This document is a Salient Characteristics specification for a 600 NMR Spectrometer. The requirements specify a comprehensive nuclear magnetic resonance system with detailed technical specifications, including a 600 MHz, 14.1 T spectrometer that must be a new system with specific capabilities and components.

Key requirements include a shielded magnet with a 150-day helium hold time, a minimum 24-auto sample changer, UPS backup system, RF generation system, site-wide NMR software license, and a variable temperature controller supporting -170°C to +250°C. The probe must support multiple isotopes (1H, 19F, 31P to 15N, 39K, and 109Ag), accept 5mm NMR tubes, and include gradient shimming and a single channel pulsed field gradient amplifier. Additional specifications mandate on-site training and installation, in-state application support, and a 1-year warranty.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SALENT CHARACTERISTICS: 600 NMR SPECTROMETER

At minimum, the NMR Spectrometer system must include or be capable of the following:

• Must be a 600 MHz, 14.1 T Spectrometer

• Must include installation and training

• Must include shielded magnet

• Must include computer, monitor and printer

• Must include startup cryogens

• Must include minimum 24 auto sample changer

• Must include auto tune probe

• Must include UPS backup system

• Must include RF generation system

• Must include site-wide NMR software license

• Must include variable temperature controller that supports range -170 °C to +250 °C

• Must have a lock and ship system

• Probe must support 1H, 19F, 31P to 15N, 39K, and 109Ag

• Must accept 5mm NMR tubes

• Must include NMR magnet with a minimum helium hold time of 150 days

• Must include pneumatic Anti-Vibration legs with a liquid cryogenic level sensor

• Must be a new NMR system

• Must include gradient shimming

• Must include single channel pulsed field gradient amplifier

• Must include probe with Z-axis pulsed field gradient coil

• 1 year warranty

• On-site training and installation; in-state (e.g. Colorado) application support