36E77619R0027-006.pdf

PDF 4 MB Posted

Attached to
Upgrade Interventional Radiology Rooms NRM Minneapolis VAMC
Solicitation number
36E77619R0027
Issued by

About this file

36E77619R0027 Attachment 5 - Final Dwgs Siemens Equipment 1R-139 - VAMC Minneapolis IR - 2-11-19.pdf

View the file

Other files

File Type Posted
36E77619C0040-000.docx DOCX document
36E77619R0027-0005000.docx DOCX document
36E77619R0027-0004001.docx DOCX document
36E77619R0027-0004000.docx DOCX document
36E77619R0027-0003005.pdf PDF
36E77619R0027-0003002.pdf PDF
36E77619R0027-0003000.docx DOCX document
36E77619R0027-0003003.pdf PDF
36E77619R0027-0003001.docx DOCX document
36E77619R0027-0003004.doc DOC document
36E77619R0027-0002001.pdf PDF
36E77619R0027-0002000.docx DOCX document
36E77619R0027-0001000.docx DOCX document
36E77619R0027-001.docx DOCX document
36E77619R0027-008.pdf PDF
36E77619R0027-002.pdf PDF
36E77619R0027-010.pdf PDF
36E77619R0027-005.pdf PDF
36E77619R0027-003.pdf PDF
36E77619R0027-011.pdf PDF
36E77619R0027-009.pdf PDF
36E77619R0027-007.pdf PDF
36E77619R0027-004.pdf PDF
36E77619R0027-012.docx DOCX document
36E77619R0027-000.docx DOCX document
Show all 25

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

MINNEAPOLIS VAMC

1801346

A-101ATTENTION:

R E

F E

R E

N C

E

D

O C

U M

E N

T

N

O T

F

R

C

N

S T

R U

C T

I O

N

WITH TILTING TABLE

AutoCAD SHX Text

ARTIS Q/Q.ZEN/ZEE BIPLANE ANGIO

AutoCAD SHX Text

REV. 19

AutoCAD SHX Text

DESIGNATION

AutoCAD SHX Text

PG NUMBER

AutoCAD SHX Text

DATE

AutoCAD SHX Text

RESOURCE LIST (SMS USE ONLY)

AutoCAD SHX Text

ARTIS Q / Q.ZEN BIPLANE ANGIO

AutoCAD SHX Text

AXAQ-070.891.02.02.02

AutoCAD SHX Text 10.15

AutoCAD SHX Text

1 VETERANS DR, MINNEAPOLIS, MN 55417

AutoCAD SHX Text

BIPLANE ROOM - ARTIS Q BIPLANE ANGIO

AutoCAD SHX Text 30193898

AutoCAD SHX Text

AS NOTED

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

M. YATZUS

AutoCAD SHX Text

THE USE OR REPRODUCTION OF

AutoCAD SHX Text

THIS TITLE BLOCK WITHOUT

AutoCAD SHX Text

PROJECT #:

AutoCAD SHX Text

SHEET:

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

REF. #:

AutoCAD SHX Text

FULL EXTENT OF THE LAW.

AutoCAD SHX Text

SIEMENS AUTHORIZATION WILL

AutoCAD SHX Text

RESULT IN PROSECUTION UNDER

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DATE:

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

CHECKED:

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

PROJECT MANAGER:

AutoCAD SHX Text

TEL:

AutoCAD SHX Text

FAX:

AutoCAD SHX Text

VMAIL:

AutoCAD SHX Text

EXT:

AutoCAD SHX Text

EMAIL:

AutoCAD SHX Text

JIM IRONSIDE

AutoCAD SHX Text

(847) 271-2567

AutoCAD SHX Text james.ironside@siemens-healthineers.com

AutoCAD SHX Text THIS SET OF FINAL DRAWINGS IS NOT REFLECTIVE OF THE LATEST NOT REFLECTIVE OF THE LATEST REFLECTIVE OF THE LATEST SALES CONFIGURATION. IT HAS BEEN MODIFIED TO REFLECT ANTICIPATED CHANGES TO YOUR CONFIGURATION. IF REQUESTED CHANGES ARE NOT IMPLEMENTED, A REVISION WILL BE REQUIRED. ADDITIONAL CHANGES TO THIS SALES CONFIGURATION MAY ALSO REQUIRE A REVISION TO THIS DRAWING SET. SIEMENS IS NOT RESPONSIBLE FOR CONSTRUCTION COSTS ASSOCIATED WITH CHANGES THAT OCCUR FROM THIS PLAN MODIFICATION.

AutoCAD SHX Text

NOTE: MIV TO REDUCE LENGTH OF MAVIG TRACK ONSITE BY 5'-4" AS SHOWN.

AutoCAD SHX Text

AutoCAD SHX Text

SPACE FOR CUSTOMER EQUIPMENT

AutoCAD SHX Text IT IS THE RESPONSIBILITY OF THE CUSTOMER/CONTRACTOR TO PROVIDE A MEANS OF MOUNTING THE SYNGO X PC TOWER OFF OF FINISHED FLOOR FOR DAMAGE PROTECTION AGAINST TIP-OVER, FLUIDS, IMPACT, ETC.

AutoCAD SHX Text

PROPOSED 2'-6" DEEP COUNTERTOP PROVIDED BY CUSTOMER/ CONTRACTOR.

AutoCAD SHX Text SINKS, COUNTERTOPS AND ALL CASEWORK SHOWN IS SUGGESTED AND MUST BE DESIGNED SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR.

AutoCAD SHX Text

RAYSAFE DISPLAY LOCATED BY CUSTOMER, INSTALLED BY RAYSAFE

AutoCAD SHX Text MEDRAD INJECTOR INSTALLED BY INJECTOR INSTALLED BY SIEMENS WITH STARTUP PERFORMED BY MEDRAD/BAYER

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

BIPLANE ROOM

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

ISOCENTER FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

ORIENTATION POINT OF PATIENT TABLE AT FLOOR

AutoCAD SHX Text

SWIVEL AREA, HEAD/LEFT SIDE POSITION (MOTORIZED MOVEMENT)

AutoCAD SHX Text

PATIENT TABLE MOVEMENT RANGE (HEAD-END)

AutoCAD SHX Text

PATIENT TABLE MOVEMENT RANGE (FOOT-END)

AutoCAD SHX Text

SAFETY/ SERVICE CLEARANCE

AutoCAD SHX Text

SERVICE CLEARANCE

AutoCAD SHX Text

NOTE: THIS OPTION REQUIRES A CS

AutoCAD SHX Text NOTE: MIV TO CUT MAVIG TRACK ONSITE IF REQUIRED TO MAINTAIN RAIL OVERHANG LIMITATION.

AutoCAD SHX Text

HEIGHT

AutoCAD SHX Text

CEILING

AutoCAD SHX Text

REQUIREMENT

AutoCAD SHX Text

9 FT. - 6 3/8 IN.

AutoCAD SHX Text

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

R-101R(C) VERSION DATED 12/03/18

AutoCAD SHX Text

APPROVED BY CUSTOMER FOR FINALS

AutoCAD SHX Text

-ISSUE BLOCK-

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

DATE

AutoCAD SHX Text

SYM

AutoCAD SHX Text IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION

AutoCAD SHX Text THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED

AutoCAD SHX Text

AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN

AutoCAD SHX Text

MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.

AutoCAD SHX Text

DOCUMENTS FOR REFERENCE.

AutoCAD SHX Text THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED

AutoCAD SHX Text EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION

AutoCAD SHX Text

PHYSICIST TO SPECIFY RADIATION PROTECTION.

AutoCAD SHX Text

ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.

AutoCAD SHX Text THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.

AutoCAD SHX Text

ALL RIGHTS ARE RESERVED.

AutoCAD SHX Text

STATE AGENCY REVIEW

AutoCAD SHX Text PRIOR TO SIEMENS EQUIPMENT INSTALLATION, APPROVAL OF CONSTRUCTION OR STRUCTURAL MODIFICATIONS UTILIZING X-RAY FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES, MUST BE OBTAINED BY THE CUSTOMER FROM THE APPROPRIATE STATE AGENCY, IF APPLICABLE.

