Attachment 1d_ SOW Attachment 3 DD2164.pdf
DoD Form PDF 131 KB Posted
- Solicitation number
- HU000124Q0012
About this file
This document is a DD Form 2164, which is an X-Ray Verification/Certification Worksheet used to inspect and test medical x-ray equipment. The form covers visual inspection of equipment components, operational testing of equipment features like interlocks and beam limiting devices, radiographic certification testing of line voltage, kilovoltage, milliamperage, and phototimer, as well as fluoroscopic certification testing of kilovoltage, brightness control, milliamperage, and timer. The form includes spaces to record the required actions, location, dates, and the inspector's name and signature.
The related federal contract opportunity is a combined synopsis/solicitation for preventative maintenance services on medical lab equipment, issued by the DOD Uniformed Services University of the Health Sciences. This is a small business set-aside solicitation seeking commercial services in accordance with FAR Parts 12 and 13. Refer to the attached Statement of Work for the full details of the requirement.
View the file
Other versions and copies of DD Form 2164
| File | Revision | Posted | Attached to |
|---|---|---|---|
| DD_Form_2164_and_2163__for_Appendix.pdf | Revision not stated | Preventive Maintenance & On-call Repair Services of Radiographic X-Ray Unit |
Who attaches DD Form 2164
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PREVIOUS EDITION IS OBSOLETE.
DD FORM 2164, NOV 78
X-RAY VERIFICATION/CERTIFICATION WORKSHEET
(Use additional sheet for remarks. Identify item by number.)
ACTION
REQUIRED NOT REQUIRED TAKEN
LOCATION (Include Building and Room Number) DATE AND TIME OF SERVICE DATE NEXT SERVICE DUE
I. EQUIPMENT IDENTIFICATION
COMPONENTS MANUFACTURER MODEL
(Include type, style, size, focal spots, etc.)
SERIAL NUMBER
(Housing)
1. CONTROL NO. 1
(Master Control)
2. CONTROL NO. 2
(Room Control)
3. RADIOGRAPHIC TUBE
4. RADIOGRAPHIC TUBE
(Auxiliary tube)
5. FLUOROSCOPIC TUBE
II. VISUAL INSPECTION OF EQUIPMENT
ITEMS FOR VISUAL INSPECTION
ACTION
NOT
REQUIRED TYPE REQUIRED ACTION
TAKEN
INITIAL AND
DATE
6. CERTIFICATION LABELS ARE AFFIXED AND VISIBLE
7. STEEL COUNTERWEIGHT CABLES
8. SHOCK-PROOF HIGH TENSION CABLES
9. TUBE HANGER ASSEMBLY AND YOKES
10. INDICATOR LIGHTS
11. X-RAY TUBES FOR OIL LEAKS
III. OPERATIONAL TESTING OF EQUIPMENT
ITEMS FOR VISUAL INSPECTION
ACTION
NOT
REQUIRED TYPE REQUIRED ACTION
TAKEN
INITIAL AND
DATE
12. INTERLOCKS
13. LOCKS
14. BACKUP SAFETY TIMERS
15. TABLE AND TUBESTAND MOTION
16. BEAM LIMITING DEVICES (Manual and automatic mode)
17. TABLE ANGULATION LIMIT SWITCHES
18. DOES TUBE OVERLOAD PROTECTION CIRCUIT DISABLE EXPOSURE CIRCUIT? YES NO
19. IS THE PRODUCTION OF X-RAYS INHIBITED UNTIL ANODE IS UP TO SPEED? YES NO
20. DOES BRAKE ON HIGH SPEED STATOR OPERATE CORRECTLY?
(Record coast down time for anode after exposure ) YES NO
IV. RADIOGRAPHIC CERTIFICATION
21. SINGLE PHASE LINE VOLTAGE AND LINE DROP
NO LOAD LINE VOLTAGE VOLTS
A. L 1 TO GROUND
B. L 2 TO GROUND
C. L 3 TO L 2
22. THREE PHASE LINE VOLTAGE AND LINE DROP
NO LOAD LINE VOLTAGE VOLTS
A. PHASE A TO B
B. PHASE B TO C
C. PHASE A TO C
LINE DROP TEST VOLTS
D. L 1 TO L 2
23. TRANSFORMER BALANCE 1
A. ANODE VOLTAGE TO GROUND AT 100 KVP
B. CATHODE VOLTAGE TO GROUND AT 100 KVP
LINE DROP TEST VOLTS
D. PHASE A TO B
E. PHASE B TO C
F. PHASE A TO C
24. EXPOSURE TIMER TEST
TIME SETTING ON CONTROL
ACTUAL TIME MEASURED
1 Only required annually.
PREVIOUS EDITION IS OBSOLETE.
