Item Description.pdf
PDF 193 KB Posted
- Attached to
- Niton XL5 Plus or Equal Handheld XRF Analyzer
- Solicitation number
- FA812626Q0056
About this file
This is a Memorandum for Record documenting equipment specifications for a federal procurement action, M26-110, issued by the Department of the Air Force Headquarters Air Force Sustainment Center (AFMC) at Tinker Air Force Base, Oklahoma on 12 March 2026.
The 76 MXSG is requesting funds to procure one Niton XL5 Plus handheld XRF (X-ray fluorescence) analyzer or equivalent for alloy analysis. The instrument must analyze 80 different elements ranging from Magnesium (Mg #12) through Uranium (U #92) on the Periodic table, with enhanced performance capabilities for light element analysis (Mg, Al, Si, P, and S) without requiring helium purge or vacuum systems. The analysis time must be approximately 1-2 seconds for most applications. The X-ray tube specifications require a minimum of 50kV and 200 µA output with a silver anode. The equipment must include the latest manufacturer design and safety features in compliance with the National Electrical Code (NEC). The analyzer must be capable of connecting to a computer for downloading scans and results, and computer software for managing scans and results must be included with the equipment.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Item Description.pdf | ||
| Solicitation.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
HEADQUARTERS AIR FORCE SUSTAINMENT CENTER (AFMC)
TINKER AIR FORCE BASE OKLAHOMA
12 March 2026
MEMORANDUM FOR RECORD
SUBJECT: Item Description, M26-110 Handheld XRF Analyzer
1. The 76 MXSG is requesting utilization of funds for the purchase of a Handheld or Equal XRF Analyzer.
2. The requested equipment descriptions/specifications are outlined below.
a. One (1) Portable or Equal XRF Spectrometer for alloy analysis, including the latest manufacturer design and safety features per the National Electrical Code (NEC).
(1) Salient Characteristics:
Must analyze elements from Magnesium (Mg) through Uranium (U) on the Periodic table. (That is 80 different elements. )
Must provide enhanced performance for light element analysis (Mg, Al, Si, P, and S) without the use of a helium purge or vacuum.
Analysis time should be approximately 1-2 seconds for most applications.
(d) The X-ray tube must be at least 50kV, 200 µA with a silver anode.
The following specifics pertain to the item listed in paragraph 2:
The instrument must be able to connect to a computer to download scans and results.
Computer software for managing scans and results should be included.