AutoCAD SHX Text

ARCHITECTURAL EQUIPMENT PLAN

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text 1 8

AutoCAD SHX Text

TRANSPORTING REQUIREMENTS

AutoCAD SHX Text LARGEST CRATE WITH PACKING: 120"(L) x 49.2"(W) x 84.7"(H), 2,458 LBS. LARGEST INDIVIDUAL PIECE WITH CARRIAGE (MIN. DOOR OPENING): 114.2"(L) x 42.6"(W) x 77.2"(H), 1,984 LBS. CEILING RAILS ARE 14 FT.(L) x 3"(W) x 3"(H) MIN. CORRIDOR WIDTH: 82.7"

AutoCAD SHX Text

ARCHITECTURAL NOTES

AutoCAD SHX Text

1) ALL PRELIMINARY EQUIPMENT LAYOUTS SUBMITTED BY SIEMENS HEALTHCARE ARE BASED ON THE RECOMMENDED SPACE NECESSARY FOR THE OPERATION AND SERVICEABILITY OF THE EQUIPMENT BEING PROPOSED. SIEMENS WILL NOT SUBMIT AN EQUIPMENT LAYOUT THAT IS NOT IN THE BEST INTEREST OF BOTH THE CUSTOMER AND SIEMENS. ALL EQUIPMENT LAYOUTS ARE BASED EITHER ON AN ACTUAL SITE SURVEY OR ARCHITECTURAL DRAWINGS SUPPLIED TO SIEMENS. SIEMENS WILL NOT BE RESPONSIBLE FOR ANY ALTERATIONS THAT ENCROACH WITHIN DESIGNATED SAFETY AND SERVICE CLEARANCE ZONES AS INDICATED ON DRAWINGS (I.E.. PIPE CHASES, VENTILATION DUCTS, CASEWORK, AND SOFFITS, ETC.) MADE BY THE CUSTOMER OR REQUIRED BY A CUSTOMER'S ARCHITECTURAL FIRM ONCE PRELIMINARY DRAWINGS HAVE BEEN SUBMITTED AND APPROVED. DO NOT ALTER ANY SPECIFICATIONS AND/OR DIMENSIONS WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM SIEMENS PROJECT MANAGER. 2) SIEMENS HEALTHCARE IS NOT AN ARCHITECTURAL OR ENGINEERING FIRM. DRAWINGS SUPPLIED BY SIEMENS ARE NOT CONSTRUCTION DRAWINGS. THEREFORE, THESE DRAWINGS ARE TO BE USED ONLY FOR INFORMATION TO COMPLEMENT ACTUAL CONSTRUCTION DRAWINGS AVAILABLE FROM A CUSTOMER APPOINTED ARCHITECTURAL REPRESENTATIVE OR A CUSTOMER'S ENGINEERING DESIGN GROUP. THE CUSTOMER'S ARCHITECT AND GENERAL CONTRACTOR SHALL BE ULTIMATELY RESPONSIBLE FOR COMPLIANCE WITH ALL APPLICABLE CODES AND PROFESSIONAL DESIGN REQUIREMENTS INCLUDING OSHA/NEC SAFETY CLEARANCE REQUIRMENTS IN ADDITION TO SIEMENS-REQUIRED SAFETY/SERVICE CLEARANCES SHOWN. 3) THE CUSTOMER IS RESPONSIBLE FOR ALL ROOM AND AREA PREPARATION COSTS, PROFESSIONAL FEES, PERMITS, REPORTS, AND INSPECTION FEES. 4) EQUIPMENT WARRANTIES, EXPRESSED OR IMPLIED ON THE PART OF SIEMENS SHALL BE CONTINGENT UPON STRICT COMPLIANCE WITH THE ARCHITECTURAL, STRUCTURAL, ELECTRICAL, MECHANICAL AND RECOMMENDATIONS AND REQUIREMENTS CONTAINED IN THESE DRAWINGS, UNLESS SPECIFIED OTHERWISE. 5) ALL DIMENSIONS SHOWN ARE FROM FINISHED SURFACES UNLESS SPECIFIED OTHERWISE. 6) THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION PHYSICIST. ACTUAL PROTECTION REQUIREMENTS SHALL BE SPECIFIED BY A REGISTERED RADIATION PHYSICIST AT CUSTOMER'S ENGAGEMENT AND EXPENSE. RESPONSIBILITY FOR ALL INFORMATION AS TO THE ROOM LOCATION, USE, AND NUMBER OF ANTICIPATED EXAMINATIONS TO BE PERFORMED PER TIME PERIOD SHALL BE PROVIDED TO THE PHYSICIST BY THE CUSTOMER. THE CUSTOMER SHALL FURTHER TAKE ALL RESPONSIBILITY IN THE COMMUNICATION AND COORDINATION OF ACTIVITIES OF THE RADIATION PHYSICIST AND THE ARCHITECTURAL REPRESENTATIVE. 7) SIEMENS HEALTHCARE SHALL BE RESPONSIBLE FOR SIEMENS EQUIPMENT INSTALLATION, CALIBRATION, CONNECTION AND INSTALLATION OF SIEMENS PROVIDED CABLES. THE CUSTOMER/ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR TERMINATIONS OF CUSTOMER/ELECTRICAL CONTRACTOR-SUPPLIED CABLES TO SIEMENS EQUIPMENT. IN THE EVENT THAT SPECIFIC TRADE RULES OR LICENSE REQUIREMENTS PROHIBIT THIS, THE CUSTOMER SHALL INITIATE THE SERVICES OF APPROVED OTHER CONTRACTORS AND PAY FOR SELECTED, APPROVED PARTIES TO PERFORM THIS WORK WITH SUPERVISION PROVIDED BY SIEMENS. CALIBRATION WHEN ACCOMPLISHED OUTSIDE OF NORMAL INSTALLATION SEQUENCES DUE TO CONTRACTOR OR TRADE RULE ACTIONS OR REQUIREMENTS SHALL BE SUPPORTED BY, CHARGED TO, AND ACCEPTED BY THE CUSTOMER AS AN ADDITIONAL INSTALLATION EXPENSE. 8) THE CUSTOMER SHALL COORDINATE WITH SIEMENS PROJECT MANAGER THE LOCATIONS AND TRAVEL OF ALL ANCILLARY EQUIPMENT TO BE CEILING OR WALL MOUNTED (I.E.: O.R. LIGHTS, MEDICAL GAS COLUMNS, PHYSIOLOGICAL MONITORING INJECTORS, CRT PLATFORMS, SPRINKLER HEADS, SMOKE DETECTORS, ELECTRICAL OUTLETS, HVAC GRILLES, SPEAKERS, AND GENERAL ROOM LIGHTING, ETC.). 9) THE GENERAL CONTRACTOR/CUSTOMER SHALL BE RESPONSIBLE FOR ALL FINAL PAINT, TOUCH-UP AND ANY COSMETIC OR TRIM WORK WHICH NEEDS TO BE OR IS REQUIRED TO BE COMPLETED AFTER THE INSTALLATION OF THE SIEMENS EQUIPMENT AND ANY ASSOCIATED SUPPORT APPARATUS.

AutoCAD SHX Text 0.5mT

AutoCAD SHX Text 1.0mT

AutoCAD SHX Text

MAGNETIC FIELD PRECAUTIONS

AutoCAD SHX Text 0.05mT

AutoCAD SHX Text 0.05mT

AutoCAD SHX Text 0.2mT

AutoCAD SHX Text

(10 GAUSS)

AutoCAD SHX Text

(5 GAUSS)

AutoCAD SHX Text

(2 GAUSS)

AutoCAD SHX Text

(0.5 GAUSS)

AutoCAD SHX Text

(0.5 GAUSS)

AutoCAD SHX Text THE PRESENCE OF MAGNETIC FIELDS IN THE VICINITY OF EQUIPMENT MAY HAVE AN ADVERSE EFFECT. IT IS THE CUSTOMER'S RESPONSIBILITY TO VERIFY THAT THE FOLLOWING VALUES ARE NOT EXCEEDED.

AutoCAD SHX Text

MAXIMUM ALLOWABLE MAGNETIC FIELD

AutoCAD SHX Text

DEVICES

AutoCAD SHX Text

COMPUTERS, MAGNETIC DISK DRIVES, OSCILLOSCOPES, PROCESSORS

AutoCAD SHX Text

X-RAY TUBES, B/W MONITORS, MAGNETIC DATA CARRIERS, DATA STORAGE DRIVES

AutoCAD SHX Text

SIEMENS CT SCANNERS

AutoCAD SHX Text

CRT MONITORS, SIEMENS LINEAR ACCELERATORS

AutoCAD SHX Text X-RAY IMAGE INTENSIFIERS, GAMMA CAMERAS, PET/CYCLOTRON, OTHER LINEAR ACCELERATORS

AutoCAD SHX Text

MAGNETIC FIELDS SHOULD BE MEASURED PRIOR TO DELIVERY

AutoCAD SHX Text TEMPERATURE RANGE: -4° F TO 158° F RELATIVE HUMIDITY: 20% TO 95% WITHOUT CONDENSATION AIR PRESSURE: 700 hPa TO 1060 hPa

AutoCAD SHX Text

ENVIRONMENTAL CONDITIONS FOR TRANSPORT/STORAGE

AutoCAD SHX Text

PARK POSITION

AutoCAD SHX Text

CEILING STAND IN

AutoCAD SHX Text *1 ROOM HEIGHT MEASURED FROM THE HIGHEST POINT OF THE FINISHED FLOOR (INCLUDING FLOOR COVERING) TO THE LOWEST POINT OF THE CEILING SUBSTRUCTURE.

AutoCAD SHX Text

HEAD SIDE POSITION

AutoCAD SHX Text

FLOOR STAND IN

AutoCAD SHX Text P2

AutoCAD SHX Text P1

AutoCAD SHX Text

ARTIS Q/Q.ZEN/ZEE BIPLANE SYSTEM ELEVATION

AutoCAD SHX Text 1/4"

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text T1

AutoCAD SHX Text THE "B10" FIXPOINT IS A THRU-FLOOR PENETRATION PROVIDED AS A MEANS OF CONNECTING A TABLE INJECTOR OR OTHER NON-SIEMENS ACCESSORIES AT THE TABLE FOR WHICH CABLE ROUTING THROUGH THE SIEMENS TABLE BASE WOULD BE PROHIBITED. THIS OPENING CAN BE ELIMINATED AS REQUIRED, OR IT CAN BE COMBINED INTO A CUSTOMER-SUPPLIED GAS/UTILITY BOX ON THE FLOOR. NOTE: THE 2'-0" DISTANCE FROM TABLE ISOCENTER SHOWN HERE APPLIES TO A CABLE PENETRATION POINT ONLY. SEE DIAGRAM FOR SEE DIAGRAM FOR POSSIBLE BOX PLACEMENT AND HEIGHT OPTIONS. VALIDATE EXACT PLACEMENT WITH SIEMENS PROJECT MANAGER.

AutoCAD SHX Text B10

AutoCAD SHX Text

PATIENT TABLE

AutoCAD SHX Text

C-ARM ISOCENTER IN HEAD-END POSITION

AutoCAD SHX Text

MAX.

AutoCAD SHX Text

DCS LARGE DISPLAY

AutoCAD SHX Text D1

AutoCAD SHX Text 2'-0"

AutoCAD SHX Text

MAX.

AutoCAD SHX Text

AutoCAD SHX Text

EQUIPMENT LEGEND

AutoCAD SHX Text

NO

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

SMS

AutoCAD SHX Text

WEIGHT

AutoCAD SHX Text

BTU/HR

AutoCAD SHX Text

DIMENSIONS (INCHES)

AutoCAD SHX Text

REMARKS

AutoCAD SHX Text

SYM

AutoCAD SHX Text

(LBS)

AutoCAD SHX Text

TO AIR

AutoCAD SHX Text W

AutoCAD SHX Text D

AutoCAD SHX Text H

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ACE (ARCHIVE CONTROL EXTENSION)

AutoCAD SHX Text

AutoCAD SHX Text N/A

AutoCAD SHX Text 12 1/4

AutoCAD SHX Text 11 3/4

AutoCAD SHX Text

AutoCAD SHX Text

ON COUNTER

AutoCAD SHX Text

AutoCAD SHX Text

CR1

AutoCAD SHX Text

CONTROL ROOM DISTRIBUTOR

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 41 1/2

AutoCAD SHX Text 8 1/4

AutoCAD SHX Text 16 1/8

AutoCAD SHX Text

WALL MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

KEYBOARD

AutoCAD SHX Text 2.2

AutoCAD SHX Text

AutoCAD SHX Text 17 1/2

AutoCAD SHX Text 6 1/8

AutoCAD SHX Text 2 1/8

AutoCAD SHX Text

ON COUNTER

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

19" LIVE DISPLAY (FLOOR PLANE)

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 16 1/2

AutoCAD SHX Text 8 1/4

AutoCAD SHX Text 13 1/2

AutoCAD SHX Text

ON COUNTER

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

19" LIVE DISPLAY (CEILING PLANE)

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 16 1/2

AutoCAD SHX Text 8 1/4

AutoCAD SHX Text 13 1/2

AutoCAD SHX Text

ON COUNTER

AutoCAD SHX Text

AutoCAD SHX Text

XWP

AutoCAD SHX Text

SYNGO X WORKSTATION TOWER, KEYBOARD, MONITOR

AutoCAD SHX Text

AutoCAD SHX Text 2,730

AutoCAD SHX Text 7 1/2

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ON FLOOR

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

TABLE CONTROL MODULES

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 8 3/4

AutoCAD SHX Text 3 1/2

AutoCAD SHX Text

ON TABLE OR TROLLEY

AutoCAD SHX Text

AutoCAD SHX Text D1

AutoCAD SHX Text

DCS EXTENDED W/ LARGE DISPLAY

AutoCAD SHX Text

AutoCAD SHX Text 1,706

AutoCAD SHX Text

AutoCAD SHX Text 60 3/4

AutoCAD SHX Text 50 3/4

AutoCAD SHX Text

CEILING SUSPENDED

AutoCAD SHX Text

AutoCAD SHX Text D2

AutoCAD SHX Text

MONITOR CART WITH (2) 18" FLAT SCREEN DISPLAY

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 32 1/2

AutoCAD SHX Text 28 1/4

AutoCAD SHX Text 60 3/8

AutoCAD SHX Text

MTD. ON CASTERS

AutoCAD SHX Text

AutoCAD SHX Text P1

AutoCAD SHX Text

ARTIS Q BIPLANE ANGIO FLOOR STAND W/ MOUNTING PLATE

AutoCAD SHX Text 1,466

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

C-ARM FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text P2

AutoCAD SHX Text

ARTIS Q BIPLANE ANGIO CEILING STAND W/ LONGITUDINAL RAILS

AutoCAD SHX Text 1,248

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

C-ARM CEILING MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text T1

AutoCAD SHX Text

PATIENT TABLE (TILT TABLE)

AutoCAD SHX Text 1,169

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

UPPER BODY RADIATION SHIELD 4 M TRACK

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

TRACK MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

UPPER BODY RADIATION SHIELD 4 M TRACK

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

TRACK MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

MAVIG LAMP

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

PU1

AutoCAD SHX Text

POLYDOROS A100 (POWER UNIT 1)

AutoCAD SHX Text

AutoCAD SHX Text 4,094

AutoCAD SHX Text 31 1/2

AutoCAD SHX Text 17 1/8

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

PU2

AutoCAD SHX Text

POLYDOROS A100 (POWER UNIT 2)

AutoCAD SHX Text

AutoCAD SHX Text 4,094

AutoCAD SHX Text 31 1/2

AutoCAD SHX Text 17 1/8

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

CC

AutoCAD SHX Text

CABLE CABINET

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 31 1/2

AutoCAD SHX Text 17 1/8

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

SC1

AutoCAD SHX Text

SYSTEM CONTROL CABINET

AutoCAD SHX Text

AutoCAD SHX Text 5,459

AutoCAD SHX Text 31 1/2

AutoCAD SHX Text 17 1/8

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

SC3

AutoCAD SHX Text

SYSTEM CONTROL CABINET 2

AutoCAD SHX Text

AutoCAD SHX Text 4,094

AutoCAD SHX Text 23 1/2

AutoCAD SHX Text 17 1/8

AutoCAD SHX Text

AutoCAD SHX Text

FLOOR MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

LDC

AutoCAD SHX Text

LARGE DISPLAY CONTAINER FOR DCS LD

AutoCAD SHX Text

AutoCAD SHX Text 1,535

AutoCAD SHX Text

AutoCAD SHX Text 37 1/2

AutoCAD SHX Text 28 3/8

AutoCAD SHX Text

MTD. ON CASTERS

AutoCAD SHX Text

AutoCAD SHX Text

IS

AutoCAD SHX Text

AXIS IMAGE SYSTEM

AutoCAD SHX Text

AutoCAD SHX Text 6,483

AutoCAD SHX Text 33 1/2

AutoCAD SHX Text 37 1/8

AutoCAD SHX Text

AutoCAD SHX Text

MTD. ON CASTERS

AutoCAD SHX Text

AutoCAD SHX Text

CU1

AutoCAD SHX Text

TUBE COOLING UNIT (PLANE A)

AutoCAD SHX Text

AutoCAD SHX Text 15,355

AutoCAD SHX Text 16 1/2

AutoCAD SHX Text 28 1/4

AutoCAD SHX Text 19 1/4

AutoCAD SHX Text

FLOOR OR SHELF MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

CU2

AutoCAD SHX Text

TUBE COOLING UNIT (PLANE B)

AutoCAD SHX Text

AutoCAD SHX Text 15,355

AutoCAD SHX Text 16 1/2

AutoCAD SHX Text 28 1/4

AutoCAD SHX Text 19 1/4

AutoCAD SHX Text

FLOOR OR SHELF MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

IN2

AutoCAD SHX Text

MEDRAD MARK 7 ARTERION INJECTOR TABLE MOUNTED

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

MEDRAD MARK 7 ARTERION INJECTOR DISPLAY CONTROL UNIT

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 12 1/2

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

IN1

AutoCAD SHX Text

MEDRAD MARK 7 ARTERION INJECTOR POWER UNIT

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 9 3/4

AutoCAD SHX Text 15 1/4

AutoCAD SHX Text 4 3/4

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

UPS

AutoCAD SHX Text

EATON 9390IT 40KVA UPS/BATTERY CABINET

AutoCAD SHX Text 1,595

AutoCAD SHX Text 10,900

AutoCAD SHX Text 35 7/8

AutoCAD SHX Text

AutoCAD SHX Text 73 3/4

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

ATS

AutoCAD SHX Text

EATON 9390IT / 9390 40KVA ATS CABINET

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 19 7/8

AutoCAD SHX Text 16 3/4

AutoCAD SHX Text

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

HMI

AutoCAD SHX Text

EATON 9390IT / 9390 40KVA ATS CONTROL PANEL

AutoCAD SHX Text 4.5

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

RMP

AutoCAD SHX Text

EATON 9390IT / 9390 40KVA REMOTE MONITOR PANEL

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 11 1/4

AutoCAD SHX Text

AutoCAD SHX Text 8 1/2

AutoCAD SHX Text

SEE MFG REQUIREMENTS

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

RAYSAFE REAL-TIME DISPLAY

AutoCAD SHX Text 2.7

AutoCAD SHX Text

AutoCAD SHX Text 11 3/4

AutoCAD SHX Text 2 1/2

AutoCAD SHX Text 9 3/4

AutoCAD SHX Text

WALL MOUNTED (PROC. RM.)

AutoCAD SHX Text

AutoCAD SHX Text

VS

AutoCAD SHX Text

VITALINQ INTERCOM CONSOLE

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

INSTALLED BY CUSTOMER

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

VITALINQ MICROPHONE

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

INSTALLED BY CUSTOMER

A-102ATTENTION:

R E

F E

R E

N C

E

D

O C

U M

E N

T

N

N

S T

R U

C T

I O

N

AutoCAD SHX Text

HEIGHT

AutoCAD SHX Text

CEILING

AutoCAD SHX Text

REQUIREMENT

AutoCAD SHX Text

9 FT. - 6 3/8 IN.

AutoCAD SHX Text IMAGER/LDC SERVICE NOTE: UNDER NORMAL CIRCUMSTANCES, NO LATERAL DISTANCE TO THE WALL OR OTHER COMPONENTS IS REQUIRED. FOR SERVICE, THE IMAGE SYSTEM/LD CONTAINER MUST BE MOVED VIA ROLLERS INTO A POSITION WHERE IT CAN BE ACCESSED FROM ALL SIDES (APPROX. 20" ON EACH SIDE).

AutoCAD SHX Text

SAFETY CLEARANCE

AutoCAD SHX Text

SWIVEL AREA, HEAD/LEFT SIDE POSITION (MOTORIZED MOVEMENT)

AutoCAD SHX Text

PATIENT TABLE MOVEMENT RANGE (FOOT-END)

AutoCAD SHX Text

PATIENT TABLE MOVEMENT RANGE (HEAD-END)

AutoCAD SHX Text

SERVICE CLEARANCE

AutoCAD SHX Text

SAFETY/ SERVICE CLEARANCE

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

BIPLANE ROOM

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

ISOCENTER FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

ORIENTATION POINT OF PATIENT TABLE AT FLOOR

AutoCAD SHX Text ATTENTION! LIGHTING AND HVAC DUCTS HAVING THE POTENTIAL TO HEAT OR COOL SIEMENS COMPONENTS MUST NOT BE LIGHTING AND HVAC DUCTS HAVING THE POTENTIAL TO HEAT OR COOL SIEMENS COMPONENTS MUST NOT BE LOCATED WITHIN SIEMENS CEILING RAIL SYSTEMS AS INDICATED BY THE SHADED AREAS ON THIS PLAN. SPRINKLER HEADS MUST NOT BE LOCATED WITHIN THE SHADED AREAS AS WELL. IF REQUIRED, LAMINAR AIRFLOW DIFFUSERS MAY BE LOCATED WITHIN SIEMENS CEILING RAIL SYSTEMS PROVIDED THEY DO NOT EXTEND BELOW THE FINISHED CEILING OR HEAT/COOL SIEMENS EQUIPMENT. IF PLACED WITHIN SIEMENS RAIL SYSTEMS, THE CUSTOMER MUST ACCEPT RESPONSIBILITY FOR THE FACT THAT, DEPENDING ON THE POSITION OF THE CARRIAGE WITHIN THE CEILING RAILS, THERE IS POTENTIAL FOR CERTAIN DIFFUSERS IN THIS AREA TO BE BLOCKED. THE CUSTOMER MUST ALSO ACCEPT RESPONSIBILITY FOR POTENTIAL DIFFICULTIES IN SERVICING THE CUSTOMER'S MECHANICAL EQUIPMENT IN THE CEILING, IF PLACED WITHIN THESE SHADED AREAS. PLEASE COORDINATE THE PLACEMENT OF THESE ITEMS WITH THE SIEMENS PROJECT MANAGER. SIEMENS SHALL BEAR NO RESPONSIBILITY FOR ANY EQUIPMENT DAMAGES RESULTING FROM THE INSTALLATION OF CUSTOMER-SUPPLIED INFRASTRUCTURE NOT ADHERING TO THE ABOVE STATED REQUIREMENTS.

AutoCAD SHX Text

CEILING STAND LONGITUDINAL RAILS

AutoCAD SHX Text

P2 CABLE OUTLET, SEE SHEET S-102. 12" DIA.

AutoCAD SHX Text

P1 CABLE OUTLET, SEE SHEET S-102. 12" DIA.

AutoCAD SHX Text

DCS MONITOR SUSPENSION SYSTEM

AutoCAD SHX Text

THYROID SHIELD TRACK

AutoCAD SHX Text NOTE: FOR ALL WEIGHTS AND SIZES OF EQUIPMENT SHOWN ON THIS PLAN PLEASE REFER TO THE EQUIPMENT LEGEND ON SHEET A-101.

AutoCAD SHX Text

FOR UNISTRUT PLACEMENT SEE SHEET S-102.

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

BIPLANE ROOM

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

ISOCENTER FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

THYROID SHIELD TRACK

AutoCAD SHX Text

MONITOR TROLLEY OUTLET

AutoCAD SHX Text

1 VETERANS DR, MINNEAPOLIS, MN 55417

AutoCAD SHX Text

BIPLANE ROOM - ARTIS Q BIPLANE ANGIO

AutoCAD SHX Text 30193898

AutoCAD SHX Text

AS NOTED

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

M. YATZUS

AutoCAD SHX Text

THE USE OR REPRODUCTION OF

AutoCAD SHX Text

THIS TITLE BLOCK WITHOUT

AutoCAD SHX Text

PROJECT #:

AutoCAD SHX Text

SHEET:

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

REF. #:

AutoCAD SHX Text

FULL EXTENT OF THE LAW.

AutoCAD SHX Text

SIEMENS AUTHORIZATION WILL

AutoCAD SHX Text

RESULT IN PROSECUTION UNDER

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DATE:

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

CHECKED:

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

PROJECT MANAGER:

AutoCAD SHX Text

TEL:

AutoCAD SHX Text

FAX:

AutoCAD SHX Text

VMAIL:

AutoCAD SHX Text

EXT:

AutoCAD SHX Text

EMAIL:

AutoCAD SHX Text

JIM IRONSIDE

AutoCAD SHX Text

(847) 271-2567

AutoCAD SHX Text james.ironside@siemens-healthineers.com

AutoCAD SHX Text

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

R-101R(C) VERSION DATED 12/03/18

AutoCAD SHX Text

APPROVED BY CUSTOMER FOR FINALS

AutoCAD SHX Text

-ISSUE BLOCK-

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

DATE

AutoCAD SHX Text

SYM

AutoCAD SHX Text

ARTIS Q/Q.ZEN/ZEE BIPLANE ANGIO

AutoCAD SHX Text

REV. 19

AutoCAD SHX Text IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION

AutoCAD SHX Text THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED

AutoCAD SHX Text

AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN

AutoCAD SHX Text

MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.

AutoCAD SHX Text

DOCUMENTS FOR REFERENCE.

AutoCAD SHX Text THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED

AutoCAD SHX Text EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION

AutoCAD SHX Text

PHYSICIST TO SPECIFY RADIATION PROTECTION.

AutoCAD SHX Text

ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.

AutoCAD SHX Text THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.

AutoCAD SHX Text

ALL RIGHTS ARE RESERVED.

AutoCAD SHX Text

CEILING NOTES

AutoCAD SHX Text

1) ALL CEILING MOUNTED LIGHT FIXTURES, MECHANICAL REGISTERS AND SPRINKLER HEADS SHALL BE FLUSH WITH FINISHED CEILING, SHALL BE OUTSIDE OF ALL HATCHED AREAS AND SHALL BE SPECIFIED BY THE ARCHITECT OF RECORD AND SUBSEQUENT CONSULTING ENGINEERS. 2) THE ACTUAL CEILING DESIGN AND COORDINATION OF LIGHTING AND MECHANICAL SYSTEMS SHALL BE THE RESPONSIBILITY OF THE ARCHITECT OF RECORD AND HIS SUBSEQUENT CONSULTING ENGINEERS. 3) THE CUSTOMER/CONTRACTOR SHALL BE RESPONSIBLE FOR FABRICATING, SUPPLYING AND INSTALLING ALL LIGHT, MECHANICAL AND STRUCTURAL SUPPORTING SYSTEMS. SIEMENS MEDICAL SYSTEMS, INC. IS ONLY RESPONSIBLE FOR THE SUPPLYING, INSTALLING AND CALIBRATION OF SMS EQUIPMENT AS SPECIFIED ON THE EQUIPMENT SCHEDULE AS SHOWN ON SHEET A-101. 4) ALL ELECTRICAL AND STRUCTURAL SYSTEMS SHOWN ON THE REFLECTED CEILING PLAN HAVE BEEN COORDINATED WITH THE EQUIPMENT LOCATIONS AS SHOWN ON THE 1/4" SCALE ARCHITECTURAL EQUIPMENT PLAN (SHEET A-101). ANY CHANGES TO THE SMS EQUIPMENT CONFIGURATION AS SHOWN, DUE TO PLACEMENT OF LIGHTING, STRUCTURAL, ELECTRICAL AND MECHANICAL SYSTEMS, MUST BE APPROVED IN WRITING BY THE SMS PROJECT MANAGER PRIOR TO THE COMPLETION OF CONSTRUCTION DOCUMENTS.

AutoCAD SHX Text

REFLECTED CEILING PLAN

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text

SAFETY/SERVICE CLEARANCE PLAN

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text

SAFETY/SERVICE CLEARANCE PLAN

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text 2 8

AutoCAD SHX Text

ATS CABINET MOUNTING

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text 1'-1"

AutoCAD SHX Text 4'-2 1/2"

AutoCAD SHX Text MOUNTING HOLES (4) RECOMMENDED 1/2-13 UNC HEX BOLTS GRADE 5 OR BETTER, WITH WASHERS, TORQUED TO 75 FT-LBS. MOUNTING HARDWARE AND WALL BACKING SUPPLIED BY CUSTOMER/CONTRACTOR

AutoCAD SHX Text

40KW/KVA ATS CABINET (WALL-MOUNTED)

AutoCAD SHX Text

AutoCAD SHX Text

40KW/KVA UPS HMI PANEL

AutoCAD SHX Text

SCALE: NONE

AutoCAD SHX Text

FRONT

AutoCAD SHX Text

SIDE

AutoCAD SHX Text

HMI DISPLAY

AutoCAD SHX Text

TOP

AutoCAD SHX Text

DRYWALL

AutoCAD SHX Text

COVER PANEL

AutoCAD SHX Text

HMI PANEL MOUNTING BOX SUPPLIED BY EATON

AutoCAD SHX Text

CONDUIT KNOCKOUTS

AutoCAD SHX Text

AutoCAD SHX Text

FINISHED FLOOR LINE

AutoCAD SHX Text

CONTROL ROOM DISTRIBUTOR (CR1)

AutoCAD SHX Text

SIDE VIEW

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

CONTROL COUNTER SUPPLIED/INSTALLED BY CUSTOMER/CONTRACTOR

AutoCAD SHX Text

10" X 3 1/2" ELECTRICAL DUCT

AutoCAD SHX Text

SIEMENS SUPPLIED/INSTALLED CONTROL ROOM DISTRIBUTOR (CR1) (64 LBS)

AutoCAD SHX Text WALL BACKING FOR CR1 BOX MOUNTING SUPPLIED/DESIGNED/INSTALLED BY CUSTOMER/CONTRACTOR.

AutoCAD SHX Text

FRONT VIEW

AutoCAD SHX Text

DETAIL A WALL MOUNTING HOLES (4)

AutoCAD SHX Text

DETAIL A

S-101ATTENTION:

R E

F E

R E

N C

E

D

O C

U M

E N

T

N

N

S T

R U

C T

I O

N

AutoCAD SHX Text

HEIGHT

AutoCAD SHX Text

CEILING

AutoCAD SHX Text

REQUIREMENT

AutoCAD SHX Text

9 FT. - 6 3/8 IN.

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text A-102

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text S-101

AutoCAD SHX Text

AutoCAD SHX Text A-102

AutoCAD SHX Text

AutoCAD SHX Text NOTE: FOR ALL WEIGHTS AND SIZES OF EQUIPMENT SHOWN ON THIS PLAN PLEASE REFER TO THE EQUIPMENT LEGEND ON SHEET A-101.

AutoCAD SHX Text CONTRACTOR TO SUPPLY AND INSTALL BACKING FLUSH MOUNTED FOR MOUNTING ELECTRONIC CABINETS. SEE ELECTRONIC CABINET INSTALLATION DETAIL THIS SHEET.

AutoCAD SHX Text CONTRACTOR TO PROVIDE BACKING IN WALL FOR MOUNTING CONTROL ROOM DISTRIBUTOR (CR1).

AutoCAD SHX Text NOTE: THIS WALL AREA BELOW THE COUNTERTOP MUST REMAIN CLEAR OF OUTLETS, COUNTER SUPPORT BRACKETS, ETC. TO ENABLE THE INSTALLATION OF THE "CR1" BOX.

AutoCAD SHX Text A-102

AutoCAD SHX Text

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

BIPLANE ROOM

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

ISOCENTER FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

ORIENTATION POINT OF PATIENT TABLE AT FLOOR

AutoCAD SHX Text

1 VETERANS DR, MINNEAPOLIS, MN 55417

AutoCAD SHX Text

BIPLANE ROOM - ARTIS Q BIPLANE ANGIO

AutoCAD SHX Text 30193898

AutoCAD SHX Text

AS NOTED

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

M. YATZUS

AutoCAD SHX Text

THE USE OR REPRODUCTION OF

AutoCAD SHX Text

THIS TITLE BLOCK WITHOUT

AutoCAD SHX Text

PROJECT #:

AutoCAD SHX Text

SHEET:

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

REF. #:

AutoCAD SHX Text

FULL EXTENT OF THE LAW.

AutoCAD SHX Text

SIEMENS AUTHORIZATION WILL

AutoCAD SHX Text

RESULT IN PROSECUTION UNDER

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DATE:

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

CHECKED:

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

PROJECT MANAGER:

AutoCAD SHX Text

TEL:

AutoCAD SHX Text

FAX:

AutoCAD SHX Text

VMAIL:

AutoCAD SHX Text

EXT:

AutoCAD SHX Text

EMAIL:

AutoCAD SHX Text

JIM IRONSIDE

AutoCAD SHX Text

(847) 271-2567

AutoCAD SHX Text james.ironside@siemens-healthineers.com

AutoCAD SHX Text

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

R-101R(C) VERSION DATED 12/03/18

AutoCAD SHX Text

APPROVED BY CUSTOMER FOR FINALS

AutoCAD SHX Text

-ISSUE BLOCK-

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

DATE

AutoCAD SHX Text

SYM

AutoCAD SHX Text

ARTIS Q/Q.ZEN/ZEE BIPLANE ANGIO

AutoCAD SHX Text

REV. 19

AutoCAD SHX Text IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION

AutoCAD SHX Text THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED

AutoCAD SHX Text

AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN

AutoCAD SHX Text

MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.

AutoCAD SHX Text

DOCUMENTS FOR REFERENCE.

AutoCAD SHX Text THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED

AutoCAD SHX Text EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION

AutoCAD SHX Text

PHYSICIST TO SPECIFY RADIATION PROTECTION.

AutoCAD SHX Text

ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.

AutoCAD SHX Text THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.

AutoCAD SHX Text

ALL RIGHTS ARE RESERVED.

AutoCAD SHX Text

STRUCTURAL NOTES

AutoCAD SHX Text

1) THE CUSTOMER/CONTRACTOR SHALL FURNISH AND INSTALL ALL STRUCTURAL SUPPORT MEMBERS AND NEEDED HARDWARE FOR THE INSTALLATION OF THE SIEMENS EQUIPMENT. 2) THE OVERHEAD STRUCTURAL SUPPORT SYSTEM SHALL BE FIXED, RIGID AND BRACED FOR SWAY. 3) ALL STRUCTURAL SUPPORT MEMBERS SHALL BE TRUE, SQUARE, LEVEL, PARALLEL AND COPLANAR WITH RESPECT TO EACH OTHER, WITH A HORIZONTAL STRUCTURAL SUPPORT MEMBER TO BE LOCATED AND SET WITH A TRANSIT. 4) ALL STRUCTURAL SUPPORT DETAILS SHOWN ARE SAMPLE DETAILS BASED UPON TYPICAL AND STANDARD BUILDING PRACTICES AND ARE NOT INTENDED AS ACTUAL CONSTRUCTION DETAILS. ALL CONSTRUCTION DETAILS AND SUPPORT CALCULATIONS SHALL BE PREPARED BY A PROFESSIONAL STRUCTURAL ENGINEER AT THE CUSTOMER'S EXPENSE. IN THE EVENT AN EXISTING SUPPORT SYSTEM IS TO BE USED, IT WILL BE THE CUSTOMER'S RESPONSIBILITY TO VERIFY THE INTEGRITY OF THAT SYSTEM. 5) MOUNTING PLATES, FRAMES, AND HARDWARE SUPPLIED BY SIEMENS AS DETAILED IN THIS DRAWING SET ARE INSTALLED BY SIEMENS UNLESS OTHERWISE REQUIRED. ANY DEVIATION FROM THE PROVIDED MATERIALS OR MOUNTING METHODS MUST BE DESIGNED AND DOCUMENTED BY THE STRUCTURAL ENGINEER OF RECORD. ALTERNATE MOUNTING MATERIALS (I.E. ANCHORS, THREADED ROD, BACKING PLATES, ETC.) MUST BE SUPPLIED BY THE CUSTOMER/CONTRACTOR. SIEMENS MAY REQUIRE ASSISTANCE FROM THE CUSTOMER/CONTRACTOR WITH INSTALLATION WHEN UTILIZING ALTERNATE MOUNTING MATERIALS. 6) ALL CEILING FIXTURES (I.E. AIR SUPPLY GRILLES, AIR RETURN GRILLES, EXHAUST GRILLES, SPRINKLER HEADS, INCANDESCENT AND FLUORESCENT LIGHT FIXTURES, INTERCOM SPEAKERS, MEDICAL GAS COLUMNS, ETC.) SHALL BE INSTALLED FLUSH MOUNTED WITH THE FINISHED CEILING TO PROVIDE FREE AND UNRESTRICTED TRAVEL OF THE SMS CEILING MOUNTED EQUIPMENT. 7) THE BOTTOM SIDE OF THE UNISTRUT CEILING GRID AND ANY CEILING MOUNTED SUPPORT PLATES ARE TO BE INSTALLED FLUSH WITH THE FINISHED CEILING. THE CUSTOMER/CONTRACTOR SHALL ALSO PROVIDE COVERSTRIPS FOR THE UNISTRUT. 8) THE STRUCTURAL PLANNING AS SHOWN ON THE 1/4" STRUCTURAL PLAN HAS BEEN COORDINATED WITH THE EQUIPMENT LOCATION AS SHOWN ON THE 1/4" EQUIPMENT LAYOUT PLAN. FOR THIS REASON, ANY DEVIATIONS FROM THE STRUCTURAL PLANNING AS SHOWN MUST BE APPROVED BY SMS PLANNING DEPARTMENT. 9) THE STRUCTURAL ENGINEER OF RECORD SHALL BE RESPONSIBLE FOR THE DESIGN AND DETAIL OF FLOOR, WALL, AND CEILING STRUCTURES IN ACCORDANCE WITH THE STRUCTURAL INFORMATION SHOWN, AND LOCAL GOVERNING BUILDING CODES.

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text

STRUCTURAL FLOOR PLAN

AutoCAD SHX Text

FLOOR LEVELNESS IN THE HATCHED AREA (AREA COVERED

AutoCAD SHX Text

BY THE ROTATION OF THE FLOOR STAND AND THE C-ARM

AutoCAD SHX Text

MOVEMENT) MUST NOT EXCEED 1/4" OVER THE WHOLE AREA.

AutoCAD SHX Text 3'-11 1/16"

AutoCAD SHX Text

FLOOR LEVELNESS

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text 1'-8 7/16"

AutoCAD SHX Text

ISOCENTER, FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

AutoCAD SHX Text 3 8

AutoCAD SHX Text

HYGIENIC-GRADE SILICONE

AutoCAD SHX Text

SEAL CONTINUOUSLY WITH

AutoCAD SHX Text

SUPPORT BASE PLATE

AutoCAD SHX Text

INSULATING PARTS

AutoCAD SHX Text

SOLID CONCRETE

AutoCAD SHX Text

DETAIL A

AutoCAD SHX Text

INSULATION PLATE

AutoCAD SHX Text *1)

AutoCAD SHX Text A

AutoCAD SHX Text A

AutoCAD SHX Text

SECTION A-A

AutoCAD SHX Text A

AutoCAD SHX Text

ORIENTATION POINT

AutoCAD SHX Text

PATIENT TABLE

AutoCAD SHX Text

IN HEAD END POSITION

AutoCAD SHX Text

ISOCENTER FLOOR STAND

AutoCAD SHX Text

INSTALLATION PLATE

AutoCAD SHX Text

INSTALLATION PLATE FLOOR STAND AND PATIENT TABLE

AutoCAD SHX Text

NONE

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

FLOOR STAND BASE PLATE SUPPLIED AND INSTALLED BY SIEMENS

AutoCAD SHX Text PATIENT TABLE INSTALLATION PLATE SUPPLIED AND INSTALLED BY SIEMENS. SEE DETAIL #2 THIS SHEET FOR MOUNTING HOLE PATTERN.

AutoCAD SHX Text

*1) INSTALLATION PLATE WITH 13 HOLES, 31/32" DIA.

AutoCAD SHX Text

- FOR INSTALLATION DIRECTLY ON CONCRETE, THE

AutoCAD SHX Text

HILTI HEAVY-DUTY EXPANSION BOLTS HSL-TZ

AutoCAD SHX Text

M12/50 (REQUIRED EMBEDMENT DEPTH OF 5 1/8")

AutoCAD SHX Text

7 HOLES EVENLY SPACED MUST BE USED. THE

AutoCAD SHX Text

MINIMUM CONCRETE QUALITY IS 3700 PSI.

AutoCAD SHX Text

- A MAXIMUM OF 2 HOLES MAY BE FREE BETWEEN

AutoCAD SHX Text

2 MOUNTING POINTS.

AutoCAD SHX Text

(13) X 27.7°

AutoCAD SHX Text

(13) X 31/32" DIA.

AutoCAD SHX Text

HILTI HEAVY DUTY EXPANSION ANCHOR OR THREADED ROD

AutoCAD SHX Text HILTI HEAVY-DUTY EXPANSION BOLTS ARE INCLUDED IN THE SHIPMENT FOR INSTALLING THE MOUNTING PLATES. IF NECESSARY, THE MOUNTING PLATES CAN ALSO BE INSTALLED USING M12 THREADED RODS (ISO 8.8 HARDNESS) SUPPLIED BY CUSTOMER/CONTRACTOR.

AutoCAD SHX Text THE TWO MOUNTING PLATES FOR THE STAND AND THE PATIENT TABLE MUST BE INSTALLED ON A SOLID BASE THAT HAS SUFFICIENT LOAD CAPACITY. CUT AWAY THE FLOOR COVERING. IF NECESSARY, ANY MATERIAL IN THE LOCATION OF THE MOUNTING PLATES THAT DOES NOTE HAVE THE REQUIRED LOAD CAPACITY MUST BE REPLACED WITH FILLED CONCRETE.

AutoCAD SHX Text

FLOOR STAND/PATIENT TABLE

AutoCAD SHX Text

AutoCAD SHX Text

STATIC REQUIREMENTS

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

MOUNTING HOLES 1-7 FOR

AutoCAD SHX Text

INSTALLATION DIRECTLY ON

AutoCAD SHX Text

THE FLOOR

AutoCAD SHX Text

SPARE MOUNTING HOLES

AutoCAD SHX Text

MOUNTING PLATE

AutoCAD SHX Text 2'-4 1/2"%%C

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text T=15

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

THIS MAXIMUM VALUE DOES NOT APPEAR SIMULTANEOUSLY

AutoCAD SHX Text

MAXIMUM TENSION FORCES: 876 LBF. ON MOUNTING

AutoCAD SHX Text

THE OPERATING POSITION AND ON THE UNIT MOVEMENT.

AutoCAD SHX Text

ON ALL MOUNTING POINTS. TENSILE FORCES DEPEND ON

AutoCAD SHX Text

FLOOR STAND

AutoCAD SHX Text

STATIC REQUIREMENTS

AutoCAD SHX Text

HOLES 1-7.

AutoCAD SHX Text THE MOUNTING PLATE IS ANCHORED IN REINFORCED CONCRETE BY MEANS OF 7 EQUALLY-SPACED HILTI HSL-3 M12/50 ANCHORS. PERMISSIBLE MINIMUM QUALITY OF THE CONCRETE 3700 PSI.

AutoCAD SHX Text

AutoCAD SHX Text

DETAIL Y

AutoCAD SHX Text

TABLE MOUNTING PLATE

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

TABLE INSTALLATION PLATE WITH THREADED RODS

AutoCAD SHX Text

FLOOR COVERING

AutoCAD SHX Text

WITH HYGIENIC-GRADE

AutoCAD SHX Text

SEAL CONTINUOUSLY

AutoCAD SHX Text

SILICONE

AutoCAD SHX Text

CABLE LED IN THROUGH

AutoCAD SHX Text

AN OPENING IN THE FLOOR

AutoCAD SHX Text

INSULATION PARTS

AutoCAD SHX Text Y

AutoCAD SHX Text

SOLID CONCRETE

AutoCAD SHX Text

INSTALLATION PLATE

AutoCAD SHX Text 4"

AutoCAD SHX Text

M12 THREADED RODS (ISO 8.8 HARDNESS) SUPPLIED BY CUSTOMER/CONTRACTOR.

AutoCAD SHX Text

AutoCAD SHX Text

PATIENT TABLE TENSION LOADS

AutoCAD SHX Text

STUDS, MINIMUM HARDNESS RATED 8.8 PER THE ISO NORM, WHICH

AutoCAD SHX Text

THE MOUNTING PLATES CAN ALSO BE INSTALLED USING M12 THREADED

AutoCAD SHX Text

SHIPMENT FOR INSTALLING THE MOUNTING PLATES. IF NECESSARY,

AutoCAD SHX Text

HILTI HEAVY DUTY EXPANSION ANCHORS ARE INCLUDED IN THE

AutoCAD SHX Text

NOT HAVE THE REQUIRED LOAD CAPACITY MUST BE REPLACED WITH

AutoCAD SHX Text

MATERIAL IN THE LOCATION OF THE MOUNTING PLATES THAT DOES

AutoCAD SHX Text

CAPACITY. CUT AWAY THE FLOOR COVERING, IF NECESSARY, ANY

AutoCAD SHX Text

MUST BE INSTALLED ON A SOLID BASE THAT HAS SUFFICIENT LOAD

AutoCAD SHX Text

THE TWO MOUNTING PLATES FOR THE STAND AND PATIENT TABLE

AutoCAD SHX Text

ARE INSERTED THROUGH THE CEILING OF THE ROOM BELOW

AutoCAD SHX Text

(THREADED STUDS, ETC. MUST BE OBTAINED LOCALLY).

AutoCAD SHX Text

FLOOR STAND/PATIENT TABLE

AutoCAD SHX Text

FOOT-END LOAD PER INSTALLATION PLATE MOUNTING POINT: MAXIMUM

AutoCAD SHX Text

MOUNTING PLATE ON SOLID CONCRETE.

AutoCAD SHX Text

TENSILE FORCE 1,012 LBS.

AutoCAD SHX Text

FILLED CONCRETE.

AutoCAD SHX Text

GENERAL FLOOR STAND

AutoCAD SHX Text

THE PREINSTALLATION KIT CONTAINS THE MOUNTING PLATE WITH

AutoCAD SHX Text

THE PREINSTALLATION PARTS ARE PART OF THE PREINSTALLATION

AutoCAD SHX Text

FOR INSTALLATION OF THE FLOOR STAND AND PATIENT TABLE

AutoCAD SHX Text

MOUNTING PLATES, THE INSTALLATION GAUGE IS REQUIRED TO

AutoCAD SHX Text

INSTALL THE MOUNTING PLATES WITH THE REQUIRED ACCURACY

AutoCAD SHX Text

THE INSTALLATION GAUGE, PART NO. 65 72 G144E, IS REQUIRED

AutoCAD SHX Text

(=SPACING AND ALIGNMENT OF THE MOUNTING PLATES TO EACH

AutoCAD SHX Text

THE PREINSTALLATION PARTS ARE PART OF THE PREINSTALLATION

AutoCAD SHX Text

INSTALLATION HARDWARE, AS WELL AS THE CEILING OUTLET.

AutoCAD SHX Text

THE PREINSTALLATION KIT CONTAINS THE MOUNTING PLATE WITH

AutoCAD SHX Text

PATIENT TABLE NOTES

AutoCAD SHX Text

INSTALLATION HARDWARE.

AutoCAD SHX Text

PATIENT TABLE

AutoCAD SHX Text

SHIPMENT.

AutoCAD SHX Text

OTHER).

AutoCAD SHX Text

FLOOR STAND

AutoCAD SHX Text

SHIPMENT.

AutoCAD SHX Text

*1 THE MOVEMENT RANGE OF THE CENTER OF GRAVITY FROM

AutoCAD SHX Text

POINT LOAD=CONCENTRATED LOAD

AutoCAD SHX Text

FLOOR STAND PLUS INSTALLATION PLATE: 1,483 LBS.

AutoCAD SHX Text

LOAD IN CASE OF THE OPTION, MULTISPACE.F.

AutoCAD SHX Text

(MULTISPACE.F OPTION CAN BE RETROFITTED).

AutoCAD SHX Text

*2 ADDITIONAL MOVEMENT RANGE OF CONCENTRATED

AutoCAD SHX Text

*3 ISOCENTER, FLOOR STAND IN HEAD SIDE POSITION.

AutoCAD SHX Text

THE HEAD (H) POSITION INTO THE LEFT SIDE (L) POSITION.

AutoCAD SHX Text

INSTALLATION PLATE

AutoCAD SHX Text

CENTER OF GRAVITY PLATE DETAIL

AutoCAD SHX Text 220%%D

AutoCAD SHX Text *2

AutoCAD SHX Text

IN HEAD END POSITION

AutoCAD SHX Text

ISOCENTER, FLOOR STAND

AutoCAD SHX Text *1

AutoCAD SHX Text

IN LEFT SIDE POSITION.

AutoCAD SHX Text

CENTER OF GRAVITY, FLOOR STAND

AutoCAD SHX Text *3

AutoCAD SHX Text

NONE

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

CENTER OF GRAVITY, FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text

AutoCAD SHX Text

DETAIL Z

AutoCAD SHX Text

SILICONE

AutoCAD SHX Text

FLOOR COVERING

AutoCAD SHX Text

INSULATION PARTS

AutoCAD SHX Text

WITH HYGIENIC-GRADE

AutoCAD SHX Text

SEAL CONTINUOUSLY

AutoCAD SHX Text

INSTALLATION PLATE

AutoCAD SHX Text

TABLE INSTALLATION PLATE WITH ANCHORS

AutoCAD SHX Text *3

AutoCAD SHX Text *2,*5

AutoCAD SHX Text *1

AutoCAD SHX Text

SECTION A-D

AutoCAD SHX Text D

AutoCAD SHX Text A

AutoCAD SHX Text

HEAD END

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

TABLE MOUNTING PLATE

AutoCAD SHX Text *4

AutoCAD SHX Text

SOLID CONCRETE

AutoCAD SHX Text *1 MOUNTING USING (4) HILTI HEAVY-DUTY EXPANSION BOLTS, HSL-3 M12/25, SUPPLIED BY SIEMENS. MINIMUM CONCRETE QUALITY IS C20/25. MAX. TENSILE FORCE PER MOUNTING POINT: 1,012 LBS. *2 AREA FOR LEADING CABLES INTO TABLE BASE *3 OPENING IN INSTALLATION PLATE *4 CABLE LED IN THROUGH AN OPENING IN THE FLOOR *5 PATIENT TABLE ORIENTATION POINT

AutoCAD SHX Text

INSTALLATION PLATE SUPPLIED AND INSTALLED BY SIEMENS

AutoCAD SHX Text Z

AutoCAD SHX Text

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

ELECTRONIC CABINET INSTALLATION

AutoCAD SHX Text

WALL ATTACH BRACKETS (9/32" MOUNTING HOLES) SUPPLIED BY SIEMENS

AutoCAD SHX Text

FLOOR LINE

AutoCAD SHX Text

12" MINIMUM CLEARANCE REQUIRED ABOVE THE CABINETS FOR EXHAUST AIR.

AutoCAD SHX Text WALL BACKING AND ANCHOR SCREWS AT TOP OF ELECTRONIC CABINETS FOR WALL ANCHORING SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR.

AutoCAD SHX Text

AIR INTAKE (MAINTAIN 6" CLEARANCE)

AutoCAD SHX Text

EXHAUST AIR

AutoCAD SHX Text

AutoCAD SHX Text 7'-3 7/8"

SEE PLAN FOR SITE-SPECIFIC CABINET CONFIGURATION

S-102ATTENTION:

R E

F E

R E

N C

E

D

O C

U M

E N

T

N

N

S T

R U

C T

I O

N

AutoCAD SHX Text

HEIGHT

AutoCAD SHX Text

CEILING

AutoCAD SHX Text

REQUIREMENT

AutoCAD SHX Text

9 FT. - 6 3/8 IN.

AutoCAD SHX Text X

AutoCAD SHX Text P1

AutoCAD SHX Text D

AutoCAD SHX Text F

AutoCAD SHX Text G

AutoCAD SHX Text C

AutoCAD SHX Text A

AutoCAD SHX Text P2

AutoCAD SHX Text E

AutoCAD SHX Text Z

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text NOTE: "P1" AND "P2" UNISTRUT PIECES TO BE MOUNTED ON TOP OF "X" UNISTRUT GRID. SEE DETAIL #'S 5 & 6, THIS SHEET

AutoCAD SHX Text S-102

AutoCAD SHX Text

AutoCAD SHX Text

MIN.

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

BIPLANE ROOM

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

ISOCENTER FLOOR STAND IN HEAD END POSITION

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text X

AutoCAD SHX Text P1

AutoCAD SHX Text P2

AutoCAD SHX Text A

AutoCAD SHX Text D

AutoCAD SHX Text F

AutoCAD SHX Text G

AutoCAD SHX Text

1 VETERANS DR, MINNEAPOLIS, MN 55417

AutoCAD SHX Text

BIPLANE ROOM - ARTIS Q BIPLANE ANGIO

AutoCAD SHX Text 30193898

AutoCAD SHX Text

AS NOTED

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

M. YATZUS

AutoCAD SHX Text

THE USE OR REPRODUCTION OF

AutoCAD SHX Text

THIS TITLE BLOCK WITHOUT

AutoCAD SHX Text

PROJECT #:

AutoCAD SHX Text

SHEET:

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

REF. #:

AutoCAD SHX Text

FULL EXTENT OF THE LAW.

AutoCAD SHX Text

SIEMENS AUTHORIZATION WILL

AutoCAD SHX Text

RESULT IN PROSECUTION UNDER

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DATE:

AutoCAD SHX Text

SHEET OF

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

CHECKED:

AutoCAD SHX Text

DRAWN BY:

AutoCAD SHX Text

PROJECT MANAGER:

AutoCAD SHX Text

TEL:

AutoCAD SHX Text

FAX:

AutoCAD SHX Text

VMAIL:

AutoCAD SHX Text

EXT:

AutoCAD SHX Text

EMAIL:

AutoCAD SHX Text

JIM IRONSIDE

AutoCAD SHX Text

(847) 271-2567

AutoCAD SHX Text james.ironside@siemens-healthineers.com

AutoCAD SHX Text

AutoCAD SHX Text 12/18/18

AutoCAD SHX Text

R-101R(C) VERSION DATED 12/03/18

AutoCAD SHX Text

APPROVED BY CUSTOMER FOR FINALS

AutoCAD SHX Text

-ISSUE BLOCK-

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

DATE

AutoCAD SHX Text

SYM

AutoCAD SHX Text

ARTIS Q/Q.ZEN/ZEE BIPLANE ANGIO

AutoCAD SHX Text

REV. 19

AutoCAD SHX Text IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION

AutoCAD SHX Text THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED

AutoCAD SHX Text

AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN

AutoCAD SHX Text

MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.

AutoCAD SHX Text

DOCUMENTS FOR REFERENCE.

AutoCAD SHX Text THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED

AutoCAD SHX Text EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION

AutoCAD SHX Text

PHYSICIST TO SPECIFY RADIATION PROTECTION.

AutoCAD SHX Text

ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.

AutoCAD SHX Text THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.

AutoCAD SHX Text

ALL RIGHTS ARE RESERVED.

AutoCAD SHX Text

STRUCTURAL CEILING PLAN

AutoCAD SHX Text

SCALE: 1/4" = 1'-0"

AutoCAD SHX Text 4 8

AutoCAD SHX Text 2,4

AutoCAD SHX Text 2,4

AutoCAD SHX Text 2,4

AutoCAD SHX Text 2,4

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text 2,4

AutoCAD SHX Text 1,3

AutoCAD SHX Text 5,6

AutoCAD SHX Text C

AutoCAD SHX Text A

AutoCAD SHX Text

CEILING STAND MOVES ALONG LONGITUDINAL RAILS

AutoCAD SHX Text

LONGITUDINAL RAILS ATTACHED TO "P-1001" UNISTRUT

AutoCAD SHX Text P1, P2

AutoCAD SHX Text

CEILING OUTLET SUPPORTS

AutoCAD SHX Text

DET

AutoCAD SHX Text

DET

AutoCAD SHX Text

CEILING PLAN LEGEND

AutoCAD SHX Text

SUPPLIED/INSTALLED BY SIEMENS

AutoCAD SHX Text

SYM

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

RAD. SHIELD RAILS ATTACHED TO "P-1001" UNISTRUT

AutoCAD SHX Text

RADIATION SHIELD SUPPORT CARRIAGE MOVES ALONG RAILS

AutoCAD SHX Text

DCS RAILS ATTACHED TO "P-1001" UNISTRUT

AutoCAD SHX Text

DCS CARRIAGE MOVES ALONG LONGITUDINAL RAILS

AutoCAD SHX Text

LONGITUDINAL RAIL SUPPORT MOUNTING POINT BOLTED

AutoCAD SHX Text

TO UNISTRUT FRAME

AutoCAD SHX Text F

AutoCAD SHX Text G

AutoCAD SHX Text D

AutoCAD SHX Text E

AutoCAD SHX Text Z

AutoCAD SHX Text

DESCRIPTION

AutoCAD SHX Text

SUPPLIED/INSTALLED BY CUSTOMER/CONTRACTOR

AutoCAD SHX Text

SYM

AutoCAD SHX Text X

AutoCAD SHX Text

"P-1001" UNISTRUT MOUNTED FLUSH WITH FINISHED

AutoCAD SHX Text

CEILING. MUST BE LEVEL AS SPECIFIED BY SIEMENS

AutoCAD SHX Text

NOTE:

AutoCAD SHX Text

ALL CONSTRUCTION DETAILS AND SUPPORT CALCULATIONS SHALL

AutoCAD SHX Text

AND ARE NOT INTENDED AS ACTUAL CONSTRUCTION DETAILS.

AutoCAD SHX Text

ALL STRUCTURAL SUPPORT DETAILS SHOWN ARE SAMPLE

AutoCAD SHX Text

DETAILS BASED UPON TYPICAL AND STANDARD BUILDING PRACTICES

AutoCAD SHX Text

TO VERIFY THE INTEGRITY OF THAT SYSTEM.

AutoCAD SHX Text

SYSTEM IS TO BE USED, IT WILL BE THE CUSTOMER'S RESPONSIBILITY

AutoCAD SHX Text

AT THE CUSTOMER'S EXPENSE. IN THE EVENT AN EXISTING SUPPORT

AutoCAD SHX Text

BE PREPARED BY A PROFESSIONAL STRUCTURAL ENGINEER

AutoCAD SHX Text

ON STRUCTURAL NOTES AND DETAILS.

AutoCAD SHX Text

ORIENTATION POINT/ISOCENTER, STANDS IN LEFT SIDE POSITION

AutoCAD SHX Text REQUIRED ROOM HEIGHT: 9 FT. - 6 3/8 IN. *1 ROOM HEIGHT MEASURED FROM THE HIGHEST POINT IN THE CROSS HATCHED AREA ON THE FINISHED FLOOR (INCLUDING FLOOR COVERING) TO THE LOWEST POINT OF THE UNISTRUT CEILING, (THE UNISTRUT MUST BE FLUSH WITH THE FINISHED CEILING). HERE, THE AREA OF THE CEILING UNISTRUT IN WHICH THE LONGITUDINAL RAILS FOR THE CEILING STAND ARE INSTALLED IS THE DETERMINING FACTOR. *2 THE CROSS HATCHED AREA IS DETERMINED BY THE MOVEMENT RANGE OF THE FLOOR STAND C-ARM.

AutoCAD SHX Text

EXAM ROOM HEIGHT

AutoCAD SHX Text

EXAM ROOM HEIGHT DETAIL

AutoCAD SHX Text

ORIENTATION POINT/ISOCENTER, STANDS IN HEAD SIDE POSITION

AutoCAD SHX Text

ROOM HEIGHT *1 9'-6 3/8"

AutoCAD SHX Text

FINISHED FLOOR (INCLUDING FLOOR COVERING)

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NONE

AutoCAD SHX Text

UNISTRUT CEILING

AutoCAD SHX Text

AutoCAD SHX Text 2'-2"

AutoCAD SHX Text

ROUND COVER DISK

AutoCAD SHX Text

OPENING IN CEILING

AutoCAD SHX Text

P1001 MAIN UNISTRUT GRID

AutoCAD SHX Text

M10 THREADED STUDS

AutoCAD SHX Text A

AutoCAD SHX Text A

AutoCAD SHX Text

OUTLET DIAMETER APPROX. 4"

AutoCAD SHX Text

P1001 MAIN UNISTRUT GRID

AutoCAD SHX Text 11 13/16"

AutoCAD SHX Text 10"

AutoCAD SHX Text 7 11/16"

AutoCAD SHX Text

SEE NOTE #1

AutoCAD SHX Text

P1 CEILING OUTLET

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NTS

AutoCAD SHX Text

FLOOR STAND CEILING OUTLET

AutoCAD SHX Text

FOR A UNISTRUT CEILING (P1)

AutoCAD SHX Text

1) CEILING ATTACHMENT WITH A PULLING FORCE ON EACH SCREW

AutoCAD SHX Text

NOTES:

AutoCAD SHX Text

OF LESS THAN 225 LBS.

AutoCAD SHX Text

SECTION A-A

AutoCAD SHX Text

FINISHED CEILING

AutoCAD SHX Text

2) UNISTRUT PIECES AND UNISTRUT HARDWARE/ACCESSORIES

AutoCAD SHX Text

SUPPLIED BY CUSTOMER/CONTRACTOR. ALL OTHER ATTACHMENT ITEMS SUPPLIED BY SIEMENS.

AutoCAD SHX Text

"P1" UNISTRUT. SEE PLAN, THIS SHEET, FOR ORIENTATION.

AutoCAD SHX Text 2"

AutoCAD SHX Text 5 1/2"

AutoCAD SHX Text

AutoCAD SHX Text

TROLLEY OUTLET

AutoCAD SHX Text

NONE

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

2" ANGLE

AutoCAD SHX Text

2" ANGLE

AutoCAD SHX Text

MOUNTING HARDWARE INCLUDED IN SHIPMENT

AutoCAD SHX Text

SLAB ABOVE

AutoCAD SHX Text

FINISHED CEILING

AutoCAD SHX Text 7 3/4"

AutoCAD SHX Text 2'-2"

AutoCAD SHX Text

UNISTRUT

AutoCAD SHX Text

CABLE OUTLET

AutoCAD SHX Text 4 3/4"

AutoCAD SHX Text

ANGLED BRACKET (DELIVERED)

AutoCAD SHX Text

AutoCAD SHX Text 7 1/2"

AutoCAD SHX Text 2'-2"

AutoCAD SHX Text 1'-0 5/8"

AutoCAD SHX Text 11"

AutoCAD SHX Text

FINISHED CEILING

AutoCAD SHX Text

OUTLET COVER DISK

AutoCAD SHX Text

OPENING IN CEILING

AutoCAD SHX Text

P1001 MAIN UNISTRUT GRID

AutoCAD SHX Text

M10 THREADED STUDS

AutoCAD SHX Text

SEE NOTE #1

AutoCAD SHX Text A

AutoCAD SHX Text A

AutoCAD SHX Text

SECTION A-A

AutoCAD SHX Text

OUTLET DIAMETER APPROX. 4"

AutoCAD SHX Text

P1001 MAIN UNISTRUT GRID

AutoCAD SHX Text

CORRUGATED HOSE HOLDER

AutoCAD SHX Text

P2 CEILING OUTLET

AutoCAD SHX Text

UNISTRUT CEILING (P2)

AutoCAD SHX Text

CEILING STAND OUTLET FOR A

AutoCAD SHX Text

1) CEILING ATTACHMENT WITH A PULLING FORCE ON EACH SCREW

AutoCAD SHX Text

NOTES:

AutoCAD SHX Text

OF LESS THAN 338 LBS.

AutoCAD SHX Text

SCALE:

AutoCAD SHX Text

NTS

AutoCAD SHX Text

2) UNISTRUT PIECES AND UNISTRUT HARDWARE/ACCESSORIES

AutoCAD SHX Text

SUPPLIED BY CUSTOMER/CONTRACTOR. ALL OTHER ATTACHMENT ITEMS SUPPLIED BY SIEMENS.

AutoCAD SHX Text 5 1/2"

AutoCAD SHX Text

"P2" UNISTRUT. SEE PLAN, THIS SHEET,…

This is the start of the file's text. The full file is on GovTribe.

File details come from the government source that posted it.