36C26320Q0176-0003001.pdf
PDF Posted
- Attached to
- Fume Hood Testing and Inspection Service Federal contract opportunity
- Solicitation number
- 36C26320Q0176
About this file
36C26320Q0176 0003 36C26320Q0176 0003 AMENDMENT SF30.pdf
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C26320Q0176-0002000.pdf | ||
| 36C26320Q0176-0001003.docx | DOCX document | |
| 36C26320Q0176-0001004.pdf | ||
| 36C26320Q0176-038.pdf | ||
| 36C26320Q0176-037.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
5. PROJECT NUMBER (if applicable)
CODE 7. ADMINISTERED BY
2. AMENDMENT/MODIFICATION NUMBER
CODE
6. ISSUED BY
8. NAME AND ADDRESS OF CONTRACTOR
4. REQUISITION/PURCHASE REQ. NUMBER 3. EFFECTIVE DATE
9A. AMENDMENT OF SOLICITATION NUMBER
9B. DATED
PAGE OF PAGES
10A. MODIFICATION OF CONTRACT/ORDER NUMBER
10B. DATED
BPA NO. 1. CONTRACT ID CODE
FACILITY CODE CODE
Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:
The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers
E. IMPORTANT:
is extended,
(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR
ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.
12. ACCOUNTING AND APPROPRIATION DATA
(REV. 11/2016)
is required to sign this document and return ___________ copies to the issuing office. is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.
15C. DATE SIGNED
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES
SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).
RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER
Contractor
16C. DATE SIGNED
14. DESCRIPTION OF AMENDMENT/MODIFICATION
16B. UNITED STATES OF AMERICA
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER 16A. NAME AND TITLE OF CONTRACTING OFFICER
15B. CONTRACTOR/OFFEROR
STANDARD FORM 30 PREVIOUS EDITION NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.243
(Type or print) (Type or print)
(Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
(Number, street, county, State and ZIP Code)
(If other than Item 6)
(Specify type of modification and authority)
(such as changes in paying office, appropriation date, etc.)
(If required)
(SEE ITEM 11)
(SEE ITEM 13)
(X)
CHECK
ONE
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
(Signature of person authorized to sign) (Signature of Contracting Officer)
1 1
0003 01-14-2020
None
Department of Veterans Affairs
NETWORK 23 CONTRACTING OFFICE
2501 W. 22nd St.
Sioux Falls SD 57105
John Becker, 605-336-3230 ext 7960 john.becker3@va.gov
Department of Veterans Affairs
NETWORK 23 CONTRACTING OFFICE
2501 W. 22nd St.
Sioux Falls SD 57105
To all Offerors/Bidders
36C26320Q0176
01-03-2020
X
X X
January 16, 2020 10:00 pm CST
See CONTINUATION Page
X one (1)
The purpose of this Amendment is to respond to Solicitation Questions.
The Electrical leakage and ground circuit resistance and polarity tests are not needed for the Clinical Lab BSCs.
The electrical leakage and ground circuit resistance test is something that is included in NSF 49-1992. Cabinets built after that are tested at the factory and do not need that test. The UV bulb testing is also not listed in NSF 49.
These tests do not need to be performed per NSF 49.
Can a copy of the 2019 BSC reports be provided. A copy of the 2019 BSC reports are attached to this solicitation amendment.
The offer due date/time is extended to January 16, 2020, 10:00 PM Central Time. Due to email file sizes, please do not wait until the last minute to submit an offer.
John Becker
Contracting Officer
VA-VHA-2019-95FFC7A6
Test Report Number
TX0242:190311:135929
BIOLOGICAL SAFETY CABINET TEST REPORT
4758 Research Drive San Antonio, TX 78240 (800)345-6094 dba ENV Services Testing & Certification, Inc.
Customer
Address
Contact Telephone Extension
ENV Services Technician Test Date
Equipment Manufacturer Model Serial Number Type Location
Testing and Certification: The purpose of field testing this equipment is to assess whether it is functioning as designed in compliance with the sspecifications, outlined in the appropriate version of NSF/ANSI-49. We perform all test procedures in accordance with these standards and specifications as detailed in ENV Services Procedures, applicable copies of which are available on request. Our testing and certification apply only to to the equipment above, and do not signify approval of the use of any hazardous agents or operational procedures.
