Contractor Questions: 75F40120Q00113
1)With regards to requirement specification in FDA-RFQ-75F40120Q00113, page 3, can you please advise minimum absorbance requirement should be 3.2 in absorbance as opposed to 2.2 in absorbance as currently listed.
Minimum Requirements for Solar Light SPF Analyzer (or equal to): 1. Light source shall provide a continuous, full spectrum radiation from 250 to 475 nanometers (nm) 2. Shall measure at least 2.2 in absorbance
2) We have available simulator to pre-irradiate 4 or 9 plates simultaneously. Can you advise on requirement (i.e. 4 or 9 plates) or should we provide quotation for each system?
Please advise
Answers from FDA:
Absorbance listed in #2 should be 3.2 instead of 2.2. Also 9 plates would be preferred.