The document is a detailed product data sheet for the Setra Model 278 Barometric Pressure Transducer, designed for remote environmental and weather monitoring applications. The transducer features a SETRACERAM™ ceramic sensor that provides high accuracy and stability across wide temperature ranges, with a pressure measurement range of 500 to 1100 hPa/mb. Key specifications include long-term stability of 0.1 hPa/mB per year, low power consumption, a sleep mode for instant startup, and a compact design with a removable terminal block for easy installation.
The device is specifically suited for automated weather stations, data buoys, ships, agriculture meteorology systems, and severe weather detection. It offers multiple output options (0-2.5 VDC or 0-5 VDC), operates across a temperature range of -40 to +140°F, and meets CE and RoHS approvals. The product can be customized through a detailed ordering matrix that allows selection of pressure range, output type, electrical connection, and pressure fitting. In the context of the associated federal contract opportunity, this specific model is intended for the National Oceanic and Atmospheric Administration (NOAA) National Weather Service, with a solicitation for 100 units for the National Data Buoy Center.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Setra’s Model 278 is the ideal solution for measuring barometric pressure for remote environmental applications. The 278 is designed using the SETRACERAM™ ceramic sensor, enabling it to meet stringent accuracy requirements over wide operating temperatures in remote applications.
The small footprint and removable terminal block on the 278 makes installation fast and easy. The 278 is ideal for solar powered applications because of its low power consumption and sleep mode feature. Under normal operation, this feature minimizes current draw when readings are not being taken.
Designed for remote sensing applications The Model 278 pressure transducer is designed to be used in remote applications that require low power consumption. Its sleep mode feature allows for instant startup and fast readings.
Improved performance with ceramic sensor The 278 utilizes a variable capacitance sensor that is made using ceramic material fused together with glass and gold to form the SETRACERAM™ pressure element. This stable material and design offers class leading thermal performance and low hysteresis, allowing it to be integrated into demanding installations. The ceramic sensor enables improved performance compared to other stainless steel sensors, enabling the 278 to give accurate measurements and better test results.
Flexibility in installation The Model 278 is designed with a compact footprint for quick installation.
The removable terminal block provides easy wiring. Its mounting holes are designed to fit industry standard grid systems to maximize the use of panel space while minimizing your time at the job site.
Model 278 Barometric Pressure Transducer
Features
• Long-term stability: 0.1 hPa/mB per year
• Sleep mode for instant startup
• Removable terminal strip module for easy wiring
• Footprint configured for easy drop-in replacement
• Calibration NIST traceable
• Wide operating voltage 9.5 to 28 VDC
• Meets CE conformance standards
• Ideal for automated weather stations
• Low power consumption
• Relied on for severe weather detection
Applications
• Automated Weather Stations (AWS)
• Data Buoys and Ships
• Agriculture Metrology System
• AWOS/ASOS Systems
• High Accuracy Barometric Pressure Measurement
800.257.3872 I www.setra.com©Setra Systems Inc. All rights reserved.
The Setra Systems name and logo are registered trademarks of Setra Systems Inc.
Model 278 I Barometric Pressure Transducer
Specifications Performance data
Pressure range hPa/mb 500 to 1100 600 to 1100 800 to 1100
Temperature at: Accuracy (hpa/mb)1
20oC (+68oF) ±0.6 ±0.5 ±0.3
0 to 40oC (+23o to +104oF) ±1.2 ±1.0 ±0.6
20 to 50oC (-4o to +122oF) ±2.0 ±1.5 ±1
-40 to 60oC (-40o to +140oF) ±2.5 ±2.0 ±1.5
Non-linearity ±0.5 ±0.4 ±0.25
Hysteresis ±0.06 ±0.05 ±0.03
Non-repeatability ±0.04 ±0.03 ±0.02
Resolution 0.01 mB
Long term stability 0.1 mB/yr
Warm-up downshift <1 Sec. from shut-mode (warm-up <0.1 mb Max.)
Response time <100 mSec
Proof pressure 1500 hPa
Burst pressure 2000 hPa
Environmental data
Operating temperature4 °F(°C) -40 to +140 (-40 to +60)
Storage temperature °F(°C) -76 to +248 (-60 to +120)
Pressure media Non condensing air or gas.
Approvals CE, RoHS
Electrical data Circuit 3 or 4-Wire
Output 2 0–2.5 VDC
0–5 VDC
Excitation3 9.5 to 28 VDC
Output impedance <10 ohms
Output noise <50 Microvolts
Current consumption 3mA Nominal (operating mode) 1uA (sleep mode)
Physical description Case Stainless steel and polyester
Pressure Fitting 1/8” (ID dia.) bardbed fitting
Electrical Connection 5-Pin terminal block
Dimensions 3.6” x 2.4” x 1.0”
Weight 4.8 ox (135g)
1The root sum squared (RSS) of end point non-linearity, hysteresis, non-repeatability, and calibration uncertainty.
2Internal regulation minimizes effect of excitation variation, with <0.02 mb output change of 9.5 VDC to 28 VDC range.
3Zero output saturates at about 20 mV.
© Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. | www.setra.com | 800.257.3872
Model 278 I Barometric Pressure Transducer
Ordering information Example part number: 2781600MA1B2BT1 Model 278, Pressure Transducer 600 to 1100 hPa/ma, Absolute Pressure, 1/8” Barbed Fitting,0 to 5 VDC Output, 5–Pin Terminal Block.
2 8 A 1 B7 1 T 1
[1] [2] [3] [5] [6]
Model
2781 Model 278
[1] [6][3] [4]
Output/Exc.
2Y 0 to 2.5 VDC/9.5 to 28 VDC
2B 0 to 5 VDC/9.5 to 28 VDC
Electrical conn.
T1 5-pin terminal block
[5]
Pressure conn.
1B 1/8” push tube fitting
[4]
Pressure type
A Absolute
[2]
Pressure range
500M 500 to 1100 hPa/mb
600M 600 to 1100 hPa/mb
800M 800 to 1100 hPa/mb
© Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. | www.setra.com | 800.257.3872
Dimensions
SS
P-
Re v.
D
/2
2.36 59.9
3.02 76.7
2X .290
2X .140
MOUNTING HOLES
FOR #6 OR #4 SCREW
CONNECTOR, 5 POSITION
BL 3.5 SCREW CLAMP
BARBED PRESSURE FITTING
1.00 25.4
2.36
3.56 90.3
2.36 59.9
3.02 76.7
2X .290
2X .140
MOUNTING HOLES
FOR #6 OR #4 SCREW
CONNECTOR, 5 POSITION
BL 3.5 SCREW CLAMP
BARBED PRESSURE FITTING
1.00 25.4
2.36
3.56 90.3
[59.9]
2.36 2X Ø.140
Mounting holes for #6 or #4 screw
Barbed pressure fitting
2X Ø.290
Connector, 5 position BL 3.5 screw clamp
[76.7] 3.02
[25.4] 1.00
[mm] in
[60] 2.36
[90.3] 3.56
File details come from the government source that posted it. Updated .