About this file
This document is a product brochure for the QWave microwave-assisted sample digestion system manufactured by Questron Lab Automation. The QWave product line consists of two configurations: the QWave 1000 (single magnetron system supporting closed and open vessel digestions) and the QWave 2000 (dual magnetron high-power system for closed vessel digestions only). Both systems feature 1800 or 900 watts of microwave power output in 1-watt increments, non-contact infrared temperature sensing, and a user-friendly touchscreen control panel with capacity to store 1000+ multi-step digestion recipes. The systems include preloaded EPA and other standard methods and provide real-time graphical display with data export and printing capabilities for documentation.
The QWave systems accommodate two vessel configurations: the eVHP Vessel Set (18 vessels with 110 ml volume, maximum control temperature of 230°C, and maximum control pressure of 69 bar/1000 psi) and the LVHT Vessel Set (36 vessels with 50 ml volume, maximum control temperature of 200°C, and maximum control pressure of 20.7 bar/300 psi). Key safety features include individual pressure control for each vessel using the patented OPGuard vessel protection system, which releases pressure vertically in a controlled manner, and automatic magnetron shutdown upon major pressure release. Additional capabilities include a QBlock pre- and post-digestion treatment system for sample concentration and acid treatment, vessel workup stations with torque-calibrated wrenches, and a two-speed exhaust system (50 or 150 CFM). The external dimensions are 22.8 x 24.8 x 25.5 inches with a weight of 69 kg (150 lbs.) and noise level under 56 dB. The system requires 200-240 VAC, 50/60 Hz, 15 Amps utility power.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 2 - PWS and Technical Specs.pdf | ||
| SF18_SP470626Q0002.PDF |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Innovative & Flexible Sample Prep Solutions
Microwave Assisted Sample Digestion
With Safety, Economy & High Throughput
Rugged Design with Enhanced Safety Features
• Ideal for digestion, extraction, evaporation, and synthesis
• Built-in temperature and pressure monitoring of each digestion vessel
• User-independent safety features ensuring safe digestions
• Controlled release of over pressure to avoid cross-contamination
• Rotating antenna and 360° carousel rotation ensuring uniform microwave distribution
• Corrosion-proof PTFE-coated reinforced cavity for corrosion resistance
• Two-900 Watt magnetrons provide continuously variable, non-pulsed, evenly distributed power in small increments for precise temperature control of every vessel.
• Two-speed, 150 CFM exhaust for fume removal and vessel cooling
• Spring-loaded door latches for enhanced safety
Configurations
Vessel Set eVHP Vessel Set (18 Vessels)
• Accurate temperature sensing
• Ease of loading carousel vessel by vessel inside the system’s cavity. Convenient manual release of residual pressure after digestion
• Compact Vessel Handling Station prepares vessel correctly every time
• Automatic protection against unusual heating of vessels
• Optimally designed carousel ensures uniform heating of all samples
LVHT Vessel Set (36 Vessels)
• Vessel for high throughput and moderate temperature requirements
• Possibility to digest 36 samples in a single run
• Temperature monitoring of each vessel
• Vessels are individually pressure controlled and equipped with our unique resealing over-pressure guard (OPGuard vessel protection system
• Ideal solution for labs running large number of samples in every batch
Innovative & Flexible Sample Prep Solutions
QWave 1000: Single Magnetron system Closed & Open Vessel Digestions Open Vessel configuration comes with External exhaust blower to take care of excessive acid fumes generated during operation
QWave 2000 Dual Magnetron High Power System for Closed Vessel Sample Digestions
Control Panel Interface
• User-friendly control panel, operated through touchscreen
• Easy navigation for method creation and storage
• Audible alarm and visual alerts during digestion runs
• Preloaded with Methods library of EPA and other standard methods
• Capable of identifying a vessel in case of over pressure release
• Data export and printing capability for tracking, tracing, and documentation
Temp. & Power
All vessels are individually pressure controlled using our unique resealing over-pressure guard (OPGuard vessel protection system. This system releases pressure vertically in a controlled manner to achieve safe pressure conditions.
Released fumes are removed quickly to avoid condensation on cavity and vessels.
If release is minor and drop in temperature is minimal, the digestion method can be continued. In an instance where pressure release is major, complete safety is achieved by automatically cutting off magnetron power, activating audible alarm, and displaying visual warnings
User may create and store
1000+ multi-step digestion recipes or select from
Questron’s existing methods library
OPGuard
System Status
Recipe Details
Digestion Progress Bar
Innovative & Flexible Sample Prep Solutions
Real time graphical display
System Specifications:
Oven
Microwave power output
1800 or 900 watts power in 1-watt increments
Magnetron frequency 2450 MHz
Temperature sensing Non-contact IR based
Utilities 200- 240 VAC, 50/60 Hz, 15 Amps
Oven Door Viewing Window
Built-in camera video displayed on dedicated LCD screen on oven door
Exhaust 2-speed, 50 cfm or 150 cfm nominal
Weight 69 kg. (150 lbs.)
Noise Level < 56 dB
Dimensions (width x depth x height)
External 22.8 x 24.8 x 25.5 in.| 58 x 63 x 65 cm.
Cavity 15.6 x 15.6 x 14 in.| 40 x 40 x 36 cm.
Wire-free Command Panel
Display Type Glass Touchscreen
Size 18.8 x 11.2 cm (7.5 X 4.5 in.)
Processor speed 1.2 GHz
Resolution 1024 x 600 pixels
Data input method Direct via touchscreen
On-board method storage 1000 +
On-board data storage 10 GB
Controller Connectivity Bluetooth
Pre & Post Digestion Treatment
Qblock with Vessel Liners
Torque calibrated wrench to ensure complete closure of vessel
Workup station pin push liner for clear access to insertion on removal eVHP and LVHT Vessels
Parameters eVHP LVHT
Liner Type TFM TFM
Volume 110 ml 50 ml
Temperature Absolute max.
temperature 300°C 300°C
Max. control temperature
230°C 200°C
Pressure Absolute max.
pressure (bar/psi) 151 /
43 /
Max. control pressure (bar/psi) 69 /
20.7 /
Sample processing capacity QWave 2000 18 36
QWave 1000 12 24
Vessel Workup Stations
Process of assembling and dissembling a digestion vessel using the eVHP Workup Station
• Option of QBlock digestion system for pre- and post-digestion treatment of samples in digestion vessel liners
• Prevent contamination from use of multiple vials
• Partly digest organic-rich samples in QBlock at low temperatures to increase sample handling capacity of microwave digestion vessels
• Concentrate samples before microwave digestion treatment
• Give boric acid treatment for excess HF acid
• Evaporate excess or unwanted acids at the end of digestion
6660 Kennedy Road, Unit 14A Mississauga, ON L5T 2M9 Canada