DEC2023 202312011130 ECN 108220 MCI AA0128 Bio_Redacted.pdf
PDF 3 MB Posted
- Attached to
- Hazardous & Non-Hazardous Clean Room Testing
- Solicitation number
- FA860124Q0111
About this file
This document is a test report for airflow velocity and smoke pattern tests conducted on a biosafety cabinet. The report includes detailed measurements and assessments of the downflow velocity, exhaust volume, airflow profile, and smoke retention tests. The cabinet was evaluated against the requirements of NSF Standard 49 and the manufacturer's specifications.
The document also references a related federal contract opportunity for Hazardous and Non-Hazardous Clean Room Testing (Solicitation FA8601-24-Q-0111) issued by the Department of the Air Force Materiel Command Lifecycle Management Center at Wright-Patterson Air Force Base. The contract is for providing testing services to ensure the 88th Medical Group's pharmacy clean rooms comply with USP 797 and USP 800 standards. The contract will be awarded on a Lowest Price Technically Acceptable (LPTA) basis to the responsible offeror submitting the lowest evaluated price. Proposals are due by June 6, 2024.
View the file
Other files
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Test Report No.: 102458822 FINAL AIRFLOW VELOCITY TESTS
Anemometer: Mfg. Model S/N Cal DueTSI 966 P23360080 February 15, 2024
Downflow Velocity
Number of Supply Readings Taken
Maximum Supply Velocity Allowable/Actual
Minimum Supply Velocity Allowable/Actual
Average Downflow Velocity
Acceptance Criteria
Diagram of Supply Velocity Profile Taken at inch(es)
39 4040 414040 42
37 4140 433842 40
41 4044 434643 46
Exhaust Volume
Velocity Conforms to:
Blower Speed Required:
Exhaust Configuration:
Exhaust Damper:
SITE INSTALLATION ASSESSMENT TESTS
Sash Alarm Verification
Airflow Alarm Verification
Type B Supply Blower Interlock Test
Exhaust Performance Test
Differential Pressure: Hood "W.C. Exhaust Duct: "W.C.
PASS FAIL NA
PASS FAIL NA
PASS FAIL NA
PASS FAIL NA
57 46
25 37
41 35 45-
NSF Std. 49 & Manufacturer's Specs
No Change
Hard Ducted
No Change
0.1 NA
No Adjustments Required
No Adjustments Required
Hard Duct Negative above 8" sash height
Pass Average Velocity
& Airflow Uniformity
Fail Average Velocity
Pass Average Velocity
Fail Airflow Uniformity
All readings within the larger of ±25% of Avg. or ±16 FPM
(cfm1 + cfm2 + ... cfm5 / 5 = Qi) Average Intake Volume (Qi)
Acceptance Criteria
270.80
100 - 110 Pass Fail
651688713 680 701
Intake Area (Ai)
(Qi / Ai = Vi) Intake Velocity (calculated) (Vi)
2.56
Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721
1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm
270271273 275 265
1) ________ cfm 2) ________ cfm 3) ________ cfm 4) ________ cfm 5) ________ cfm
90533004
(cfm1 + cfm2 + ... cfm5 / 5 = Qe) Average Exhaust Volume (Qe) cfm (INFORMATION ONLY)686.60
Note: Readings were taken with flowhood at intake with supply blower off and supply intake sealed. Exhaust volumes are for information only.
Intake Volume
Flowhood: Mfg. Model January 31, 2024S/N Cal DueTSI EBT721 90533004
Note: Readings were taken with flowhood at intake with supply blower on and intake side edge slots sealed.
(Blower must shut off with ~20% reduction of exhaust cfm.)
Pass-No Adjustment Req'd cfm sq.ft.
fpm fpm fpm fpm fpm fpm
B2DMINT Rev. 05MAR2016 R1.0Copyright © 2006 - All rights reserved
Test Report No.:
A smoke source shall be passed from one end of the cabinet to the other, along the centerline of the work surface at a height of 4 inches (10 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux (upward flow). No smoke shall escape from the cabinet.
102458822 AIRFLOW SMOKE PATTERN TESTS
1. DOWNFLOW TEST:
PASS FAIL
Comments: NA
A smoke source shall be passed from one end of the cabinet to the other, 1 inch (2.5 cm) behind the viewscreen, at a height of 6 inches (15 cm) above the top of the access opening. The smoke shall show smooth downward flow with no dead spots or reflux. No smoke shall escape from the cabinet.
PASS FAIL MEETS MANUFACTURER'S SPECIFICATIONS
3. WORK OPENING EDGE RETENTION TEST:
Comments: NA
A smoke source shall be passed along the entire perimeter of the work opening edges, approximately 1.5'' (3.8 cm) outside the cabinet. Particular attention should be paid to corners and vertical edges. No smoke shall reflux out of the cabinet once drawn in, nor shall smoke billow over the work surface or penetrate onto it.
PASS FAIL
Comments: NA
PASS FAIL
4. SASH/WINDOW SEAL TEST:
A smoke source shall be passed up the inside of the window 2'' (5 cm) from the sides and along the top of the cabinet interior work area. No smoke shall escape out of the cabinet.
Comments: NA
1-18 x 48 x 3.54 P/N: 41978
2-11.15 x 22.08 x 8 P/N: 42061
Supply HEPA Filter(s) Size:
Exhaust HEPA Filter(s) Size:
HEPA Filter(s) Replaced:
Prefilter(s) Replaced:
U.V. Lamp(s) Replaced:
REMARKS
No
No
NA
2. VIEW SCREEN RETENTION TEST:
MK41978
MK372G002
Pre Filter(s) Size:
B2DMINT Rev. 05MAR2016 R1.0Copyright © 2006 - All rights reserved
File details come from the government source that posted it. Updated .