Attachment A Bid Specifications.pdf
PDF 89 KB Posted
- Solicitation number
- DOH24-039P
- Issued by
- Adams County, Asotin County, Benton County, Chelan County, Clallam County, Clark County, Columbia County, Cowlitz County, Douglas County, Ferry County, Franklin County, Garfield County, Grant County, Grays Harbor County, Island County, Jefferson County, King County, Kitsap County, Kittitas County, Klickitat County, Lewis County, Lincoln County, Mason County, Okanogan County, Pacific County, Pend Oreille County, Pierce County, San Juan County, Skagit County, Skamania County, Snohomish County, Spokane County, Stevens County, Thurston County, Wahkiakum County, Walla Walla County, Whatcom County, Whitman County, Yakima County, Asotin City, Clarkston City, Clarkston Heights-Vineland CDP, West Clarkston-Highland CDP, Benton City, Chelan City, Chelan Falls CDP, Clallam Bay CDP, Lewisville CDP, Rock Island City, Pacific Beach CDP, Whidbey Island Station CDP, Mercer Island City, Pacific City, Bainbridge Island City, Kingston CDP, Kitsap Lake CDP, Kittitas City, Klickitat CDP, Okanogan City, Anderson Island CDP, Fort Lewis CDP, Fox Island CDP, Herron Island CDP, Ketron Island CDP, North Fort Lewis CDP, Raft Island CDP, Stevenson City, Hat Island CDP, Lake Stevens City, Snohomish City, Spokane City, Spokane Valley City, Puget Island CDP, Garfield Town, Yakima City, Washington
About this file
This document is an Attachment A - Bid Specification for a solicitation issued by the Washington State Department of Health (DOH) for an Ultra-High Performance Liquid Chromatography (UHPLC) instrument. The DOH is seeking bids from qualified firms to provide the UHPLC instrument, along with service, training, and a two-year extended warranty with one-year preventative maintenance. The instrument must meet specific technical requirements, including detection capabilities, sample capacity, reliability, and compatibility with existing laboratory equipment. The contract will be awarded to a single vendor, and the successful bidder must be able to provide on-site service within three business days.
The bid specification outlines detailed requirements for the UHPLC instrument, including the pump, column oven, autosampler, and diode array detector. It also specifies software and computer requirements, as well as training and support expectations. The contract is expected to be for a five-year period, with the option for the DOH to extend the service agreement in one-year increments after the initial warranty period. The solicitation is open to all qualified vendors, and there are no set-asides or incumbent information provided.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| DOH24-039P Invitation for Bid.pdf | ||
| Appendix B-DOH Bid submittal.doc | DOC document | |
| Addendum 1 DOH24-039P.pdf | ||
| Appendix C-DOH PO Terms and Conditions.pdf | ||
| Appendix A DOH Solicitation Standards.pdf | ||
| DOH Notice of Solicitation DOH24-039P.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment A - Bid Specification for DOH24-039P # Item Description Yes No
0 UHPLC-DAD UHPLC and Diode Array Detector capable of detecting domoic acid in shellfish tissues.
[ ] Yes [ ] No (fail)
1 Installation Requirements
Installation, Setup, Calibration, and Performance Verification shall be performed at PHL at no cost to the DOH.
[ ] Yes [ ] No (fail)
1 Discrimination UHPLC-DAD must be able to achieve baseline separation between the peaks of L-tryptophan and domoic acid (g)
[ ] Yes [ ] No (fail)
2 Sample Capacity Auto sampler with suitable injector capable of holding minimum 100 sample vials
[ ] Yes [ ] No (fail)
3 LOQ A limit of quantitation of 1 ppm of domoic acid in shellfish
[ ] Yes [ ] No (fail)
4 Sensitivity The instrument must be able to achieve a signal to noise ratio of 10:1 for a 40 ppb injection of domoic acid
[ ] Yes [ ] No (fail)
5 Reliability The instrument is expected to routinely perform sample measurements around the clock for months at a time without failure or required service. Routine recommended maintenance intervals should not exceed once every 6 months.
[ ] Yes [ ] No (fail)
6 Reliability (Up-time)
Instrument IS NOT prone to frequent breakdown AND is fully operational at least 90% of the time.
A reasonable standard applies.
