This Performance Work Statement and Technical Specifications document describes the government's requirement to acquire a Questron QWave 1000-Watt Microwave Digestion System for the DLA Product Test Center in Philadelphia, PA. The system is needed to prepare chemically resistant and complex sample matrices—including titanium alloys, cobalt, tungsten ores, and specialty steels—for elemental analysis, supplementing the laboratory's existing Questron Vulcan S Uno digestor. The acquisition seeks to standardize the laboratory's digestion platforms for operational interoperability. The contractor must deliver a complete turnkey system with an integrated 1000-Watt magnetron (2450 MHz frequency), 56-liter stainless steel cavity, rotating carousel accommodating 12 Enhanced Volume High-Pressure (eVHP) vessels, and dual temperature control (infrared sensor and fiber optic probe) capable of reaching 260°C maximum with ±1°C accuracy. The system must include 16 total eVHP vessels (12 operational plus 4 spares) with TFM liners, over-pressure release valves, non-contact pressure monitoring (42 bar operating / 83 bar maximum), camera display, and a separated controller with color touchscreen interface utilizing the same common software control platform as the existing Questron Vulcan S Uno.
The contractor shall deliver all hardware, software, and accessories within 90 days from award, including perpetual software licensing, on-site installation and Operational Qualification (IQ/OQ) by a factory-certified Field Service Engineer (approximately 4 hours), on-site operator training for up to five personnel, and a one-year comprehensive warranty covering all parts and labor. Deliverables also include operator and service manuals in hardcopy and electronic formats, along with NIST Traceable ANSI/NCSL Z540.1 calibration certificates. Pricing must include all shipping and delivery costs (FOB Destination) and be submitted through Wide Area Workflow (WAWF). The DoDAAC for invoicing is SL0701. Delivery shall be directed to Edward Francoeur at DLA Product Test Center, 700 Robbins Street, Building 5-D, Philadelphia, PA 19111, with advance notice required. The government will provide a suitable laboratory location with a certified fume hood and required electrical utilities (208 VAC, 50/60Hz, 15A primary or 120 VAC, 20A alternative).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SF18_SP470626Q0002.PDF | ||
| Attachment 1 - Tech Specs QWave System.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Performance Work Statement (PWS) & Technical Specifications
Performance Work Statement (PWS)
Questron QWave 1000-Watt Microwave Digestion System
1.0 Background
The Product Test Center requires a high-power, high-performance microwave digestion system to prepare chemically resistant and complex sample matrices, including but not limited to, titanium alloys, cobalt, tungsten ores, and various steels for elemental analysis.
This system will supplement the laboratory's existing Questron Vulcan S Uno digestor, providing a necessary capability for samples that require the high pressures and temperatures unique to microwave technology. This acquisition seeks to standardize the laboratory's digestion platforms to ensure operational interoperability.
2.0 Scope
The contractor shall provide a complete, "turnkey" Questron QWave 1000 Watt Microwave Digestion System. This includes the core instrument, all necessary high-pressure vessels and accessories, software, delivery, on-site installation by a certified technician, system verification, comprehensive operator training, and a full warranty.
3.0 Performance Requirements
The system shall meet the following minimum performance requirements:
• 3.1 Digestion Performance: The system must be capable of achieving complete and reproducible digestion of difficult-to-digest materials, specifically including Titanium 6-4, tungsten ores, and specialty steels. The resulting solution must be clear, free of particulates, and suitable for trace metals analysis.
• 3.2 Throughput: The system shall be capable of digesting a minimum of 12 samples simultaneously using the specified high-pressure vessels.
• 3.3 Safety: The system must provide maximum operator safety through multiple layers of monitoring and control, including real-time temperature and pressure control for all vessels, automatic microwave power adjustment to prevent over-pressurization, and a robust, blast-resistant cavity door.
• 3.4 Interoperability: The system must be fully compatible with the laboratory’s existing Questron equipment ecosystem. This requires a common software control platform and the ability to use interchangeable sample racks and vials post-digestion to streamline the workflow with the existing Questron Vulcan S Uno (s/n
VN24-1005).
4.0 Deliverables
The contractor shall provide the following:
Deliverable No.
Description Quantity
Questron QWave 1000 Watt Time-to-Temperature Microwave Digestion System
Complete Set of twelve (12) Enhanced Volume, High- Pressure (eVHP), HF-Compatible Vessels, plus four (4) additional spare vessels (total of 16 complete vessels).
3 eVHP Vessel Liner, TFM 12
4 eVHP Vessel TFM Top for standard Vessel 12
5 eVHP / LVHT Over Pressure Release Valve 12
Perpetual Software License (common platform with Vulcan S Uno)
On-Site Installation and successful IQ/OQ (Installation/Operational Qualification)
On-Site Operator Training for up to five (5) personnel (minimum 4 hours).
1 Session
9 Operator and Service Manuals 1 Hardcopy, 1 Electronic
10 Certificate of One-Year Comprehensive Warranty 1
5.0 Government-Furnished Items (GFI)
The government will provide a suitable and secure location for the instrument, including a certified laboratory fume hood prepared to the contractor's pre-installation specifications, and all required electrical utilities.
6.0 Administrative Requirements
• 6.1 Period of Performance: Delivery of all hardware, software, and accessories shall be completed within 90 days from the date of award. Installation and training shall be coordinated with the technical point of contact upon delivery.
• 6.2 Pricing: Pricing shall include all costs for shipping and delivery (FOB Destination) to the address specified below.
• 6.3 Delivery and Marking Instructions:
o 6.3.1 Notice of Delivery: Contractor shall give notice to Edward Francoeur (Edward.Francoeur@dla.mil, 445-737-0829) and Mike McClain (James.McClain@dla.mil, 445-737-1764) when the instrument is to be delivered.
o 6.3.2 Delivery Address / Mark For:
Attn: Edward Francoeur DLA Product Test Center 700 Robbins St., Bldg. 5-D Philadelphia, PA 19111
• 6.4 Invoicing: All invoices must be submitted through Wide Area Workflow (WAWF).
See DFARS Clause 252.232-7006 Wide Area Workflow Payment Instructions for detailed instructions. The DoDAAC for this action is SL0701.
Technical Specifications
Questron QWave 1000 Watt Time-to-Temperature Configuration
The proposed Microwave Digestion System must meet or exceed the following technical specifications:
1.0 General
1.1 The system shall be a new, current-production model Questron QWave 1000 Watt Time-to-Temperature Configuration.
2.0 Main Unit & Microwave Cavity
2.1 Microwave Power: Shall have a magnetron output power of at least 1000 Watts. Power shall be un-pulsed and continuously adjustable in 1 Watt increments.
2.2 Microwave Frequency: Shall operate at a frequency of 2450 MHz.
2.3 Microwave Cavity: The cavity shall be constructed of 316L reinforced stainless steel https://www.google.com/url?q=mailto%3AEdward.Francoeur%40dla.mil https://www.google.com/url?q=mailto%3AJames.McClain%40dla.mil and coated with a multi-layer, corrosion-resistant PFA or similar fluoropolymer. Cavity volume shall be a minimum of 56 liters.
2.4 Camera Display: The system shall feature a camera display on the front door to allow visual monitoring of the microwave cavity state.
2.5 Rotating Carousel: Shall include a 360° rotating carousel and antenna to ensure uniform microwave heating.
3.0 Temperature and Pressure Control
3.1 Temperature Control: Must utilize a "Time-to-Temperature" control methodology. The system must include:
* 3.1.1 A non-contact infrared (IR) sensor for real-time temperature monitoring of the solution in all vessels simultaneously.
* 3.1.2 A direct, immersion-style fiber optic temperature probe for precise reference control in one vessel.
* 3.1.3 The system must have a maximum operating temperature of at least 260°C and a normal operating temperature of 230°C, with an accuracy of ±1°C.
3.2 Pressure Control: Must utilize a non-contact pressure monitoring system that measures the pressure in all vessels in real-time. The system must be capable of uniquely identifying over-pressure and leaking vessels.
* 3.2.1 Operating pressure shall be controlled at approximately 42 bar / 610 psi.
* 3.2.2 Maximum system pressure shall be at least 83 bar / 1200 psi.
4.0 Rotor and Digestion Vessels
4.1 Rotor: Shall accommodate a minimum of 12 eVHP vessel positions. The design shall allow for loading the carousel vessel-by-vessel within the system.
4.2 Vessel Type: Shall be Questron-manufactured "Enhanced Volume, High-Pressure" (eVHP) vessels, equipped with a resealing OPGuard™ vessel protection system. A minimum of 4 vessels per run is required.
4.2.1 Quantity: A total of sixteen (16) complete eVHP vessels shall be provided, each including a liner, top, and over-pressure valve (12 standard capacity, plus 4 spares).
4.3 Spare Liner Materials: Liners/tubes must be constructed of high-purity, HF-compatible TFM fluoropolymer. The outer pressure-retaining casing must be made of PEEK, composite, or other high-strength material.
4.3.1 Quantity: Twelve (12) eVHP vessel liners, twelve (12) tops for standard vessel, and twelve (12) eVHP / LVHT over pressure release valves.
4.4 Vessel Performance: Vessel volume must be 100 mL.
4.5 Integrated Workflow: The system shall support Pre- and Post-Digestion Treatment (sample concentration, HF neutralization, acid evaporation) within the microwave vessel liners using an external block (e.g., QBlock) to minimize contamination.
5.0 Control Software
5.1 User Interface: The system shall utilize a separated controller with a built-in, color touch-screen interface capable of displaying real-time graphical digestion parameters for each vessel.
5.2 Interoperability: The operating software must be the same common control platform used by the Government's existing Questron Vulcan S Uno (s/n VN24-1005).
5.3 Preloaded Methods: Shall include preloaded EPA methods 3015, 3051, and 3052, and provide extensive storage for custom recipes and data.
6.0 Installation and Site Requirements
6.1 Electrical Requirements: 208 VAC, 50/60Hz, 15A. Power cord(s) must include a NEMA 6-15P plug end. Alternatively, 120 VAC, 50/60Hz, 20A is acceptable, if available.
6.2 Physical Dimensions: Required bench space shall not exceed 25" W x 23" D x 26" H (63 x 58 x 65 cm).
6.3 Ventilation: The instrument must be equipped with an integrated 150 CFM exhaust blower. It requires a 3” hose connection to the laboratory exhaust system (an 8’ hose shall be included).
7.0 Installation, Training, and Warranty
7.1 A factory-certified Field Service Engineer must perform the complete on-site installation, totaling approximately 4 working hours.
7.2 Installation shall include on-site training on operation and maintenance for up to five (5) personnel.
7.3 A minimum one-year comprehensive warranty covering all parts, labor, travel, food, and lodging for one service technician must be included.
8.0 Certification and Verification
8.1 The contractor shall provide instrument certification or verification of instrument performance upon completion of installation (IQ/OQ).
8.2 A NIST Traceable ANSI/NCSL Z540.1 certificate (with data) shall be provided for calibrated components or sub-components, as applicable.