This is a testing and certification specification document for a Pennsylvania state and local contract opportunity related to chemical fume hoods and biological safety cabinets. The contract covers testing and certification services for multiple types of laboratory equipment including ducted and ductless chemical fume hoods, fuming chambers, biological safety cabinets, drying cabinets, and laminar flow hoods. The contract period runs from July 1, 2026 through June 30, 2027. The award will be made to the lowest responsive and responsible bidder, and sellers must quote all items to be eligible for consideration. Bidders are instructed not to deliver or provide services until notified by the department.
The vendor performing these services must maintain ISO 17025 accreditation for controlled environment testing and calibration. Testing requirements vary by equipment type but include airflow smoke pattern tests, airflow velocity tests, HEPA filter leak tests, down-flow velocity profile tests, inflow velocity tests, and workers comfort and safety tests. All testing must meet manufacturer specifications and applicable state and federal regulations, including NSF/ANSI Z9.5 and 49-2002 standards. The vendor is responsible for adjusting equipment to meet specifications at no additional charge where applicable. If equipment cannot be recertified, the vendor must provide a repair cost estimate to bring the hood within required specifications. The contract may be awarded as a whole or by line item, whichever serves the best interest of the city.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Seller SOP Checklist B2727471.pdf | ||
| Price Increase Language B2727471.docx | DOCX document | |
| SOP Terms and Conditions FY27 B2727471.pdf | ||
| LGBTQ Applicant Data Form B2727471.pdf | ||
| PHL-Contracts Consent Authorization B2727471.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The following requirements must be met:
a. The vendor must be ISO 17025 Accredited for controlled environment testing and calibration
b. The following minimum test must be performed on Chemical Fume Hoods (Ducted)
i. Airflow Smoke Pattern Test
II. Airflow Velocity Test
c. The following minimum test must be performed on Chemical Fume Hoods (Ductless)
i. Airflow Smoke Pattern Test
ii. Airflow Velocity Test
iii. Filter Check
d. The following minimum test must be performed on Chemical Fume Hoods (Fuming Chambers)
i. Airflow Smoke Pattern Test
ii. Airflow Velocity Test
iii. Filter Check
e. The following minimum test must be performed on Biological Safety Cabinets:
i. HEPA Filter Leak Test
ii. Airflow Smoke pattern Test
iii. Down-flow Velocity Profile Test
iv. Inflow Velocity Test
v. Workers Comfort and Safety Tests
f. The following minimum test must be performed on Drying Cabinets:
i. Airflow Smoke Pattern Test
ii. Airflow Velocity Test with velocity profile recorded
g. The following minimum test must be performed on Laminar Flow Hoods:
i. HEPA Filter Leak Test
ii. Airflow Smoke pattern Test
iii. Airflow Velocity Test
iv. Workers Comfort and Safety Tests
Testing must meet manufacturer’s specifications, as well as any state or federal specifications such as those outlined in NSF/ANSI Z9.5 and 49-2002 or other regulations which may apply.
Supplier will adjust equipment to meet specifications at no additional charge where applicable. If hood cannot be recertified the supplier will supply an estimate on the repair cost for bring the hood within the set specifications.