This document outlines the salient characteristics for Microbiology Autoclaves, a medical equipment specification for MAMC (Military/Medical Application). The autoclaves must have a 250L capacity with a chamber size of 20x20x38 inches, featuring a touch-sensitive interactive screen with an 18-bit color graphic display and >600 resolution color-active matrix. The equipment requires a single vertical sliding door, loading rack, and two shelves for optimum capacity.
The specification mandates specific technical and regulatory compliance, including noise level requirements (68Dba for water ejector units, 74Dba for vacuum pump units) and adherence to multiple standards such as CE Mark Governing Directives, Pressure Equipment Directive, UL/EN/CSA 61010-1, UL/EN/CSA 61010-2-040, and ASME Code Section VIII, Division 1 for unfired pressure vessels. Additionally, the autoclaves must be environmentally green and provide cycle data records, indicating a focus on both performance and sustainability.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| W81K0225QA084_Steam Sterilizing Units_RFQ Final.pdf | ||
| Attachment 0001 Performance Work Statement Steam Serilizer Unit.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics Microbiology Autoclaves
Medical Equipment
MAMC
• Have a 250L capacity
• Shall have a chamber size no larger than 20x20x38 in dimension
• Have a touch-sensitive interactive screen with 18-bit color graphic display
• Have display features >600 resolution color-active matrix
• Shall contain a single vertical sliding door, loading rack and two shelves for optimum capacity
• Noise level for units with water ejector 68Dba, units with vacuum pump 74Dba
• Shall meet the following regulatory standards: Governing Directives for Affixing the CE Mark, Pressure Equipment Directive, UL/EN/CSA 61010-1, UL/EN/CSA
61010-2-040, ASME Code, Section VIII, Division 1 for unfired pressure vessels.
• Be environmentally Green
• Provide Cycle data Records
Salient Characteristics Microbiology Autoclaves