Attach 3 Sysmex Checklist.pdf
PDF 231 KB Posted
- Solicitation number
- HT941024Q2022
- Issued by
-
Defense Health Agency
About this file
This request for quote solicits preventive maintenance and calibration services for various laboratory analyzers at the U.S. Naval Hospital in Okinawa, Japan. The Defense Health Agency Contracting Activity seeks annual preventive maintenance and calibration to be provided under a firm-fixed-price contract from April 1, 2024 through March 31, 2025. Quotes are due by a specified date in 2024 and must demonstrate technical capability to perform the required services, along with past performance on two relevant projects within the past three years. Technical capability will be rated on an acceptable/unacceptable basis, and past performance is more important than price in the evaluation. The awarded contract results from a tradeoff between past performance and price allowing selection of other than the lowest price.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
#13 & #14
Luis.Caburian Highlight
I i
Clinitek Advantus Maintenance Checklist-
Product Name: Uririalysis Analyzerj "Clinitek Advantus" Serial Numb.er: j Requested Content; Regular lnspet:tion l '
Inspection content i
1. Sterilization
2. Confirmation .of Accessories, eti:.
3. Confirmation of initial Crinditio�
3.1 Jnstrumen� Setting/Conteht
3.2 Confirmatibn of i:he initial!condition of Lead Head Test
l ;
4: Confirmation of replaced parts :
5. Confirmation of 1:he operation after repair
5.1 Confirmati�n of Strip Centiering
* Judgment criteria: Betw�en 4 a:nd 40
5.2 Confirmation of)Strip Walking ;
* Judgment criteria: The jlldgment result shall be A or 8.
5.3 Confirmation ofjTouch Screen i
* Judgment crit�ria:·"+" mark shall beldisplayed in a 4mm radius from Touch· point.
i . ! .
5.4 Confirmati6n of the sensitivity of Strip Detector
* JudgmJnt criteria: Thelsensitfvity of each detectOr shall be the range of 10 to 254.' ' ' . .
* Confirm to sound;the answ_er to?e (ble�p sound) at any posi tion when moving step by step from the left side to the right side Of Reagent loading area.
5.5 Confirmation of Display
* VisuallVi coryfirm that e�ch pixel of display does i-iot have abnormality.
5.6 ConfirmatiOn of Lead Head
* Judgr:n�nt criteria: CAii and.CAl values shall be the range of 600 to 1820.
i
DKl and 0:K2 values shall be the range of 1 to 400.
RH2 t;:AL Edge Corr. Shall be-1, 0 or +l.
5.6 Confirmatitjn of Engaging
Result:
Result:·
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
* Judgmeint criteria: No �bnorn:iality shall be occurred after running the c�ntinuous operating of 500 cydes with Regent feeding motion. (Approx. for·l hour) Result:
6. Preparation of.Sh)pment
6.1 ConfirmatiOn of lnstrumerit settilig
6.2 ConfirmatiJn of Appearanbe
6.3 Confirmaticin of Accessori�s
Result:
Result:
Result:
Line Item # 10 and # 11
File details come from the government source that posted it. Updated .