Exhibit B Perf Requirements Specs.pdf
PDF 204 KB Posted
- Attached to
- Nitrogen Generators
- Solicitation number
- DOH25-014P
- Issued by
- Adams County, Asotin County, Benton County, Chelan County, Clallam County, Clark County, Columbia County, Cowlitz County, Douglas County, Ferry County, Franklin County, Garfield County, Grant County, Grays Harbor County, Island County, Jefferson County, King County, Kitsap County, Kittitas County, Klickitat County, Lewis County, Lincoln County, Mason County, Okanogan County, Pacific County, Pend Oreille County, Pierce County, San Juan County, Skagit County, Skamania County, Snohomish County, Spokane County, Stevens County, Thurston County, Wahkiakum County, Walla Walla County, Whatcom County, Whitman County, Yakima County, Asotin City, Clarkston City, Clarkston Heights-Vineland CDP, West Clarkston-Highland CDP, Benton City, Chelan City, Chelan Falls CDP, Clallam Bay CDP, Lewisville CDP, Rock Island City, Pacific Beach CDP, Whidbey Island Station CDP, Mercer Island City, Pacific City, Bainbridge Island City, Kingston CDP, Kitsap Lake CDP, Kittitas City, Klickitat CDP, Okanogan City, Anderson Island CDP, Fort Lewis CDP, Fox Island CDP, Herron Island CDP, Ketron Island CDP, North Fort Lewis CDP, Raft Island CDP, Stevenson City, Hat Island CDP, Lake Stevens City, Snohomish City, Spokane City, Spokane Valley City, Puget Island CDP, Garfield Town, Yakima City, Washington
About this file
This is an Exhibit B Performance Requirements/Specifications document for the Washington Department of Health (DOH) procurement of nitrogen generators. The document outlines 14 mandatory pass/fail requirements for bidders, including specifications for self-contained compressor nitrogen generator units with a minimum quantity of 1 unit up to 4 units. Key technical requirements include nitrogen purity greater than 95%, output pressure of 100 PSI or greater, capacity of 3500+ liters per hour, specific dimensional constraints (height ≤66 inches, width ≤26", depth ≤26"), and noise level ≤62 db. Installation, setup, calibration, and performance verification must be performed at PHL at no cost to DOH.
The specifications require bidders to provide a minimum one-year warranty and service response within two weeks of request, with customer support available within 48 hours. The system must use specific electrical requirements (NEMA L6-15 receptacle with 208/240 volts or NEMA L6-30 receptacle with 208 volts, 60 Hz frequency) and be portable with wheels. Bidders must indicate whether they offer annual service agreements and post-maintenance options with multi-year cost lock-in pricing, though this is not marked as a mandatory requirement.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
EXHIBIT B - PERFORMANCE REQUIREMENTS/SPECIFICATIONS
INSTRUCTIONS: Bidders submitting a proposal must complete and submit an Exhibit B for evaluation purposes.
1. Review all requirements, priorities and provided definitions:
• Mandatory Pass/Fail (M): Minimum requirement; Bidder that does not meet this requirement will not be considered any further.
# Item Description Yes No Notes 1 Nitrogen generator Self contained compressor nitrogen generator unit. Minimum of 1 unit, up to 4
[ ] Yes [ ] No (fail)
2 Installation requirements Installation, Setup, Calibration, and
Performance Verification shall be performed at PHL at no cost to the DOH.
[ ] Yes [ ] No (fail)
3 Self contained nitrogen generator.
The nitrogen generator must be self-contained (compressor units included in the nitrogen generator system).
[ ] Yes [ ] No (fail)
4 Nitrogen purity Oil free nitrogen, with a purity of greater than 95% nitrogen. Nitrogen must be suitable for operation of mass spectrometry instrumentation.
[ ] Yes [ ] No (fail)
5 Supplied pressure
Out put pressure must be 100 PSI or greater.
[ ] Yes [ ] No (fail)
6 Capacity The Nitrogen generator must supply 3500 (or greater) liters per hour.
[ ] Yes [ ] No (fail)
7 Fittings The nitrogen generator must supply multiple (two or more) 1/4" BSPP fittings. Or a single fitting for which the flow can be split to multiple instruments.
[ ] Yes [ ] No (fail)
8 Dimensions Height <= 66 inches, width <=26", depth <=26"
[ ] Yes [ ] No (fail)
9 Noise level Noise level of <= 62 bd
[ ] Yes [ ] No (fail)
10 Electrical requirments Must use a NEMA L6-15 receptacle and run off either 208 or 240 volts. Or use a NEMA L6-30 receptacle and run off 208 volts.
Voltage frequency 60 Hz. Necessary power cables to be provided by vendor.
[ ] Yes [ ] No (fail)
11 Porability Must be mounted on wheels for portability of unit and not require permanet pluming for waste.
[ ] Yes [ ] No (fail)
12 Warranty The Vendor must be able to provide an warranty for one (1) year minimum.
[ ] Yes [ ] No (fail)
13 Service.
The vendor must be able to provide service as needed within two (2) weeks of request.
Vendor must provide customer support, on call within 48 hours from knowledgeable staff to reduce any down time and insure maximum productivity.
[ ] Yes [ ] No (fail)
14 Service Agreements Are there annual service agreement and
Post Maintenance options? If so do you offer multi-year costs lock-in pricing?
Please list the options and pricing for each.
[ ] Yes [ ] No
SPECIFICATIONS: Bidder shall complete all lines of the Specification. Failure to complete each line may result in bid disqualification.
Information for any non-yes/no response can be included on additional pages when there is insufficient room provided for your response.
General System Requirements The required technical features of the system include: