Contractor Questions with the Government Responses
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Tech_Eval_Criteria_-_Network_Analyzers_Revision_1_(13-November_2015).docx | DOCX document | |
| REVISION_1_Questions_and_Answers_-_Solicitation_FA8126-16-Q-0022-0001_Network_Analyzers__(13-November_2015)_.pdf | ||
| Amendment_-_Solicitation_FA8126-16-Q-0022-0001_Network_Analyzers.doc | DOC document | |
| SOW_Network_Analyzers_Revision_1_(13-November_2015).docx | DOCX document | |
| SOLICITATION_FA8126-16-Q-0022.doc | DOC document | |
| SOW_Network_Analyzers.docx | DOCX document | |
| Tech_Eval_Criteria_-_MB-2016-0037_Network_Analyzers.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Solicitation PR F3YCFH5275A002 Contractor Questions and Government Responses
Contractor Question 1:
Do you require 2 or 4 ports?
Government Response:
Only two ports are required.
Contractor Question 2:
Do you want an automated calibration kit or manual?
The Calibration kit can be automated or manual.
Contractor Question 3:
Network Analyzer use test port cables but none were listed - can you share with me if they were just mistakenly not included or just not needed?
The Test Port Cables will be purchased under a separate acquisition.
Contractor Question 4:
What connector type is needed?
Government Response:
The connector type is dependent on the test port cables which will be purchased under a separate acquisition.
Contractor Question 5:
Are bi-directional S-parameters (S11, S21, S12 and S22) measurements required or uni-directional S-parameters (S11 and S21 only)?
Bi-directional S-parameters are required.
Contractor Question 6:
Is a display and/or control panel required? Or would the unit mainly be used remotely?
Government Response:
Display or control panel doesn't have to be built-in if the network analyzer can be controlled from a Windows based PC.
Contractor Question 7:
How large is the typical (expected) VSWR and what accuracy is expected for the reflection measurement?
Government Response:
VSWR tolerances range from 1 to 8 for various tests. The network analyzer needs to exceed this range.
Accuracy should be +/- 0.1 or better.
Contractor Question 8:
How much loss do you have in your system and what accuracy do you need for the transmission measurement?
We need accuracy of 0.01 dB or better.
Contractor Question9:
What is the required dynamic range?
100 dB
Contractor Question 10:
Is measurement speed important?
Yes. We need 0.5 ms/point or better.