SERVICE SUMMARY
Containment tests performed:
HEPA Filter Leak Test Airflow Smoke Patterns Downflow Velocity Profile Test Inflow Velocity Test Site Installation Assessment Tests
Corrected noise level meets NSF criteria for Class II BSCs.
Worker comfort and safety tests performed:
OVERALL CERTIFICATION
X
X X X
PASS FAIL N/A
Repairs performed:
None Filter Leaks Repaired Structural Repairs Made Filters Replaced Airflow Adjusted Pressure Gauge Adusted Warning Alarm System Electrical System Repaired Other:
X
X
EQUIPMENT UTILIZATION SURVEY
Use Biological Agent Used Biosafety Classification Tissue/Cell Cultures
Chemicals Isotopes Surface Decontamination Performed Formaldehyde Gas Decontamination Performed
Using
COMMENTS AND RECOMMENDATIONS
SITE INSTALLATION
Exhaust Alarm or Interlock Functional Cabinet Exhaust
Clinical Non-Carcinogens
NO
Level 2
HUMAN:
70% EtOH
N/A Vented to Room N/A
UNIT CERTIFIED SAFE TO USE
Exhaust Connection PressurizationSash Alarm Function PASS
No Yes
Decontamination Report # N/A
Please print name
Service Technician:Customer:
Signature
ONE VETRANS DRIVE
MINNEAPOLIS, MN 55417
VA MIDWEST HEALTH CENTER
MARGARET KRISHNAN
612/467-2647
BAKER
SG404
127706 Class II, Type A2 BSC
03/11/2019
BA-113
Test Frequency Certification - Annual
Light intensity meets NSF criteria for Class II BSCs.
Electrical results: 115 VAC
0.000 ohms Polarity was correct GFI is functional
Tim Hanson
X
Signature Date
Cabinet Leak Test X
ENV TLBSC- 04/23/12 Page 1 of 2ENV SERVICES, INC. HEADQUARTERS; HATFIELD, PA (800)883-3681
Tim Hanson 11-MAR-2019
For:
Model:
S/N:
Test Report Number
TX0242:190311:135929
VA MIDWEST HEALTH
SG404
127706(800)345-6094
BIOLOGICAL SAFETY CABINET TEST REPORT
Leaks Detected Media: Gasket: Structural:
Supply Filter Quantity/Size:
Approximate Filter Obstruction Caused by Repairs 0.0 % Aerosol Challenge Concentration:
HEPA FILTER LEAK TEST
Exhaust Filter Quantity/Size:
Overall Test Result:
AIRFLOW SMOKE PATTERNS
Smooth downward flow with no dead spots or reflux No outside smoke penetrates onto the work surface
No upward refluxing toward the sash wiper seal No escape of smoke through the sash wiper seal No external turbulence at front access opening
Overall Test Result:Cause of external turbulence, if any _______________________________________________
AIR VELOCITY PROFILES:
Pressure/Flow Gauge Reading Gauge Zeroed
Fan Speed Adjustment Made Final Fan Speed
Specification Used
Overall Inflow Velocity Test Result:
Overall Downflow Velocity Profile Test Result:
ft ft
Area Volume cfm Inflow Exhaust Downflow
Avg. Vel. Specification Uniformity
N/A N/A
N/A
DOWNFLOW VELOCITY PROFILE TEST AND INFLOW VELOCITY TEST
Dens. Corr. 1.000 0 elevation
Downflow Grid:
55 61 57 56 57 52 53 52
48 52 51 49 50 47 47 49
55 52 50 48 48 47 52 52
Exhaust Grid:
272 269 267 273 281 289 287
0.9600 2.56
266 No Change 65%
0.30
NO
SG404
(R) (C) (F)
49 fpm 51 fpm
277 fpm
47 - 57 fpm 42 - 52 fpm
6% 9%
10%55 fpm 47 - 57 fpm
PASS
PASS
Inflow Exhaust
Velometer Model: S/N: Cal. Date:
SECONDARY TESTS AND TEST EQUIPMENT
NOISE Total dbA Ambient CorrecteddbA dbA Background FC 7Average FC 79
Line Voltage VAC ohms
Microwatts/square centimeterAt the work levelUV LIGHT
Criteria <= 70 dbA Corrected Avg. 45 FC > Bkgd, Bkgd <= 15 FC
N/A <= 0.15 ohms or meter min. sensitivity Must be correct Trips when tested None
N/A
N/A
Photometer: S/N: Cal. Date:
Leaks Detected Media: Gasket: Structural:
20X48X3
0 0 0 16X30X6
0 0 0 >= 10 MicroGms/Liter PAO 2 inches per second
PASS
PASS
PASS
PASS
PASS
PASS
PASS
62 61 55
PASS
0.000
PASS
101432 05/16/2018AVM430 2-H 101275 03/05/2018
Filter Scanning Rate:
No Smoke escapes from the cabinet No outside smoke billows over the work surface
LIGHTING
ELECTRICAL
Ground Resistance Polarity Ground Fault Interrupter
Meets
Meets Meets
Meets
PASS
Leakage Repaired with NO LEAKS
Meets
Inflow Method Calculated
PASS
266 cfm100 - 110104 fpm
ENV TLBSC-05/22/09 Page 2 of 2
TX0242:190311:094036
Inventory Number
70-HD-008
BIOLOGICAL SAFETY CABINET TEST REPORT
4758 Research Drive San Antonio, TX 78240 (800)345-6094 dba ENV Services Testing & Certification, Inc.
Customer
Address
Contact Telephone Extension
ENV Services Technician Test Date
Equipment Manufacturer Model Serial Number Type Location
Testing and Certification: The purpose of field testing this equipment is to assess whether it is functioning as designed in compliance with the sspecifications, outlined in the appropriate version of NSF/ANSI-49. We perform all test procedures in accordance with these standards and specifications as detailed in ENV Services Procedures, applicable copies of which are available on request. Our testing and certification apply only to to the equipment above, and do not signify approval of the use of any hazardous agents or operational procedures.
SERVICE SUMMARY
Containment tests performed:
HEPA Filter Leak Test Airflow Smoke Patterns Downflow Velocity Profile Test Inflow Velocity Test Site Installation Assessment Tests
Corrected noise level meets NSF criteria for Class II BSCs.
Worker comfort and safety tests performed:
OVERALL CERTIFICATION
X
X X X
PASS FAIL N/A
Repairs performed:
None Filter Leaks Repaired Structural Repairs Made Filters Replaced Airflow Adjusted Pressure Gauge Adusted Warning Alarm System Electrical System Repaired Other:
X
X
EQUIPMENT UTILIZATION SURVEY
Use Biological Agent Used Biosafety Classification Tissue/Cell Cultures
Chemicals Isotopes Surface Decontamination Performed
Exhaust Alarm or Interlock Functional Cabinet Exhaust
Clinical Non-Carcinogens
NO
Level 2 Other
PASS Thimble Unit NegativeExhaust Connection PressurizationSash Alarm Function N/A
No Yes
Decontamination Report # N/A
Please print name
Service Technician:Customer:
Signature
ONE VETERANS DRIVE
MINNEAPOLIS, MN 55417
VA MIDWEST HEALTH CARE SYTEM
MARGRET KRISHNAN
612/467-2647
NUAIRE
NU-407-600
12026RS
Class II, Type A2 BSC
03/11/2019
BB-130
Test Frequency Certification - Annual
Light intensity does not meet NSF criteria for Class II BSCs.
