DEC2023 202312011130 ECN 108220 MCI AA0128 Bio_Redacted.pdf

PDF 3 MB Posted

Attached to
Hazardous & Non-Hazardous Clean Room Testing
Solicitation number
FA860124Q0111
Issued by

About this file

This document is a test report for airflow velocity and smoke pattern tests conducted on a biosafety cabinet. The report includes detailed measurements and assessments of the downflow velocity, exhaust volume, airflow profile, and smoke retention tests. The cabinet was evaluated against the requirements of NSF Standard 49 and the manufacturer's specifications.

The document also references a related federal contract opportunity for Hazardous and Non-Hazardous Clean Room Testing (Solicitation FA8601-24-Q-0111) issued by the Department of the Air Force Materiel Command Lifecycle Management Center at Wright-Patterson Air Force Base. The contract is for providing testing services to ensure the 88th Medical Group's pharmacy clean rooms comply with USP 797 and USP 800 standards. The contract will be awarded on a Lowest Price Technically Acceptable (LPTA) basis to the responsible offeror submitting the lowest evaluated price. Proposals are due by June 6, 2024.

View the file

Other files

Show all 15

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Test Report No.: 102458822 FINAL AIRFLOW VELOCITY TESTS

Anemometer: Mfg. Model S/N Cal DueTSI 966 P23360080 February 15, 2024

Downflow Velocity

Number of Supply Readings Taken

Maximum Supply Velocity Allowable/Actual

Minimum Supply Velocity Allowable/Actual

Average Downflow Velocity

Acceptance Criteria

Diagram of Supply Velocity Profile Taken at inch(es)

39 4040 414040 42

37 4140 433842 40

41 4044 434643 46

Exhaust Volume

Velocity Conforms to:

Blower Speed Required:

Exhaust Configuration:

Exhaust Damper:

SITE INSTALLATION ASSESSMENT TESTS

Sash Alarm Verification

Airflow Alarm Verification

Type B Supply Blower Interlock Test

Exhaust Performance Test

Differential Pressure: Hood "W.C. Exhaust Duct: "W.C.

PASS FAIL NA

PASS FAIL NA

PASS FAIL NA

PASS FAIL NA

57 46

25 37

41 35 45-

NSF Std. 49 & Manufacturer's Specs

No Change

Hard Ducted

No Change

0.1 NA

No Adjustments Required

No Adjustments Required

Hard Duct Negative above 8" sash height

Pass Average Velocity

& Airflow Uniformity

Fail Average Velocity

Pass Average Velocity

Fail Airflow Uniformity

All readings within the larger of ±25% of Avg. or ±16 FPM

(cfm1 + cfm2 + ... cfm5 / 5 = Qi) Average Intake Volume (Qi)

Acceptance Criteria

270.80

100 - 110 Pass Fail

651688713 680 701

Intake Area (Ai)

(Qi / Ai = Vi) Intake Velocity (calculated) (Vi)

2.56

Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721

1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm

270271273 275 265

1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm

90533004

(cfm1 + cfm2 + ... cfm5 / 5 = Qe) Average Exhaust Volume (Qe) cfm (INFORMATION ONLY)686.60

Note: Readings were taken with flowhood at intake with supply blower off and supply intake sealed. Exhaust volumes are for information only.

Intake Volume

Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721 90533004

Note: Readings were taken with flowhood at intake with supply blower on and intake side edge slots sealed.

(Blower must shut off with ~20% reduction of exhaust cfm.)

Pass-No Adjustment Req'd cfm sq.ft.

fpm fpm fpm fpm fpm fpm

B2DMINT Rev. 05MAR2016 R1.0Copyright © 2006 - All rights reserved

Test Report No.:

A smoke source shall be passed from one end of the cabinet to the other, along the centerline of the work surface at a height of 4 inches (10 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux (upward flow). No smoke shall escape from the cabinet.

102458822 AIRFLOW SMOKE PATTERN TESTS

1. DOWNFLOW TEST:

PASS FAIL

Comments: NA

A smoke source shall be passed from one end of the cabinet to the other, 1 inch (2.5 cm) behind the viewscreen, at a height of 6 inches (15 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux. No smoke shall escape from the cabinet.

PASS FAIL MEETS MANUFACTURER'S SPECIFICATIONS

3. WORK OPENING EDGE RETENTION TEST:

Comments: NA

A smoke source shall be passed along the entire perimeter of the work opening edges, approximately 1.5'' (3.8 cm) outside the cabinet. Particular attention should be paid to corners and vertical edges. No smoke shall reflux out of the cabinet once drawn in, nor shall smoke billow over the work surface or penetrate onto it.

PASS FAIL

Comments: NA

PASS FAIL

4. SASH/WINDOW SEAL TEST:

A smoke source shall be passed up the inside of the window 2'' (5 cm) from the sides and along the top of the cabinet interior work area. No smoke shall escape out of the cabinet.

Comments: NA

1-18 x 48 x 3.54 P/N: 41978

2-11.15 x 22.08 x 8 P/N: 42061

Supply HEPA Filter(s) Size:

Exhaust HEPA Filter(s) Size:

HEPA Filter(s) Replaced:

Prefilter(s) Replaced:

U.V. Lamp(s) Replaced:

REMARKS

No

No

NA

2. VIEW SCREEN RETENTION TEST:

MK41978

MK372G002

Pre Filter(s) Size:

B2DMINT Rev. 05MAR2016 R1.0Copyright © 2006 - All rights reserved