DD FORM 2164 Reverse, NOV 78
25. KILOVOLTAGE AND MILLIAMPERAGE VERIFICATION
CONTROL
SETTINGS
KILOVOLTAGE
20 40 60 80 100 120 140 150
MA
MA
MA
MA
MA
MA
MA
MA
26. PENETROMETER FILM DENSITY SATISFACTORY UNSATISFACTORY
27. RADIOGRAPHIC PHOTOTIMER TEST (Record MAS)
BUCKY CHEST OTHER
A. NORMAL
B. LIGHT
C. DARK
V. FLUOROSCOPIC CERTIFICATION
28. KILOVOLTAGE VERIFICATION
MA
SETTING
KILOVOLTAGE PEAK
FLUORO ACTUAL
1.0 40
1.0 60
1.0 80
1.0 100
1.0 120
1.0 MAXIMUM
29. AUTOMATIC BRIGHTNESS CONTROL
SATISFACTORY
UNSATISFACTORY
30. SPOT FILM KILOVOLTAGE VERIFICATION
KILOVOLTAGE PEAK
SPOT FILM SETTING ACTUAL
60 @
80 @
100 @
120 @
MA
MA
MA
MA
31. SPOT FILM MILLIAMPERAGE AND SPACE CHARGE VERIFICATION
FIXED MA
STATION
ACTUAL MILLIAMPERAGE AT
LOW KVP NEUTRAL KVP HIGH KVP
32. FLUOROSCOPIC MILLIAMPERAGE VERIFICATION
MA
STATION
NEUTRAL KVP
SETTING
ACTUAL
MA
33. FLUORO TIMER TEST (Timer set at 5 minutes)
A. WARNING DEVICE ALARMED AT MINUTES
B. TIMER TERMINATED AT MIN SEC
C. DID TIMER TERMINATE EXPOSURE?
YES NO
D. IS TIMER TIMING CORRECTLY WHEN CHECKED AGAINST
CALIBRATED STOP WATCH? YES NO
34. SPOT FILM TIMER TEST SATFY UNSATFY
A. SHORT TIME
B. MEDIUM TIME
C. LONG TIME
35. PHOTOTIMER TEST RECORD MAS
A. NORMAL SETTING
B. LIGHT SETTING (-)
C. DARK SETTING (+)
INSPECTED BY (Type or print name and grade) SIGNATURE
DRAFT
SAMPLE
PREVIOUS EDITION IS OBSOLETE.
DD FORM 2164, NOV 78
PREVIOUS EDITION IS OBSOLETE.
DD FORM 2164 Reverse, NOV 78
X-RAY VERIFICATION/CERTIFICATION WORKSHEET
(Use additional sheet for remarks. Identify item by number.)
ACTION
I. EQUIPMENT IDENTIFICATION
COMPONENTS
MANUFACTURER
MODEL
(Include type, style, size, focal spots, etc.)
SERIAL NUMBER
(Housing)
1. CONTROL NO. 1
(Master Control)
2. CONTROL NO. 2
(Room Control)
3. RADIOGRAPHIC TUBE
4. RADIOGRAPHIC TUBE
(Auxiliary tube)