Electrical results: 116 VAC
0.100 ohms Polarity was correct
Tim Hanson
X
S/N:
Test Report Number
TX0242:190311:094036
VA MIDWEST HEALTH CARE
NU-407-600
12026RS(800)345-6094
BIOLOGICAL SAFETY CABINET TEST REPORT
Leaks Detected Media: Gasket: Structural:
Supply Filter Quantity/Size:
Approximate Filter Obstruction Caused by Repairs 0.0 % Aerosol Challenge Concentration:
HEPA FILTER LEAK TEST
Exhaust Filter Quantity/Size:
Overall Test Result:
AIRFLOW SMOKE PATTERNS
Smooth downward flow with no dead spots or reflux No outside smoke penetrates onto the work surface
No upward refluxing toward the sash wiper seal No escape of smoke through the sash wiper seal No external turbulence at front access opening
Overall Test Result:Cause of external turbulence, if any _______________________________________________
AIR VELOCITY PROFILES:
Pressure/Flow Gauge Reading Gauge Zeroed
Fan Speed Adjustment Made Final Fan Speed
Specification Used
Overall Inflow Velocity Test Result:
Overall Downflow Velocity Profile Test Result:
ft ft
Area Volume cfm Inflow Exhaust Downflow
Avg. Vel. Specification Uniformity
N/A N/A
N/A
DOWNFLOW VELOCITY PROFILE TEST AND INFLOW VELOCITY TEST
Dens. Corr. 1.000 1 elevation
Downflow Grid:
81 81 84 81 87 82 82 84 83 79 80 82
82 84 84 82 82 86 87 81 82 84 89 84
84 86 85 85 81 82 87 81 83 84 86 87
Exhaust Grid:
123 123 119 124 121
4.2800 4.89
522 No Change 80%
0.60
NO
U285E
122 fpm 7%83 fpm 76 - 86 fpm
PASS
PASS
Inflow Exhaust
Velometer Model: S/N: Cal. Date:
SECONDARY TESTS AND TEST EQUIPMENT
NOISE Total dbA Ambient CorrecteddbA dbA Background FC 12Average FC 50
Line Voltage VAC ohms
Microwatts/square centimeterAt the work levelUV LIGHT
Criteria <= 70 dbA Corrected Avg. 45 FC > Bkgd, Bkgd <= 15 FC
N/A <= 0.15 ohms or meter min. sensitivity Must be correct Trips when tested None
N/A
N/A
Photometer: S/N: Cal. Date:
Leaks Detected Media: Gasket: Structural:
(1) 24x72x6 Double Gasket
0 0 0
(1) 24x30x6 Superflow DG
0 0 0 >= 10 MicroGms/Liter PAO 2 inches per second
PASS
PASS
PASS
PASS
PASS
PASS
PASS
65 58 64
PASS
0.100
N/A
101432 05/16/2018AVM430 2-H 101408 12/07/2018
Filter Scanning Rate:
No Smoke escapes from the cabinet No outside smoke billows over the work surface
LIGHTING
ELECTRICAL
Ground Resistance Polarity Ground Fault Interrupter
Meets
Meets
Doesn't meet
PASS
Leakage Repaired with N/A
Meets
N/A
Inflow Method Calculated
PASS
All downflow readings must be within the greater of 25% or 16 fpm of the average.
522 cfm100 - 109107 fpm
TX0242:190311:103235
BIOLOGICAL SAFETY CABINET TEST REPORT
4758 Research Drive San Antonio, TX 78240 (800)345-6094 dba ENV Services Testing & Certification, Inc.
Customer
Address
Contact Telephone Extension
ENV Services Technician Test Date
Equipment Manufacturer Model Serial Number Type Location
Testing and Certification: The purpose of field testing this equipment is to assess whether it is functioning as designed in compliance with the sspecifications, outlined in the appropriate version of NSF/ANSI-49. We perform all test procedures in accordance with these standards and specifications as detailed in ENV Services Procedures, applicable copies of which are available on request. Our testing and certification apply only to to the equipment above, and do not signify approval of the use of any hazardous agents or operational procedures.
SERVICE SUMMARY
Containment tests performed:
HEPA Filter Leak Test Airflow Smoke Patterns Downflow Velocity Profile Test Inflow Velocity Test Site Installation Assessment Tests
Corrected noise level meets NSF criteria for Class II BSCs.
Worker comfort and safety tests performed:
OVERALL CERTIFICATION
X
X X X
PASS FAIL N/A
Repairs performed:
None Filter Leaks Repaired Structural Repairs Made Filters Replaced Airflow Adjusted Pressure Gauge Adusted Warning Alarm System Electrical System Repaired Other:
X
X
EQUIPMENT UTILIZATION SURVEY
Use Biological Agent Used Biosafety Classification Tissue/Cell Cultures
Chemicals Isotopes Surface Decontamination Performed
Exhaust Alarm or Interlock Functional Cabinet Exhaust
Clinical Non-Carcinogens
NO
Level 2 Other
PASS Thimble Unit NegativeExhaust Connection PressurizationSash Alarm Function N/A
No Yes
Decontamination Report # N/A
Please print name
Service Technician:Customer:
Signature
ONE VETERANS DRIVE
MINNEAPOLIS, MN 55417
VA MIDWEST HEALTH CARE SYSTEM
MARGARET KRISHNAN
612/467-2647
NUAIRE
NU-407FM-400
12613RW
Class II, Type A2 BSC
03/11/2019
BB-129
Test Frequency Certification - Annual
Ambient light too bright for NSF test for Class II BSCs.
Electrical results: 116 VAC
0.100 ohms Polarity was correct
Tim Hanson
X
S/N:
Test Report Number
TX0242:190311:103235
VA MIDWEST HEALTH CARE
NU-407FM-400
12613RW(800)345-6094
BIOLOGICAL SAFETY CABINET TEST REPORT
Leaks Detected Media: Gasket: Structural:
Supply Filter Quantity/Size:
Approximate Filter Obstruction Caused by Repairs 0.0 % Aerosol Challenge Concentration:
HEPA FILTER LEAK TEST
Exhaust Filter Quantity/Size:
Overall Test Result:
AIRFLOW SMOKE PATTERNS
Smooth downward flow with no dead spots or reflux No outside smoke penetrates onto the work surface
No upward refluxing toward the sash wiper seal No escape of smoke through the sash wiper seal No external turbulence at front access opening
Overall Test Result:Cause of external turbulence, if any _______________________________________________
AIR VELOCITY PROFILES:
Pressure/Flow Gauge Reading Gauge Zeroed
Fan Speed Adjustment Made Final Fan Speed
Specification Used
Overall Inflow Velocity Test Result:
Overall Downflow Velocity Profile Test Result:
ft ft
Area Volume cfm Inflow Exhaust Downflow
Avg. Vel. Specification Uniformity
N/A N/A
N/A
DOWNFLOW VELOCITY PROFILE TEST AND INFLOW VELOCITY TEST
Dens. Corr. 1.000 1 elevation
Downflow Grid:
76 78 81 84 84 82 81 79
85 82 82 85 84 83 88 81
82 86 84 86 85 82 81 83
Exhaust Grid:
142 139 142 141 138
2.4400 3.23
342 No Change 80%
0.40
NO
U273E
140 fpm 8%83 fpm 78 - 88 fpm
PASS
PASS
Inflow Exhaust
Velometer Model: S/N: Cal. Date:
SECONDARY TESTS AND TEST EQUIPMENT
NOISE Total dbA Ambient CorrecteddbA dbA Background FC 28Average FC 61
Line Voltage VAC ohms
Microwatts/square centimeterAt the work levelUV LIGHT
Criteria <= 70 dbA Corrected Avg. 45 FC > Bkgd, Bkgd <= 15 FC
N/A <= 0.15 ohms or meter min. sensitivity Must be correct Trips when tested None
N/A
N/A
Photometer: S/N: Cal. Date:
Leaks Detected Media: Gasket: Structural:
(1) 24x48x6 Double Gasket
0 0 0
(1) 18x24x6 Superflow DG
0 0 0 >= 10 MicroGms/Liter PAO 2 inches per second
PASS
PASS
PASS
PASS
PASS
PASS
PASS
65 55 65
PASS
0.100
N/A
101432 05/16/2018AVM430 2-H 101406 12/07/2018
Filter Scanning Rate:
No Smoke escapes from the cabinet No outside smoke billows over the work surface
LIGHTING
ELECTRICAL
Ground Resistance Polarity Ground Fault Interrupter
Meets
Meets
PASS
Leakage Repaired with N/A
Meets
N/A
N/A
Inflow Method Calculated
PASS
All downflow readings must be within the greater of 25% or 16 fpm of the average.
