This is a Memorandum from the 97th Logistics Readiness Squadron (97LRS) at Altus Air Force Base, Oklahoma, documenting the Air Force's requirement to purchase an OptiFlash Flash Point Tester for critical fuels analysis on all JAA (Joint Army/Air Force) receipts per Technical Order 42B-1-1.
The OptiFlash Flash Point Tester must comply with ASTM D93 Methods A, B, and C and includes 25 detailed specifications covering heating systems, temperature measurement capabilities, automatic stirring mechanisms, and intelligent ignition sources. Key technical requirements include a heating system with rates from 0.5 to 12 degrees Celsius per minute, an intelligent Pt 100 temperature probe with a range of -50 to +450 degrees Celsius, automatic stirring from 0 to 300 RPM, and a unique thermal flash detection system capable of detecting flash points for all sample types. The device features a 7-inch colored touch screen user interface with barcode reader capability, storage for 500 results and 200 products, USB data export functionality with Excel compatibility, and automatic quality control sample handling with QC charts. The tester includes built-in fire sensors and fire extinguishing systems with external inert gas and alarm relay capability for lab integration. Additional specifications address safety pre-testing, alarm functions, password-protected access levels, automatic calibration and diagnostics, electrical requirements (115VAC or 230VAC, 50-60Hz, 1100W), operating conditions (10°C to 35°C), and physical dimensions of 25 cm wide by 51 cm deep by 56 cm tall, weighing 25 kg. Optional features include an automatic lift arm and ultra-fast optical sensor fire extinguishing system, with available accessories including a printer, barcode reader, and metal temperature sensor.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combo.pdf | ||
| Attachment 4 - Evaluation Factors.pdf | ||
| Attachment 2 - Provisions and Clauses.pdf | ||
| Attachment 3 - Instructions to Offerors.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
97TH AIR MOBILITY WING (AETC)
ALTUS AIR FORCE BASE OKLAHOMA
VICTORY BEGINS HERE!
MEMORANDUM FOR 97LRS/LGLOR
FROM: 97LRS/LGRF
SUBJECT: OptiFlash Flash Point Tester
1. The 97th Logistics Readiness Squadron has a requirement to purchase an OptiFlash Flash Point Tester. The Flash Point tester is required to perform critical fuels analysis on all JAA receipts per TO 42B-1-1. Following is list of the salient characteristics the OptiFlash Flash Point Tester must have.
2. Specifications:
1. ASTM D93 Methods A, B, & C
2. Heating System: Heater band for uniform heating of the heater block. Test method or user-defined heat rate from 0.5 to 12 Celsius per minute.
3. Temperature measurement: Intelligent Pt 100 probe with built in calibration, 10 calibration points glass or metal Pt 100, temperature range -50 degrees Celsius to +450 degrees Celsius, resolution 0.1 degrees Celsius.
4. Sample Stirrer Automatic stirrer, test method or user-defined RPM from 0 to 300
RPM.
5. Ignition Source Intelligent electric igniter with automatic power management over lifetime or automatic gas ignition with gas flame monitoring, test method or user-defined Test Interval from 0.5°C to 5°C.
6. Barometric Pressure Sensor: Built-in sensor for automatic correction of flash point, pressure unit’s mbar, hPa, kPa, mmHg or TorrFour wheel high angle steer.
7. Flash Detection System: Unique thermal flash point detection which can detect flash point for all sample types.
8. User Interface: 7” colored touch screen, alpha numeric data input, barcode reader.
9. Result Documentation: 500 results, 200 products, user defined result reports for printer and LIMS.
10. Printer: Any printer with USB interface and PCL5 or higher. Automatic print out.
11. LIMS interface: Ethernet or RS232, used defined data string, automatic LIMS transfer.
12. Data Export: USB memory stick, import into Excel.
13. QC-Functions: Automatic QC-sample handling and QC-chart.
14. Fire Extinguisher: 2 built-in fire sensors for detection of fire or just flash outside of the cup.
15. Fire extinguishing system with external inert gas, alarm relay to link
OptiFlash to a lab alarm system.
VICTORY BEGINS HERE!
16. Safety pre-test: Safety pre-dips can detect high volatile contamination in normally high flash point samples and avoid a risk of fire.
17. Alarm Functions: Automatic detection of method or safety violations. User selects test termination or alarm message.
18. Password protection: Different access levels for operator, service or lab manager.
19. Calibration and Diagnostics: User defined calibration intervals. Automatic diagnostic functions.
20. Electrical: 115VAC ±10% or 230VAC ±10% with automatic switching; 50-60Hz;
1100 W.
21. Ambient Conditions: Operating temperature: 10°C to 35°C; Humidity: 80% rel. at
35°C.
22. Storage Temperature: -15°C to +55°C.
23. Dimensions & Weight: 25 cm (9.85”) wide, 51 cm (20.1”) deep, 56 cm (22”) tall, 25kg, (55 lb).
24. Options: Automatic lift arm, fire extinguishing system with ultra-fast optical sensor.
25. Accessories: Printer, barcode reader, metal temperature sensor.