[ ] Yes [ ] No (fail)
7 Service Agreement
Contractor shall provide an as needed option to the DOH a service agreement for the five (5) years in one year increments following expiration of the original warranty period.
[ ] Yes [ ] No (fail)
General System Requirements The required technical features of the system include:
8 Service Availability
The Contractor must be able to provide a service agreement that includes on-site response to service calls within three (3) business days.
[ ] Yes [ ] No (fail)
9 Extended Warranty
The Contractor must be able to provide an extended warranty for two years that also includes one preventative maintenance in the second year.
[ ] Yes [ ] No (fail)
10 Compatibility The UHPLC components must be compatible with an Agilent LC/MS Ultivo and 6475A in order to provide redundancy in the lab.
[ ] Yes [ ] No (fail)
11 Size Requirements
Due to laboratory space for installation, the footprint of the instrument is not to exceed a height of 35 inches (not including solvent bottles), a width of 18 inches, and a depth of 19 inches.
[ ] Yes [ ] No (fail)
1 Type: Quaternary [ ] Yes [ ] No (fail) 2 Delay volume Must have as low as possible gradient delay volume of the pump = 350 μL [ ] Yes [ ] No (fail)
3 Flow Rate Must be able to maintain an isocratic 1.0 mL/min up to 5.0/ml/min maximum flow rate.
[ ] Yes [ ] No (fail)
4 Operation Pressure Range
Must have operating range up to 1300 bar at 2 ml/min, and 800 bar at 5 mL/min.
[ ] Yes [ ] No (fail)
5 Solvent Channels
At least 4 Solvent Channels [ ] Yes [ ] No (fail)
6 pH Range Can tolerate pH range of 1.0 — 12.5 [ ] Yes [ ] No (fail) 7 Flow Precision Must have ≤0.07% RSD or ≤0.01 min SD, whatever is greater. This is included to provide a more accurated compound identification through precise retention time.
[ ] Yes [ ] No (fail)
8 Hydraulic System Dual pistons in series. [ ] Yes [ ] No (fail)
UUHPLC/UHPLC Specific Pump
Column Oven/Thermostat
1 Temperature control
• Column compartment with oven capable of accurate temperature control at 40°C for domoic acid analysis.
• Temperature accuracy must be ± 0.5°C with calibration
• Must have a temperature range from 10°C below ambient to 85°C.
• The module must enable pre column heating and post column cooling.
[ ] Yes [ ] No for any item (fail)
2 Independent Temperature Zones
Must have at least two independent temperature zones.
[ ] Yes [ ] No (fail)
3 Capacity Must be able to contain up to 4 columns with a length of 30 cm each.
[ ] Yes [ ] No (fail)
1 Mobile phase degassing unit. [ ] Yes [ ] No (fail)
1 Refrigerated Autosampler Sample in vials must be maintained at a temperature of 0-4 °C.
[ ] Yes [ ] No (fail)
2 Capacity Must be able to have various and scalable drawer types from 2-16 well plates all within the LC-stack footprint.
[ ] Yes [ ] No (fail)
3 Injection Volume • Loop suitable for twenty (20) μL injections.
• Adjustable range: 0.1 to 900μL volumes or 1800µL by seat loop extension .
[ ] Yes [ ] No for any item (fail)
4 Carryover Prevention
Must be capable of performing a multi-wash option.
[ ] Yes [ ] No (fail)
1 Type Diode Array- 1024 element preferred for highest spectral resolution.
[ ] Yes [ ] No (fail)
2 Wavelength range
• Wavelength range must be at least 190-950 nm, settable in 1 nm increments.
• Must be sensitive at wavelengths of 242 nm and 350 nm for Domoic acid detection.
[ ] Yes [ ] No for any item (fail)
3 Noise < ± 0.7 x 10-5 AU at 254 nm and at 750 nm (cell path length 10 mm, response time 2 s, flow 1 ml/min LC-grade.
[ ] Yes [ ] No (fail)
4 Light Source Dual lamp design (tungsten and deuterium).
[ ] Yes [ ] No (fail)
5 Drift Signal drift must be equal or smaller 0.9 x 10-3 AU/h after adequate warm up.
[ ] Yes [ ] No (fail)
Autosampler
DAD Detector
Degassing Unit
6 Maximum sampling rate
120 Hz [ ] Yes [ ] No (fail)
7 Linearity > 2 AU (upper limit). Important to allow the simultaneous determination of high and low levels of the analytes.
