Exhibit D Analyzer, Calibrator and Training Specifications.pdf

PDF 142 KB Posted

Attached to
Air Monitoring Instrumentation, Replacement Parts and Servic
Solicitation number
RFP FAAA 2026000164 1
Issued by
Denver County, Denver City, Colorado

About this file

This document is a specification exhibit for the Colorado Department of Public Health & Environment Air Pollution Control Division outlining equipment, parts, repairs, and training requirements for a procurement opportunity. The specification details two categories of equipment: FRM/FEM-designated analyzers and calibrators for ambient air quality monitoring, including ozone analyzers, carbon monoxide analyzers, nitrogen oxide analyzers, sulfur dioxide analyzers, and particulate matter analyzers (PM10 and PM2.5). Additional required equipment includes zero air generators at both standard and trace levels, ozone calibrators, multi-gas dynamic dilution calibrators, ambient filter samplers, a multi-port carbonyl sampler, and a solar-powered remote air quality monitoring platform with meteorological sensors. All equipment must meet Federal Reference Method or Federal Equivalency Method standards and be listed on the EPA's most recently published list of designated methods as of December 15, 2025.

Bidders must demonstrate the capability to supply available replacement parts specific to each manufacturer's equipment and must possess full in-house repair capability for all equipment being bid. The specification mandates comprehensive factory-level training including hands-on instruction for complete equipment operation, routine maintenance procedures (weekly, monthly, quarterly, and annual), troubleshooting and diagnosis of equipment faults, and repair of all equipment components. All training must include reference documentation and result in certification documentation for participants. The document does not specify pricing terms, quantities, delivery dates, contract duration, or funding amounts.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Exhibit D

SUMMARY OF ANALYZER, CALIBRATOR and TRAINING SPECIFICATIONS

FOR

COLORADO DEPARTMENT OF PUBLIC HEALTH & ENVIRONMENT

AIR POLLUTION CONTROL DIVISION

The items on the following list must have a Federal Reference Method (FRM) or Federal Equivalency Method

(FEM) designation, and be listed on the EPA’s List of Designated Reference and Equivalent Methods, most recently published on December 15, 2025. Equipment must be able to operate in a quality control environment that meets FEM standards.

FRM/FEM Equipment:

1. Ambient UV Adsorption Ozone Analyzer

2. Ambient Chemiluminescence Ozone Analyzer

3. Ambient Trace Level Carbon Monoxide Analyzer

4. Ambient Chemiluminescence Oxides of Nitrogen (NO/NO2/NOx) Analyzer

5. Ambient Trace Level Chemiluminescence Oxides of Nitrogen (NO/NO2/NOx) Analyzer

6. Ambient Trace Level “True” Nitrogen Dioxide (NO2) Analyzer with or without NO

7. Ambient Trace Level Total Oxides of Nitrogen (NO/NOy-NO/NOy) Analyzer

8. Ambient Trace Level Sulfur Dioxide Analyzer

9. Ambient Continuous PM10, PM2.5, PM10-2.5 Analyzer

10. Zero Air Generator meeting the following requirements:

a. less than 0.1 ppb SO2

b. less than 0.1 ppb H2S

c. less than 0.1 ppb NO

d. less than 0.1 ppb NO2

e. less than 0.1 ppb NH3

f. less than 0.4 ppb O3

g. less than 20 ppb CO

h. less than 5 ppb Hydrocarbons

11. Trace Level Zero Air Generator meeting the following requirements:

a. less than 0.025 ppb SO2

b. less than 0.025 ppb H2S

c. less than 0.025 ppb NO

d. less than 0.025 ppb NO2

e. less than 0.025 ppb NH3

f. less than 0.30 ppb O3

g. less than 10 ppb CO

h. less than 0.25 ppb Hydrocarbons

12. Ozone Calibrator

13. Trace Level Ozone Calibrator

14. Multi-Gas Dynamic Dilution Calibrator without Photometer (2 MFC) https://www.epa.gov/system/files/documents/2025-12/amtic-list-december-2025_508-compliant.pdf

Exhibit D

15. Multi-Gas Dynamic Dilution Calibrator with Photometer (2 MFC)

16. Trace Level Multi-Gas Dynamic Dilution Calibrator (3 MFC)

17. Ambient filter samplers (with FRM designation) for PM10 and PM2.5

Other Equipment:

1. Multi-port carbonyl sampler meeting PAMS sampling specifications.

2. Solar-powered, remote air quality monitoring platform offering meteorological monitoring, cellular service with online data collection/storage, and sensors, including, but not limited to: CO, NO, NO2, O3, SO2, PM10, and PM2.5.

3. Portable, battery-powered, mass flow controllers and mass flow meters designed for remote use in a quality control environment, at a variety of different ranges.

Parts:

Available replacement parts specific to each manufacturer’s piece of equipment being bid on.

Repairs:

Bidders must be fully capable of repairing their own equipment, in their own facility, when the need for such repairs arises.

Training:

1. Factory Level Operation, Maintenance, and Advanced Troubleshooting and Repair Training to include at a minimum:

a. Hands-on training for the complete operation of all provided equipment, to include reference documentation

b. Hands-on training for all routine (weekly, monthly, quarterly, and annual) maintenance procedures, to include reference documentation

c. Hands-on training for the troubleshooting and diagnosis of common and difficult equipment faults, to include reference documentation

d. Hands-on training for the repair of all equipment components, to include reference documentation

e. Certification documentation for all above mentioned training https://www.epa.gov/system/files/documents/2023-05/TAD%20R3%20May%202023.pdf