ATTACHMENT 2 A04 - Salient Characteritics Getinge Autoclave.pdf
PDF 118 KB Posted
- Solicitation number
- 12319826Q5020
About this file
This document is a specification and requirements package for the procurement of a Getinge Lancer LSS-275 Steam Sterilizer with broiler system to replace an outdated Getinge Lancer LSS-130 unit. The objective is to ensure continuous material sterilization capability for station research operations while eliminating delays caused by mechanical malfunction, lack of replacement parts, and safety risks associated with the existing three-decade-old equipment. The replacement system will provide modern safety features and operational efficiency to protect research staff.
The required equipment consists of a standalone single-door steam sterilizer system with a 21"x21"x38" chamber, microcomputer controls, 8" or larger touch screen with password protection, 19 preprogrammed cycles with configurable temperature setpoints in 0.1°F increments and timing functions in one-second increments. The system includes a 480V, AC, 60Hz, 3-phase carbon steel electric boiler with automatic blowdown and automatic feed water pump, interior load car rails, stainless steel open wire shelves, and a single-door enclosure cabinet. The pressure vessel must meet ASME code requirements for working pressures up to 45 psig with air removal through water-ejector method and automatic cooling of wastewater to below 140°F. The contractor shall deliver the system within 60 days of contract award, provide removal of the old unit, complete installation, initial system validation, staff training, ongoing technical support, troubleshooting services, warranty coverage, and comprehensive documentation and standard operating procedures. The equipment includes a 12-month equipment warranty and 15-year pressure vessel warranty.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Copy of ATTACHMENT 3 - RFQ Question Answers.xlsx | XLSX spreadsheet | |
| ATTACHMENT 3 - RFQ Question Format.xlsx | XLSX spreadsheet | |
| Solicitation 12319826Q5020.pdf | ||
| ATTACHMENT 1 A04 - Getinge Autoclave Statement-of-Work.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Required Documents for Getinge Lancer LSS-275 Steam Sterilizer with broiler
Salient Characteristics
The following characteristics make an item equal to the specific item requested:
- One standalone single door steam sterilizer system
- One carbon steel electric boiler with automatic blowdown and automatic feed water pump
- One single door enclosure cabinet
Specifications
The following specifications are required. Currently Getinge Lancer LSS-275 Steam Sterilizer with broiler meets these requirements.
- 21”x21”x38” Chamber.
- Single Manual Door.
- Microcomputer controlled.
- Interior load car rails.
- Stainless steel open wire shelves.
- 8” or larger touch screen.
- Password protection and administrative override.
- Temperature setpoints configurable in 0.1°F (0.1°C) increments.
- Timing functions selectable in one-second increments and accurate to within
0.04%.
- 19 selectable preprogramed cycles.
- Choice of pre-vacuum or gravity displacement cycles.
- 480V, AC, 60Hz, 3-phase carbon steel electric boiler.
- automatic blowdown for carbon steel electric boiler.
- automatic feed water pump for carbon steel electric boiler.
- air removal through water-ejector method.
- pressure vessel meets ASME code requirements for working pressures up to 45 psig
(3.12 bar).
- Standard process piping for chamber and jacket connections.
- Freestanding.
- Single door enclosure cabinet.
- automatic cooling of wastewater to below 140°F (60°C).
- Uniform Plumbing Code compliant.
- Disconnection and removal of the outmoded unit (LSS-130).
- Installation of the new steam sterilizer, boiler, and containment cabinet.
- Connection of utilities to new steam sterilizer system.
- 12 month equipment warranty.
- 15 year new sterilizer pressure vessel warranty.
Statement of Objectives
The objective of this procurement is to acquire a Getinge Lancer LSS-275 Steam Sterilizer with broiler to replace the current outmoded first floor Getinge Lancer LSS-130 steam sterilizer with boiler and ensure smooth continuation of material sterilization for station research. This sterilizer will guarantee that there will be no extended delays in preparation of research materials due to mechanical malfunction or lack of existing replacement parts.
Malfunctioning autoclaves can result in research time and materials lost, costly repairs, and severe injury. The current sterilizer is over three decades old and replacing it would save time, funds, and labor; as well as ensure more modern safety features to protect the research staff using it.
Performance Work Statement
The contractor shall supply, deliver, install, and verify the operational functionality of Getinge Lancer LSS-275 Steam Sterilizer with broiler. The contractor shall provide removal of the old unit, installation, complete initial system validation, and deliver training to designated staff on system operation. The contractor shall ensure that the system has all the necessary components specified (, etc.). Ongoing technical support shall be available for troubleshooting and warranty services during the coverage period. The contractor shall provide documentation and SOPs for system use and maintenance.
Statement of Work
The contractor shall deliver LSS-275 Steam Sterilizer with broiler within 60 days of contract award. The contractor shall provide removal of the old unit, installation, support on setup, operation, and troubleshooting procedures of the system. All deliverables must comply with the specifications in the purchase order and meet the operational standards defined by Getinge. Any deviations must be addressed promptly and corrected at no additional cost.
File details come from the government source that posted it. Updated .