Attachment 1 Specifications-Salient Characteristics.pdf

PDF 140 KB Posted

Attached to
Roche Cobas C 303 (Brand Name or Equal) Clinical Chemistry Analyzer
Solicitation number
12805B23Q0874
Issued by

About this file

This document outlines specifications for a clinical chemistry analyzer required by the Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Plains Area. The analyzer must be able to measure a broad testing menu including hs-CRP, cholesterol, HDL, LDL, triglycerides, cysteine, creatinine, glucose, HbA1c, albumin, alkaline phosphate, urea, nitrites, and total protein. It requires high throughput capabilities, liquid ready-to-use reagents with no preparation needed, auto-calibration and long reagent stability for cost savings, ability to measure sample volumes under 20 microliters, long on-board reagent stability over 100 days for additional cost savings, carryover evasion, ultrasonic mixing, reduced solid waste and carbon footprint from packaging and shipping, and computer independence from internet connectivity. The space constraints for the analyzer are also specified as 4.2 inches wide, 2.9 inches deep, and 3.5 inches high.

The related federal contract opportunity is solicitation number 12805B23Q0874 for a Roche Cobas C 303 clinical chemistry analyzer or equal brand to meet the specifications.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Clinical Chemistry Analyzer

Specifica�ons (Salient/Generic Characteris�cs or Compa�bility Requirements)

• have a broad tes�ng menu to handle tests from different human study requirements. Ability to measure at minimum, hs-CRP, total cholesterol, HDL direct, LDL direct, triglycerides, cysta�n-C, crea�nine, glucose, HbA1c, albumin, alkaline phosphate, urea, nitrites, and total protein.

• high-throughput capabili�es

• no reagent prepara�on; reagents should be all liquid and ready to use

• AutoCal for chemistry tests and long calibra�on stability, which means cost savings

• ability to measure sample volumes less than 20 microliters; the less sample volume required, the less blood we need to draw from study par�cipants

• long on-board stability of reagent >100 days, which means cost savings

• carryover evasion

• ultrasonic mixing

• commitment to reducing solid waste from packaging, water in reagents, and carbon footprint by shipping less reagents because of their stability

• computer does not need to be connected to the internet/network; so�ware allows for predic�ve load list that makes loading reagents and supplies much easier and faster.

• includes ion selec�ve electrode

• historical reliability of reagents, equipment, and service support based on the long partnership with

WHNRC

• analyzer must fit in the space provided: 4.2 � W x 2.9 � D x 3.5 � H

File details come from the government source that posted it. Updated .