Pharmacy_Trailer_Salients_(002).pdf
PDF 320 KB Posted
- Attached to
- Upgrade Site for USP800 Compliant
- Solicitation number
- 36C26119R0002
About this file
36C26119R0002 0001 Pharmacy Trailer Salients (002).pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26119C0111-000.docx | DOCX document | |
| 36C26119R0002_0004.docx | DOCX document | |
| 36C26119R0002_0003.docx | DOCX document | |
| 36C26119R0002_0002.docx | DOCX document | |
| 36C26119R0002_0001.docx | DOCX document | |
| 36C26119R0002-008.doc | DOC document | |
| 36C26119R0002-004.docx | DOCX document | |
| 36C26119R0002-013.docx | DOCX document | |
| 36C26119R0002-010.pdf | ||
| 36C26119R0002-005.docx | DOCX document | |
| 36C26119R0002-011.pdf | ||
| 36C26119R0002-006.docx | DOCX document | |
| 36C26119R0002-009.doc | DOC document | |
| 36C26119R0002-014.pdf | ||
| 36C26119R0002-001.docx | DOCX document | |
| 36C26119R0002-003.docx | DOCX document | |
| 36C26119R0002-007.docx | DOCX document | |
| 36C26119R0002-012.docx | DOCX document | |
| 36C26119R0002-002.pdf | ||
| 36C26119R0002-000.docx | DOCX document |
Show all 20
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALIENT CHARACTERISTICS/PRICING SCHEDULE
Company is to design and build a rectangular semi-trailer with two separate enclosed ante rooms, one sterile hazardous drug compounding room, one sterile non-hazardous drug compounding room, one hazardous drug storage room, one receiving room, one IT closet, and one cleaning closet that meets USP 800 code compliance. If the rooms listed above do not fit in a single semi-trailer, then company needs to design a multi trailers system that have sufficient space for the required rooms above. Access between trailers need to be sterile to comply with USP 800 code.
Company to provide 3 conceptual drawings for selection and obtain approval from the VA prior to manufacturing. The Drawings/Semi-trailer shall comply with the features listed below:
Salient Characteristics USP 800 Hazardous Drug & Non-Hazardous Drug Pharmaceutical Compounding Pharmacy Trailer
ROOMS & AREAS
• Ante Rooms ISO 7
• Housekeeping Sterile Compounding Room located within Ante Room
• Receiving Room
• Sterile Hazardous Drug Buffer Room ISO 7 that is vented to the outside
• Sterile Hazardous Drug Storage Room ISO 7
• Sterile Non-Hazardous Drug Buffer Room ISO 7
• Isolated, secured, and fully function IT closet
• Self-Supporting Modular Design inclusive of walls, roof / ceiling, flooring, lighting, electrical & filtering equipment.
• Exterior partitions between the pharmacy and surrounding spaces shall be 15-minute forced entry construction and extend from slab to roof.
• All surrounding partitions and walls, floor, and ceiling shall be permanently constructed and attached to each other. To provide visual evidence of attempted entry, all construction, to include above the false ceiling and below a raised floor, shall be done in such a manner as to provide visual evidence of unauthorized penetration.
• Partitions and walls surrounding the perimeter of the pharmacy shall be reinforced, slab to roof, with
9-gauge expanded metal. The expanded metal shall be spot welded every 6 inches (150 mm) to vertical and horizontal metal supports of 16-gauge or greater thickness that has been solidly and permanently attached to the true floor and true ceiling.
• Windows or skylights below 18 feet (4 m) from ground level or the roof of a lower abutment, or less than 18 feet (4 m) from windows of an adjoining building, or accessible by a building ledge leading to windows of other floor rooms, shall have forced entry construction of stainless steel woven security mesh.
• All lighting, security devices, and power shall be backed by emergency electrical system in accordance to VA Electrical Design Manual. Refer to VA Electrical Design Manual for specific loads to be connected to the EES.
• All doors to be equipped with sensors and handles for easy access and during emergency.
• Main door to have security system preferably card reader and intercom.
• A security system with following features o Fixed cameras with recording capability at drug storage room and manufacturing area o Motion detectors at drug storage room, manufacturing area, and entrances o Glass break sensor at all windows o Door contact sensors at all exterior doors and entrances o Duress / panic alarm at manufacturing area and storage areas o Card reader and intercom at drug storage room and manufacturing area.
o Exterior lighting to help with camera visibility
• All vents, ducts, and similar openings more than 96 square inches (620 cm2) that enter or pass through a perimeter partition to the pharmacy shall be protected with either bars or grills. Where bars are used, they shall be 1/2-inch (12.7 mm) diameter steel welded vertically and horizontally six inches (150 mm) on center; where grills are used, they shall be of 9-gauge expanded steel.
• All ducts shall be treated to provide sound attenuation.
• Openings in construction above ceilings or below raised access floor shall be protected as above.
• All temperature and pressure monitoring equipment to be capable of producing data output for viewing or downloading.
• Modular Clean Room seismic integrity provisions where necessary.
• Finishes - Non-porous and cleanable surfaces, ceilings, walls, and floors subject to wet cleaning. Solid, nonporous countertop with a smooth seamless appearance. Easy to clean and maintain and with proper cleaning does not support the growth of mold. An acrylic-based solid surface product. Standard thickness of 1”, and a 4” butt backsplash/curb. Also referred to as a work surface or work top. Used in lab and other hospital areas requiring optimum physical and chemical resisting properties.
