This document details specifications for the annual calibration and preventative maintenance of Landtec GEM 5000 Portable Gas Analyzers. The specifications outline a comprehensive service process that includes instrument inspection, calibration across multiple gas measurements (methane, carbon dioxide, oxygen, carbon monoxide, and hydrogen sulfide), and potential component replacements. The service will cover atmospheric pressure and gas flow calculations, with additional provisions for oxygen cell programming, water trap kit adjustments, firmware updates, and replacement of various instrument components as needed.

The specifications require that all calibration be performed according to manufacturer's protocols using traceable equipment that meets Landtec's calibration specifications. Each calibrated instrument must be accompanied by a unique and signed Statement of Calibration. In cases where mechanical issues are discovered during inspection that might prevent successful calibration, the service provider is required to provide a written quotation to the customer before proceeding with any repairs. The document emphasizes a thorough and transparent approach to maintaining the accuracy and functionality of the gas analyzers, with return freight included in the service.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers

MINIMUM SPECIFICAITONS

Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Portable Gas Analyzers against Landtec/QED Environmental Systems equipment or equivalent.

This calibration and assessment process shall include the following:

1. Instrument Inspection

2. Instrument Calibration:

• Methane: 0-100% (volume)

• Carbon Dioxide: 0-100% (volume)

• Oxygen: 0-25% (volume)

• Carbon Monoxide: 0-2000 parts per million

• Hydrogen Sulfide: 0-1000 parts per million

• Atmospheric pressure, differential pressure and calculates gas flow

3. Oxygen cell programming (as needed)

4. Water trap kit (as needed)

5. Firmware Updates (as needed)

6. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)

7. Return freight

8. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.

9. A unique and signed Statement of Calibration must accompany each newly calibrated

Landtec GEM 5000 instrument.

10. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.

Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers