This document presents specifications for the annual calibration of Landtec GEM 5000 Plus Portable Gas Analyzers, targeting equipment maintenance through LandTech/QED Environmental Systems or an equivalent service provider. The specifications outline a comprehensive calibration and preventative maintenance process to be performed annually, which includes a detailed instrument inspection and calibration for multiple gas measurements including methane, carbon dioxide, oxygen, carbon monoxide, and hydrogen sulfide, with specific measurement ranges for each gas type.
The calibration process requires adherence to manufacturer protocols using traceable equipment that meets specified calibration standards. Each calibrated instrument must be accompanied by a unique, signed Statement of Calibration. The service includes potential replacement of various instrument components such as sensors, transducers, pump, keypad, lens, tubing, USB cable, and battery as needed. If any mechanical issues are discovered during inspection that could prevent successful calibration, the service provider is required to provide a written quotation to the customer before proceeding with instrument servicing.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers
MINIMUM SPECIFICAITONS
Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers
Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Plus Portable Gas Analyzers against LandTech/QED Environmental Systems equipment or equivalent. This calibration and assessment process shall include the following:
1. Instrument Inspection
2. Instrument Calibration:
• Methane: 0-100% (volume)
• Carbon Dioxide: 0-100% (volume)
• Oxygen: 0-25% (volume)
• Carbon Monoxide: 0-2000 parts per million
• Hydrogen Sulfide: 0-1000 parts per million
• Atmospheric pressure, differential pressure and calculates gas flow
3. Oxygen cell programming (as needed)
4. Firmware Updates (as needed)
5. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)
6. Return freight
7. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.
8. A unique and signed Statement of Calibration must accompany each newly calibrated Landtec GEM 5000 Plus instrument.
9. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.
Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers