6640--Total Chlorine Monitoring System
Solicitation number
36C26225Q1144
Issued by

About this file

This document is a Statement of Work (SOW) for the VA Greater Los Angeles Healthcare System's procurement of chlorine monitoring systems. The VA requires four continuous chlorine monitoring systems to be delivered and installed in Buildings 500 and 213, specifically in dialysis water treatment areas. The systems must have salient characteristics including real-time chlorine measurements, reagent-free operation, integrated flow cell and sensor, user-friendly interface, alarm and output capabilities, and durability for harsh environments.

The related federal contract opportunity (Solicitation Number 36C26225Q1144) is a Request for Quote (RFQ) for commercial items, seeking a single Firm-Fixed Price award. Key dates include a site visit on July 25th, 2025, at 10:00 AM PDT, and quote submissions due by July 31st, 2025, at 10:00 AM PDT. The procurement is open to businesses with NAICS Code 334516 and limited to 1,000 employees. Evaluation will be based on the lowest-priced quotation that meets all requirements, with a focus on brand name or equivalent products that satisfy specified technical characteristics.

View the file

Other files

File Type Posted
Amendment 0002.pdf PDF
S06 - Attachment A v3.xlsx XLSX spreadsheet
S06 - Attachment B v3.docx DOCX document
36C26225Q1144 0002.docx DOCX document
Attachment A v2.xlsx XLSX spreadsheet
Amendment 0001.pdf PDF
36C26225Q1144 0001.docx DOCX document
Attachment B v2.docx DOCX document
Attachment A.xlsx XLSX spreadsheet
36C26225Q1144.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

BACKGROUND and NEED

Medical Service-Dialysis Unit of the VA Greater Los Angeles Healthcare System required 4 continuous chlorine monitoring system with installation at bldgs. 500 and 213 dialysis water treatment areas.

SCOPE

The Contractor shall deliver in total of 4 units of continuous chlorine monitory system and install 2 units of the said system in Building 500, 6 North, Room 6004 and 2 units of the said system in Building 213, Room 044, located at 11301 Wilshire Blvd., Los Angeles, CA 90073.

Hours of Service: Work is to be done between the hours of 8:00 a.m. and 4:00 p.m. Monday through Friday, excluding federal holidays. Request for weekend work must be submitted in writing.

Federal Holidays:

New Year’s DayJanuary 1st
Martin Luther King’s Birthday3rd Monday in January
President’s Day3rd Monday in February
Memorial DayLast Monday in May
JuneteenthJune 19th
Independence DayJuly 4th
Labor Day1st Monday in September
Columbus Day2nd Monday in October
Veteran’s DayNovember 11th
Thanksgiving DayLast Thursday in November
Christmas DayDecember 25th

When one of the above designated legal holidays falls on a Sunday, the following Monday will be observed as a legal holiday. When a legal holiday falls on a Saturday, the preceding Friday is observed as a holiday by U.S. Government agencies.

BRAND NAME OR EQUAL DESCRIPTION

Item #
Description/Part Number*
Qty
1
Chlorine Monitoring System for Dialysis
4
2
ee MONTHLY MAINTENANCE KIT, CM130. Standard lead time 3 days
4
3
CM130 Periodic Maintenance Bundle. Standard lead time 30 days.
4

SALIENT CHARACTERISTICS:

1. Continuous, Online Monitoring

· Provide real-time measurements of free Chlorine (or Total Chlorine depending on configuration).

· Ensures accurate process control and compliance with regulations

2. Reagent-Free Operation

· Uses amperometric sensor technology, which does not require reagents or chlorimetric chemicals.

· Reduce maintenance and operating costs.

3. Integrated Flow Cell and Sensor

· Comes with a flow cell and pre-calibrated sensor

· Allows easy instruction and startup.

4. Easy Maintenance

· Must incorporate a user-friendly sensor design that promotes consistent and reliable chlorine monitoring while minimizing operational downtime

· Designed to extend calibration and cleaning intervals beyond those of older systems.

5. Intuitive User Interface

· Includes a color display and user-friendly menus for setup, calibration and diagnostics.

6. Alarm and Output Capabilities

· Offers analog (4-20 mA) and digital communication options for integration with SCADA or PLC systems.

· Can trigger alarms or control processes automatically based on chlorine levels.

7. Measurements Ranges

· Free Chlorine : 0-2 mg/L up to 0-20 mg/L (depending on sensor and configuration)

· Response Time : Rapid, usually within seconds to minutes depending on flow and sample condition

8. Durability and Environmental Suitability

· Designed for harsh environments and continuous use.

· Waterproof and corrosion-resistant housing (IP65/67 rated, depending on installation).

DELIVERY / INSTALLATION REQUIREMENTS

· The Contractor shall deliver and install 4 units of continuous chlorine monitory system and install 2 units of the said system in Building 500, 6 North, Room 6004, 1 unit post city water, 1 unit post primary carbon tank. Alarm needs to be installed in Rm.6034 near the RN station that is about 120 feet from the water room (6004). 2 units of the said system in Building 213, Room 044, located at 11301 Wilshire Blvd., Los Angeles, CA 90073, 1 unit post city water, 1 unit post primary carbon tank. Alarm needs to be installed in B-44 near the RN station that is about 100 feet from the water room (B-48).

· The Contractor shall ensure all equipment delivered are fully operational to manufacturer’s specification and free from defect.

· The Contractor shall provide delivery, set up, complete installation, and technical support for clinical staff and clinician training for use the system.

· The Contractor shall report all recalls, removals, modifications and safety notices from the manufacturer, distributor and FDA, and provide immediate notices (within 48 hours) to the Contracting Officer and the facility point of contact and the VHA Product Recall Office for any notices identified for medical devices.

· The Contractor shall coordinate with VA Supply Chain Management or facility appropriate contact agree upon delivery day and arrival to the facility.

· The Contractor shall provide digital format operation manual include the care, handling, set up, and cleaning instruction on the new continuous chlorine monitoring system.