ESBD_406910_1743614238673_582-25-08663 Attachment A.pdf
PDF 83 KB Posted
- Attached to
- Two Channel Canister Samplers
- Solicitation number
- 582-25-08663
- Issued by
- Travis County, Texas
About this file
This document is a solicitation for a Dual Channel Canister Sampler issued by the Texas Commission on Environmental Quality (TCEQ) under Solicitation Number 582-25-08663. The solicitation seeks a microprocessor-controlled dual-channel air sampler capable of collecting whole ambient air samples into evacuated six-liter canisters over a 24-hour period. The sampler must meet specific technical requirements, including a color touchscreen display, two independent collection channels, independent mass flow controllers, EPA Method TO-15 cleanliness standards, and programmable sampling capabilities. The routine delivery response is set at 120 calendar days after receipt of order.
Additional terms and conditions include requirements for new equipment with manufacturer's standard parts and accessories, a minimum two-year warranty covering onsite system, parts, and labor, and an electronic manual with comprehensive operating instructions. Technical support must be available Monday through Friday, 8 a.m. to 5 p.m. Central Standard Time, via telephone and email. The sampler's specifications are detailed, requiring specific technical components such as microprocessor-based digital circuitry, analog outputs for canister pressures and flowrates, specific power requirements, and sample inlet/outlet configurations with fused silica lined Swagelok fittings.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Solicitation Name: Dual Channel Canister Sampler Solicitation Number: 582-25-08663
Attachment A Specifications for a Dual Channel Canister Sampler
Description: Microprocessor controlled dual-channel sampler to collect whole ambient air samples into evacuated (negative pressure) canisters. The unit must be capable of collecting time-integrated samples into a sub ambient pressure six-liter canister over a 24-hour period.
MINIMUM GENERAL SPECIFICATIONS
1. The canister sampler must meet the following requirements:
A. The system must contain the following components:
i. Microprocessor based digital circuitry.
ii. A SVGA color touchscreen display.
iii. Two independent collection channels.
iv. Independent mass flow controllers for each channel.
v. Analog outputs of 0-5 volts for canister pressures and flowrates.
vi. A 0-15 pound per square inch pressure gauge with 0-5-volt output.
vii. Alternating current (AC) power: 115 volts, 60 hertz (Hz).
viii. Rear located sample inlet with ¼-inch fused silica lined Swagelok fittings, shipped in a manner to maintain cleanliness (capped).
ix. Two front located sample outlets with 1/8-inch fused silica lined Swagelok fittings, shipped in a manner to maintain cleanliness (capped).
B. Each of the two sampler channels must meet U.S. Environmental Protection Agency (EPA) Method TO-15 requirements for cleanliness (less than 0.2 parts per billion by volume) for all measured compounds using a humid (relative humidity >30%) air or nitrogen matrix. Documentation of the analytical verification will be provided with each sampler.
C. Programmable for a recurring automated 24-hour sample collection over 6-day collection cycle.
D. Active sample line purge prior to sample collection.
E. Manual leak check during setup.
F. Automated leak check prior to sample collection.
G. Canister flow rate and pressure logging during sampling.
H. Software calibration of mass flow controllers and pressure transducers.
I. On screen operating procedures.
J. Real-time editing for programmable parameters.
ADDITIONAL TERMS AND CONDITIONS
1. All items must be bid new, of current production, and include the manufacturer’s standard parts and accessories.
2. Routine delivery response to TCEQ will be within 120 calendar days after receipt of order.
Solicitation Name: Dual Channel Canister Sampler Solicitation Number: 582-25-08663
3. The warranty period should consist of complete coverage of system onsite, parts, and labor for a minimum of two years from the date of acceptance.
4. At least one electronic manual must be supplied with the samplers. The manual will include complete operating instructions, circuit descriptions, theory of operation, location of controls and switches, software commands, maintenance, and service instructions. The manual will also include schematic diagrams, pneumatic flow charts, lists of commonly replaced parts, warnings, and safety precautions.
5. Technical support must be available between the hours of 8 a.m. and 5 p.m. Central Standard Time (CST) during regular business days, Monday through Friday, by telephone and electronic mail (e-mail).
Specifications for a Dual Channel Canister Sampler
File details come from the government source that posted it. Updated .