CUTSHEET Serta Room Pressure Monitor Cutsheet.pdf

PDF 214 KB Posted

Attached to
Z1DA--Installation of Additional Environmental Monitors Federal contract opportunity
Solicitation number
36C25621R0131
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16

About this file

This document includes a product cutsheet for the Setra Systems SRPM Room Pressure Monitor and a presolicitation notice for the installation of additional environmental monitors at the Southeast Louisiana Veterans Health Care System.

The Setra Systems SRPM Room Pressure Monitor cutsheet provides specifications for a BACnet capable differential pressure sensor for measuring pressure relationships in critical environments like protection and isolation rooms. The touchscreen monitor can switch between room modes and features configurable alarms, password protection, and one-time zero calibration for ease of installation and use. The presolicitation notice seeks proposals from service-disabled veteran-owned small businesses to install additional environmental monitors at the Southeast Louisiana Veterans Health Care System in New Orleans for between $500,000 to $1,000,000. Interested offerors must meet SDVOSB eligibility requirements and the prime contractor must perform a minimum of 15-25% of the contract cost using its own employees or other eligible SDVOSB employees. The solicitation will be issued on June 4, 2021 on beta.SAM.gov and questions can be submitted to the contracting officer prior to the scheduled site visit.

View the file

Other files for this federal contract opportunity

Other files attached to Z1DA--Installation of Additional Environmental Monitors, newest first.
File Type Posted
S04 Site Visit Sign-In Sheet (06.17.2021).pdf PDF
36C25621R0131 0001.docx DOCX document
EXHIBIT C SUBCONTRACTOR INFORMATION AND CONSENT.docx DOCX document
MANUAL Setra Installation and Operating Manual.pdf PDF
PHOTO 4CC01A.jpg JPG image
PHOTO 4BC02.jpg JPG image
DRAWING 1AI102 M1 IP OVERALL SECOND FLOOR PLAN.pdf PDF
DRAWING 1A104-QUESTION MARK M3 IP OVERALL FOURTH FLOOR PLAN.pdf PDF
26 05 11 Requirements for Electrical Installations.pdf PDF
01 45 00 Quality Control.pdf PDF
GENERAL DECISION NUMBER LA20210041.pdf PDF
36C25621R0131 SOL.docx DOCX document
EXHIBIT B PAST AND PRESENT PERFORMANCE QUESTIONNAIRE.docx DOCX document
DRAWING 1AI104-B2 IP Fourth Wall Types Building B2.pdf PDF
CHART M11 Environmental Monitor Room Location List.pdf PDF
26 05 19 Low Voltage Electrical Power Conductors and Cables 600 Volts and Below.pdf PDF
23 05 11 Common Work Results for HVAC.pdf PDF
09 91 00 Painting.pdf PDF
02 41 00 Demolition.pdf PDF
01 33 23 Shop Drawings Product Data and Samples.pdf PDF
01 00 00 General Requirements.pdf PDF
00 01 10 Table of Contents.pdf PDF
SOW (Final).pdf PDF
EXHIBIT G LIST OF RETURNABLES.docx DOCX document
EXHIBIT F REQUEST FOR INFORMATION FORM.docx DOCX document
EXHIBIT A PERFORMANCE RELEVANCY SURVEY.docx DOCX document
DRAWING 2AI104-D2 M4 DNT FOURTH FLOOR PLAN - AREA D1 AND D2.pdf PDF
DRAWING 2AI103-D2 M5 DNT THIRD FLOOR PLAN - AREA D1 AND D2.pdf PDF
DRAWING 2AI102-G2 M8 DNT SECOND FLOOR PLAN - AREA G1 AND G2.pdf PDF
DRAWING 1AI103-QUESTION MARK M2 IP OVERALL THIRD FLOOR PLAN.pdf PDF
26 05 33 Raceway and Boxes for Electrical Systems.pdf PDF
23 09 23 Direct Digital Control System For HVAC.pdf PDF
01 42 19 Reference Standards.pdf PDF
Justification and Approval (Brand Name).pdf PDF
DRAWING 2AI104-E2 M7 DNT FOURTH FLOOR PLAN - AREA E1 AND E2.pdf PDF
EXHIBIT E SURETY FORM.docx DOCX document
EXHIBIT D CONTRACTOR EVALUATION FORM CONSTRUCTION SAFETY.docx DOCX document
JCI Temperature Humidity Sensor Cutsheet.pdf PDF
DRAWING 8AS103 M9 RESEARCH THIRD FLOOR PLAN.pdf PDF
DRAWING 2AI102-F2 M10 DNT SECOND FLOOR PLAN - AREA F1 AND F2.pdf PDF
DRAWING 1AI104-C1 IP Fourth Wall Types Building C1.pdf PDF
DRAWING 1AI104-B1 IP Fourth Wall Types Building B1.pdf PDF
2AI102-E2 M6 DNT SECOND FLOOR PLAN -AREA E1 AND E2.pdf PDF
09 29 00 Gypsum Board.pdf PDF
07 84 00 Firestopping.pdf PDF
01 35 26 Safety Requirements.pdf PDF
ICRA ILSM.xlsx XLSX spreadsheet
Show all 47

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.

Model SRPM Room Pressure Monitor

The SRPM is Setra’s standard single room BACnet capable room pressure monitor for measuring low differential pres-sure in critical applications. The SRPM’s backlit touchscreen LCD provides an intuitive graphic user interface for ease of setup. The SRPM has a built-in calibration feature and only requires zeroing when installed, significantly reducing the cost of ownership. The SRPM monitors and alarms while pro-viding a digital input for a door alarm. The SRPM is a simple, cost-effective solution which combines state-of-the-art elec-tronics with Setra’s superior true differential pressure sens-ing technology to ensure safety in critical environments. The SRPM also incorporates two-level password protection.

Monitor & Alarm Critical Rooms The SRPM is most user friendly room pressure monitor on the market today. It has an intuitive touchscreen interface that allows the user to easily configure alarm set points, passwords and audible alarming conditions. With its bi-di-rectional sensor, the unit can switch between protection and isolation room modes, or be put into standby mode when the room is not in use.

On-Board Dead-Ended Pressure Sensor Protection and isolation rooms are designed to adhere to strict standards in order to provide a proper barrier between the room and reference space. Unlike a flow-through design, the SRPM utilizes an on-board dead-ended low differential pressure sensor. This technology provides the user with a trusted solution & peace of mind that the sensor will prevent contaminated air from passing through it.

Save Time and Money on Installation & Calibration The SRPM is designed with both the installer and end user in mind. The BACnet enabled unit can be installed in an off-the-shelf electrical box, improving the ease of installation instead of having to use a custom electrical box that is not typically available at the rough stage of the project. The SRPM offers push button zero and span calibration that is easily per-formed by any low differential pressure calibrator and can be calibrated in minutes.

n Maximize Patient Safety n Save on Installation Costs n Low-Cost BACnet Solution

Model SRPM Features:

• On-board Sensor - Industry Best Accuracy

• LCD Touch Screen for Easy Setup and Room Display

• Monitor Single Pressure Relationship and Door Status

• Configurable Audible & Visual Alarms to Avoid Nuisance

• Easy Surface Mounting - Wall Thickness is Irrelevant

• Increased Safety with 2 Layer Password Protection

• Calibration: Only Requires Zeroing Once Installed

Where We’re Installed:

• Brigham and Women’s Hospital

• Emory University Medical Center

• Memorial Sloan Kettering Cancer Center

• Stanford University Medical Center

• Veterans Affairs (VA) Medical Center

Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.

Model SRPM Room Pressure Monitor

SSP-SRPM REV. 12/13

DIMENSIONS GENERAL SPECIFICATIONS

ORDERING INFORMATION

Model Range Code Excitation/Output Accuracy

SRPM = SRPM

RANGE CODE INCHES W.C.

A1 24 VAC/4-20 mA or 0-5 and 0-10 VDC

E ±0.5% FS

005WB ±5

V1 120/240 VAC/4-20 mA or 0-5 and 0-10 VDC

V ±0.25% FS

2R5WB ±2.5 A2 24 VAC w/ BACnet®

001WB ±1.0 V2 120/240VAC BACnet®

0R5WB ±0.50

R25WB ±0.25

0R1WB ±0.1

S R P M

Ordering Example: Part No. SRPM005WBA1E for a SRPM, ±5 in. W.C. Range, 24 VAC EXC. with 4 to 20 mA output, and ±0.5% FS Accuracy.* Please contact factory for versions not shown..

Performance Data Environmental Data Standard Optional Operating Temp.3 °F (°C) 32 to +120 (0 to +50)

Accuracy RSS1 ±0.5% ±0.25% Storage Temp. °F (°C) -20 to +160 (-30 to +170)

Non-Linearity (BFSL) ±0.49% ±0.24% Operating Humidity 5 to 95% RH (Non-Condensing)

Hysteresis ±0.05% ±0.05% Electrical Data (Voltage) Non-Repeatability ±0.05% ±0.05% Circuit 3-Wire (Exc, Out, Com)

Span Setting Tol.5 ±0.5% Rdg. ±0.5% Rdg. Output4 0 to 5 VDC, 0 to 10 VDC

Thermal Effects2 Alarm Output SPDT Relay: 1A @ 24 VDC, 1A @ 120 VDC

Compensated Range (oFoC) 40 to 120 (4.5 to 50) Power Consumption 5W

Zero/Span Shift %FS ±0.03% FS (±0.05%FS) Excitation:

Code V1 Code A1 Code V2 Code A2

85-265 VAC, 50-60 Hz 18-32 VAC, 50-60 Hz 85-265 VAC, BACnet® 18-32 VAC, BACnet®

Overpressure ±15”W.C.

Physical Description

Case Fire-Retardant Plastic (NEMA1, IP20 Rated for Indoor Applications)

Electrical Data (Current)

Dimensions 8”H x 5.1”W x 1.8”D (203 x 130 x 46 mm)

Circuit 2-Wire

Electrical Connection Removable Terminal Block Output 4 to 20 mA

Pressure Fittings Barbed Fittings 1/4” O.D.

Tubing

External Load 0 to 510 ohms

Weight (approx.) 1.5lbs (680g) Excitation:

Code Vi:

Code A1

85-265 VAC, 50-60 Hz1 8-32 VAC, 50-60 Hz

5.1

1.8

8.0

2.17

4.61 in.

mm

"BACnet" is a registered trademark of ASHRAE"

1RSS of Non-Linearity, Hystereis, and Non-Repeatability.

2Units calibrated at nominal 70oF. Max thermal error computer from this datum.

3Operating temperature limits of the electronics only.

4Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater 5 Zero setting tol. negated by zero push button Specifications subject to change.

File details come from the government source that posted it. Updated .