This Statement of Work (SOW) details requirements for a Temperature and Humidity Controlled Glove Box for the Air Force Institute of Technology (AFIT). The specialized scientific equipment is needed to support student training and research in engineering physics, with precise environmental control specifications including temperature accuracy of ± 1 degrees Celsius, humidity control to ± 3% Relative Humidity, and specific dimensional requirements. Key features include a clear acrylic top section, inert gas hookup capability, thermoelectric assembly, 8" color touch screen, 24-hour data logging, four pre-set conditions, and a serial port for external computer logging. The glove box will be delivered to Wright-Patterson AFB, Ohio, with a standard 24-month warranty (excluding gloves and molecular sieve), supporting AFIT students' laboratory and thesis research activities.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Vendor Q and A 01 MAY 2025.pdf | ||
| Vendor Q and A 30 APR 2025.pdf | ||
| Solicitation Amendment 30 APR 2025.pdf | ||
| AFIT Glove Box SOW 10 Apr 2025.pdf | ||
| Solicitation.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
For
Temp & Humidity Controlled Glove Box
Title: Temp & Humidity Controlled Glove Box
Document Date: 29 APR 2025
1. Scope:
The Air Force Institute for Technology has a need for a Temp & Humidity Controlled Glove Box to train students in fundamental scientific concepts. Experiments with this item will complement AFIT coursework in the Departments of Engineering Physics enabling invaluable lab-based study of the concepts examined in lecture. The Glove Box will also support thesis and dissertation projects for AFIT students.
1.1 Background:
Glove Boxes have been a fundamental tool of nuclear science since the beginnings of the field. The Glove Box accurately maintains temperature and humidity inside a roomy single-operator glove box.
2. Applicable Documents N/A
3. Requirements/Salient Characteristics:
The following specification lists the requirements for this system:
1. Glove Box with the following dimensions:
a. Inside area 41 inches wide by 28 inches deep by 26 inches.
b. Outside Dimensions with Transfer chamber – 55 inches wide by 35 inches deep by
38 inches high.
c. Maximum working pressure 1 psi.
2. Temperature controlled to an accuracy of ± 1 degrees Celsius.
3. Platinum RTD with sealing gland.
4. Humidity Sensor with humidity controlled to an accuracy of ± 3% Relative Humidity.
5. Humidity (Rh) transmitter with 8 feet of sensor cable.
6. Drying Train System
7. 3 Canisters containing Molecular Sive
8. Nebulizer for the Rh system.
9. Room temperature and Rh read outs.
10. Oxygen Sensor
11. Circulation vacuum pump
12. 2 sets of ambidextrous Hypalon gloves.
13. Inert gas hookup capability
14. Thermoelectric assembly (Peltier device).
15. The top section is a one-piece clear acrylic plastic.
16. Four (4) key cock valves for purging. Two (2) on the main chamber and two (2) on the transfer chamber
17. Inert gas hookup capability
18. 24-hour one touch, on screen data logging
19. Four (4) pre-set conditions.
20. 8” (203 mm) Color panel touch screen.
21. Serial port for external computer data logging & remote access.
22. Standard 24-month warranty except for gloves and Molecular Sive.
4. Delivery/Period of Performance:
4.1 Place of Performance:
Air Force Institute of Technology (AFIT/EN) ATTN: Gregory Smith 2950 Hobson Way, Building 640, Room 265 Wright-Patterson AFB, OH 45433-7765