[ ] Yes [ ] No (fail)
8 Operating Temperature
The temperature operating range of instrument must be from 4°C to 55°C.
[ ] Yes [ ] No (fail)
9 Compatibility Compatibility with wide range of flow cells [ ] Yes [ ] No (fail) 10 RFID tags • Radio Frequency Identification Tags for flow cells and UV lamp to record parameters.
• The information needs to be saved with each individual raw data file for data traceability, minimizing the risk of false data interpretation
[ ] Yes [ ] No for any item (fail)
11 Number of signals
Simultaneous acquisition of up to 8 compound specific wavelength to provide increased sensitivity and selectivity.
[ ] Yes [ ] No (fail)
12 Slit Width • Programmable: 1, 2, 4, 8, 16 nm
• Must allow performance optimization regarding:
o Spectral resolution o Signal to noise (or limit of detection)
[ ] Yes [ ] No for any item (fail)
1 PC The PC is the latest available and is capable of running 64 bit MS Windows or greater.
[ ] Yes [ ] No (fail)
2 Network Interface Card
(NIC)
Is there a manufacturer installed NIC card?
Type: ______________________
[ ] Yes [ ] No (fail) No answer = fail
3 Analytical Software Requirements
Instrument must have software for both qualitative and quantitative analysis
[ ] Yes [ ] No (fail)
4 Instrument Control Software
The software that controls the operations of the instrument must have the statuses and control of individual modules displayed in the same window.
[ ] Yes [ ] No (fail)
System Computer
5 Quantitative Software
Quantitative software allows user to quickly and easily inspect chromatograms and area counts for multiple analytes for each calibration, quality control, and unknown sample. Software must meet this condition.
[ ] Yes [ ] No (fail)
6 Peak Integration Analysts must be able to manually integrate peaks on chromatograms quickly and easily.
Software must meet this condition.
[ ] Yes [ ] No (fail)
7 Method Editing The software must allow analyst to quickly reprocess batches and edit calibration concentrations and analyte retention times.
[ ] Yes [ ] No (fail)
8 Data Format The software must display the run data in a format that displays all quality control criteria for an analyte in one window for facilitated inspection and printing purposes.
[ ] Yes [ ] No (fail)
9 Data File Management
Data folder management must be efficient and organized in a way that all files from individual runs are placed into the same subfolder.
[ ] Yes [ ] No (fail)
10 Software Roadmap
Bidder must have a software road map for major/minor updated versions of the software. Please describe how the enhancements are handled.
[ ] Yes [ ] No (fail) No answer = fail
11 Service Packs Bidder must provide service packs. Please describe what is included in the service pack for the instrument.
[ ] Yes [ ] No (fail) No answer = fail
12 Data Back-up • The software must automatically backup the data on a reasonable periodic schedule.
• Please describe if it is it a full back-up or incremental back up and what is the back-up frequency?
[ ] Yes [ ] No (fail) No answer = fail
13 Data Overwriting Instrument must allow manual overwriting of the dataset.
[ ] Yes [ ] No (fail)
14 Internet Access • Does your customer support need access to the instrument via the internet?
• If Yes, Please describe.
[ ] Yes [ ] No (fail) If Yes, No answer = fail
15 Monitor • Does the software allow for touch screen capability?
Note: If yes, the bidder must provide a monitor that also has touchscreen.
• Is the monitor an LED type
• Is the size is 24 inch or larger?
[ ] Yes [ ] Yes [ ] Yes
[ ] No aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaa [ ] No aaaaaaaa [ ] No (fail)
16 LIMS System Capability
Is the instrument capable of connecting to a Laboratory Information System (LIMS) to be able to transfer data to a lab reporting system?
[ ] Yes [ ] No
1 Hands -on Training
Hands-on Training is provided at the customer site and includes at least 8 hours of instructor led and hands-on training provided at the customer site by a service engineer for 2 to 4 personnel
• Instrument orientation
• UHPLC/DAD Operation
• Method Set-up
• Instrument Preventative maintenance.
• Trouble shooting
• Method development
• Quick reference guides or other training materials
[ ] Yes [ ] No (fail)
2 Software Training A minimum 6 hrs. of Instructor Led Software orientation & training for up to 4 personnel.
[ ] Yes [ ] No (fail)
Training
Notes
Sheet1