342 cfm102 - 112106 fpm
TX0242:190311:105339
Inventory Number
70-HD-010
BIOLOGICAL SAFETY CABINET TEST REPORT
4758 Research Drive San Antonio, TX 78240 (800)345-6094 dba ENV Services Testing & Certification, Inc.
Customer
Address
Contact Telephone Extension
ENV Services Technician Test Date
Equipment Manufacturer Model Serial Number Type Location
Testing and Certification: The purpose of field testing this equipment is to assess whether it is functioning as designed in compliance with the sspecifications, outlined in the appropriate version of NSF/ANSI-49. We perform all test procedures in accordance with these standards and specifications as detailed in ENV Services Procedures, applicable copies of which are available on request. Our testing and certification apply only to to the equipment above, and do not signify approval of the use of any hazardous agents or operational procedures.
SERVICE SUMMARY
Containment tests performed:
HEPA Filter Leak Test Airflow Smoke Patterns Downflow Velocity Profile Test Inflow Velocity Test Site Installation Assessment Tests
Corrected noise level meets NSF criteria for Class II BSCs.
Worker comfort and safety tests performed:
OVERALL CERTIFICATION
X
X X X
PASS FAIL N/A
Repairs performed:
None Filter Leaks Repaired Structural Repairs Made Filters Replaced Airflow Adjusted Pressure Gauge Adusted Warning Alarm System Electrical System Repaired Other:
X
X
EQUIPMENT UTILIZATION SURVEY
Use Biological Agent Used Biosafety Classification Tissue/Cell Cultures
Chemicals Isotopes Surface Decontamination Performed
Exhaust Alarm or Interlock Functional Cabinet Exhaust
Clinical Non-Carcinogens
NO
Level 2 Other
PASS Thimble Unit NegativeExhaust Connection PressurizationSash Alarm Function N/A
No Yes
Decontamination Report # N/A
Please print name
Service Technician:Customer:
Signature
ONE VETERANS DRIVE
MINNEAPOLIS, MN 55417
VA MIDWEST HEALTH CARE SYSTEM
MARGARET KRISHNAN
612/467-2647
NUAIRE
NU-407FM-400
12615RW
Class II, Type A2 BSC
03/11/2019
BB-128
Test Frequency Certification - Annual
Ambient light too bright for NSF test for Class II BSCs.
Electrical results: 116 VAC
0.100 ohms Polarity was correct
Tim Hanson
X
S/N:
Test Report Number
TX0242:190311:105339
VA MIDWEST HEALTH CARE
NU-407FM-400
12615RW(800)345-6094
BIOLOGICAL SAFETY CABINET TEST REPORT
Leaks Detected Media: Gasket: Structural:
Supply Filter Quantity/Size:
Approximate Filter Obstruction Caused by Repairs 0.0 % Aerosol Challenge Concentration:
HEPA FILTER LEAK TEST
Exhaust Filter Quantity/Size:
Overall Test Result:
AIRFLOW SMOKE PATTERNS
Smooth downward flow with no dead spots or reflux No outside smoke penetrates onto the work surface
No upward refluxing toward the sash wiper seal No escape of smoke through the sash wiper seal No external turbulence at front access opening
Overall Test Result:Cause of external turbulence, if any _______________________________________________
AIR VELOCITY PROFILES:
Pressure/Flow Gauge Reading Gauge Zeroed
Fan Speed Adjustment Made Final Fan Speed
Specification Used
Overall Inflow Velocity Test Result:
Overall Downflow Velocity Profile Test Result:
ft ft
Area Volume cfm Inflow Exhaust Downflow
Avg. Vel. Specification Uniformity
N/A N/A
N/A
DOWNFLOW VELOCITY PROFILE TEST AND INFLOW VELOCITY TEST
Dens. Corr. 1.000 1 elevation
Downflow Grid:
79 77 79 81 81 84 79 81
83 83 83 81 86 83 84 83
77 77 76 78 78 81 80 77
Exhaust Grid:
137 134 137 138 137
2.4400 3.23
334 No Change 80%
0.45
NO
U273E
137 fpm 7%80 fpm 78 - 88 fpm
PASS
PASS
Inflow Exhaust
Velometer Model: S/N: Cal. Date:
SECONDARY TESTS AND TEST EQUIPMENT
NOISE Total dbA Ambient CorrecteddbA dbA Background FC 21Average FC 45
Line Voltage VAC ohms
Microwatts/square centimeterAt the work levelUV LIGHT
Criteria <= 70 dbA Corrected Avg. 45 FC > Bkgd, Bkgd <= 15 FC
N/A <= 0.15 ohms or meter min. sensitivity Must be correct Trips when tested None
N/A
N/A
Photometer: S/N: Cal. Date:
Leaks Detected Media: Gasket: Structural:
(1) 24x48x6 Double Gasket
0 0 0
(1) 18x24x6 Superflow DG
0 0 0 >= 10 MicroGms/Liter PAO 2 inches per second
PASS
PASS
PASS
PASS
PASS
PASS
PASS
64 56 63
PASS
0.100
N/A
101432 05/16/2018AVM430 2-H 101406 12/07/2018
Filter Scanning Rate:
No Smoke escapes from the cabinet No outside smoke billows over the work surface
LIGHTING
ELECTRICAL
Ground Resistance Polarity Ground Fault Interrupter
Meets
Meets
PASS
Leakage Repaired with N/A
Meets
N/A
N/A
Inflow Method Calculated
PASS
All downflow readings must be within the greater of 25% or 16 fpm of the average.
334 cfm102 - 112103 fpm
TX0242:190311:090032
BIOLOGICAL SAFETY CABINET TEST REPORT
4758 Research Drive San Antonio, TX 78240 (800)345-6094 dba ENV Services Testing & Certification, Inc.
Customer
Address
Contact Telephone Extension
ENV Services Technician Test Date
Equipment Manufacturer Model Serial Number Type Location
Testing and Certification: The purpose of field testing this equipment is to assess whether it is functioning as designed in compliance with the sspecifications, outlined in the appropriate version of NSF/ANSI-49. We perform all test procedures in accordance with these standards and specifications as detailed in ENV Services Procedures, applicable copies of which are available on request. Our testing and certification apply only to to the equipment above, and do not signify approval of the use of any hazardous agents or operational procedures.
SERVICE SUMMARY
Containment tests performed:
HEPA Filter Leak Test Airflow Smoke Patterns Downflow Velocity Profile Test Inflow Velocity Test Site Installation Assessment Tests
Corrected noise level meets NSF criteria for Class II BSCs.
Worker comfort and safety tests performed:
OVERALL CERTIFICATION
X
X X X
PASS FAIL N/A
Repairs performed:
None Filter Leaks Repaired Structural Repairs Made Filters Replaced Airflow Adjusted Pressure Gauge Adusted Warning Alarm System Electrical System Repaired Other:
X
X
EQUIPMENT UTILIZATION SURVEY
Use Biological Agent Used Biosafety Classification Tissue/Cell Cultures
Chemicals Isotopes Surface Decontamination Performed
Exhaust Alarm or Interlock Functional Cabinet Exhaust
Clinical Non-Carcinogens
NO
Level 2 Other
PASS Direct Connection NegativeExhaust Connection PressurizationSash Alarm Function PASS
No Yes
Decontamination Report # N/A
Please print name
Service Technician:Customer:
Signature
ONE VETERANS DRIVE
MINNEAPOLIS, MN 55417
VA MIDWEST HEALTH CARE SYSTEM
JOHN HOLTER
612/467-2405
BAKER
BCG601
120556 Class II, Type B2 BSC
03/11/2019
BB122
Test Frequency Certification - S/A
Light intensity meets NSF criteria for Class II BSCs.
