Glove box with purification system for storing high value and oxygen-sensitive materials
Solicitation number
RFP-A26-1148
Issued by

About this file

This document is a specification sheet for a glove box system with integrated purification capabilities designed for storing high-value and oxygen-sensitive materials.

The system features a stainless steel 304 chamber (3 mm thickness) with minimum dimensions of 1200 mm length × 740 mm width × 900 mm height, equipped with an openable tempered glass front window panel (8 mm thick) and optional hinged design. The glove box operates on switchable 110 VAC or 220 VAC, 50/60 Hz single-phase power with maximum consumption of 1.5 kW. The system includes two antechambers (approximately 360 mm ID × 600 mm) and a small tool chamber (approximately 150 mm ID × 220 mm length), with a vacuum pump capacity exceeding 8.5 CFM. The chamber maintains a leak rate below 0.05 vol%/hr and achieves final purity specifications of less than 1 ppm oxygen and less than 1 ppm water concentration at 20°C and 1 atmosphere.

The integrated gas purification system incorporates automated water and oxygen removal using purification cylinders containing BASF R3-11 Copper Catalyst and Lindy Molecular Sieve materials, with moisture removal capacity of approximately 2 kg and oxygen removal capacity of approximately 60 liters. The system includes a built-in regeneration temperature controller for automatic filter regeneration, an automatic purging function to rapidly reduce oxygen below 200 ppm and water below 50 ppm before recycling, and an automatic pressure control system for maintaining constant internal pressure. The equipment requires NRTL certification for compliance.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Glove box with purification system for storing high value and oxygen-sensitive materials

Specifications:

1. Glove box system: Equip with H2O and O2 purification system

2. Working voltage: 110 VAC or 220 VAC switchable, 50/60 Hz, single-phase. Max.

powder: 1.5 kW

3. Glove box chamber:

Case Material: Stainless steel 304 with 3 mm thickness Chamber dimensions: not small than 1200 mm L x 740 mm W x 900 mm H Openable front window panel (tempered glass, 8 mm thick) Hinged-type front window preferred, but optional

4. Chamber condition:

The regeneration Temperature Controller is built-in for automatically regenerating the filters An automatic purging function is built-in for quickly reducing O2 < 200 ppm &

H2O < 50 ppm before recycling.

The automatic pressure control system is built in to keep the pressure at a constant level.

Water concentration: <1 ppm (20 °C, 1 atm) Oxygen concentration: <1 ppm (20 °C, 1 atm)

5. Chamber Leak rate: <0.05 vol%/hr

6. Antechamber: Two antechambers

Antechamber dimension: ~360 mm ID x 600 mm Small chamber dimension (for tools): ~150 mm ID x 220 mm L

7. Pump: vacuum rate: >8.5 CFM

8. Gas purification system:

Automated regeneration process Automated Water and oxygen removal Purification cylinder: made of BASF R3-11 Copper Catalyst and Lindy

Molecular Sieve Moisture removal capacity: ~2 kg Oxygen removal capacity: ~60L

9. Compliance: NRTL certification

File details come from the government source that posted it. Updated .