APS-U Radiation Monitor Spec Sheet.xlsx

XLSX spreadsheet 22 KB Posted

Attached to
RFI: Area Radiation Monitor Procurement for APS
Solicitation number
6-S176-I--05497-00
Issued by

About this file

This is a specifications sheet for an Area Radiation Monitor (ARM) system required by Argonne National Laboratory's Advanced Photon Source (APS) facility. The specification document outlines mechanical, electrical, and operational requirements that vendors must meet, categorized as preferred, acceptable, or required capabilities. The single detector unit must not exceed the dimensions of an ADM-610 model, with total configuration specifications of 11 inches wide by 16 inches high by 8 inches deep, including hardware for interlock interface, local audio indicator, light stack, and test button. Weight constraints specify less than 30 pounds per detector and less than 100 pounds for the total configuration. The enclosure must meet NEMA 12 or IP 54 standards minimum, include physical security with a built-in key lock, and operate on 120 VAC ±5%/-10% at 60 Hz through a DC source with wall outlet adapter that is NRTL listed.

This specification sheet is paired with a Request for Information (RFI) issued by the Department of Energy Office of Science through UChicago Argonne, LLC, seeking vendor information for procuring, maintaining, and supporting a system of approximately 50 area radiation monitors for the APS facility. The current ARM system is nearing end-of-life, and each unit must interface with a control system to monitor signal output, current readbacks, and limit trips to enable appropriate facility control responses. Vendors with aligned capabilities must submit responses to sections 1-6 of the "ANL-651 RFI for APS ARMs" document and complete the vendor capability column on this specification sheet by March 3, 2026 at 12:00 PM CST to mollyhunter@anl.gov. This is a pre-solicitation RFI for planning and informational purposes only and does not commit ANL to any payment or procurement action.

View the file

Other files

File Type Posted
Addendum - RFI Questions 1.doc DOC document
ANL-651 RFI for APS ARMs.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1 APS U Radiation Monitor Specifications Preferred Acceptable Required Vendor Capability (Please provide a response to all fields)

Mechanical
DimensionsSingle unit no larger than the ADM-610Total configuration: 11" (wide) x 16" (high) x 8" (deep) which includes any hardware required for interlock interface, local audio indicator, local light stack, and "press to test button".Same as acceptable
WeightLess than 30 pounds per detector, less than 100 pounds total configuration
EnclosureNEMA 12 or betterIP 54 or betterNEMA 12, IP54 or better
Physical securitybuilt-in key lockPadlockLockable
Electrical Power
Power RequiredDC source with wall outlet size adapter (must be NRTL listed)120 vac +5%/-10% 60 Hz (<2 amps)120 vac +5%/-10% 60 Hz or DC
AC ConnectorDC input connector preferredIEC 60320 C14 inlet male inletNot hardwired for quick removal for calibration or replacement
Electrical safety testingUL listedother ANL approved NTRL listing or passes DEEI inspectionANL approved NTRL listing or passes DEEI inspection
Electrical safety designGuarding required for voltage above 50 volts (inside or out)
Detection Capabilities
Particle typesNeutrons, PhotonsNeutrons, PhotonsNeutrons, Photons
Measurement readoutAmbient dose equivalent
Energy Range - Neutronthermal - 1 GeVthermal - 100 MeV
Energy Range - Photon100 keV-100 MeV100 keV - 10 MeV
Measured Dose Range - Neutron.1-1000 mrem/h
Measured Dose Range - Photon.1-1000 mrem/h
Detected Dose Range - Neutron.1 mrem/hr - 25 rem/hr
Detected Dose Range - Photon.1 mrem/hr - 25 rem/hr
Pulsed energy detectionYesYes, mention mode of operation (current or pulse)
Pulsed energy widthns - us
Rate meter channels2 (one neutron, one photon)2 (one neutron, one photon)
Integrated doseYes
Interference Sensitivity
Smart DetectorYes
Gamma RejectionComparable to a system with moderated BF3/3He detectorsYes (state parameters)
Angular dependenceIsotropic (Spherical)Cylindrical geometry
Temperature Operability Range
Humidity Operability Range
Probes capable of remote mountingno repeater required for probes located 130 ft awayrepeater required for probes located 130 ft awayCapable of normal operation with probes located 130 ft away
Remote communication requirements
Communication interfaceEthernetRS485 or RS232CEthernet, RS232C or RS485
EPICS driver supportYesANL will develop with vendor supportYes, will be needed for installation and interface to APS control systems
Menu/setting securitylocal keypad and remote protocol is password protected for calibrations/functional settingsKey switch input instead of password for local protectionlocally and remotely tamper proof from a radalogical safety stand point
Communication ProtocolEPICS IOC application availableEPICS driver availablecommunication protocol documentation available
Data required for remote accessGamma instanaeous level, high and high-high alarm set points. Neutron instanaeous level, high, high-high alarm set points, Alarm A, Alarm B, and probe/system Failure status
Operational Requirements
Local Display2 line LCD - Neutron and Photon mrem/h readoutssingle line LCD auto scrolling - Neutron and Photon mrem/h readoutboth Neutron and Photon readouts
Alarm relay outputs2 isolate - Form A or C contacts for combined Neutron/Photon2 isolate - SSR outputs (needs review) for combined Neutron/Photon2 isolate outputs for combined Neutron/Photon alarm
Probe/Unit failure relay output1 isolate - Form A or C contact1 isolate - SSR output (needs review)1 isolate output for probe/unit failure
Digital I/O connectionsLocal light stack, audible alarm, test button. Remote Local light stack & audible alarm. Alarm/Failure relays on dedicated connector or terminal capable of accepting 18 awg conductorsAll preferred requirements accomplished with external interface enclosure. Some connections may be hardwired as long as main unit can be quickly removed for probe calibration or probe/unit replacement.Local light stack, audible alarm, press to test button. Connections for remote Local light stack, audible alarm. Connections for Alarm/Failure relays. Connections capable of accepting 18 awg conductors. Main unit can be quickly removed for probe calibration or probe/unit replacement.
Visual alarm/failure IndicationOther colors same function visible at 100 metersOK and Alarm/Failure light, visible at 100 meters
Local audible alarm/failure indicationPulse toneSiren, Ringer, Warble, Buzzer, Bell, BeepAudible heard at 100 meters
Remote audible and visual alarm/failure interfaceseparate output relay contacts (SSR maybe acceptable but needs review)may use same outputs as local audible and visual indicator if relibability is not compromisedCapable of having a remote OK and Alarm/Failure visual indicators and audible alarm with a dedicated external connector for cable to remote Light stack and Sonalert
Local ResetNoneNoneNon-latching alarm and failure (latch is done by interlock system)
Test buttonmembrane switch (momentary)flush mount momentary pushbuttonPress to test must simulate an actual radiation event by opening all alarm and failure contacts and causing appropriate local and remote light stack and audible alarm response.