Electrical results: 116 VAC
0.100 ohms Polarity was correct GFI is functional
Tim Hanson
X
S/N:
Test Report Number
TX0242:190311:090032
VA MIDWEST HEALTH CARE
BCG601
120556(800)345-6094
BIOLOGICAL SAFETY CABINET TEST REPORT
Leaks Detected Media: Gasket: Structural:
Supply Filter Quantity/Size:
Approximate Filter Obstruction Caused by Repairs 0.0 % Aerosol Challenge Concentration:
HEPA FILTER LEAK TEST
Exhaust Filter Quantity/Size:
Overall Test Result:
AIRFLOW SMOKE PATTERNS
Smooth downward flow with no dead spots or reflux No outside smoke penetrates onto the work surface
No upward refluxing toward the sash wiper seal No escape of smoke through the sash wiper seal No external turbulence at front access opening
Overall Test Result:Cause of external turbulence, if any _______________________________________________
AIR VELOCITY PROFILES:
Pressure/Flow Gauge Reading Gauge Zeroed
Fan Speed Adjustment Made Final Fan Speed
Specification Used
Overall Inflow Velocity Test Result:
Overall Downflow Velocity Profile Test Result:
ft ft
Area Volume cfm Inflow Exhaust Downflow
Avg. Vel. Specification Uniformity
N/A N/A
N/A
DOWNFLOW VELOCITY PROFILE TEST AND INFLOW VELOCITY TEST
Dens. Corr. 1.000 0 elevation
Downflow Grid:
36 38 39 42 43 43 43 45 46 48 52 53
45 43 44 45 47 46 48 49 49 51 46 49
39 41 39 42 41 40 42 45 44 46 47 53
Exhaust Grid:
985 981 982 981 981
1.0000 4.01
N/A
BCG60
982 fpm 19%45 fpm 35 - 45 fpm
PASS
PASS
Inflow Exhaust
Velometer Model: S/N: Cal. Date:
SECONDARY TESTS AND TEST EQUIPMENT
NOISE Total dbA Ambient CorrecteddbA dbA Background FC 0Average FC 46
Line Voltage VAC ohms
Microwatts/square centimeterAt the work levelUV LIGHT
Criteria <= 70 dbA Corrected Avg. 45 FC > Bkgd, Bkgd <= 15 FC
N/A <= 0.15 ohms or meter min. sensitivity Must be correct Trips when tested None
N/A
N/A
Photometer: S/N: Cal. Date:
Leaks Detected Media: Gasket: Structural:
18X72X3.54
0 0 0 11X22X8
0 0 0 >= 10 MicroGms/Liter PAO 2 inches per second
PASS
PASS
PASS
PASS
PASS
PASS
PASS
59 59 52
PASS
0.100
PASS
101432 05/16/2018AVM430 2-H 101408 12/07/2018
Filter Scanning Rate:
No Smoke escapes from the cabinet No outside smoke billows over the work surface
LIGHTING
ELECTRICAL
Ground Resistance Polarity Ground Fault Interrupter
Meets
Meets Meets
PASS
Leakage Repaired with N/A
Meets
Inflow Method DIM
PASS
982 cfm100 - 110110 fpm
| 2019 BSC Inspection Results.pdf |
| Tim, VA Mar 2019 |
| VA |
| VA |
| VA |
| VA |
| 70-HD-085 |
| 70-HD-086 |
| 70-HD-090-signed |
| 70-HD-091 |
| 70-HD-102 |
| 70-HD-103 |
| 70-HD-104 |
| 70-HD-110 |
| 70-HD-113 |
| 70-HD-119 |
| 70-HD-120 |
| 70-HD-122 |
| 70-HD-123 |
| 70-HD-126 |
| 70-HD-130 |
| va 2 |
| 49-HD-167 |
| 49-HD-169 |
| 70-HD-006 |
| 70-HD-072 |
| 70-HD-125 |
| 70-HD-129 |
| 127706 |
| va 3 |
| 49-HD-158 |
| 49-HD-171 |
| 70-HD-069 |
| 70-HD-108 |
| 70-HD-109 |
| 70-HD-111 |
| 70-HD-115 |
| 70-HD-124 |
| 70-HD-127 |
| 70-HD-137 |
| 70-HD-187 |
| 127706-2 |
| bsc |
| 12026RS |
| 12026RS-2 |
| 12613RW |
| 12613RW-2 |
| 12615RW |
| 12615RW-2 |
| VA.pdf |
| 49-HD-150 |
| 49-HD-163 |
| 49-HD-169 |
| 120556 |
| 120556-2 |
| VA, 13 MAR |
| 49-HD-150 |
| 61EE59786 |
| 61EE95785 |
| 70-HD-016 |
| 70-HD-030 |
| 70-HD-092 |
| 70-HD-098 |
| 70-HD-099 |
| 70-HD-201 |
| 70-HD-202 |
| 618EE93201 |
| PROSTHETICS 2 |
| PROSTHETICS 3 |
File details come from the government source that posted it. Updated .