A.2 2026_SOW_XRR_703.pdf
PDF 383 KB Posted
- Attached to
- X-ray Reflectometer
- Solicitation number
- 481221
- Issued by
-
Department of Energy Office of Science
About this file
This is a Statement of Work (SOW) for the procurement of an X-ray diffractometer system by Brookhaven National Laboratory. The project has a 9-month performance duration, with a Lead Principal Investigator of Nathalie Bouet (bouet@bnl.gov, 631-344-3564). The objective is to procure a complete experimental system capable of measuring the reflectivity and band-pass characteristics of high-resolution multilayers. The system must include an enclosure, chiller, copper X-ray tube, computer, control and analysis software, incident and diffracted beam optics, detector, and all associated installation, commissioning, training, and documentation.
The technical requirements specify a copper source with long-fine focus energy-selected to CuKα1 (8.048 keV), operating at maximum 3kW continuous power with 60kV and 80mA maximum settings. The system requires a θ-θ geometry high-resolution goniometer with angular range of -1.5°/3° to +45°/90°, smallest step size of 0.0001° or better, and reproducibility of ±0.0001° or better. The sample stage must accommodate at least a 100mm wafer with five motorized axes of motion. The detector must be a pixel array detector suitable for x-ray reflectivity and grazing incidence diffraction with 2.5° angular coverage minimum. Additional requirements include modular diffracted beam optics, computer-controlled absorber, enclosed beam path system with radiation safety interlocks, software with XRD analysis capabilities, and compliance with NFPA 70 electrical standards. The system must be installed in Building 703's Deposition Laboratory, which is 47' x 23'9" with 8'5" ceilings. Deliverables include Factory Acceptance Test Procedure (6 months ARO), Factory Acceptance Test report (before shipment), system delivery (8 months ARO), installation and commissioning (9 months ARO), Site Acceptance Test including NIST SRM 1976a (after installation), training, and final documentation. The vendor must provide a two-year warranty from final acceptance and supply training for BNL staff. Intellectual property rights are to be determined.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment F W-9.pdf | ||
| Attachment C AMS-FORM-050.pdf | ||
| Attachment G W-8.pdf | ||
| Quotation Sheet Attachment D.pdf | ||
| B.1 RFQ 481221 Rev1.pdf | ||
| Attachment E ACH Authorization.pdf | ||
| Draft PO Attachment A.PDF |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
• � Brookhaven·" ..._ National Labarat□r-y
Statement of Work
Title X-ray Diffractometer acquisition Sponsor [insert Sponsor's Name]
PIMS [insert# if applicable] Lead Pl Nathalie Bouet
& contact info bouet@bnl.gov, 631-344-3564 Duration of 9 months ARO
Performance SOW Version 2
Project Objective
Procure an X-ray diffractometer system capable of measuring the reflectivity and band-pass characteristics of high-resolution multilayers.
Project Scope
This SOW covers the procurement of a complete experimental system including enclosure, chiller, Cu X ray tube, computer, control and analysis software, incident and diffracted beam optics, detector, installation, commissioning, training, and documentation.
Technical requirements
Requirement Description
Copper source Cu source, long-fine focus, energy selected to CuKa1 (8.048 keV) only. 3kW maximum continuous power, with 60kV and 80mA maximum settings
High resolution optics configuration for measuring the reflectivity and band- Optics pass characteristics of high-resolution multilayers
9 - 9 geometry (horizontal sample) high-resolution goniometer:
- On the fly (while moving) as well as step-by-step measurement capability required
Goniometer
- Angular range (9/29): -1.5°/3° to +45°/90° or better with accessories needed mounted
- Smallest step size (9/29): 0.0001 ° or better, reproducibility (9/29): +-
0.0001° or better.
- Capable of accommodating at least a 100mm wafer lying flat and other optics of varying shape with thicknesses up to at least 4mm with five axes of motion
- All axes of sample translation shall be motorized and capable of use for
Sample stage scanning
- X-axis: 100mm travel with 0.01 mm step size or better
- Y-axis: (parallel to goniometer) 100mm travel with 0.01 mm step size or better
- Z-axis: 10mm travel with 0.01 mm step size or better
- Tilt-axis (Chi): +/-10° rotation, 0.01 ° step size or better
July 2026 Statement of Work