5. FLUOROSCOPIC TUBE
II. VISUAL INSPECTION OF EQUIPMENT
ITEMS FOR VISUAL INSPECTION
ACTION
NOT REQUIRED
TYPE REQUIRED
ACTION TAKEN
INITIAL AND DATE
6. CERTIFICATION LABELS ARE AFFIXED AND VISIBLE
7. STEEL COUNTERWEIGHT CABLES
8. SHOCK-PROOF HIGH TENSION CABLES
9. TUBE HANGER ASSEMBLY AND YOKES
10. INDICATOR LIGHTS
11. X-RAY TUBES FOR OIL LEAKS
III. OPERATIONAL TESTING OF EQUIPMENT
ITEMS FOR VISUAL INSPECTION
ACTION
NOT REQUIRED
TYPE REQUIRED
ACTION TAKEN
INITIAL AND DATE
12. INTERLOCKS
13. LOCKS
14. BACKUP SAFETY TIMERS
15. TABLE AND TUBESTAND MOTION
16. BEAM LIMITING DEVICES (Manual and automatic mode)
17. TABLE ANGULATION LIMIT SWITCHES
18. DOES TUBE OVERLOAD PROTECTION CIRCUIT DISABLE EXPOSURE CIRCUIT?
19. IS THE PRODUCTION OF X-RAYS INHIBITED UNTIL ANODE IS UP TO SPEED?
20. DOES BRAKE ON HIGH SPEED STATOR OPERATE CORRECTLY?
IV. RADIOGRAPHIC CERTIFICATION
21. SINGLE PHASE LINE VOLTAGE AND LINE DROP
NO LOAD LINE VOLTAGE
VOLTS
A. L 1 TO GROUND
B. L 2 TO GROUND
C. L 3 TO L 2
22. THREE PHASE LINE VOLTAGE AND LINE DROP
NO LOAD LINE VOLTAGE
VOLTS
A. PHASE A TO B
B. PHASE B TO C
C. PHASE A TO C
LINE DROP TEST
VOLTS
D. L 1 TO L 2
23. TRANSFORMER BALANCE 1
A. ANODE VOLTAGE TO GROUND AT 100 KVP
B. CATHODE VOLTAGE TO GROUND AT 100 KVP
LINE DROP TEST
VOLTS
D. PHASE A TO B
E. PHASE B TO C
F. PHASE A TO C
24. EXPOSURE TIMER TEST
TIME SETTING ON CONTROL
ACTUAL TIME MEASURED
1 Only required annually.
25. KILOVOLTAGE AND MILLIAMPERAGE VERIFICATION
CONTROL SETTINGS
KILOVOLTAGE
MA
MA
MA
MA
MA
MA
MA
MA
26. PENETROMETER FILM DENSITY
27. RADIOGRAPHIC PHOTOTIMER TEST (Record MAS)
BUCKY
CHEST
OTHER
A. NORMAL
B. LIGHT
C. DARK
V. FLUOROSCOPIC CERTIFICATION
28. KILOVOLTAGE VERIFICATION
MA SETTING
KILOVOLTAGE PEAK
FLUORO
ACTUAL
1.0 1.0 1.0 1.0 1.0 1.0
MAXIMUM
29. AUTOMATIC BRIGHTNESS CONTROL
30. SPOT FILM KILOVOLTAGE VERIFICATION
KILOVOLTAGE PEAK
SPOT FILM SETTING
ACTUAL
60 @ 80 @ 100 @ 120 @
31. SPOT FILM MILLIAMPERAGE AND SPACE CHARGE VERIFICATION
FIXED MA STATION
ACTUAL MILLIAMPERAGE AT
32. FLUOROSCOPIC MILLIAMPERAGE VERIFICATION
MA STATION
NEUTRAL KVP SETTING
ACTUAL MA
33. FLUORO TIMER TEST (Timer set at 5 minutes)
MINUTES
SEC
C. DID TIMER TERMINATE EXPOSURE?
D. IS TIMER TIMING CORRECTLY WHEN CHECKED AGAINST
CALIBRATED STOP WATCH?
34. SPOT FILM TIMER TEST
SATFY
UNSATFY
A. SHORT TIME
B. MEDIUM TIME
C. LONG TIME
35. PHOTOTIMER TEST
RECORD MAS
A. NORMAL SETTING
B. LIGHT SETTING (-)
C. DARK SETTING (+)
9.0.0.2.20120627.2.874785
| Click this button to save this form.: |
| Click this button to print this form.: |
| Click this Reset button to erase data from all fields.: |
| ResetButton1: |
| Select if action is required.: |
| Select if action is not required.: |
| Select if action is taken.: |
| Enter location (include building and room number).: |
| Enter date and time of service.: |
| Enter date next service due.: |
| Radiographic certification. Exposure timer test. Enter actual time measured.: |
| Spot film timer test. Select if long time unsatisfactory.: |
| Enter coast down time for anode after exposure.: |
| Kilovoltage and Milliamperage verification. MA. Control setting 150. Enter kilovoltage.: |
| Phototimer test. Dark setting. Enter record MAS.: |
| Spot film milliamperage and space charge verification. Actual milliamperage. Enter low KVP.: |
| Spot film milliamperage and space charge verification. Actual milliamperage. Enter neutral KVP.: |
| Spot film milliamperage and space charge verification. Actual milliamperage. Enter high KVP.: |
| Fluoro timer test (timer set at 5 minutes). Enter minutes warning device alarmed at.: |
| Fluoro timer test (timer set at 5 minutes). Enter minutes timer terminated at.: |
| Fluoro timer test (timer set at 5 minutes). Enter seconds timer terminated at.: |
| Enter inspected by (type or print name and grade).: |
| Enter signature.: |