ANTE ROOMS
• 2 Scrub sinks with eye wash station positioned at least 3’ from the Sterile HD Buffer Room door
• Clean Personal Protective Equipment (PPE) Storage Bin
• Shelves
• Soiled PPE Waste Bin
• Staff Changing Bench
• Environmental Services Closet
• Interlocking HEPA filtered Pass-Thru Chamber from Receiving Room
STERILE HAZARDOUS DRUG BUFFER ROOM Equipment
• 2 ISO 5 Class II Type B2 Biological Safety Cabinets (BSC) with 100% Dedicated Outside Air Supply and Exhaust - 3ft each
• Interlocking HEPA filtered Pass-Thru Chamber from Sterile Hazardous Drug Storage Room
• Stainless Steel Workstation
• RFID Refrigerator for inventory control
• Soiled PPE Waste Bin
• Storage Cabinet for sterile hazardous drugs
HAZARDOUS DRUG STORAGE ROOM Equipment
• USP 800 Compliant Refrigerator
• Shelves
STERILE NON-HAZARDOUS DRUG BUFFER ROOM Equipment
• 2 ISO 5 Horizontal Laminar Airflow Workbench (LAFW) - 4ft
• Stainless Steel Workstation
• USP 800 Compliant Refrigerator
• Soiled PPE Waste Bin
• Non-Hazardous Drug Storage Cabinet
• Wall mounted double locked cabinet to store controlled substances
RECEIVING ROOM
• Interlocking HEPA filtered Pass-Thru Chamber to HD Storage and Ante Room
• Stainless steel work station with network capability
IT ROOM
• One IT relay rack seismically anchored to the floor
• A vertical cable management raceway and horizontal cable management between patch panels
• A grounding bus bar. Electrically ground relay racks using wire # 6
• All horizontal cable to be terminated end to end, label and tested. Company to provide all equipment necessary to fully terminate including but not limited to CAT 6 cable, patch panels, RJ45 ice cubes, modular keystone jacks, faceplates, cable tray, j-hooks.
• Provide and install APC 1500 switch with network card at bottom of relay rack
• Comply with VA Palo Alto OIT Code of Practice
MECHANICAL ROOM
• Back-up Generator with Automatic Transfer Switch and by pass switch to turn on emergency power.
• Water Storage Tanks (Fresh & Gray) with gravity drained that could be hooked up to drainage system
UTILITY REQUIREMENTS
• Electrical: 208V 3-Phase 200 Amps. Emergency panel ready to be hooked up to facility emergency power.
• Water: 1" Potable Water Line
• Sewer: 4" Standard Sewer Line
WHEELCHAIR RAMP
• ADA Compliant
• Aluminum ramp with guardrails, landing, and stairs
• Width: 4'
• Installation on site
ENVIRONMENTAL CONTROLS
• All Facility & Engineering Controls Compliant with new USP 800
• HVAC requirements shall comply with the HVAC Design manual and Pharmacy Design Manual including humidity, noise, cooling and heating temperatures, and humidity requirements.
• Physically separated Ante Room and Sterile Hazardous Drug Buffer Room and Sterile Non-Hazardous Drug
Buffer Room and Hazardous Drug Storage Room and Receiving Room
• C-PEC Containment Primary Environmental Control: (BSC or LAFW) Controls) o ISO 5 Biological Safety Cabinet (BSC) located in the Sterile HD Buffer Room o ISO 5 LAFW located in Sterile Nonhazardous Buffer Room
• C-SEC Containment Secondary Environmental Controls (Room Controls) o ISO 7 Ante Room Environmental Controls:
• Positive pressure (less positive than adjacent sterile non hd buffer room)
• Minimum of 35 Air Exchanges Per Hour (ACPH) compliant with USP 800
• High Efficiency Particulate Air (HEPA) filtered supply air
• 24/7 Real-time Environmental Room Monitor providing:
❖ Temperature
❖ Humidity
❖ Differential pressure
❖ Airborne particle count o ISO 7 Sterile HD Buffer Room Environmental Controls:
• Negative pressure between 0.01” to 0.03”
• Minimum of 35 Air Exchanges Per Hour (ACPH) compliant with USP 800
• 100% Outside Supply Air (into room)
• High Efficiency Particulate Air (HEPA) filtered supply air
• Externally Vented 100% of Air Exhausted Directly Outside
• Zero recirculation of HD Buffer Room air
• 24/7 Real-time Environmental Room Monitor providing:
❖ Temperature
❖ Humidity
❖ Differential pressure
❖ Airborne particle count
• 24/7 Real-time Monitoring of RFID Refrigerator interior temperature to ensure drug safety o ISO 7 Sterile Non-HD Buffer Room Environmental Controls:
• Positive pressure (more positive than the adjacent Ante Room)
• Minimum of 35 Air Exchanges Per Hour (ACPH) compliant with USP 800
• High Efficiency Particulate Air (HEPA) filtered supply air
• 24/7 Real-time Environmental Room Monitor providing:
❖ Temperature
❖ Humidity
❖ Differential pressure
❖ Airborne particle count
❖ 24/7 Real-time Monitoring of RFID Refrigerator interior temperature to ensure drug safety o Hazardous drug storage room requires 35ACPH and negative airflow requirements o Receiving area has a negative or neutral airflow requirement
COMMISSIONING & CERTIFICATION
• Delivery
• Set-up
• On-site Test & Balance Certification
• HEPA Filtration
• Certification Supervision
• On-site Training
• Follow VA Commissioning Manual