Solicitation PR F3YCFH5275A002 Contractor Questions and Government Responses

Contractor Question 1:

Do you require 2 or 4 ports?

Government Response:

Only two ports are required.

Contractor Question 2:

Do you want an automated calibration kit or manual?

The Calibration kit can be automated or manual.

Contractor Question 3:

Network Analyzer use test port cables but none were listed - can you share with me if they were just mistakenly not included or just not needed?

The Test Port Cables will be purchased under a separate acquisition.

Contractor Question 4:

What connector type is needed?

Government Response:

The connector type is dependent on the test port cables which will be purchased under a separate acquisition.

Contractor Question 5:

Are bi-directional S-parameters (S11, S21, S12 and S22) measurements required or uni-directional S-parameters (S11 and S21 only)?

Bi-directional S-parameters are required.

Contractor Question 6:

Is a display and/or control panel required? Or would the unit mainly be used remotely?

Government Response:

Display or control panel doesn't have to be built-in if the network analyzer can be controlled from a Windows based PC.

Contractor Question 7:

How large is the typical (expected) VSWR and what accuracy is expected for the reflection measurement?

Government Response:

VSWR tolerances range from 1 to 8 for various tests. The network analyzer needs to exceed this range.

Accuracy should be +/- 0.1 or better.

Contractor Question 8:

How much loss do you have in your system and what accuracy do you need for the transmission measurement?

We need accuracy of 0.01 dB or better.

Contractor Question9:

What is the required dynamic range?

100 dB

Contractor Question 10:

Is measurement speed important?

Yes. We need 0.5 ms/point or better.