group, David Grant Medical Center, Travis AFB, CA4740252024081696000000000000000
Weed Army Community Hospital, Fort Irwin, CA500400100200000000000000000
Naval Medical Center San Diego, CA
- Core Lab140008001240038003720000000300000000
Tripler Army Medical Center, Fort Shafter, HI5891483550811840001120161241124439165
56th Medical Group, Luke AFB, AZ43329600000000000000000
673rd Medical Group, Joint Base Elmendorf-Richardson (JBER), AK14629838468700000000000000
Totals:5096329010515711768355918802433137931471251717391630