The certification form requires documentation of booth specifications including location, manufacturer, serial number, wall configuration, operational settings (lights and fan status), and equipment details (sound level meter, microphone, octave band filter, and calibrator with calibration dates). The form establishes maximum permissible sound pressure levels (SPL) in decibels across seven octave band center frequencies (125 Hz through 8000 Hz) for three distinct types of audiometric testing: Medical Surveillance Testing with ears covered, Diagnostic Audiology Testing with ears covered using headphones or insert phones, and Diagnostic Audiology Testing with ears not covered using sound field or bone conduction testing. Each testing type has progressively lower SPL limits, requiring increasingly quieter booth environments. Certifiers must document pre- and post-field calibration results, measure SPL inside and outside the booth under normal operational conditions, and verify that all equipment has been professionally calibrated within one year. Annual recertification is mandatory, with recertification also required following any significant new noise sources or booth relocation. The procedures specify use of Type I sound level meters capable of measuring at least 3 dB below maximum SPL criteria, measurement in linear/all-pass setting with slow response mode, and testing at head height from the patient's chair position. For mobile medical helicopter (MOHCAV/MOHCAT) and shipboard booths, special considerations apply including worst-case condition testing and validation that pierside-certified booths meet representative underway conditions before underway use. Troubleshooting guidance addresses common certification failures from fan noise, electrical hum, jack panel leaks, and door seal deterioration.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Audiometric Test Booth Certification Form

Command Owning Booth: _________________________ Date: __________________________

Shore Stationary Booth MOHCAT/MOHCAV Booth Shipboard Booth Pier-side/Cold Iron Pier-side/Lit-Off Underway Speed: ____ (knots)

Booth Location (Bldg/Rm/Space): _________________________ Single/Double Wall: __________ Booth Manufacturer: ___________________________________ Serial/Prop #: ________________ Booth Lights (On/Off): _______________ Booth Fan (On/Off): ___________

Types of Audiometric Testing

Octave Band Center Frequency (Hz) *Max SPL allowed (dB)

**Certified to conduct this type of audiometric testing (Yes/No) 125 250 500 1000 2000 4000 8000

Medical Surveillance Testing (Ears Covered) - HCP (2215/16), physical exams, PHA’s, etc

N/A N/A 27 29 34 39 41

Diagnostic Audiology Testing (Ears Covered) - Headphones or insert phones

39 25 21 26 34 37 37

Diagnostic Audiology Testing (Ears Not Covered) - Sound field testing or bone conduction testing

35 21 16 13 14 11 14

SPL Measured Inside Booth (dB)

SPL Measured Outside the Booth (Info Only): (dBA): _____________ (dBC) : _____________

Field Pre-Calibration (Ref dB/Measured dB): _________/_________ PASS FAIL Field Post-Calibration (Ref dB/Measured dB) : _________/_________ PASS FAIL

EQUIPMENT DATA Manufacturer Model # Serial # Cal Date

SLM

Microphone

Octave Band Filter

Calibrator

Comments/Notes:

Certifier Name (Print) Signature Command Phone

* Max permissible ambient noise level (MPANL) criteria per ANSI S3.1, 1999 (R2008) and DODI 6055.12 (03 Dec 2010) ** Any significant new noise (inside or outside the booth) or relocation of the booth requires recertification

Basic Procedures for Audiometric Booth Certification

Refs: (a) NMCPHC TM6290.91–2 Rev. B IHFOM, Chapter 5

(b) NMCPHC TM6260.51-99-2 Navy Medical Department HCP Procedures dtd 15 Sep 2008

Background Information

• All audiometric booths require, at minimum, annual certification (365 day interval).

• Coordination with an Audiologist is critical to clearly identify what type(s) of audiometric testing is conducted in the booth, as there are three distinct approval criteria depending on booth purpose.

• If audiology testing will be conducted down to -10 dB hearing level (HL), then ambient noise levels in the booth must be at least 3 dB lower than maximum permissible ambient noise levels (MPANLs).

Procedures

• At a minimum, a Type I sound level meter with octave band filter/analyzer is required. The SLM, OBA and microphone must be capable of measuring at least 3 dB below the applicable criteria “Max SPL” values listed in the table on the certification form. Frequently, SLM, OBA and microphone ensemble will not meet this stringent criteria specified for “Diagnostic Audiology Testing, Ears Not Covered (sound field & bone conduction testing).

The SLM, OBA, microphone, and calibrator must each have been professionally calibrated within one year.

• Obtain measurements inside the booth during normal operational conditions during activity levels that are representative of anticipated use conditions, including internal conditions (lights and ventilation turned on).

