DEC2023 202312011131 ECN 108221 MCI AA0128 RPT_Redacted.pdf
PDF 3 MB Posted
- Attached to
- Hazardous & Non-Hazardous Clean Room Testing
- Solicitation number
- FA860124Q0111
About this file
This document is a test report related to a federal contract opportunity. The report details the results of airflow velocity tests and other site installation assessment tests conducted on a piece of equipment. It includes details on the equipment, test methods, and test results.
The related federal contract opportunity is for Hazardous and Non-Hazardous Clean Room Testing services. The contract is a firm-fixed-price award to be made to the lowest priced, technically acceptable offeror. The 12-month contract period will require testing of 797 USP and 800 USP standards-compliant clean rooms at an 88th Medical Group facility that supports 55,000 beneficiaries and performs 20,000 sterile IV admixtures annually. Offerors must submit proposals by June 6, 2024 in accordance with the instructions provided.
View the file
Other files
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Test Report No.: 102458839 FINAL AIRFLOW VELOCITY TESTS
Anemometer: Mfg. Model S/N Cal DueTSI 966 P23360080 February 15, 2024
Downflow Velocity
Number of Supply Readings Taken
Maximum Supply Velocity Allowable/Actual
Minimum Supply Velocity Allowable/Actual
Average Downflow Velocity
Acceptance Criteria
Diagram of Supply Velocity Profile Taken at inch(es)
45 4144 454144 43
46 4444 434542 41
37 4136 393739 41
Exhaust Volume
Velocity Conforms to:
Blower Speed Required:
Exhaust Configuration:
Exhaust Damper:
SITE INSTALLATION ASSESSMENT TESTS
Sash Alarm Verification
Airflow Alarm Verification
Type B Supply Blower Interlock Test
Exhaust Performance Test
Differential Pressure: Hood "W.C. Exhaust Duct: "W.C.
PASS FAIL NA
PASS FAIL NA
PASS FAIL NA
PASS FAIL NA
58 46
26 36
42 35 45-
NSF Std. 49 & Manufacturer's Specs
No Change
Hard Ducted
No Change
.01 NA
No Adjustments Required
No Adjustments Required
Hard Duct Negative above 8" sash height
Pass Average Velocity
& Airflow Uniformity
Fail Average Velocity
Pass Average Velocity
Fail Airflow Uniformity
All readings within the larger of ±25% of Avg. or ±16 FPM
(cfm1 + cfm2 + ... cfm5 / 5 = Qi) Average Intake Volume (Qi)
Acceptance Criteria
257.20
100 - 110 Pass Fail
689656678 677 662
Intake Area (Ai)
(Qi / Ai = Vi) Intake Velocity (calculated) (Vi)
2.56
Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721
1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm
257260258 256 255
1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm
90533004
(cfm1 + cfm2 + ... cfm5 / 5 = Qe) Average Exhaust Volume (Qe) cfm (INFORMATION ONLY)672.40
Note: Readings were taken with flowhood at intake with supply blower off and supply intake sealed. Exhaust volumes are for information only.
Intake Volume
Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721 90533004
Note: Readings were taken with flowhood at intake with supply blower on and intake side edge slots sealed.
(Blower must shut off with ~20% reduction of exhaust cfm.)
Pass-No Adjustment Req'd cfm sq.ft.
fpm fpm fpm fpm fpm fpm
B2DMINT Rev. 05MAR2016 R1.0Copyright © 2006 - All rights reserved
Test Report No.:
A smoke source shall be passed from one end of the cabinet to the other, along the centerline of the work surface at a height of 4 inches (10 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux (upward flow). No smoke shall escape from the cabinet.
102458839 AIRFLOW SMOKE PATTERN TESTS
1. DOWNFLOW TEST:
PASS FAIL
Comments: NA
A smoke source shall be passed from one end of the cabinet to the other, 1 inch (2.5 cm) behind the viewscreen, at a height of 6 inches (15 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux. No smoke shall escape from the cabinet.
PASS FAIL MEETS MANUFACTURER'S SPECIFICATIONS
3. WORK OPENING EDGE RETENTION TEST:
Comments: NA
A smoke source shall be passed along the entire perimeter of the work opening edges, approximately 1.5'' (3.8 cm) outside the cabinet. Particular attention should be paid to corners and vertical edges. No smoke shall reflux out of the cabinet once drawn in, nor shall smoke billow over the work surface or penetrate onto it.
PASS FAIL
Comments: NA
PASS FAIL
4. SASH/WINDOW SEAL TEST:
A smoke source shall be passed up the inside of the window 2'' (5 cm) from the sides and along the top of the cabinet interior work area. No smoke shall escape out of the cabinet.
Comments: NA
1-18 x 48 x 3.54 P/N: 41978
2-11.15 x 22.08 x 8 P/N: 42061
Supply HEPA Filter(s) Size:
Exhaust HEPA Filter(s) Size:
HEPA Filter(s) Replaced:
Prefilter(s) Replaced:
U.V. Lamp(s) Replaced:
REMARKS
No
No
NA
2. VIEW SCREEN RETENTION TEST:
MK41978
MK372G002
Pre Filter(s) Size: