36E77618R0054-007.pdf

PDF 2 MB Posted

Attached to
Prepare Digital Radiography Rooms Site Prep Design Build - Fargo
Solicitation number
36E77618R0054
Issued by

About this file

36E77618R0054 ATTACHMENT 3 - SOW Appendix 4 - Samsung Equipment Drawings - 6-6-17.pdf

View the file

Other files

File Type Posted
36E77618R0054-00001001.pdf PDF
36E77618R0054-00001000.docx DOCX document
36E77618R0054-008.zip ZIP file
36E77618R0054-010.pdf PDF
36E77618R0054-013.docx DOCX document
36E77618R0054-003.docx DOCX document
36E77618R0054-005.pdf PDF
36E77618R0054-006.pdf PDF
36E77618R0054-002.docx DOCX document
36E77618R0054-011.docx DOCX document
36E77618R0054-004.docx DOCX document
36E77618R0054-009.pdf PDF
36E77618R0054-012.docx DOCX document
36E77618R0054-001.docx DOCX document
36E77618R0054-000.docx DOCX document
Show all 15

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

- SITE PROGRESS CHECKLIST -

REVIEW EQUIPMENT ORDER FOR EXACT ITEMS PURCHASED. OPTIONAL/-

FUTURE ITEMS NOT ON ORDER MAY BE INDICATED ON THESE PLANS.

ALL ROOM DIMENSIONS ARE CRITICAL! IMMEDIATELY CONTACT GENE NOWACZYK

AT RADSOURCE IF CHANGES OCCUR OR DIMENSIONS ARE NOT CORRECT.

CONTACT A RADIATION PHYSICIST OR CONSULTANT TO SPECIFY

REQUIREMENTS FOR RADIATION CONTAINMENT.

PROVIDE A LOCKABLE EQUIPMENT HOLDING AREA CLOSE TO THE

INSTALLATION FOR STORING TOOLS AND TEST EQUIPMENT.

MAKE SURE A DUST FREE, TEMPERATURE AND HUMIDITY CONTROLLED

ENVIRONMENT IS AVAILABLE FOR STORING THE EQUIPMENT IF YOUR SITE

IS NOT READY FOR INSTALLATION AT THE TIME OF DELIVERY. ONCE THE

SITE IS PREPARED, YOU ARE THEN RESPONSIBLE FOR DELIVERING THE

MAKE ARRANGEMENTS FOR ANY RIGGING, SPECIAL HANDLING, OR FACILITY

MODIFICATIONS THAT MUST BE MADE IN ORDER FOR THE EQUIPMENT TO BE

DELIVERED TO THE INSTALLATION SITE. IF DESIRED, YOUR LOCAL RADSOURCE

TEAM REPRESENTATIVE CAN SUPPLY A REFERENCE LIST OF RIGGERS.

EQUIPMENT TO THE SITE.

CUSTOMER

ALL REQUIREMENTS FOR THIS

EQUIPMENT ARE NOT NOTED ON THIS

SHEET. IT IS SUGGESTED THAT THESE

DRAWINGS BE REVIEWED BY QUALIFIED

PROFESSIONALS WHO CAN ASSIST

WITH MAKING DECISIONS REGARDING

RADIATION CONTAINMENT, MAGNETIC

FIELD CONTAINMENT, ELECTRICAL,

STRUCTURAL AND MECHANICAL

REQUIREMENTS. ALTHOUGH THE

EQUIPMENT MAY BE INSTALLED IN AN

EXISTING ROOM OF SIMILAR FUNCTION,

REQUIREMENTS STILL NEED TO BE

CHECKED.

SITE VISIT

NORMAL PROCEDURE IS TO ATTEND AN

INITIAL CONSTRUCTION MEETING AND

THEN PERFORM AN INSPECTION OF

THE ELECTRICAL BUILD OUT AT OR

ABOUT THE SAME TIME AS THE

ELECTRICAL ROUGH IN (PRIOR TO

SHEETROCK OR LEAD BEING

INSTALLED), ADDITIONAL VISITS MAY BE

REQUIRED

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT MANAGER

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

RADIOLOGY MANAGER

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE:

DRAWING SHEET #

NOTED

A.W.

COVER

6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

INSTALL LOCATION MAP

SAMSUNG

GC85

~ DRAWING INDEX ~

A1 EQUIPMENT LAYOUT

AN NOTES

E1 ELECTRICAL LAYOUT GC85 (1B-35)

E2 ELECTRICAL LAYOUT GC85 (1B-37)

E3 ELECTRICAL LAYOUT GC85 (1B-31)

E1.1 ELECTRICAL DETAILS

E1.3 MAIN PANEL DETAILS

S1 OVERHEAD STRUCTURAL

LAYOUT (1B-35)

S1.1 OVERHEAD STRUCTURAL

LAYOUT (1B-37)

S1.2 OVERHEAD STRUCTURAL

LAYOUT (1B-31)

S2 STRUCTURAL NOTES GENERAL

CONNECTION DETAILS

Q1 EQUIPMENT DETAILS GC85

QE1 EQUIPMENT CABLE LENGTH GC85

PROJECT MANAGER: SHAWN SALANDER

5257 27TH ST S. SUITE 101

FARGO,NORTH DAKOTA

58104

OFFICE PHONE 701-297-0305

CELL PHONE 701-371-0505

E-MAIL SSALANDER@IMAGINGSOLUTIONSINC.COM

RECEIPT OF DRAWING FINAL

THIS SIGNATURE PRINTED OR OTHERWISE

REPRESENTS RECEIPT OF THIS SET OF

PLANS, IT IS UNDERSTOOD THAT ANY

DEVIATION FROM THESE DRAWINGS,

DETAILS AND SPECIFICATIONS MAY

ENCROACH UPON THE EQUIPMENT

OPERATION, SERVICEABILITY, OR SAFETY

GUIDELINES. ALSO I AM AWARE THAT ANY

CHANGES MADE AFTER 8 June 2017

COULD RESULT IN ADDITIONAL EXPENSES

BEING INCURRED IF SAMSUNG

NEUROLOGICA IS NOT NOTIFIED IN

WRITING i.e. E-MAIL ETC... NEW SHEETS

WILL BE ISSUED AS ADDITIONS AND

REVISIONS WILL HAVE CURRENT DATES

ADDED.

NAME:______________________DATE:_______

PRINTED:________________________________

COMPANY:______________________________

NAME:______________________DATE:_______

PRINTED:________________________________

COMPANY:______________________________

NAME:______________________DATE:_______

PRINTED:________________________________

COMPANY:______________________________

NAME:______________________DATE:_______

PRINTED:________________________________

COMPANY:______________________________

NAME:______________________DATE:_______

PRINTED:________________________________

COMPANY:______________________________

US VETERANS MEDICAL CTR FARGO,ND

2101 ELM ST N, FARGO, ND 58102

LAST PLOT DATE: 8 June 2017

LAST REVIEW: 6/6/2017 11:59:53 PM

CUT DOWN

LONG RAILS

3'-6"

4'-5"

3' -8

3 4"

4'-6"

1'-01 2"

1'-13 4"

'-8

1 2"

4'-73 4"

3 4"

8'-111 2"

4' -2

3 4" 3'-

6"

20'-1"

3'-0"

4' -0

'-8

1 4"

3'-81 2"

4 1

31 2"

ROOM LAYOUT - RADIOLOGY 1B-31

SCALE: 1/2" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT

PATIENT TABLE

WALL / CHEST STAND

440LBS

330 LBS

2,046 BTU

511 BTU

3 1 GENERATOR CABINET 220 LBS 11,798 BTU

4 1 CEILING RAIL 140 LBS 3400 BTU

5 1 CEILING MOUNTED TUBE HEAD

6 1 WORK STATION 35 LBS 1,364 BTU

ENVIRONMENT

AMBIENT OPERATION TEMPERATURE: 55-75° (F)

ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR

HUMIDITY: 20-80 PERCENT NON CONDENSING

ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR

EQUIPMENT LEGEND SAMSUNG XGEO-GC85

PROVIDED BY INSTALLED BY

OWNER

OWNER

NOTE:

SPECIAL DELIVERY TO BUILDING NOT COVERED BY SAMSUNG , TO BE

COORDINATED BETWEEN "US VETERANS MEDICAL CTR FARGO,ND

AND/OR N/A" AND SAMSUNG

OWNER

OWNER

OWNER

OWNER

616 LBS 1,909 BTU

7 1 STITCHING STAND 235 LBS N/A BTU OWNER

OWNER: US VETERANS MEDICAL CTR FARGO,ND

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

3'1' 2' 4'

1m

5' 7'

2m

6' 8' 10'

3m

9' DRAWING SHEET #

A1.2

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

NOTE:

ARCHITECT / CONTRACTOR IS TO VERIFY ALL

DIMENSIONS ON THIS DRAWING AGAINST AS BUILT.

DIMENSIONS ILLUSTRATED WITH IN HAVE BEEN

USED FOR PROPER EQUIPMENT PLACEMENT AND

MAY BE USED IN THE FUTURE FOR RADIATION

SHIELDING PROTECTION BY THE CUSTOMERS

SELECTED PHYSICIST.

EACH RESPECTIVE ROOM WILL HAVE AN ENLARGED

VIEW ILLUSTRATING THE PLACEMENT OF THE

RADIOLOGY EQUIPMENT FOR THE ASSOCIATED

ROOM IDENTIFICATION AS SHOWN ON SHEETS

LABELED WITH THE LETTER "AQ-?".

SUPPLEMENTAL NOTES:

ALL ITEMS LISTED BELOW ARE TO BE PROVIDED /

DESIGNED BY THE ARCHITECT N/A. THE ARCHITECT

N/A SHALL CONSULT WITH US VETERANS MEDICAL

CTR FARGO,ND FOR ANY ADDITIONAL INFORMATION.

· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,

MAMMO, R/F, CT, ETC., SHALL HAVE " LED" CAN

LIGHTING AND/OR FLORESCENT LIGHT

FIXTURES. THE LED FIXTURES ARE TO BE

DISTRIBUTED AROUND THE ROOM AND SHOULD

BE PROVIDED WITH DIMMER CONTROLS. THE

DIMMER CONTROLS SHOULD BE LOCATED NEAR

THE ENTRANCES AND IN THE CONTROL AREAS.

· IT IS RECOMMENDED THAT ALL DIAGNOSTIC

ROOMS SHOULD HAVE TILE OR VINYL

FLOORING.

· A TILE DROP CEILING IS RECOMMENDED FOR

ANY DIAGNOSTIC ROOMS. MINIMUM CEILING

HEIGHT 9'-2"

· ALL CABINETRY WORK IS TO BE DESIGNED BY

THE ARCHITECT AND PROVIDED BY US

VETERANS MEDICAL CTR FARGO,ND AND OR N/A

. RECOMMEND ALL COUNTERS SHALL HAVE

ROUNDED CORNERS.

· PROVIDE FOR REFUSE REMOVAL AND DISPOSAL

(E.G. CRATES, CARTONS, PACKING)

· PROVIDE A WORKING RESTROOM WITH

SUPPLIES WITH IN THE FACILITY

NOTE:

THIS SAMSUNG GC85 ROOM CONFIGURATION

UTILIZES A STANDARD THREE (3) METER TRANSVERSE

BRIDGE AND STANDARD FOUR (4) METER LONGITUDE

RAILS OTHER VARIATIONS ARE AVAILABLE AT AN

ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT

YOUR SALES PERSONAL FOR INFORMATION.

COUNTER TOP IN OPERATOR CONTROL AREA SHOULD

BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

RADIOLOGY MANAGER

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

A.W.

NOTED

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

DVS

IB

5 METER LONG

RAILS, 3

METER

BRIDGE

4' -9

4'

-4

81 2" 10

5' -0

7'-11 4"

2"

16'-43 4"

9' -1

01 2"

18'-71 4" 5'-91

2"

4'-0"

3'-0"3'-0"

3' -3

3 4"

2 5

3'-3

ROOM LAYOUT - RADIOLOGY 1B-35

SCALE: 1/2" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT

PATIENT TABLE

WALL / CHEST STAND

440LBS

330 LBS

2,046 BTU

511 BTU

3 1 GENERATOR CABINET 220 LBS 11,798 BTU

4 1 CEILING RAIL 140 LBS 3400 BTU

5 1 CEILING MOUNTED TUBE HEAD

6 1 WORK STATION 35 LBS 1,364 BTU

ENVIRONMENT

AMBIENT OPERATION TEMPERATURE: 55-75° (F)

ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR

HUMIDITY: 20-80 PERCENT NON CONDENSING

ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR

EQUIPMENT LEGEND SAMSUNG XGEO-GC85

PROVIDED BY INSTALLED BY

OWNER

OWNER

NOTE:

SPECIAL DELIVERY TO BUILDING NOT COVERED BY SAMSUNG , TO BE

COORDINATED BETWEEN "US VETERANS MEDICAL CTR FARGO,ND

AND/OR N/A" AND SAMSUNG

OWNER

OWNER

OWNER

OWNER

616 LBS 1,909 BTU

7 1 STITCHING STAND 235 LBS N/A BTU OWNER

OWNER: US VETERANS MEDICAL CTR FARGO,ND

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

3'1' 2' 4'

1m

5' 7'

2m

6' 8' 10'

3m

9' DRAWING SHEET #

A1.1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

NOTE:

ARCHITECT / CONTRACTOR IS TO VERIFY ALL

DIMENSIONS ON THIS DRAWING AGAINST AS BUILT.

DIMENSIONS ILLUSTRATED WITH IN HAVE BEEN

USED FOR PROPER EQUIPMENT PLACEMENT AND

MAY BE USED IN THE FUTURE FOR RADIATION

SHIELDING PROTECTION BY THE CUSTOMERS

SELECTED PHYSICIST.

EACH RESPECTIVE ROOM WILL HAVE AN ENLARGED

VIEW ILLUSTRATING THE PLACEMENT OF THE

RADIOLOGY EQUIPMENT FOR THE ASSOCIATED

ROOM IDENTIFICATION AS SHOWN ON SHEETS

LABELED WITH THE LETTER "AQ-?".

SUPPLEMENTAL NOTES:

ALL ITEMS LISTED BELOW ARE TO BE PROVIDED /

DESIGNED BY THE ARCHITECT N/A. THE ARCHITECT

N/A SHALL CONSULT WITH US VETERANS MEDICAL

CTR FARGO,ND FOR ANY ADDITIONAL INFORMATION.

· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,

MAMMO, R/F, CT, ETC., SHALL HAVE " LED" CAN

LIGHTING AND/OR FLORESCENT LIGHT

FIXTURES. THE LED FIXTURES ARE TO BE

DISTRIBUTED AROUND THE ROOM AND SHOULD

BE PROVIDED WITH DIMMER CONTROLS. THE

DIMMER CONTROLS SHOULD BE LOCATED NEAR

THE ENTRANCES AND IN THE CONTROL AREAS.

· IT IS RECOMMENDED THAT ALL DIAGNOSTIC

ROOMS SHOULD HAVE TILE OR VINYL

FLOORING.

· A TILE DROP CEILING IS RECOMMENDED FOR

ANY DIAGNOSTIC ROOMS. MINIMUM CEILING

HEIGHT 9'-2"

· ALL CABINETRY WORK IS TO BE DESIGNED BY

THE ARCHITECT AND PROVIDED BY US

VETERANS MEDICAL CTR FARGO,ND AND OR N/A

. RECOMMEND ALL COUNTERS SHALL HAVE

ROUNDED CORNERS.

· PROVIDE FOR REFUSE REMOVAL AND DISPOSAL

(E.G. CRATES, CARTONS, PACKING)

· PROVIDE A WORKING RESTROOM WITH

SUPPLIES WITH IN THE FACILITY

NOTE:

THIS SAMSUNG GC85 ROOM CONFIGURATION

UTILIZES A STANDARD THREE (3) METER TRANSVERSE

BRIDGE AND STANDARD FOUR (4) METER LONGITUDE

RAILS OTHER VARIATIONS ARE AVAILABLE AT AN

ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT

YOUR SALES PERSONAL FOR INFORMATION.

COUNTER TOP IN OPERATOR CONTROL AREA SHOULD

BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

RADIOLOGY MANAGER

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

T

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

4 X 32'-63 4"

4' -4

4' -4

1 2"

9'-43 4"

4' -4

1 2"

13'-13 4"

21 2"

4"

18'-71 2"

2' -6

1'-81 4"

'-9

1 4" 4'

-0

2'-6"3'-0"

2' -8

2 5

ROOM LAYOUT - RADIOLOGY 1B-37

SCALE: 1/2" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT

PATIENT TABLE

WALL / CHEST STAND

440LBS

330 LBS

2,046 BTU

511 BTU

3 1 GENERATOR CABINET 220 LBS 11,798 BTU

4 1 CEILING RAIL 140 LBS 3400 BTU

5 1 CEILING MOUNTED TUBE HEAD

6 1 WORK STATION 35 LBS 1,364 BTU

ENVIRONMENT

AMBIENT OPERATION TEMPERATURE: 55-75° (F)

ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR

HUMIDITY: 20-80 PERCENT NON CONDENSING

ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR

EQUIPMENT LEGEND SAMSUNG XGEO-GC85

PROVIDED BY INSTALLED BY

OWNER

OWNER

NOTE:

SPECIAL DELIVERY TO BUILDING NOT COVERED BY SAMSUNG , TO BE

COORDINATED BETWEEN "US VETERANS MEDICAL CTR FARGO,ND

AND/OR N/A" AND SAMSUNG

OWNER

OWNER

OWNER

OWNER

616 LBS 1,909 BTU

7 1 STITCHING STAND 235 LBS N/A BTU OWNER

OWNER: US VETERANS MEDICAL CTR FARGO,ND

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

SAMSUNG

3'1' 2' 4'

1m

5' 7'

2m

6' 8' 10'

3m

9' DRAWING SHEET #

A1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

NOTE:

ARCHITECT / CONTRACTOR IS TO VERIFY ALL

DIMENSIONS ON THIS DRAWING AGAINST AS BUILT.

DIMENSIONS ILLUSTRATED WITH IN HAVE BEEN

USED FOR PROPER EQUIPMENT PLACEMENT AND

MAY BE USED IN THE FUTURE FOR RADIATION

SHIELDING PROTECTION BY THE CUSTOMERS

SELECTED PHYSICIST.

EACH RESPECTIVE ROOM WILL HAVE AN ENLARGED

VIEW ILLUSTRATING THE PLACEMENT OF THE

RADIOLOGY EQUIPMENT FOR THE ASSOCIATED

ROOM IDENTIFICATION AS SHOWN ON SHEETS

LABELED WITH THE LETTER "AQ-?".

SUPPLEMENTAL NOTES:

ALL ITEMS LISTED BELOW ARE TO BE PROVIDED /

DESIGNED BY THE ARCHITECT N/A. THE ARCHITECT

N/A SHALL CONSULT WITH US VETERANS MEDICAL

CTR FARGO,ND FOR ANY ADDITIONAL INFORMATION.

· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,

MAMMO, R/F, CT, ETC., SHALL HAVE " LED" CAN

LIGHTING AND/OR FLORESCENT LIGHT

FIXTURES. THE LED FIXTURES ARE TO BE

DISTRIBUTED AROUND THE ROOM AND SHOULD

BE PROVIDED WITH DIMMER CONTROLS. THE

DIMMER CONTROLS SHOULD BE LOCATED NEAR

THE ENTRANCES AND IN THE CONTROL AREAS.

· IT IS RECOMMENDED THAT ALL DIAGNOSTIC

ROOMS SHOULD HAVE TILE OR VINYL

FLOORING.

· A TILE DROP CEILING IS RECOMMENDED FOR

ANY DIAGNOSTIC ROOMS. MINIMUM CEILING

HEIGHT 9'-2"

· ALL CABINETRY WORK IS TO BE DESIGNED BY

THE ARCHITECT AND PROVIDED BY US

VETERANS MEDICAL CTR FARGO,ND AND OR N/A

. RECOMMEND ALL COUNTERS SHALL HAVE

ROUNDED CORNERS.

· PROVIDE FOR REFUSE REMOVAL AND DISPOSAL

(E.G. CRATES, CARTONS, PACKING)

· PROVIDE A WORKING RESTROOM WITH

SUPPLIES WITH IN THE FACILITY

NOTE:

THIS SAMSUNG GC85 ROOM CONFIGURATION

UTILIZES A STANDARD THREE (3) METER TRANSVERSE

BRIDGE AND STANDARD FOUR (4) METER LONGITUDE

RAILS OTHER VARIATIONS ARE AVAILABLE AT AN

ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT

YOUR SALES PERSONAL FOR INFORMATION.

COUNTER TOP IN OPERATOR CONTROL AREA SHOULD

BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

RADIOLOGY MANAGER

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

MINIMUM SITE PREPARATION REQUIREMENTS

WALLS TO BE PAINTED OR COVERED, BASEBOARDS INSTALLED, FLOORS TO BE

TITLED AND OR COVERED, CEILING SHALL HAVE GRID TILES AND LIGHTING

FIXTURES INSTALLED.

DOORS AND WINDOWS, ESPECIALLY RADIATION PROTECTION BARRIERS,

INSTALLED AND FINISHED WITH LOCK SETS OPERATIONAL.

ALL ELECTRICAL CONVENIENCE, CONDUIT, RACEWAY AND JUNCTION BOXES

INSTALLED.

INCOMING MAINS POWER OPERATIONAL AND CONNECTED TO X-RAY ROOM

BREAKER.

115 VOLTS CONVENIENCE OUTLETS OPERATIONAL.

ALL SUPPORT STRUCTURES CORRECTLY INSTALLED.

ALL CHANNELS, PIPES, BEAM'S AND OR OTHER SUPPORTING DEVICES SHOULD

BE LEVEL, PARALLEL AND FREE OF LATERAL OR LONGITUDINAL

MOVEMENTS.

ALL CONTRACTOR SUPPLIED CABLES PULLED AND TERMINATED.

A DUST FREE ENVIRONMENT IN AND AROUND THE PROCEDURE ROOM.

ALL HEATING AND VENTILATION / AIR-CONDITIONING INSTALLED AND

OPERATIONAL.

A CLEAR DOOR OPENING FOR MOVING EQUIPMENT INTO THE BUILDING MUST

BE 42 IN. X 82 IN. OR LARGER CONTINGENT ON AN 8' CORRIDOR WIDTH.

NOTICE:

THIS DRAWING SET IS THE SOLE PROPERTY OF

NOWACZYK / PEMBER & ASSOCIATES

MEP / ARCHITECTURAL ENGINEERS

IT'S USE IS AUTHORIZED ONLY FOR THE CUSTOMER US

VETERANS MEDICAL CTR FARGO,ND/CONTRACTOR N/A TO

DESIGN AND INCORPORATE OUR CONCEPT INTO

CONSTRUCTION AND PREPARATION FOR IMAGING

EQUIPMENT INSTALLATION. THESE DRAWINGS SHALL NOT

BE USED FOR ANY PURPOSE OTHER THAN THE AGREED

UPON DESIGN BETWEEN NEUROLOGICA AND THEIR

CUSTOMER.

THE CUSTOMER US VETERANS MEDICAL CTR FARGO,ND IS

RESPONSIBLE FOR ALL ROOM PREPARATION COSTS, FEES,

PERMITS AND INSPECTIONS UNLESS OTHERWISE

SPECIFIED IN THE GENERAL ORDER FOR THE EQUIPMENT

PURCHASE.

RADIATION SHIELDING IS NOT SHOWN ON THESE PLANS.

HOWEVER, THE CUSTOMER US VETERANS MEDICAL CTR

FARGO,ND SHALL, AT THEIR OWN EXPENSE, HAVE

SHIELDING CALCULATIONS FOR THE ROOM PREPARED BY A

PREFACE

THESE DRAWINGS HAVE BEEN PREPARED BY

NEUROLOGICA , THEIR PURPOSE IS TO PROVIDE THE

SITING, ELECTRICAL, MECHANICAL AND ENVIRONMENTAL

SPECIFICATIONS REQUIRED TO ACCOMMODATE THE

INSTALLATION AND OPERATION OF THE DIAGNOSTIC

IMAGING EQUIPMENT AND SUB-COMPONENTS

ILLUSTRATED.

THE LAYOUT(S) PROVIDED FOR ALL COMPONENTS

PURCHASED FROM AND/OR PROVIDED BY THE EQUIPMENT

MANUFACTURER SHALL SERVE AS A GUIDE FOR

INSTALLATION BY THE LOCAL SERVICE/INSTALLATION

REPRESENTATIVES.

INFORMATION IN THESE DRAWINGS RELATING TO

BUILDING/FACILITY SPECIFICATIONS THAT WILL SUPPORT

IMAGING EQUIPMENT COMPONENTS SUCH AS ELECTRICAL,

STRUCTURAL, MECHANICAL AND ENVIRONMENTAL

REQUIREMENTS SHALL BE UTILIZED AS A GUIDE BY THE

CUSTOMER'S ARCHITECT N/A AND/OR CONTRACTOR N/A.

FACILITY CONDITIONS MAY DICTATE ACTUAL

CONSTRUCTION METHODS AND MATERIALS APPLIED.

HOWEVER, ALL METHODS AND MATERIALS MUST COMPLY

WITH EQUIPMENT MANUFACTURER SPECIFICATIONS, AS

WELL AS LOCAL AND/OR NATIONAL BUILDING CODES.

THE REQUIRED CEILING HEIGHT OF 9'-2" INDICATED ON

THESE PLANS IS TO INSURE EQUIPMENT FUNCTION IS NOT

INHIBITED. CONSULT WITH YOUR EQUIPMENT

INSTALLATION SPECIALIST REGARDING ACCEPTABILITY OF

THE OTHER CEILING HEIGHTS. CHECK ALL DOOR

OPENINGS FROM DELIVERY LOCATION TO WHERE

EQUIPMENT IS TO BE INSTALLED.

THE QUALITY OF CONSTRUCTION METHODS, MATERIALS

AND CONFORMANCE TO EQUIPMENT MANUFACTURER

SPECIFICATIONS AND TOLERANCES, WILL AFFECT

EQUIPMENT PERFORMANCE.

MAGNETIC INTERFERENCE SPECIFICATIONS

DIGITAL FLAT PANEL MUST BE LOCATED IN AMBIENT STATIC

MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO

GUARANTEE SPECIFIED IMAGING PERFORMANCE.

X-RAY TUBES AND EQUIPMENT MUST BE LOCATED IN

AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN

GAUSS TO GUARANTEE SPECIFIED PERFORMANCE DATA

INTEGRITY

SYSTEM ELECTRONICS / EQUIPMENT MUST BE LOCATED IN

AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN

GAUSS TO GUARANTEE DATA INTEGRITY

CONSOLE EQUIPMENT MUST BE LOCATED IN AMBIENT

STATIC MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO

OBTAIN SPECIFIED GEOMETRIC LINEARITY.

SITE ENVIRONMENT SPECIFICATIONS

AMBIENT OPERATING TEMPERATURE: 55⁰ TO 75⁰ F (16⁰ TO

24⁰ C) MAXIMUM

ALLOWABLE TEMPERATURE CHANGE OF 15⁰ F, (5⁰ C) /

HOUR MAXIMUM ROOM

TEMPERATURE GRADIENT 5⁰ F, (3⁰ C)

HUMIDITY: 20 TO 80 PERCENT NON-CONDENSING, MAXIMUM

ALLOWABLE CHANGE OF 5 PERCENT / HOUR

ALTITUDE: NOT TO EXCEED 10,000 FT (3050M) ABOVE SEA

LEVEL

THE ENVIRONMENT FOR THE ELECTRONICS CABINET MUST

BE CONTROLLED SO THE ABOVE RESTRICTIONS ARE NOT

EXCEEDED.

DO NOT RESTRICT THE AIR INTAKE OR AIR EXHAUST OF THE

SYSTEM COMPONENTS.

ENVIRONMENTAL CONDITIONS LISTED ABOVE MUST BE

MAINTAINED AT ALL TIMES, INCLUDING OVERNIGHT AND

HOLIDAYS.

DRAWING SHEET #

AN

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

RADIOLOGY MANAGER

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

ACR

WL

UB

OC

ST1

CBE1

1 5

XG XT

9' -0

3 4"

12'-01 2"

3' -7

1 2"

3'-101 2"

8' -1

1 2"

10'-63 4"

12'-61 4"

1' -2

3'

-4 1 4"

ELECTRICAL LEGEND

MAIN DISCONNECT CONTRACTOR SUPPLIED, PART #

( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE

OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE

NOWACZYK AT 816-223-6378

480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC

CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM.

FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1)

NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM

BREAKER PER N.E.C.

INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'

OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT

LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,

TERMINATE FLEX WITH A 90DEG BOTH ENDS.

SEE SHEET E-3 FOR DETAILS

1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING,

PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING

ABOVE CEILING OUTLET.

ACR

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM

OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE

BASE

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

UB

SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX,

SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP

BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT

MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY

ELECTRICAL CONTRACTOR

ST1

WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND

INSTALLED BY CONTRACTORWL

XG

CBE1

1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT

COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"

PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)

INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER

LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND

CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.

1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT

MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET

(E1.3)XT

1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER

OF BOX TO 18.0" A.F.F.

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

OC

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF

BOX FLUSH TO FINISHED FLOOR

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

FB

110/115VAC 20 AMP

PHONE / DATA

PHONE

DATA

WALL MOUNTED PHONE 48"A.F.F.W

GROUND FAULT INTERRUPT RECEPTACLE

G.F.I.

SYMBOLS LEGEND

ABOVE CEILING CONDUIT

BELOW GRAD CONDUIT

3 x 10 VERTICAL WALL DUCT

3 x 10 FLUSH TO FINISHED FLOOR DUCT

ELECTRICAL NOTES

ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,

FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT

LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR

STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER,

SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR

DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING

OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS

MUST BE CONTINUOUS COPPER STRANDED AND FREE

FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN.

WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER

SIZES MAY BE REQUIRED BY LOCAL CODES.

IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,

AS REQUIRED IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO

THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT

AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE

HOSPITAL APPROVED OUTLET OR EQUIVALENT.

GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.

CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT

FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO

CEILING MOUNTING COMPONENTS AND WIRING IF HIGH

WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,

USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT

LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING

MOUNTED ACCESSORIES WILL BE PARKED.

ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST

RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE

NEED FOR GREATER THAN STANDARD CABLE LENGTHS

(REFER TO THE INTERCONNECTION DIAGRAM FOR

MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1

CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH

MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG

EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL

CONTRACTOR XXX WITH THE SUPERVISION OF A

RADSOURCE IMAGING REPRESENTATIVE. THE

RADSOURCE IMAGING REPRESENTATIVE WOULD BE

REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION

LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG

EQUIPMENT.

ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE

OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED

FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL

PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW

CONNECTIONS.

LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT

ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED

INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT

OR CONDUCTORS WILL BE ALLOWED.

ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL

CONDUIT (FMC)

ELECTRICAL LAYOUT SAMSUNG GC85 (1B-35)

SCALE: 3/8" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"

REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR

1 CBE1 TO XG

CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @

#3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS

AT "XG". LABEL "FOR X-RAY POWER"

2 CBE1 T0 ST1

CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__

CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE

3 XG TO OC

2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND

5 N/A TO N/A

1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,

LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR

WARNING LIGHT"

6 XG TO FB 2.0" CONDUIT BELOW GRAD

7 XG TO UB 2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

8 INCOMING

MAINS

TO CBE1 CONDUIT AS REQUIRED

SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED

NEA-43080-5W-ST24DC-SPD-WLC-FC

MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT

TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA

SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH

COVER 15.4"w x 32.4"h

NEA-ESTOP22 EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS

NEA-ESTOPCV EMERGENCY STOP COVER PVC

NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY

XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT

W1035-KIT-GC

WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES

10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X

12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED

JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

DRAWING SHEET #

E1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

TT

ACR

WL

UB

OC

ST1

CBE1

1 5

XG XT

9'-83 4"

3' -6

10'-61 4"

7' -1

03 4"

6' -4

3 4"

3 4"

3' -5

1 2"

ELECTRICAL LEGEND

MAIN DISCONNECT CONTRACTOR SUPPLIED, PART #

( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE

OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE

NOWACZYK AT 816-223-6378

480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC

CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM.

FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1)

NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM

BREAKER PER N.E.C.

INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'

OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT

LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,

TERMINATE FLEX WITH A 90DEG BOTH ENDS.

SEE SHEET E-3 FOR DETAILS

1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING,

PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING

ABOVE CEILING OUTLET.

ACR

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM

OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE

BASE

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

UB

SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX,

SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP

BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT

MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY

ELECTRICAL CONTRACTOR

ST1

WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND

INSTALLED BY CONTRACTORWL

XG

CBE1

1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT

COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"

PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)

INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER

LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND

CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.

1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT

MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET

(E1.3)XT

1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER

OF BOX TO 18.0" A.F.F.

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

OC

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF

BOX FLUSH TO FINISHED FLOOR

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

FB

110/115VAC 20 AMP

PHONE / DATA

PHONE

DATA

WALL MOUNTED PHONE 48"A.F.F.W

GROUND FAULT INTERRUPT RECEPTACLE

G.F.I.

SYMBOLS LEGEND

ABOVE CEILING CONDUIT

BELOW GRAD CONDUIT

3 x 10 VERTICAL WALL DUCT

3 x 10 FLUSH TO FINISHED FLOOR DUCT

ELECTRICAL NOTES

ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,

FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT

LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR

STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER,

SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR

DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING

OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS

MUST BE CONTINUOUS COPPER STRANDED AND FREE

FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN.

WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER

SIZES MAY BE REQUIRED BY LOCAL CODES.

IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,

AS REQUIRED IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO

THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT

AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE

HOSPITAL APPROVED OUTLET OR EQUIVALENT.

GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.

CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT

FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO

CEILING MOUNTING COMPONENTS AND WIRING IF HIGH

WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,

USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT

LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING

MOUNTED ACCESSORIES WILL BE PARKED.

ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST

RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE

NEED FOR GREATER THAN STANDARD CABLE LENGTHS

(REFER TO THE INTERCONNECTION DIAGRAM FOR

MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1

CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH

MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG

EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL

CONTRACTOR XXX WITH THE SUPERVISION OF A

RADSOURCE IMAGING REPRESENTATIVE. THE

RADSOURCE IMAGING REPRESENTATIVE WOULD BE

REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION

LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG

EQUIPMENT.

ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE

OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED

FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL

PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW

CONNECTIONS.

LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT

ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED

INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT

OR CONDUCTORS WILL BE ALLOWED.

ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL

CONDUIT (FMC)

ELECTRICAL LAYOUT SAMSUNG GC85 (1B-37)

SCALE: 3/8" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"

ELECTRICAL LAYOUT SAMSUNG GC85 (1B-31)

REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR

1 CBE1 TO XG

CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @

#3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS

AT "XG". LABEL "FOR X-RAY POWER"

2 CBE1 T0 ST1

CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__

CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE

3 XG TO OC

2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND

5 N/A TO N/A

1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,

LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR

WARNING LIGHT"

6 XG TO FB 2.0" CONDUIT BELOW GRAD

7 XG TO UB 2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

8 INCOMING

MAINS

TO CBE1 CONDUIT AS REQUIRED

SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED

NEA-43080-5W-ST24DC-SPD-WLC-FC

MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT

TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA

SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH

COVER 15.4"w x 32.4"h

NEA-ESTOP22 EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS

NEA-ESTOPCV EMERGENCY STOP COVER PVC

NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY

XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT

W1035-KIT-GC

WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES

10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X

12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED

JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

DRAWING SHEET #

E2

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

ACR

WL

UB

OC

ST1

CBE1 XG XT

7'-63 4"

2' -0

3 4" 1' -3

1' -1

01 4"

3'-31 4"

4' -3

9'-01 2"

6' -7

1 4"

ELECTRICAL LEGEND

MAIN DISCONNECT CONTRACTOR SUPPLIED, PART #

( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE

OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE

NOWACZYK AT 816-223-6378

480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC

CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM.

FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1)

NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM

BREAKER PER N.E.C.

INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'

OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT

LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,

TERMINATE FLEX WITH A 90DEG BOTH ENDS.

SEE SHEET E-3 FOR DETAILS

1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING,

PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING

ABOVE CEILING OUTLET.

ACR

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM

OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE

BASE

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

UB

SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX,

SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP

BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT

MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY

ELECTRICAL CONTRACTOR

ST1

WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND

INSTALLED BY CONTRACTORWL

XG

CBE1

1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT

COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"

PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)

INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER

LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL

CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND

CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.

1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT

MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET

(E1.3)XT

1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER

OF BOX TO 18.0" A.F.F.

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

OC

1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF

BOX FLUSH TO FINISHED FLOOR

1 @ COVER

1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,

INSTALLED BY CONTRACTOR (CENTER IN COVER)

FB

110/115VAC 20 AMP

PHONE / DATA

PHONE

DATA

WALL MOUNTED PHONE 48"A.F.F.W

GROUND FAULT INTERRUPT RECEPTACLE

G.F.I.

SYMBOLS LEGEND

ABOVE CEILING CONDUIT

BELOW GRAD CONDUIT

3 x 10 VERTICAL WALL DUCT

3 x 10 FLUSH TO FINISHED FLOOR DUCT

ELECTRICAL NOTES

ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,

FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT

LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR

STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER,

SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR

DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING

OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS

MUST BE CONTINUOUS COPPER STRANDED AND FREE

FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN.

WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER

SIZES MAY BE REQUIRED BY LOCAL CODES.

IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,

AS REQUIRED IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO

THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT

AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE

HOSPITAL APPROVED OUTLET OR EQUIVALENT.

GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.

CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT

FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO

CEILING MOUNTING COMPONENTS AND WIRING IF HIGH

WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,

USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT

LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING

MOUNTED ACCESSORIES WILL BE PARKED.

ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST

RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE

NEED FOR GREATER THAN STANDARD CABLE LENGTHS

(REFER TO THE INTERCONNECTION DIAGRAM FOR

MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1

CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH

MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND

LOCAL ELECTRICAL CODES.

PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG

EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL

CONTRACTOR XXX WITH THE SUPERVISION OF A

RADSOURCE IMAGING REPRESENTATIVE. THE

RADSOURCE IMAGING REPRESENTATIVE WOULD BE

REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION

LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG

EQUIPMENT.

ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE

OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED

FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL

PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW

CONNECTIONS.

LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT

ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED

INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT

OR CONDUCTORS WILL BE ALLOWED.

ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL

CONDUIT (FMC)

ELECTRICAL LAYOUT SAMSUNG GC85 (1B-37)

ELECTRICAL LAYOUT SAMSUNG GC85 (1B-31)

SCALE: 3/8" = 1'-0"

MINIMUM CEILING HEIGHT: 9'-2"

REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR

1 CBE1 TO XG

CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @

#3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS

AT "XG". LABEL "FOR X-RAY POWER"

2 CBE1 T0 ST1

CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__

CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE

3 XG TO OC

2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND

5 N/A TO N/A

1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,

LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR

WARNING LIGHT"

6 XG TO FB 2.0" CONDUIT BELOW GRAD

7 XG TO UB 2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW

GRAD IS ACCEPTABLE

8 INCOMING

MAINS

TO CBE1 CONDUIT AS REQUIRED

SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED

NEA-43080-5W-ST24DC-SPD-WLC-FC

MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT

TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA

SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH

COVER 15.4"w x 32.4"h

NEA-ESTOP22 EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS

NEA-ESTOPCV EMERGENCY STOP COVER PVC

NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY

XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT

W1035-KIT-GC

WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES

10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X

12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED

JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL

DRAWING SHEET #

E3

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

UB

DAT. 0'-0"

FINISHED FLOOR

DAT. 9'-4"

FINISHED CEILING

1'-117 8"

2 UPRIGHT ELEVATION

SCALE: 3/8" = 1'-0"

1 JUNCTION BOX ELEVATION

SCALE: 3/8" = 1'-0"

XG

XT

CBE1

DAT. 0'-0"

FINISHED FLOOR

DAT. 9'-4"

FINISHED CEILING

1' -0

"1'-4"

FIT INTO NEAREST

STUD POCKET

6"

14.5" 3.5"

1' -0

SUPPLIED

DIVIDER

TO

A

B O

V E

F

IN

IS

H

E D

C E

IL

IG

SUPPLIED

DIVIDER

10."3.5"

XG

XT

CONNECT 2.0" CND

FOR INCOMING LINE

FEED TO CBE1

JUNCTION BOX/DUCT DETAILS

SCALE: NTS

DRAWING SHEET #

E1.1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

AC

In Use Sign

X-RAY

Generator Digital

X-RAY

PDU

Imaging Interface CabinetInternal

Controls Local Indication

Emergency Power Off(EPO)

DC

Ethernet

Main Circuit

Breaker

A

B

C

GND

A

B

C

GND

MAIN DISCONNECT ( NEA-43080-4W-ST24DC-SPD-WLC-FC)

Class 2

24VDC

Class 2

24VDC

Contractor Supplied 480V ac, 3 Phase +

GND.

50/60Hz

Phase Type

SPD

ST24DC

(Dry Contact)

24VDC

INCHESDUAL DIMENSIONS:

MILLIMETERS

H150 F OR S

J250 F OR S AND CJ250NS

NEMA/EEMAC TYPE 1

CONDUCTORS

PER

TERMINAL

CIRCUIT

BREAKER

CATALOG NO

PREFIX

HD

JD

WIRING

DIA

A,B A,B

AMPERE

RATING

15-150 150-250

FIELD INSTALLABLE CIRCUIT BREAKER DATA

UL LISTED INTERRUPTING RATING

RMS SYMMETRICAL AMPERES

AC VOLTAGE DC VOLTAGE

240 480Y/277

25K 25K

480 600Y/

600 125 250 500

HG

JG

HJ

JJ

HL

JL

A,B A,B A,B A,B A,B A,B

15-150 150-250 15-150

150-250 15-150

150-250 125K 125K

65K 65K

100K 100K

18K 18K 35K 35K 65K 65K 65K 65K

65K 65K 65K 65K

35K 35K

18K 18K

14K 14K 18K 18K 25K 25K 25K 25K

25K 25K 25K 25K

18K 18K

14K 14K

20K 20K

20K 20K 20K 20K

20K 20K

20K 20K

20K 20K

20K 20K

20K 20K

WIRING DIAGRAM

A B S N N

S

FIELD INSTALLABLE SOLID NEUTRAL ASSEMBLIES AND GROUND KIT

CATALOG

NUMBER

SN100FA 1

MAXIMUM

AMPERE

RATING

NUMBER OF

TERMINALS

100A

PK0GTJ250

WIRE SIZE

(AWG/KCMIL)

MAX MIN

SN400LA 250A

250A

2 1

OR 2

#4

#1/0

#1/0 #14

#12

#1 #1

#600 #250

#6

#300

#300

MM

IN MM

DIAMETER

1.38 35

2.00 51

2.50

3.63 92

3.00 76

IN

SYMBOL

1.00C

1.50

2.00

2.50

3.00

E

F

G

H

CONDUIT SIZE

KNOCKOUTS

J250S 14.35

J250F 32.50

J250S 31.43

27.00

11.90

J250F 15.42

5.96

1.85

31.00

2.85

25.30

2.85

14.00

9.25.25

1.82

2.37

C,E,F,G,H

C,E,F,G,H

C,E,F,G,H

DASHED LINES

INDICATE

SURFACE COVER

(2 HOLES)

.38

[10].25 DIA

[6]

.50 DIA

[12]

.28 DIA

[7]

(2 HOLES)

AA

SEISMIC NOTES:

USE (4) 1/4" DIA GRADE 5 STEEL MOUNTING BOLTS @ HOLES AA

MAX CONFIGURED WEIGHT 41 LB FOR THE PURPOSE OF DETERMINING

SEISMIC ANCHORAGE REQUIREMENTS. FOR ALL OTHER APPLICATIONS,

CONTACT SCHNEIDER ELECTRIC.

2.98

FROM

MTG

SURFACE

5.95

13.50

CG

CG

2.18

13.50

1.23

CIRCUIT BREAKER TERMINAL LUG DATA

15-150

150-175

RATING

AMPERE

JD, JG

CATALOG NO.

HD, HG

PREFIX

CIRCUIT

BREAKER

#3/01

#4/0 #4

MAX

(AWG/KCMIL)

WIRE SIZE

PER PHASE

CONDUCTORS

#14

MIN

TYPE

HJ, HL, HR

AL/CU

JJ, JL, JR AL/CU

200-250

350 #3/0

PK0GTH150 150A 2 1 #14#2

HR

JR

A,B A,B

15-150 150-250

125K 125K

65K 65K

65K 65K

25K 25K

25K 25K

20K 20K

20K 20K

CJ250NS AND H150 COVER OPENING

J250 COVER OPENING

WITH TRIP SETTING ACCESS

3.102.08

.25

.08

4.00 2.08

.25 .08

1.00

4.25

1.25

HAS

OFFSET

FOR 200% NEURALS USE SN400LA (AMP RANGE 100 TO 250,

CU WIRE ONLY).

SEE COVER

OPENING DETAILS

TYPE

CU

AL

AL OR CU

AL OR CU

AL OR CU

AL OR CU

AL OR CU

DRAWING SHEET #

E1.3

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

WALL STAND

TABLE

CABINET

CEILING

WORK STATION

POWER

CABINET WALLSTAND CABINET CABINETTABLE

WIRING : WALL or FLOOR WIRING : FLOOR WIRING : FLOOR

CEILINGCABINET

WIRING : WALL AND CEILING

WIRING : FLOOR

WORK-CABINET

STATION

3.9 m

& CIB

13.1 m

11.2 m

STANDTABLE

WIRING : FLOOR

13 m

WALL-

8.5 m

WALLSTAND CEILING

WIRING : FLOOR

13.65 m

7.85 m

CONTROL ROOM

EXAM ROOM

DISTRIBUTION

POWER

DISTRIBUTION

DRAWING SHEET #

QE1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

TABLE DETAIL DRAWING

FRONT VIEW SIDE VIEW

MIN MAX

TOP VIEW

95.73

67.52

31.73

.4

6 35 .9

10.71

95.58 18.98

5.

.6

WALL STAND DETAIL DRAWING

FRONT VIEW SIDE VIEW

TOP VIEW

MIN

MAX

21.97

.5

.1

.4

20°

31.74

.9

CABINET DETAIL DRAWING

FRONT VIEW SIDE VIEW TOP VIEW

19.29

.7

29.53

.7

29.53

.7

THU DETAIL DRAWING

FRONT VIEW SIDE VIEW

TOP VIEW

Focal Spot

Rotation Conter

Focal Spot

118.50

.3

.3

DRAWING SHEET #

Q1

OFFICE: 816-741-5558

CELL: 816-223-6378

FAX: 816-587-2423

EMAIL: GNOWACZYK@RADSOURCE.NET

8121 N.W. 97TH TERR

KANSAS CITY, MISSOURI 64153

GENE NOWACZYK

MEP

SITE PLANNING &

PROJECT DEVELOPMENT

SAMSUNG GC85

US VETERANS MEDICAL CTR

FARGO,ND

2101 ELM ST N, FARGO, ND 58102

PROJECT

ARCHITECT:

N/A p.N/A N/A

ELECTRICAL CONTRACTOR:

XXX

p.XXX-XXX-XXXX m.XXX-XXX-XXXX

.NET/.COM

GENERAL CONTRACTOR:

N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A

PROJECT MANAGER FOR CUSTOMER:

DENNIS LANGEVIN

p.XXX

DENNIS.LANGEVIN@VA.GOV

IT

KEN LAMMERS

p. XXX

MAILTO:KENNETH.LAMMERS@VA.GOV

INFECTIOUS DISEASE

RYAN WEISER

p.XXX

MAILTO:RYAN.WEISER@VA.GOV

GENE N.DRAWN BY:

CHK'D BY:

SCALE:

DATE: 6/6/2017 12:10:00 PM

FILE NAME:

FARGO VA 06-01-2017 F1.dwg

REVISION: DATE:

CUSTOMER :

STRUCTURAL LAYOUT SAMSUNG GC85 (1B-35)

SCALE: 1/2" = 1'-0"

RECOMMENDED CEILING HEIGHT 9'-2"

A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO WHICH THE STATIONARY CEILING TRACKS

ARE FASTENED.

THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH INSTALLATION.

THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE SPECIFIED ON SITE

PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.

IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS, ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED

TO MEET SPECIFIC SITE REQUIREMENTS BY THE ENGINEER OF RECORD.

CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN. STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS. THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT MOVEMENT IN ANY DIRECTION.

THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE BUILDING…