• Take readings at all frequencies listed on the form, measuring one octave below the lowest frequency to be tested. For example, record the sound pressure level at 125 hertz (Hz) if the clinician tests at 250 Hz and above.

• Perform pre and post-field calibration of the sound level meter.

• Obtain octave band readings in the “Linear” or “All Pass” setting, slow response mode. Significant errors occur if the “A” weighting network is engaged.

• Sit in the patient’s chair with sound level meter held away from your body at head height.

• Select the desired octave band, dial in slow response, and take the reading. Record results for each required octave band.

• For multiple station booths, check levels at seats closest and furthest from the door, and record the higher values.

• Have someone talk outside the booth to see if the booth is certifiable under that condition. If external conversation precludes valid testing, annotate this on the certification form.

• Record all values, and document all equipment data on the form. Sign, date, and post the certification on the exterior of the booth or on an adjacent wall. Retain a file copy.

• For MOHCAV/MOHCAT booths, follow same procedures as above. Conduct certification procedure at the location most often used (the major customer). Realistic external noise/activity should be in effect for an accurate and meaningful certification.

o It is impractical to re-certify mobile booths each time they are moved to a different location, however, readings may be taken at each of the primary customer locations.

o Occasional cross-traffic, generators, flyovers, and small crafts pier-side all have the potential to invalidate test results. Some alternatives to ensure test validity:

Conduct/document booth certification at each prospective test location in worst case conditions Do a test audiogram (on a normal listener) at each location prior to beginning patient care

• Shipboard booths that were certified pierside cannot be used underway until they have been evaluated under representative underway conditions. Type of conditions should be annotated.

• Any significant new noise (inside or outside of the booth) or relocation of the booth requires recertification.

Troubleshooting Non-Certifiable Booths

• If low frequencies fail certification, re-check ambient levels with the fan turned off. If fan noise is determined to be the problem, then initiate fan repair or replacement.

• Electrical lighting may be a source for low frequency noise in the form of 60-cycle hum, with harmonics migrating into the 500 Hz test range. Initiate repair or replacement

• Leaks may be occurring around the jack panel. Sound attenuating material should be carefully packed around the wiring to seal the opening. Contact Biomedical Repair staff to conduct continuity checks and clean/replace jacks and plugs as needed.

• Door seal problems may occur due to hardened or worn out foam seals. These can be replaced. The door may also be hung improperly.

• If the above actions do not solve the problem, consider removing/relocating external noise sources, relocating the booth, adding vibration dampers aboard ship, or obtain a replacement.

• Evaluate internal noise sources (chairs/stools, privacy curtains, etc.). Carpeting/rubber mats may dampen noise.

Command Owning Booth:
Shore Stationary Booth: Off
Shipboard Booth: Off
Underway: Off
Speed:
Booth Manufacturer:
Model SLM:
Model Microphone:
Model Octave Band Filter:
Model Calibrator:
MOHCAT/MOHCAV Booth: Off
Pierside/Cold Iron: Off
Pierside/LitOff: Off
Booth Location:
Single/Double Wall:
Serial/Property #:
Booth Lights:
Booth Fan:
Date:
Certified to conduct this type of audiometric testing 1:
Certified to conduct this type of audiometric testing 2:
Certified to conduct this type of audiometric testing 3:
SPL Measured Inside Booth 125 Hz:
SPL Measured Inside Booth 250 Hz:
SPL Measured Inside Booth 500 Hz:
SPL Measured Inside Booth 1000 Hz:
SPL Measured Inside Booth 2000 Hz:
SPL Measured Inside Booth 4000 Hz:
SPL Measured Inside Booth 8000 Hz:
SPL Measured Outside the Booth (dBA):
SPL Measured Outside the Booth (dBC):
Field Pre-Calibration (Measured dB):
Field Pre-Calibration (Reference dB):
Field Post-Calibration (Reference dB):
Field Post-Calibration (Measured dB):
Field Pre-Calibration Pass: Off
Field Post-Calibration Pass: Off
Field Post-Calibration Fail: Off
Field Pre-Calibration Fail: Off
Manufacturer SLM:
Manufacturer Microphone:
Manufacturer Octave Band Filter:
Manufacturer Calibrator:
Serial # SLM:
Serial # Microphone:
Serial # Octave Band Filter:
Serial # Calibrator:
Cal Date SLM:
Cal Date Microphone:
Cal Date Octave Band Filter:
Cal Date Calibrator:
Certifier Name (Print):
Certifier Name (Signature):
Certifier Command:
Certifier Phone:
Comments/Notes: