Attachment A - SOW - NNMC DZHC Biohood Cabinet 11-14-2025w.docx
DOCX document 38 KB Posted
- Attached to
- Bio Hood Certifications and Inspections
- Solicitation number
- IHS1521344
- Issued by
- Not on record
About this file
This is a Statement of Work (SOW) for biological safety cabinet (biohood) recertification and inspection services at the Indian Health Service's Northern Navajo Medical Center (NNMC) in Shiprock, New Mexico, and the Dzilth-Na-O-Dith-Hle Health Center (DZHC) facility. The SOW outlines the requirement for four total biohood recertifications and inspections, consisting of three units at NNMC and one unit at DZHC. The biohoods must be certified semiannually (twice yearly) to ensure proper operation and compliance with applicable standards. The contractor shall be responsible for performing comprehensive inspections, testing, and certification of each biological safety cabinet in accordance with NSF/ANSI 49 standards and applicable federal regulations.
The SOW specifies that the contractor must provide detailed documentation of all inspection findings, certifications, and any deficiencies identified during the recertification process. The work includes on-site inspection and certification services at both the NNMC and DZHC locations. The contractor is required to coordinate scheduling with facility personnel and provide timely completion of all recertification work to maintain compliance with laboratory safety and operational requirements. All certifications must be documented and provided to the facilities upon completion of the work.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Amendment 1 RFQ 1521344 - Bio Hood Certifications and Inspections.pdf | ||
| Service Contract Act WD 2015-5445.pdf | ||
| RFQ IHS1521344 - Bio Hood Certifications and Inspections.docx | DOCX document | |
| Attachment B - Past Performance Questionnaire.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
BACKGROUND
The Indian Health Service, Northern Navajo Medical Center (NNMC) in Shiprock, NM is requesting for four (4) bio hood (biological safety cabinet) recertifications and inspections. There are Three (3) bio hoods at NNMC and one 1 (one) at Dzilth-Na-O-Dith-Hle Health Center (DZ) facility. The bio hoods are required to be certified semiannual (twice-yearly) for proper operation.
OBJECTIVES
The objectives of the biohood certifications are to ensure the equipment provides a safe, sterile environment for the laboratory work involving infectious or hazard materials. The certification also protests the lab work, the product and the environment.
The bio hood certifications are required to meet National Sanitation Foundation (NSF), the Joint Commission Accreditation for Healthcare Organization (JCAHO) and Occupational Safety and Health Administration (OSHA) regulations.
PERIOD OF PERFORMANCE
The contract shall include a base period for 1 year and 4 Option years. A total of 5 years.
TASK
Services for the following Hood Fume Bacteriological:
| Item | Make Model | Serial Number | Location | |||||||||
| Hood Fume Bacteriological | Nuaire NU-540-300 | 179393032717 | NNMC – GC128 Hood Fume Bacteriological | Nuaire NU-540-600 | 179394032717 | NNMC - GC134 Hood Fume Bacteriological | Nuaire NU-301-436 | 185905051418 | NNMC – LA141C Hood Fume Bacteriological | Nuaire NU-540-400 | 186350060818 | DZ Lab – B127 |
Statement of Work Certification and Inspection for Bio Hoods Technical Point of Contac: Robert Whitehorse, Jr., Equipment Specialist robert.witehorsejr@ihs.gov (505) 368-6140
· The certification process must meet the critical test to ensure it meets the required standards America National Standard - NSF/ANSI 49
· Testing will confirm to Manufacturer’s Specifications, OSHA, NFPA 99-2005, ASSE 6030,
· Federal Standard 209E, ISO 14644-1, USP 797 and U.S.
· A certified bio hood technician is required to perform the services
· Contractor is responsible for per-inspection prior to award
· Contractor will notify the TPC at least two weeks in advance prior to conducting services
· Contractor will conduct Certification in accordance to re-certification criteria:
· Airflow testing (rate)
· Air exchange - percentage on air count and inflow
· Perform minor repairs necessary such as repairing small gaskets leaks or defects
· Electrical safety test
· Provide inflow and downflow velocity test in the cabinet to measure the velocity of air flow
· Immediately notify TPC if certification fails and replace damaged and perform re-certification after repairs are completed.
· Contractor shall update parts replacement, operator instructions or any procedural change.
· Contractor shall provide a comprehensive service report that indicates the test result and maintained as a permanent equipment history file. This report will include description of work performed, man hours parts replacement numbers or nomenclature
· Affix a certification sticker on each bio hood with clear certification date and expiration date
· Telephone numbers for service calls and contract administration shall be provided to the Contracting Officer Representative (COR) and Technical Point of Contact (TPC). The COR and TPC will monitor the contract by review of the information provided with the bids, evaluation of actual work performed, response, timeliness, checking-in, service reports, general compliance with contract specifications and technical requirements.
· The Contractor must see the COR/POC to get their signature on the service report and to be a legible photo copy of the services report; this constitutes “acceptance” by the COR/POC and a required for payment.
GOVERNMENT-FURNISHED (PROPERTY)
No government furnish supplies, equipment, etc. will be provided. The Contractor shall be responsible for safeguarding all government equipment, information and property provided to Contractor personnel while performing the services described herein.
CONTRACT-FURNISHED (PROPERTY)
The Contractor/Field Engineer or Technician, upon arrival at the facility, shall have all tools, test equipment, parts and manpower required for completion of the certification, inspection and repair services.
SECURITY REQUIREMENTS
Contractor shall check in with Facility Maintenance Staff and will escorted to security to obtain a Contractor’s visitor badge. All Contractors personnel are required to ware identification badges during the entire time they on in the hospital.
Smoking is not permitted within or around the Healthcare facilities.
Work shall be performed Monday through Friday from 8:00 am to 4:30 pm only.
PLACE OF PERFORMANCE
Northern Navajo Medical Center, U.S. Hwy, 491 North, Shiprock, NM 87420 – Laboratory and Pharmacy rooms Dzilth-Na-O-Dith-Hle Health Center, U.S. 6 Road 7586, Bloomfield, NM 87413. Laboratory Room
File details come from the government source that posted it. Updated .