This document is a Salient Characteristics specification for a 600 NMR Spectrometer. The requirements specify a comprehensive nuclear magnetic resonance system with detailed technical specifications, including a 600 MHz, 14.1 T spectrometer that must be a new system with specific capabilities and components.
Key requirements include a shielded magnet with a 150-day helium hold time, a minimum 24-auto sample changer, UPS backup system, RF generation system, site-wide NMR software license, and a variable temperature controller supporting -170°C to +250°C. The probe must support multiple isotopes (1H, 19F, 31P to 15N, 39K, and 109Ag), accept 5mm NMR tubes, and include gradient shimming and a single channel pulsed field gradient amplifier. Additional specifications mandate on-site training and installation, in-state application support, and a 1-year warranty.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALENT CHARACTERISTICS: 600 NMR SPECTROMETER
At minimum, the NMR Spectrometer system must include or be capable of the following:
• Must be a 600 MHz, 14.1 T Spectrometer
• Must include installation and training
• Must include shielded magnet
• Must include computer, monitor and printer
• Must include startup cryogens
• Must include minimum 24 auto sample changer
• Must include auto tune probe
• Must include UPS backup system
• Must include RF generation system
• Must include site-wide NMR software license
• Must include variable temperature controller that supports range -170 °C to +250 °C
• Must have a lock and ship system
• Probe must support 1H, 19F, 31P to 15N, 39K, and 109Ag
• Must accept 5mm NMR tubes
• Must include NMR magnet with a minimum helium hold time of 150 days
• Must include pneumatic Anti-Vibration legs with a liquid cryogenic level sensor
• Must be a new NMR system
• Must include gradient shimming
• Must include single channel pulsed field gradient amplifier
• Must include probe with Z-axis pulsed field gradient coil
• 1 year warranty
• On-site training and installation; in-state (e.g. Colorado) application support