ESBD_454755_1758136939578_GEM_5000_Spec.pdf
PDF 108 KB Posted
- Attached to
- GEM Calibrations
- Solicitation number
- 582-26-10358
- Issued by
- Texas
About this file
This document details specifications for annual factory calibration and preventative maintenance of Landtec GEM 5000 Portable Gas Analyzers. The specifications outline a comprehensive calibration process to be performed against Landtec/QED Environmental Systems equipment or equivalent. The calibration requirements include a detailed instrument inspection and calibration across multiple gas measurements: methane (0-100% volume), carbon dioxide (0-100% volume), oxygen (0-25% volume), carbon monoxide (0-2000 parts per million), and hydrogen sulfide (0-1000 parts per million).
The specifications further require calibration according to manufacturer's protocols using traceable equipment, with provisions for potential component replacements such as sensors, transducers, pump, keypad, lens, tubing, USB cable, and battery. Additional requirements include oxygen cell programming, water trap kit adjustments, firmware updates, and return freight. A unique and signed Statement of Calibration must accompany each calibrated instrument, and if any mechanical issues are discovered during inspection that could impede calibration, a written quotation must be provided to the customer prior to servicing the instrument.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers
MINIMUM SPECIFICAITONS
Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Portable Gas Analyzers against Landtec/QED Environmental Systems equipment or equivalent.
This calibration and assessment process shall include the following:
1. Instrument Inspection
2. Instrument Calibration:
• Methane: 0-100% (volume)
• Carbon Dioxide: 0-100% (volume)
• Oxygen: 0-25% (volume)
• Carbon Monoxide: 0-2000 parts per million
• Hydrogen Sulfide: 0-1000 parts per million
• Atmospheric pressure, differential pressure and calculates gas flow
3. Oxygen cell programming (as needed)
4. Water trap kit (as needed)
5. Firmware Updates (as needed)
6. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)
7. Return freight
8. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.
9. A unique and signed Statement of Calibration must accompany each newly calibrated
Landtec GEM 5000 instrument.
10. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.
Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers
File details come from the government source that posted it. Updated .