This document is an Operating Manual for the Structurix M ECO machine, specifically focusing on a detailed checklist for major maintenance. The maintenance checklist is comprehensive and divided into four key sections: Cleaning, Inspection, Functional Capacity, and Settings. The Cleaning section provides instructions for washing tanks, racks, dryer/distribution rollers, filters, and various machine components. The Inspection section outlines critical checks for shipping units, film surface detectors, mechanical components like drive axles and racks, electronic systems, and dryer functionality. The Functional Capacity section requires verification of pump operations, drive motors, fans, sensors, and other machine-critical systems. The Settings section provides space to record specific operational parameters like bath temperatures, replenishing pump rates, and washing volumes. The final step in the maintenance process is a processing test using both large and small films to ensure complete system functionality.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Structurix M ECO Operating Manual |
9.3 CHECKLIST FOR MAJOR MAINTENANCE
Cleaning
Tanks and racks Drain tanks and wash with water. Also wash the racks with water. If appropriate, clean the tanks and racks with cleaning chemicals. Clean the rollers and baffles.
Dryer/distribution rollers Remove the dust from the reflectors of the IR heater. Clean the rollers with a wet cloth.
Filter Remove the dust from the reflectors of the IR heater. Clean the rollers with a wet cloth.
Miscellaneous Clean the caver and side panels of the machine, replenishment tank, and mixer.
Inspection
Shippinq units Film surface detector The rolls must roll freely. Check with display.
Check for strong connections ond cable breaks Rack / Crossover / Dryer Check the torque and wear of the brackets, gear wheels, rollers, and baffles.
Drive axles (horizontal / vertical)
Check the play between the warm gears, gear wheels, brackets and couplings, and possibly lubricate.
Input trag Check for damage and positionina Cover and side panels Check for damage.
Dryer ~ -------------------------- Fan Check blower for function and noise IR heater Check heater, safety control, and reflector
Tank and tube system “ ^ ~ Tanks/ tubes/ connections . Check for leaks, breaks, and blockages.
Circulating and replenishing pumps Check for noise, jams, and leaks.
Electronics General Check the detectors, display block, switches, and plugs to ensure that all contacts are in good condition. Check the cover and the safety switch.
Functional capacity
Circulating and replenishing pumps Chprk thp fi inrtinnnl (“npririty Drive motor and axles (horizontal / vertical) of: Fans and blowers
(possibly with service Bath heating, level sensor and temperature sensor program) Dryer function and dryer levels
Film surface detector, magnetic valves. OK sianal Mixer, accessories
Settings Bath temp, of developer; °C Bath temp, of fixer: °C Developer replenishing pump; ml / min Fixer replenishing pump: ml / min Final washing: 1 / min Intermediate washing: l/min
Final test of maintenance Processing test with large and small films
Revision 02.08 Structurix M ECO (Type 5203/300) GE Insoection Tpchnnlnnip’; GmhH Rnv1^fi7 ‘^hX7n uiirth.